Search Results
Found 1 results
510(k) Data Aggregation
(324 days)
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is intended for healthcare professionals to deliver epinephrine 1 mg/mL via intramuscular or subcutaneous injection for pediatric patients.
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is comprised of (A) a 3/10mL Syringe with an attached 29 Gauge x ½ inch Needle, and (B) a plastic overlay which provides barrel markings. The items on the overlay include (1) a zero line, (2) 0.01 mL unit graduation markings, (3) secondary graduation markings at 0, 0.05, 0.10, 0.15, 0.20, 0.25, 0.30 mL, (4) color bars, and (5) the name of the drug for which the syringe is indicated, epinephrine 1 mg/mL.
The Certa Dose™ PD Epinephrine 1 mg/mL IM/SC Syringe is a pre-filled syringe designed for healthcare professionals to deliver epinephrine to pediatric patients.
Acceptance Criteria and Reported Device Performance
Acceptance Criteria / Performance Claim | Reported Device Performance (from the document) | Study Supporting Performance |
---|---|---|
Drug-Device Compatibility | Short contact times do not adversely impact drug quality. | Drug-device compatibility studies |
Syringe Accuracy | Markings are applied correctly and remain accurate. | Syringe accuracy testing after sterilization and accelerated aging |
Human Factors / Usability | The device's use is validated. | Human factors study in accordance with "Guidance for Industry and Food and Drug Administration Staff: Applying Human Factors and Usability Engineering to Medical Devices." dated Feb. 3, 2016. |
Biocompatibility | All materials of the syringe are acceptable per ISO10993-1: 2009/(R) 2013 for Cytotoxicity, Irritation, Sensitization, Systemic Toxicity, and Hemacompatibility. | Biocompatibility testing per ISO10993-1: 2009/(R) 2013 |
Study Information (Based on the Provided Text)
The provided text focuses on demonstrating substantial equivalence to predicate devices and describes the rationale for design choices and conducted tests rather than a standalone clinical study with detailed statistical results. Therefore, many of the requested details about a "study" in the traditional sense of a diagnostic device's performance are not explicitly available. However, based on the information provided, we can infer the following:
- 1. Sample size used for the test set and the data provenance: This information is not provided in the document for any of the described tests. The studies mentioned (drug-device compatibility, syringe accuracy, biocompatibility, human factors) are engineering and laboratory tests, not clinical studies with patient data.
- 2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This information is not provided as these were not studies involving expert consensus on patient data.
- 3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: This information is not provided as these were not studies involving expert consensus on patient data.
- 4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: An MRMC comparative effectiveness study was not performed. The device is a syringe and does not involve AI or human readers for interpretation. The human factors study would assess usability, not diagnostic accuracy improvement.
- 5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The concept of "standalone performance" typically applies to AI algorithms. This device is a physical syringe, so this question is not applicable. The performance tests focus on the physical and material integrity and accuracy of the syringe itself.
- 6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Drug-Device Compatibility: The ground truth would be the expected chemical stability and quality of epinephrine.
- Syringe Accuracy: The ground truth would be established by metrological standards for volume measurement.
- Biocompatibility: The ground truth is defined by the ISO10993-1 standard.
- Human Factors Study: The "ground truth" would be the identified usability issues and user errors during tasks performed by healthcare professionals, evaluated against predefined success criteria.
- 7. The sample size for the training set: This information is not applicable as this is not an AI/machine learning device.
- 8. How the ground truth for the training set was established: This information is not applicable as this is not an AI/machine learning device.
Ask a specific question about this device
Page 1 of 1