K Number
K202453

Validate with FDA (Live)

Date Cleared
2021-03-24

(209 days)

Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine500ng/mL
Oxazepam300 ng/mL
Cocaine150ng/mL
Marijuana50 ng/mL
Methamphetamine500ng/mL
Morphine300ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine. Oxazepam. Marijuana. Methamphetamine, Morphine. Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine500ng/mL
Oxazepam300 ng/mL
Cocaine150ng/mL
Marijuana50 ng/mL
Methamphetamine500ng/mL
Morphine300ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

Device Description

The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

The provided document describes the performance characteristics and studies for the SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup. It does not describe an AI/ML device but rather an in-vitro diagnostic device (IVD) for drug screening. Therefore, several of the requested categories for AI/ML device evaluation are not applicable (e.g., number of experts, adjudication method, MRMC study, standalone performance, training set).

Here's the information extracted from the document, tailored to the nature of the device:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for qualitative drug tests are typically defined by precision around the cutoff concentration. The device is expected to consistently classify samples below the cutoff as negative and above the cutoff as positive. For samples near the cutoff, some variability in classification is expected.

Test Parameter / Drug (Cut-off)Acceptance Criteria (Implicit from study design)Reported Device Performance (Precision Study - Example for Dip Card, Lot 1)
PrecisionSamples < -25% Cut-off: 100% negative calls. Samples > +25% Cut-off: 100% positive calls. Samples within +/-25% of Cut-off: Expected variability.Amphetamine 500: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Cocaine 150: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Methamphetamine 500: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive) Morphine 300: -100% Cut-off: 50-/0+ (100% negative) -75% Cut-off: 50-/0+ (100% negative) -50% Cut-off: 50-/0+ (100% negative) -25% Cut-off: 50-/0+ (100% negative) Cut-off: 24-/26+ (48% negative, 52% positive) +25% Cut-off: 50+/0- (100% positive) +50% Cut-off: 50+/0- (100% positive) +75% Cut-off: 50+/0- (100% positive) +100% Cut-off: 50+/0- (100% positive)
Lay-User Study (Accuracy)For samples far from cutoff (e.g., -50% and +50%), approximately 100% correct results. For samples near cutoff (e.g., -25% and +25%), high accuracy (e.g., >90%).AMP500: -50% Cutoff: 100% correct (0 Pos, 170 Neg)+50% Cutoff: 100% correct (40 Pos, 0 Neg)-25% Cutoff: 90% correct (2 Pos, 18 Neg)+25% Cutoff: 95% correct (19 Pos, 1 Neg)(Similar results reported for COC150, THC, BAR, BZO, MET500, MTD, MOP300, MDMA, OXY, BUP, PCP, TCA, PPX, EDDP)

2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Precision Study (Test Set):

    • For each drug (Amphetamine, Cocaine, Methamphetamine, Morphine), 9 concentrations were tested around the cutoff (-100%, -75%, -50%, -25%, Cut-off, +25%, +50%, +75%, +100%).
    • For each concentration, tests were performed two runs per day for 25 days, using 3 different lots of the device. This equates to 50 replicates per concentration per lot, totaling 450 tests per lot per drug.
    • Total replicates for 4 drugs (AMP, COC, MET, MOP): 9 concentrations x 50 replicates/concentration x 3 lots = 1350 tests per drug. (Some data for other drugs were referenced from prior 510(k)s: K182654, K181968, K153646, K201120).
    • Data Provenance: The document states "in-house" for comparison studies and "urine samples were prepared by spiking drug in negative samples" for precision studies. This suggests a controlled laboratory setting (likely prospective, artificial samples). The document does not specify the country of origin of the data.
  • Method Comparison Study (Clinical/Test Set):

    • 80 "unaltered clinical samples" were used for each drug. These samples were split into categories: 10 negative, 10 low negative, 20 near cutoff negative, 20 near cutoff positive, 20 high positive for each drug.
    • Total samples per device type (Dip Card or Cup) for 4 drugs mentioned: 80 samples x 4 drugs = 320 samples per device type.
    • Data Provenance: "in-house" and "unaltered clinical samples," implying real-world samples but processed within the manufacturer's lab. The document does not specify the country of origin or whether these clinical samples were retrospective or prospectively collected for the study.
  • Lay-User Study (Test Set):

    • 310 lay persons participated for each device format (Dip Card and Cup).
    • Urine samples were prepared at 7 concentrations: negative, +/-75%, +/-50%, +/-25% of the cutoff.
    • Total samples: For each drug, the number of samples varied across concentrations. For example, for AMP500, 20 samples at -100% cutoff, 20 at -75%, 170 at -50%, 20 at -25%, 20 at +25%, 40 at +50%, 20 at +75%. This totals 310 samples per drug per device format.
    • Data Provenance: "at three intended user sites." Samples were "prepared by spiking drugs into drug free-pooled urine specimens," making them artificial but intended to mimic a range of concentrations. This suggests a prospective study design, mimicking real-world use conditions but with controlled samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Precision Study: Ground truth was established by preparing urine samples with known drug concentrations confirmed by LC/MS. No human experts were involved in establishing ground truth, as it was an analytical study.
  • Method Comparison Study: The ground truth for clinical samples was established by LC/MS results. No mention of human experts for ground truth.
  • Lay-User Study: Ground truth was established by LC/MS results of the prepared spiked urine samples.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. This device is a qualitative diagnostic test read directly by users, not an AI/ML imaging device requiring expert adjudication. In the method comparison study, three laboratory assistants "ran" the samples, implying they performed the test, but the ground truth was LC/MS, not their consensus.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/ML device, and no MRMC study comparing human readers with and without AI assistance was mentioned. The lay-user study evaluated the device's performance with lay users, not an "AI assistance" scenario.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. The device is a physical, lateral flow immunochromatographic assay. Its performance inherently involves a human interpreting the result line, even if it's a simple positive/negative visual interpretation. It is not an algorithm-only device. The precision and method comparison studies evaluate the device's analytical performance, which is its inherent "standalone" capability.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • LC/MS (Liquid Chromatography-Mass Spectrometry) was used as the ground truth method to confirm drug concentrations in both spiked samples (for precision and lay-user studies) and clinical samples (for method comparison studies). This is a highly accurate and quantitative analytical method.

8. The sample size for the training set

  • Not applicable. This is not an AI/ML device that requires a "training set" in the machine learning sense. The device is based on immunoassay principles.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

March 24, 2021

Safecare Biotech(Hangzhou)Co., Ltd. % Joe Shia, Manager LSI International 504 E Diamond Ave., Suite I Gaithersburg, MD 20877

Re: K202453

Trade/Device Name: SAFECARE Multi-Drug Urine Test Cup SAFECARE Multi-Drug Urine Test Dip-Card Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Code: NFT, NFW, NFY, NGG, NGI, NFV, NGL, PTH, PTG, QAW, QBF, NGM Dated: August 25, 2020 Received: August 27, 2020

Dear Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kellie Kelm, Ph.D. Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202453

Device Name

SAFECARE® Multi-Drug Urine Test Dip Card SAFECARE® Multi-Drug Urine Test Cup

Indications for Use (Describe)

SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine500ng/mL
Oxazepam300 ng/mL
Cocaine150ng/mL
Marijuana50 ng/mL
Methamphetamine500ng/mL
Morphine300ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine. Oxazepam. Marijuana. Methamphetamine, Morphine. Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine500ng/mL
Oxazepam300 ng/mL
FORM FDA 3881 (7/17)Page 1 of 2

{3}------------------------------------------------

Cocaine150ng/mL
Marijuana50 ng/mL
Methamphetamine500ng/mL
Morphine300ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

510(k) Number: K202453

    1. Date: March 18, 2021
    1. Submitter: Safecare Biotech (Hangzhou) Co. Ltd. 18 Haishu Road, Yuhang District Hangzhou, China
    1. Contact person: Joe Shia LSI International Inc. 504E Diamond Ave., Suite I Gaithersburg, MD 20877 Telephone: 240-505-7880 Email: shiajl@yahoo.com
  • SAFECARE® Multi-Drug Urine Test Dip Card 4. Device Name: SAFECARE® Multi-Drug Urine Test Cup

Classification:

Class 2

Product CodeClassificationRegulation SectionPanel
NFTAmphetamineII21 CFR § 862.3100, Amphetamine Test SystemToxicology (91)
NFWCannabinoidsII21 CFR § 862.3870, Cannabinoids Test SystemToxicology (91)
NFYCocaineII21 CFR § 862.3250, Cocaine and Cocaine Metabolites Test SystemToxicology (91)
NGGMethamphetamineII21 CFR § 862.3610, Methamphetamine Test SystemToxicology (91)
NGIMorphineII21 CFR § 862.3640, Morphine Test SystemToxicology (91)
NFVOxazepamII21 CFR § 862.3170, Benzodiazepine Test SystemToxicology (91)
NGLOxycodoneII21 CFR § 862.3650, Opiate Test SystemToxicology (91)
PTHSecobarbitalII21 CFR § 862.3150, Barbiturate Test SystemToxicology (91)
NGLBuprenorphineII21 CFR § 862.3650, Opiate Test SystemToxicology (91)
NGGMethylenedioxy-methamphetamineII21 CFR § 862.3610, Methamphetamine Test SystemToxicology (91)
NGMPhencyclidineunclassifiedEnzyme Immunoassay PhencyclidineToxicology (91)
PTGMethadoneII21 CFR § 862.3620, Methadone Test SystemToxicology (91)
QAWNortriptylineII21 CFR, 862.3910 Tricyclic Antidepressant Drugs Test SystemToxicology (91)
QBFPropoxypheneII21 CFR, 862.3700 Propoxyphene Test SystemToxicology (91)

{5}------------------------------------------------

Product CodeClassificationRegulation SectionPanel
PTG2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidineII21 CFR § 862.3620, Methadone TestSystemToxicology(91)
    1. Predicate Device
      The SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup (K201120)
    1. Intended Use
      SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:
Drug (Identifier)Cut-off level
Amphetamine500ng/mL
Oxazepam300 ng/mL
Cocaine150ng/mL
Marijuana50 ng/mL
Methamphetamine500ng/mL
Morphine300ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d

{6}------------------------------------------------

Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine500ng/mL
Oxazepam300 ng/mL
Cocaine150ng/mL
Marijuana50 ng/mL
Methamphetamine500ng/mL
Morphine300ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxy-methamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

7. Device Description

The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

8. Substantial Equivalence Information

A summary comparison of features of the SAFECARE® Dip Card Tests and SAFECARE® Cup Tests and the predicate devices is provided in following table.

{7}------------------------------------------------

Features Comparison of SAFECARE® Dip Card Tests and the Predicate Device

ItemDevicePredicate - K201120
Indication(s)for UseFor the qualitative determination ofAmphetamine, Buprenorphine, Secobarbital,Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine,Methamphetamine,Methylenedioxymethamphetamine,Morphine, Methadone, Oxycodone,Phencyclidine, Propoxyphene, Nortriptylineand Cannabinoids in human urine.Same
Similarities
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsDip CardSame
Differences
Calibrator and Cut-OffValuesAmphetamine (AMP): 500 ng/mlOxazepam (BZO):300 ng/mlCocaine (COC): 150 ng/ml11-Nor-△9-Tetrahydrocannabinol-9-COOH(THC):50 ng/mlMethamphetamine (MET): 500 ng/mlMorphine (MOP): 300ng/mlSecobarbital (BAR): 300 ng/mlMethadone (MTD): 300 ng/mlMethylenedioxymethamphetamine(MDMA): 500 ng/mlOxycodone (OXY): 100 ng/mlBuprenorphine (BUP): 10 ng/mlPhencyclidine (PCP): 25 ng/mlNortriptyline (TCA): 1000 ng/mlPropoxyphene (PPX): 300 ng/ml2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP): 300 ng/mLSame as candidate devicewith exception ofAmphetamine: 1000ng/mLCocaine: 300ng/mLMethamphetamine:1000ng/mLMorphine: 2000ng/mL

{8}------------------------------------------------

Features Comparison of SAFECARE® Cup Tests and the Predicate Devices

ItemDevicePredicate - K201120
Indication(s)for UseFor the qualitative determination ofAmphetamine, Buprenorphine,Secobarbital, Oxazepam, Cocaine, 2-ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine, Methamphetamine,Methylenedioxymethamphetamine,Morphine, Methadone, Oxycodone,Phencyclidine, Propoxyphene,Nortriptyline and Cannabinoids inhuman urine.Same
Similarities
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsCupCup
Differences
Calibrator and Cut-OffValuesAmphetamine (AMP): 500 ng/mlOxazepam (BZO):300 ng/mlCocaine (COC): 150 ng/ml11-Nor-△9-Tetrahydrocannabinol-9-COOH(THC):50 ng/mlMethamphetamine (MET): 500 ng/mlMorphine (MOP): 300ng/mlSecobarbital (BAR): 300 ng/mlMethadone (MTD): 300 ng/mlMethylenedioxymethamphetamine(MDMA): 500 ng/mlOxycodone (OXY): 100 ng/mlBuprenorphine (BUP): 10 ng/mlPhencyclidine (PCP): 25 ng/mlNortriptyline (TCA): 1000 ng/mlPropoxyphene (PPX): 300 ng/ml2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP): 300 ng/mLSame as candidate devicewith exception ofAmphetamine: 1000ng/mLCocaine: 300ng/mLMethamphetamine:1000ng/mLMorphine: 2000ng/mL

{9}------------------------------------------------

9. Test Principle

The SAFECARE® Dip Card Tests, and SAFECARE® Cup Tests are rapid tests for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3diphenylpyrrolidine in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration. it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off. -75% cut off, -50% cut off, -25% cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables for Amphetamine 500, Cocaine 150, Methamphetamine 500 and Morphine 300. Please refer to K182654 for precision data for Methylenedioxy-Methamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene, and to K181968 for precision data for Oxazepam, Secobarbital and Methadone, and to K153646 for precision data for Cannabinoids and to K201120 for precision data for EDDP. The candidate device uses the same assays (same reagent pads) for these drugs as was cleared in K182654/K181968/K153646/K201120.

ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Amphetamine 500 Dip Card

{10}------------------------------------------------

Cup
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Cocaine 150

Dip Card

Results-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Cup
Results-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

Methamphetamine 500

50-/0+

Lot 3

Dip Card
Results
-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-

50-/0+

50-/0+

Cup
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

50-/0+

25-/25+

50+/0-

50+/0-

Cut-off

+100%

50+/0-

50+/0-

50+/0-

Cut-off

+75%

50+/0-

50+/0-

50+/0-

{11}------------------------------------------------

Morphine 300
Dip Card
Results
-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Cup
Results
-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

The cut-off values of 500 ng/mL, 500 ng/mL, and 300 ng/mL for Amphetamine, Cocaine, Methamphetamine and Morphine are verified.

  • b. Linearity
    Not applicable.

  • c. Stability
    The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 50 ℃ and real time stability studies at 4℃ and 30 ℃.

  • d. Interference
    Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100ug/mL, with the exception of albumin with no interference at a concentration of 100mg/dL and ethanol at 1% volume, are summarized in the following tables. There were no differences observed between the SAFECARE® Cup and Dip Card formats.

Acetominophen (4-Acetamidophenol)Ecgonine methyl esterD,L-Octopamine
AcetophenetidinErythromycinOxalic acid
N-Acetylprocainamideβ-EstradiolOxolinic acid
Acetylsalicylic acidEthanol (1% vol)Oxymetazoline
Albumin (100mg/dL)FenoprofenPapaverine
AminopyrineFurosemidePenicillin-G
AmoxicillinGentisic acidPerphenazine
AmpicillinHemoglobinPhenelzine
ApomorphineHydralazinePrednisone
Ascorbic acidHydrochlorothiazideDL-Propranolol
AspartameHydrocortisoneD-Pseudoephedrine
AtropineO-Hydroxyhippuric acidQuinine
Benzilic acid3-HydroxytyramineRanitidine
Benzoic acidIbuprofenSalicylic acid

{12}------------------------------------------------

BilirubinD,L-IsoproterenolSerotonin (5- Hydroxytyramine)
ChloralhydrateIsoxsuprineSulfamethazine
ChloramphenicolKetamineSulindac
ChlorothiazideKetoprofenTetrahydrocortisone, 3-acetate
ChlorpromazineLabetalolTetrahydrocortisone 3-(β-
Dglucuronide)
CholesterolLoperamideTetrahydrozoline
ClonidineMeperidineThiamine
CortisoneMeprobamateThioridazine
(-) CotinineMethoxyphenamineTriamterene
CreatinineNalidixic acidDL-Tyrosine
DeoxycorticosteroneNaloxoneTrifluoperazine
DextromethorphanNaltrexoneTrimethoprim
DiclofenacNaproxenD L-Tryptophan
DiflunisalNiacinamideTyramine
DigoxinNifedipineUric acid
DiphenhydramineNorethindroneVerapamil
DisopyramideNoscapineZomepirac

e. Specificity

To test specificity, drug metabolites and other structure related compounds that are likely to cross-react in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below for Amphetamine 500, Cocaine 150, Methamphetamine 500 and Morphine 300. Please refer to K182654 for Methylenedioxy-Methamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene, and to K181968 for Oxazepam, Secobarbital and Methadone, and to K153646 for Cannabinoids and to K201120 for EDDP. There were no differences observed between the SAFECARE® Cup and Dip Card formats. The candidate device uses the same assays (same reagent pads) for these drugs as was cleared in K182654/K181968/K153646/K201120.

CompoundsResultPositive at(ng/ml)% Cross-Reactivity
l-Amphetamine250002%
dl- Amphetamine150033%
3,4-Methylenedioxyamphetamine (MDA)250020%
Phentermine150033%
Hydroxyamphetamine80006%
β-Phenylethylamine1000000.5%
Tyramine1000000.5%
p-Hydroxynorephedrine1000000.5%
p-Hydroxyamphetamine1000000.5%
d/l-Norephedrine1000000.5%
d-Methamphetamine>100,000<0.5%
l-Methamphetamine>100,000<0.5%
Methylenedioxyethylamphetamine (MDE)>100,000<0.5%
Methylenedioxymethamphetamine(MDMA)>100,000<0.5%
Ephedrine>100,000<0.5%
Phenylpropanolamine>100,000<0.5%
(±)Phenylpropanolamine>100,000<0.5%
Benzphetamine>100,000<0.5%

Amphetamine

{13}------------------------------------------------

Cocaine
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Cocaine37540%
Cocaethylene62502.4%
Ecgonine160001%
Ecgonine methyl ester>100,000<0.2%
Norcocaine>100,000<0.2%
Methamphetamine
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
D-Amphetamine250002%
L- Amphetamine375001%
Chloroquine100005%
(+/-)-Ephedrine250002%
D/L-Methamphetamine500100%
L-Methamphetamine100005%
(+/-)3,4-Methylenedioxy-n-ethylamphetamine (MDEA)500100%
3,4-Methylenedioxyamphetamine(MDA)500100%
Methylenedioxymethamphetamine(MDMA)200025%
β-Phenylethylamine250002%
Trimethobenzamide500010%
d/l-Amphetamine750001%
p-Hydroxymethamphetamine150003%
Mephentermine250002%
(1R,2S)-(-)-Ephedrine500001%
1-Phenylephrine1000000.5%
(-)-Methamphetamine125002%
Morphine
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Normorphine300100%
s-Monoacetylmorphine300100%
Codeine300100%
Ethyl Morphine100300%
Heroin300100%
Hydrocodone50006%
Hydromorphone100030%
Morphinie-3-β-d-glucuronide100030%
Oxycodone>100000<0.3%
Oxymorphone1000000.3%
Thebaine300010%
Levorphanol100003%
6-Monoacetylmorphine (6-MAM)150200%
Norcodeine62505%
Procaine1500000.2%

{14}------------------------------------------------

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed between the SAFECARE® Cup and Dip Card formats.

2. Comparison Studies

Method comparison studies for the SAFECARE® Dip Card Tests and the SAFECARE® Cup Tests were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results. The results are presented in the tables below for Amphetamine 500, Cocaine 150, Methamphetamine 500 and Morphine 300. Please refer to K182654 for Methylenedioxy-Methamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene, and to K181968 for Oxazepam, Secobarbital and Methadone, and to K153646 for Cannabinoids and to K201120 for EDDP. The candidate device uses the same assays (same reagent pads) for these drugs as was cleared in K182654/ K181968/ K153646 and K201120.

DipCardNegativeLow Negative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0012020
Negative10101900
ViewerBPositive0001920
Negative10102010
ViewerCPositive0012020
Negative10101900

Amphetamine 500

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer ALM8554498Positive
Viewer CLM8554498Positive
Viewer BLM1169502Negative

{15}------------------------------------------------

CupNegativeLow Negative by LC/MS(less than -50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between the cutoff and +50%)High Positive by LC/MS(greater than +50%)
Viewer A Positive0001920
Viewer A Negative10102010
Viewer B Positive0012020
Viewer B Negative10101900
Viewer C Positive0001920
Viewer C Negative10102010
ViewerSample NumberLC/MS ResultCup Viewer Results
Viewer BLM8561498Positive
Viewer ALM1497502Negative
Viewer CLM1497502Negative

Cocaine 150

Dip CardNegativeLow Negative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0012020
Negative10101900
ViewerBPositive0001920
Negative10102010
ViewerCPositive0012020
Negative10101900
ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer ALM8739148Positive
Viewer CLM8739148Positive
Viewer BLM7119152Negative

{16}------------------------------------------------

CupNegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and +50%)High Positiveby LC/MS(greater than +50%)
ViewerPositive0001920
ANegative10102010
ViewerPositive0012020
BNegative10101900
ViewerPositive0002020
CNegative10102000
ViewerSample NumberLC/MS ResultCupViewer Results
Viewer BLM4418148Positive
Viewer ALM1485152Negative

Methamphetamine 500

DipCardNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
ViewerANegative10101920
ViewerBPositive0021920
ViewerBNegative10101810
ViewerCPositive0001820
ViewerCNegative10102020
ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer ALM6106467Positive
Viewer BLM7710464Positive
Viewer BLM6106467Positive
Viewer ALM7165513Negative
Viewer ALM6198511Negative
Viewer BLM7165513Negative
Viewer CLM7165513Negative
Viewer CLM6198511Negative

{17}------------------------------------------------

CupNegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and +50%)High Positiveby LC/MS(greater than +50%)
ViewerAPositive0021920
ANegative10101810
ViewerBPositive0011820
BNegative10101920
ViewerCPositive0011920
CNegative10101910
ViewerSample NumberLC/MS ResultCup Viewer Results
Viewer ALM3480464Positive
Viewer ALM8436467Positive
Viewer BLM3480464Positive
Viewer CLM8436467Positive
Viewer ALM5908513Negative
Viewer BLM8727511Negative
Viewer BLM5908513Negative
Viewer CLM5908513Negative

Morphine 300

Dip CardNegativeLow Negative by LC/MS(less than -50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between the cutoff and +50%)High Positive by LC/MS(greater than +50%)
Viewer APositive0011920
Viewer ANegative10101910
Viewer BPositive0012020
Viewer BNegative10101900
Viewer CPositive0021920
Viewer CNegative10101810
ViewerSample NumberLC/MS ResultDip Card Viewer Results
Viewer ALM4260280Positive
Viewer BLM4260280Positive
Viewer CLM4260280Positive
Viewer CLM9616281Positive

{18}------------------------------------------------

Viewer ALM2412305Negative
Viewer CLM4067306Negative
CupNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0012020
ANegative10101900
ViewerPositive0021920
BNegative10101810
ViewerPositive0001920
CNegative10102010
ViewerSample NumberLC/MS ResultCupViewer Results
Viewer ALM8592280Positive
Viewer BLM8592280Positive
Viewer BLM1718281Positive
Viewer BLM7813305Negative
Viewer CLM1916306Negative

Lay-user study:

A lay user study was performed at three intended user sites with 310 lay persons for each device format. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested. Typical results are shown below.

AMP500:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS (ng/mL)Lay person ResultsThe percentage
No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff20125020100
-50% Cutoff1702500170100
-25% Cutoff2037521890
+25% Cutoff2062519195
+50% Cutoff40750400100
+75% Cutoff20878200100

{19}------------------------------------------------

COC150:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correctresults (%)
-100% Cutoff200No. ofPositiveNo. ofNegative100
-75% Cutoff2038020100
-50% Cutoff170750170100
-25% Cutoff2011321890
+25% Cutoff2018819195
+50% Cutoff40225400100
+75% Cutoff20263200100

THC:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2013020100
-50% Cutoff170250170100
-25% Cutoff203711995
+25% Cutoff206219195
+50% Cutoff4076400100
+75% Cutoff2087200100

BAR:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2074020100
-50% Cutoff1701510170100
-25% Cutoff2022621890
+25% Cutoff2037318290
+50% Cutoff40450400100
+75% Cutoff20526200100

{20}------------------------------------------------

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentage
No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff1701510170100
-25% Cutoff2022421890
+25% Cutoff2037519195
+50% Cutoff40452400100
+75% Cutoff20524200100

BZO:

MET500:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff200No. ofPositive 0No. ofNegative 20100
-75% Cutoff20124No. ofPositive 0No. ofNegative 20100
-50% Cutoff170251No. ofPositive 0No. ofNegative 170100
-25% Cutoff20375No. ofPositive 2No. ofNegative 1890
+25% Cutoff20625No. ofPositive 19No. ofNegative 195
+50% Cutoff40750No. ofPositive 40No. ofNegative 0100
+75% Cutoff20877No. ofPositive 20No. ofNegative 0100

MTD:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2074020100
-50% Cutoff1701520170100
-25% Cutoff2022521890
+25% Cutoff2037618290
+50% Cutoff40451400100
+75% Cutoff20524200100

{21}------------------------------------------------

MOP300:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentage
No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff1701510170100
-25% Cutoff2022511995
+25% Cutoff2037618290
+50% Cutoff40452400100
+75% Cutoff20525200100

MDMA:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff200020100
-75% Cutoff20124020100
-50% Cutoff1702500170100
-25% Cutoff2037721890
+25% Cutoff2062218290
+50% Cutoff40749400100
+75% Cutoff20874200100

OXY:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2025020100
-50% Cutoff170500170100
-25% Cutoff207511995
+25% Cutoff2012418290
+50% Cutoff40151400100
+75% Cutoff20175200100

{22}------------------------------------------------

BUP:

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff203020100
-50% Cutoff17050170100
-25% Cutoff20821890
+25% Cutoff201318290
+50% Cutoff4015400100
+75% Cutoff2018200100

PCP:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff206020100
-50% Cutoff170130170100
-25% Cutoff201921890
+25% Cutoff203118290
+50% Cutoff4038400100
+75% Cutoff2044200100

TCA:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20252020100
-50% Cutoff1705010170100
-25% Cutoff2075211995
+25% Cutoff20125519195
+50% Cutoff401503400100
+75% Cutoff201751200100

{23}------------------------------------------------

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentage
No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff1701500170100
-25% Cutoff2022521890
+25% Cutoff2037418290
+50% Cutoff40452400100
+75% Cutoff20523200100

PPX:

EDDP:

% of CutoffNumber of samplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff1701500170100
-25% Cutoff2022411995
+25% Cutoff2037418290
+50% Cutoff40450400100
+75% Cutoff20524200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are substantially equivalent to the predicate.

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).