K Number
K182654

Validate with FDA (Live)

Date Cleared
2018-11-19

(55 days)

Regulation Number
862.3610
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamine, Phencyclidine, Methadone, Nortriotyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL
Morphine2000 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxymethamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine. Oxazepam, Secobarbital. Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunichromatographic assays for qualitative and simultaneous detection of Amphetanine, Oxazepam, Methamphetamine, Morghine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine, Phencyclidine, Methadone, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL
Morphine2000 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxymethamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Cup contination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine. Oxazenam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

Device Description

The SAFECARE Dip Card Tests and SAFECARE Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

The document describes the performance characteristics and studies for the SAFECARE Multi-Drug Urine Test Dip Card and SAFECARE Multi-Drug Urine Test Cup.

Here's an analysis of the acceptance criteria and study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a separate section. However, the performance data presented implies a very high standard for accuracy, particularly around the established cutoff concentrations. The "precision" studies demonstrate the device's ability to consistently produce correct results for samples at varying concentrations relative to the cutoff. The "comparison studies" compare the device's qualitative results to a quantitative gold standard (LC/MS) to establish concordance. The "lay-user study" assesses the ability of untrained individuals to correctly use and interpret the device.

Based on the provided data, the implied acceptance criteria for qualitative immunoassay devices for drug testing would typically be:

  • 100% agreement for samples significantly below the cutoff (Negative).
  • 100% agreement for samples significantly above the cutoff (Positive).
  • High percentage agreement for samples near the cutoff (typically >90%).

For the Precision Study (Analytical Performance - "Test Set" Data):

The table below summarizes the reported performance for Methylenedioxymethamphetamine (MDMA), Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, and Propoxyphene (results from 3 lots, 50 tests per concentration per lot, per device type). The format is (Number of Negative Results)/(Number of Positive Results).

Drug (Identifier)Cut-off LevelConcentrationReported Performance (Typical across 3 lots)Implied Acceptance Criteria Met?
Methylenedioxymethamphetamine (MDMA) (Dip Card & Cup)500 ng/mL-100% Cut-off50-/0+Yes (100% negative)
-75% Cut-off50-/0+Yes (100% negative)
-50% Cut-off50-/0+Yes (100% negative)
-25% Cut-off50-/0+Yes (100% negative)
Cut-off25-/25+ or 26-/24+Yes (near 50/50 split at cutoff)
+25% Cut-off50+/0-Yes (100% positive)
+50% Cut-off50+/0-Yes (100% positive)
+75% Cut-off50+/0-Yes (100% positive)
+100% Cut-off50+/0-Yes (100% positive)
Oxycodone (Dip Card & Cup)100 ng/mL-100% Cut-off50-/0+Yes (100% negative)
-75% Cut-off50-/0+Yes (100% negative)
-50% Cut-off50-/0+Yes (100% negative)
-25% Cut-off50-/0+Yes (100% negative)
Cut-off24-/26+ or 26-/24+ or 25-/25+Yes (near 50/50 split at cutoff)
+25% Cut-off50+/0-Yes (100% positive)
+50% Cut-off50+/0-Yes (100% positive)
+75% Cut-off50+/0-Yes (100% positive)
+100% Cut-off50+/0-Yes (100% positive)
Buprenorphine (Dip Card & Cup)10 ng/mL-100% Cut-off50-/0+Yes (100% negative)
-75% Cut-off50-/0+Yes (100% negative)
-50% Cut-off50-/0+Yes (100% negative)
-25% Cut-off50-/0+Yes (100% negative)
Cut-off27-/23+, 25-/25+, 24-/26+, 26-/24+Yes (near 50/50 split at cutoff)
+25% Cut-off50+/0-Yes (100% positive)
+50% Cut-off50+/0-Yes (100% positive)
+75% Cut-off50+/0-Yes (100% positive)
+100% Cut-off50+/0-Yes (100% positive)
Phencyclidine (Dip Card & Cup)25 ng/mL-100% Cut-off50-/0+Yes (100% negative)
-75% Cut-off50-/0+Yes (100% negative)
-50% Cut-off50-/0+Yes (100% negative)
-25% Cut-off50-/0+Yes (100% negative)
Cut-off26-/24+, 25-/25+, 24-/26+Yes (near 50/50 split at cutoff)
+25% Cut-off50+/0-Yes (100% positive)
+50% Cut-off50+/0-Yes (100% positive)
+75% Cut-off50+/0-Yes (100% positive)
+100% Cut-off50+/0-Yes (100% positive)
Nortriptyline (Dip Card & Cup)1000 ng/mL-100% Cut-off50-/0+Yes (100% negative)
-75% Cut-off50-/0+Yes (100% negative)
-50% Cut-off50-/0+Yes (100% negative)
-25% Cut-off50-/0+Yes (100% negative)
Cut-off25-/25+, 26-/24+, 24-/26+Yes (near 50/50 split at cutoff)
+25% Cut-off50+/0-Yes (100% positive)
+50% Cut-off50+/0-Yes (100% positive)
+75% Cut-off50+/0-Yes (100% positive)
+100% Cut-off50+/0-Yes (100% positive)
Propoxyphene (Dip Card & Cup)300 ng/mL-100% Cut-off50-/0+Yes (100% negative)
-75% Cut-off50-/0+Yes (100% negative)
-50% Cut-off50-/0+Yes (100% negative)
-25% Cut-off50-/0+Yes (100% negative)
Cut-off23-/27+, 26-/24+, 24-/26+, 25-/25+Yes (near 50/50 split at cutoff)
+25% Cut-off50+/0-Yes (100% positive)
+50% Cut-off50+/0-Yes (100% positive)
+75% Cut-off50+/0-Yes (100% positive)
+100% Cut-off50+/0-Yes (100% positive)

For the Lay-user Study:

The acceptance criteria for the lay-user study would typically be a high percentage of correct results, especially for samples clearly negative or positive, and a reasonable proportion around the cutoff. The document shows the "percentage of correct results (%)" for each concentration level. The implied acceptance criteria for OTC urine drug screens are usually very high concordance (e.g., >90%) with the ground truth for samples away from the cutoff, and some level of agreement for near-cutoff samples (where variability is expected). The device performance meets these generally implied high standards, especially for samples clearly below or above the cutoff (100% for most categories). For samples near +/-25% of cutoff, the accuracy ranges from 85-95%, which is typical for qualitative tests and acceptable for over-the-counter use where confirmation is recommended for positive results.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Studies (Test Set):
    • For each drug, device type (Dip Card/Cup), and each of the 9 concentration levels: 50 tests were performed per concentration.
    • This was repeated for 3 different lots for each device type.
    • Therefore, for a single drug and device type, the test set size for precision was: 9 concentrations * 50 tests/concentration * 3 lots = 1350 tests.
    • Data Provenance: The samples were prepared by "spiking drug in negative samples" and confirmed by LC/MS. This suggests controlled, prospective laboratory-prepared samples rather than retrospective clinical samples from specific countries. The testing was "in-house."
  • Comparison Studies (Test Set):
    • For each drug and device type (Dip Card/Cup): 80 unaltered clinical samples (40 negative and 40 positive) were used.
    • Data Provenance: "unaltered clinical samples." The document does not specify the country of origin, but given the manufacturer (China) and submission to US FDA, the samples could be from various global sources or acquired for the study. It's a retrospective analysis against independently confirmed samples.
  • Lay-user Study (Test Set):
    • For each device format (presumably Dip Card and Cup, though only combined results are shown in the given table): 300 lay persons participated.
    • The total number of samples processed for each drug across the different concentration levels in the lay-user study tables is: 20 + 20 + 160 + 20 + 20 + 40 + 20 = 300 samples. (It seems 300 samples were tested for each drug type, probably with each of the 300 lay users testing one sample)
    • Data Provenance: Urine samples were "prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens." These were laboratory-prepared samples. The study involved "three intended user sites."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Precision Studies: Ground truth was established by LC/MS (Liquid Chromatography-Mass Spectrometry), which is a definitive quantitative analytical method. It's an instrumental method, so human "experts" in interpretation are typically not specified in the same way as for imaging or clinical diagnosis. The samples were "blindly labeled by the person who prepared the samples," which indicates blinding to the LC/MS results during the device testing.
  • Comparison Studies: Ground truth was established by LC/MS results. The document states, "The samples were blind labeled and compared to LC/MS results." These are definitive chemical analysis results.
  • Lay-user Study: Ground truth for the samples was established by LC/MS confirmation of the spiked drug concentrations.

4. Adjudication Method for the Test Set

  • Precision Studies: No explicit human adjudication method mentioned, as the device results are quantitative categories (positive/negative) and compared to definitive LC/MS values. The results are reported as counts (e.g., 25-/25+).
  • Comparison Studies: The device results were read by "three laboratory assistants." For discordant results, the individual viewer's result is listed against the LC/MS truth, but no specific human adjudication process (e.g., 2+1 majority voting) is described for resolving discrepancies among the assistants or between the assistants and LC/MS. The LC/MS is the definitive ground truth here.
  • Lay-user Study: The results are presented as aggregated counts of positive/negative readings by the lay users compared to the known ground truth (LC/MS confirmed concentrations). No human adjudication process is mentioned among the lay users.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic devices where human readers interpret images or complex data, and the AI serves as an aid.
  • This device is a qualitative, over-the-counter urine drug test. Its performance is evaluated against chemical analytical ground truth, and its usability is assessed with lay users, not by evaluating how it assists expert readers.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

  • This device is a physical, qualitative immunoassay test kit, not a standalone algorithm/AI for analysis. The "performance" of the device itself (the reaction on the dip card/cup) is evaluated.
  • However, the Comparison Studies and Precision Studies could be considered the "standalone" performance of the device when read by trained laboratory personnel (3 laboratory assistants), as it's the device's output itself being compared to LC/MS.
  • The Lay-user study evaluates the device performance when interpreted by the intended "human-in-the-loop" (a lay user).

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

  • The primary ground truth used for all performance studies (Precision, Comparison, Lay-user) is quantitative chemical analysis by LC/MS (Liquid Chromatography-Mass Spectrometry). This is a highly accurate and widely accepted method for confirming drug concentrations in urine.
  • For the precision and lay-user studies, urine samples were spiked with known concentrations of drugs, and these concentrations were confirmed by LC/MS.
  • For the comparison studies, "unaltered clinical samples" were used, with LC/MS providing the definitive truth for the presence and concentration of drugs.

8. The Sample Size for the Training Set

  • This document describes premarket notification (510(k)) for a drug test kit, which is a physical diagnostic device based on immunochromatography, not an AI/Machine Learning algorithm.
  • Therefore, there is no "training set" in the context of machine learning model development. The device's performance relies on its physical and chemical design, not on a trained algorithm.

9. How the Ground Truth for the Training Set Was Established

  • As there is no AI/ML algorithm or training set, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the text "ADMINISTRATION".

November 19, 2018

Safecare Biotech (Hangzhou) Co., Ltd. % Joe Shia Regulatory Consultant LSI International Inc. 504 E Diamond Ave., Suite I Gaithersburg, Maryland 20877

Re: K182654

Trade/Device Name: SAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip Card Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: Class II Product Code: NGG, NGL, NGM, QAW, QBF, NFT, NFW, NFY, NGI, NFV, PTH, PTG Dated: September 21, 2018 Received: September 25, 2018

Dear Joe Shia:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name

SAFECARE® Multi-Drug Urine Test Dip Card SAFECARE® Multi-Drug Urine Test Cup

Indications for Use (Describe)

SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamine, Phencyclidine, Methadone, Nortriotyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL
Morphine2000 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxymethamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine. Oxazepam, Secobarbital. Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunichromatographic assays for qualitative and simultaneous detection of Amphetanine, Oxazepam, Methamphetamine, Morghine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxymethamphetamine, Phencyclidine, Methadone, Nortriptyline and d-Propoxyphene in human urine at the cutoff concentrations of:

Drug(Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Marijuana50 ng/mL
Methamphetamine1000 ng/mL

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020

See PRA Statement below.

{3}------------------------------------------------

Morphine2000 ng/mL
Oxycodone100 ng/mL
Secobarbital300 ng/mL
Buprenorphine10 ng/mL
Methylenedioxymethamphetamine500 ng/mL
Phencyclidine25 ng/mL
Methadone300 ng/mL
Nortriptyline1000 ng/mL
d-Propoxyphene300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Cup contination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine. Oxazenam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GCMS or LC/MS is the preferred confirmatory method.

The tests are intended for over-the-counter use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

  • November 1, 2018 1. Date:
  • Safecare Biotech (Hangzhou) Co. Ltd. 2. Submitter: 18 Haishu Road, Yuhang District Hangzhou, China
  • Alex Qiu 3. Contact person: Safecare Biotech 18 Haishu Road, Yuhang District Hangzhou, China Telephone: 86 571-89712897 Fax: 86 571-80389223 Email: alexqiu@safecare.com.cn
  • SAFECARE Multi-Drug Urine Test Dip Card 4. Device Name: SAFECARE Multi-Drug Urine Test Cup
Product CodeClassificationRegulation SectionPanel
NFTAmphetamineII21 CFR § 862.3100, AmphetamineTest SystemToxicology (91)
NFWCannabinoidsII21 CFR § 862.3870, CannabinoidsTest SystemToxicology (91)
NFYCocaineII21 CFR § 862.3250, Cocaine andCocaine Metabolites Test SystemToxicology (91)
NGGMethamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology (91)
NGIMorphineII21 CFR § 862.3640, MorphineTest SystemToxicology (91)
NFVOxazepamII21 CFR § 862.3170,Benzodiazepine Test SystemToxicology (91)
NGLOxycodoneII21 CFR § 862.3650, Opiate TestSystemToxicology (91)
PTHSecobarbitalII21 CFR § 862.3150, BarbiturateTest SystemToxicology (91)
NGLBuprenorphineII21 CFR § 862.3650,Opiate Test SystemToxicology (91)
NGGMethylenedioxy-methamphetamineII21 CFR § 862.3610,Methamphetamine Test SystemToxicology (91)
NGMPhencyclidineunclassifiedEnzyme ImmunoassayPhencyclidineToxicology (91)
PTGMethadoneII21 CFR § 862.3620, MethadoneTest System
QAWNortriptylineII21 CFR, 862.3910 TricyclicAntidepressant Drugs Test SystemToxicology (91)
QBFPropoxypheneII21 CFR, 862.3700 PropoxypheneTest SystemToxicology (91)
  1. Predicate Devices: K142396

{5}------------------------------------------------

    1. Intended Use
      SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene in human urine at the cutoff concentrations of:
Drug (Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Cannabinoids50 ng/mL
Methamphetamine1000 ng/mL
Morphine2000 ng/mL
Secobarbital300 ng/mL
Methadone300 ng/mL
Methylenedioxymethamphetamine500 ng/mL
Oxycodone100 ng/mL
Buprenorphine10 ng/mL
Phencyclidine25 ng/mL
Nortriptyline1000 ng/mL
Propoxyphene300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene in human urine at the cutoff concentrations of:

Drug (Identifier)Cut-off level
Amphetamine1000 ng/mL
Oxazepam300 ng/mL
Cocaine300 ng/mL
Cannabinoids50 ng/mL
Methamphetamine1000 ng/mL
Morphine2000 ng/mL
Secobarbital300 ng/mL

{6}------------------------------------------------

Methadone300 ng/mL
Methylenedioxymethamphetamine500 ng/mL
Oxycodone100 ng/mL
Buprenorphine10 ng/mL
Phencyclidine25 ng/mL
Nortriptyline1000 ng/mL
Propoxyphene300 ng/mL

Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.

    1. Device Description
      The SAFECARE Dip Card Tests and SAFECARE Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone,

Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.

    1. Substantial Equivalence Information
      A summary comparison of features of the SAFECARE Dip Card Tests and SAFECARE Cup Tests and the predicate devices is provided in following tables.
ItemDevicePredicate - K142396
Indication(s)for UseFor the qualitative determination of drugs ofabuse in human urine.Same (but the number ofdrugs detected isdifferent)
Calibrator andCut-Off ValuesAmphetamine (AMP): 1,000 ng/mlOxazepam (BZO):300 ng/mlCocaine (COC): 300 ng/ml11-Nor-Δ9-Tetrahydrocannabinol-9-COOH(THC):50 ng/mlMethamphetamine (MET): 1,000 ng/mlMorphine (OPI): 2000ng/mlSecobarbital (BAR): 300 ng/mlMethadone (MTD): 300 ng/mlSame

Table 1: Features Comparison of SAFECARE Dip Card Tests and the Predicate Devices

{7}------------------------------------------------

Methylenedioxymethamphetamine(MDMA): 500 ng/ml
Oxycodone (OXY): 100 ng/ml
Buprenorphine (BUP): 10 ng/ml
Phencyclidine (PCP): 25 ng/ml
Nortriptyline (TCA): 1000 ng/ml
Propoxyphene (PPX): 300 ng/ml
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame
Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsDip CardDip Card
Table 2: Features Comparison of SAFECARE Cup Tests and the Predicate Devices
ItemDevicePredicate - K142396
Indication(s)for UseFor the qualitative determination ofdrugs of abuse in human urine.Same (but the number ofdrugs detected is different)
Calibrator andCut-Off ValuesAmphetamine (AMP): 1,000 ng/mlOxazepam (BZO):300 ng/mlCocaine (COC): 300 ng/ml11-Nor-Δ9-Tetrahydrocannabinol-9-COOH(THC):50 ng/mlMethamphetamine (MET): 1,000 ng/mlMorphine (OPI): 2000ng/mLSecobarbital (BAR): 300 ng/mlMethadone (MTD): 300 ng/mlMethylenedioxymethamphetamine(MDMA): 500 ng/mlOxycodone (OXY): 100 ng/mlBuprenorphine (BUP): 10 ng/mlPhencyclidine (PCP): 25 ng/mlNortriptyline (TCA): 1000 ng/mlPropoxyphene (PPX): 300 ng/mlSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Type of TestQualitativeSame

{8}------------------------------------------------

Specimen TypeHuman UrineSame
Intended UseFor over-the-counterSame
ConfigurationsCupDip Card

9. Test Principle

The SAFECARE Dip Card Tests, and SAFECARE Cup Tests are rapid tests for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenediox ymethamphetamine, Ox ycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propox yphene in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut off. -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables for Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine. Nortriptyline and Propoxyphene. Please refer to K181968 for precision data for Oxazepam, Methamphetamine, Morphine, Secobarbital and Methadone, and to K153646 for precision data for Amphetamine, Cocaine, and Cannabinoids.

ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Methylenedioxymethamphetamine (MDMA)

Dip Card

{9}------------------------------------------------

Cup
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Oxycodone

Dip Card

Dip Card Results
LotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Cup
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

Buprenorphine

Dip Card

Dip Card
LotNumberResults-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Cup
ResultsLot-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Number
Lot 150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

{10}------------------------------------------------

Phencyclidine

Dip Card

Dip Card
LotNumberResults-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Cup

Results-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

Nortriptyline

Dip Card

Dip Card
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
r)
--------
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Propoxyphene

Dip Card

Dip Card

Results-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
LotNumber
Lot 150-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

{11}------------------------------------------------

Cup
ResultsLotNumber-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%
Lot 150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

The following cut-off values are verified.

Drug (Identifier)Cut-off level
Methylenedioxy-Methamphetamine (MDMA)500 ng/mL
Oxycodone (OXY)100 ng/mL
Buprenorphine (BUP)10 ng/mL
Phencyclidine (PCP)25 ng/mL
Nortriptyline (TCA)1000 ng/mL
Propoxyphene (PPX)300 ng/mL

b. Linearity

Not applicable.

  • c. Stability
    The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 50 °C and real time stability studies at 4°C and 30 °C.

d. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration of 100ug/mL are summarized in the following tables. There were no differences observed between the SAFECARE Cup and Dip Card formats.

Acetominophen (4-Acetamidophenol)Ecgonine methyl esterD,L-Octopamine
AcetophenetidinEMDPOxalic acid
N-AcetylprocainamideErythromycinOxolinic acid
Acetylsalicylic acidβ-EstradiolOxymetazoline
AlbuminFenoprofenPapaverine
AminopyrineFurosemidePenicillin-G
AmoxicillinGentisic acidPerphenazine
AmpicillinHemoglobinPhenelzine
ApomorphineHydralazinePrednisone
Ascorbic acidHydrochlorothiazideDL-Propranolol
AspartameHydrocortisoneD-Pseudoephedrine
AtropineO-Hydroxyhippuric acidQuinine

{12}------------------------------------------------

Benzilic acid3-HydroxytyramineRanitidine
Benzoic acidIbuprofenSalicylic acid
BilirubinD,L-IsoproterenolSerotonin (5- Hydroxytyramine)
ChloralhydrateIsoxsuprineSulfamethazine
ChloramphenicolKetamineSulindac
ChlorothiazideKetoprofenTetrahydrocortisone, 3-acetate
ChlorpromazineLabetalolTetrahydrocortisone 3-(β-Dglucuronide)
CholesterolLoperamideTetrahydrozoline
ClonidineMaprotilineThiamine
CortisoneMeperidineThioridazine
(-) CotinineMeprobamateTriamterene
CreatinineMethoxyphenamineDL-Tyrosine
DeoxycorticosteroneNalidixic acidTrifluoperazine
DextromethorphanNaloxoneTrimethoprim
DiclofenacNaltrexoneD L-Tryptophan
DiflunisalNaproxenTyramine
DigoxinNiacinamideUric acid
DiphenhydramineNifedipineVerapamil
DisopyramideNorethindroneZomepirac
EDDPNoscapine

e. Specificity

To test specificity, drug metabolites and other structure related compounds that are likely to cross-react in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below for Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene. Please refer to K181968 for specificity data for Oxazepam, Methamphetamine, Morphine, Secobarbital and Methadone, and to K153646 for specificity data for Amphetamine, Cocaine, and Cannabinoids. There were no differences observed between the SAFECARE Cup and Dip Card formats.

CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Methylenedioxymethamphetamine (MDMA)500100%
3,4-Methylenedioxyamphetamine (MDA)500010%
3,4-Methylenedioxyethylamphetamine (MDEA)300166%
d-methamphetamine>100000<1%
d-amphetamine>100000<1%
l-amphetamine>100000<1%
l-methamphetamine>100000<1%
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Oxycodone100100%
Oxymorphone25040%
Hydrocodone31253.2 %
Dihydrocodeine>100000<0.1%
Codeine>100000<0.1%
Hydromorphone>100000<0.1%
Morphine>100000<0.1%
Buprenorphine>100000<0.1%
Ethylmorphine>100000<0.1%

{13}------------------------------------------------

CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Buprenorphine10100%
Buprenorphine -3-D-Glucuronide10100%
Norbuprenorphine5020%
Norbuprenorphine-3-D-Glucuronide10010%
Morphine>100000<0.01%
Oxymorphone>100000<0.01%
Hydromorphone>100000<0.01%
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Phencyclidine25100%
4-Hydroxyphencyclidine7533%
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
Nortriptyline1000100%
Amitriptyline750133.30%
Clomipramine1000010%
Desipramine200500%
Doxepin125080%
Imipramine625160%
Maprotiline200050%
Nordoxepin1000100%
Promazine150066.70%
Promethazine250004%
Trimipramine300033.30%
Cyclobenzaprine Hydrochloride500020%
Norclomipramine300033.30%
CompoundsResultPositive at(ng/ml)% Cross-Reactivity
d-Propoxyphene300100%
D-Norpropoxyphene33091%

f. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed between the SAFECARE Cup and Dip Card formats.

2. Comparison Studies

Method comparison studies for the SAFECARE Dip Card Tests and the SAFECARE Cup Tests were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results. The results are presented in the tables below for Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene. Please refer to K181968 for method comparison data for Oxazepam, Methamphetamine, Morphine, Secobarbital and Methadone, and to K153646 for method

{14}------------------------------------------------

comparison data for Amphetamine, Cocaine, and Cannabinoids.

DipCardNegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between the cutoff and+50%)High Positiveby LC/MS(greater than +50%)
ViewerAPositive0011820
Negative10101920
ViewerBPositive0011820
Negative10101920
ViewerCPositive0021820
Negative10101820

Methylenedioxymethamphetamine (MDMA)

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AAM1822491Positive
Viewer BAM5121497Positive
Viewer CAM1822491Positive
Viewer CAM5121497Positive
Viewer AAM2281517Negative
Viewer AAM3339509Negative
Viewer BAM2281517Negative
Viewer BAM6217525Negative
Viewer CAM6217525Negative
Viewer CAM3339509Negative
CupNegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and +50%)High Positive byLC/MS(greater than +50%)
ViewerAPositive0021820
ViewerANegative10101820
ViewerBPositive0011820
ViewerBNegative10101920
ViewerCPositive0021820
ViewerCNegative10101820

{15}------------------------------------------------

Discordant Results
ViewerSample NumberLC/MS ResultCupViewer Results
Viewer AAM5135491Positive
Viewer AAM6745497Positive
Viewer BAM5135491Positive
Viewer CAM5135491Positive
Viewer CAM6745497Positive
Viewer AAM8048517Negative
Viewer AAM4472509Negative
Viewer BAM4472509Negative
Viewer BAM4835525Negative
Viewer CAM8048517Negative
Viewer CAM4472509Negative

Oxycodone

DipCardNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
Negative10101920
ViewerBPositive0011820
Negative10101920
ViewerCPositive0011720
Negative10101930

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AAM859295Positive
Viewer BAM386798Positive
Viewer CAM749592Positive
Viewer AAM1177104Negative
Viewer AAM3128107Negative
Viewer BAM9817106Negative
Viewer BAM9785109Negative
Viewer CAM0928108Negative
Viewer CAM1177104Negative
Viewer CAM3128107Negative

{16}------------------------------------------------

CupNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
ViewerANegative10101920
ViewerBPositive0011720
ViewerBNegative10101930
ViewerCPositive0011720
ViewerCNegative10101930

Discordant Results

ViewerSample NumberLC/MS ResultCupViewer Results
Viewer AAM062398Positive
Viewer BAM535795Positive
Viewer CAM062398Positive
Viewer AAM7016105Negative
Viewer AAM3335106Negative
Viewer BAM7525104Negative
Viewer BAM0654107Negative
Viewer BAM3335106Negative
Viewer CAM7016105Negative
Viewer CAM7525104Negative
Viewer CAM3335106Negative

Buprenorphine

DipCardNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
Negative10101920
ViewerBPositive0011820
Negative10101920
ViewerCPositive0011920
Negative10101910

{17}------------------------------------------------

Discordant Results
ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AAM74939Positive
Viewer BAM10939Positive
Viewer CAM10939Positive
Viewer AAM693611Negative
Viewer AAM090211Negative
Viewer BAM649711Negative
Viewer BAM090211Negative
Viewer CAM693611Negative
CupNegativeLow Negative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
Negative10101920
ViewerBPositive0011820
Negative10101920
ViewerCPositive0011820
Negative10101920

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AAM83389Positive
Viewer BAM41309Positive
Viewer CAM41309Positive
Viewer AAM305911Negative
Viewer AAM560511Negative
Viewer BAM560511Negative
Viewer BAM276311Negative
Viewer CAM305911Negative
Viewer CAM276311Negative

{18}------------------------------------------------

Phencyclidine

Dip CardNegativeLow Negative by LC/MS (less than -50%)Near Cutoff Negative by LC/MS (Between -50% and cutoff)Near Cutoff Positive by LC/MS (Between the cutoff and +50%)High Positive by LC/MS (greater than +50%)
Viewer APositive0011920
Viewer ANegative10101910
Viewer BPositive0011820
Viewer BNegative10101920
Viewer CPositive0011920
Viewer CNegative10101910

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AAM155323Positive
Viewer BAM115924Positive
Viewer CAM115924Positive
Viewer AAM298225.3Negative
Viewer BAM590126Negative
Viewer BAM298225.3Negative
Viewer CAM298225.3Negative
CupNegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between thecutoff and +50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
ViewerANegative10101920
ViewerBPositive0011920
ViewerBNegative10101910
ViewerCPositive0011820
ViewerCNegative10101920

{19}------------------------------------------------

Discordant Results
ViewerSample NumberLC/MS ResultCup Viewer Results
Viewer AAM746724Positive
Viewer BAM663723Positive
Viewer CAM746724Positive
Viewer AAM144926Negative
Viewer AAM872725.3Negative
Viewer BAM872725.3Negative
Viewer CAM144926Negative
Viewer CAM872725.3Negative

Nortriptyline

DipCardNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerAPositive0011820
Negative10101920
ViewerBPositive0011820
Negative10101920
ViewerCPositive0011820
Negative10101920

Discordant Results

ViewerSample NumberLC/MS ResultDip CardViewer Results
Viewer AAM9403982Positive
Viewer BAM2316998Positive
Viewer CAM9403982Positive
Viewer AAM78121004Negative
Viewer AAM01701031Negative
Viewer BAM43731023Negative
Viewer BAM01701031Negative
Viewer CAM78121004Negative
Viewer CAM43731023Negative

{20}------------------------------------------------

CupNegativeLow Negative by LC/MS(less than -50%)Near Cutoff Negative by LC/MS(Between -50% and cutoff)Near Cutoff Positive by LC/MS(Between the cutoff and +50%)High Positive by LC/MS(greater than +50%)
ViewerAPositive0011820
ViewerANegative10101920
ViewerBPositive0011820
ViewerBNegative10101920
ViewerCPositive0011820
ViewerCNegative10101920

Discordant Results

ViewerSample NumberLC/MS ResultCup Viewer Results
Viewer AAM0528998Positive
Viewer BAM0528998Positive
Viewer CAM0528998Positive
Viewer AAM60841004Negative
Viewer AAM61571031Negative
Viewer BAM60841004Negative
Viewer BAM98691023Negative
Viewer CAM98691023Negative
Viewer CAM61571031Negative

Propoxyphene

DipCardNegativeLow Negative byLC/MS(less than -50%)Near Cutoff Negative byLC/MS(Between -50% andcutoff)Near Cutoff Positive byLC/MS(Between the cutoff and+50%)High Positiveby LC/MS(greater than +50%)
ViewerAPositive0011820
Negative10101920
ViewerBPositive0011820
Negative10101920
ViewerCPositive0001720
Negative10102030

{21}------------------------------------------------

Discordant Results
ViewerSample NumberLC/MS ResultDip Card Viewer Results
Viewer AAM9748294Positive
Viewer BAM8770290Positive
Viewer AAM6660315Negative
Viewer AAM8794306Negative
Viewer BAM6660315Negative
Viewer BAM2236309Negative
Viewer CAM6660315Negative
Viewer CAM8794306Negative
Viewer CAM2236309Negative
CupNegativeLowNegative byLC/MS(less than-50%)Near CutoffNegative byLC/MS(Between-50% andcutoff)Near CutoffPositive byLC/MS(Between thecutoff and+50%)High Positiveby LC/MS(greater than+50%)
ViewerPositive0011820
ANegative10101920
ViewerPositive0011720
BNegative10101930
ViewerPositive0011820
CNegative10101920

Discordant Results

ViewerSample NumberLC/MS ResultCup Viewer Results
Viewer AAM7095290Positive
Viewer BAM1226294Positive
Viewer CAM7095290Positive
Viewer AAM9825306Negative
Viewer AAM6247309Negative
Viewer BAM8797315Negative
Viewer BAM9825306Negative
Viewer BAM6247309Negative
Viewer CAM8797315Negative
Viewer CAM6247309Negative

Lay-user study

A lay user study was performed at three intended user sites with 300 lay persons for each device format. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert,

{22}------------------------------------------------

one blind labeled sample and a device. Each device was tested. Representative results are shown below.

AMP:

Number ofsamplesDrugConcentration byLC/MS (ng/mL)Lay person ResultsThe percentage
% of CutoffNo. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff20248020100
-50% Cutoff1605050160100
-25% Cutoff2075421890
+25% Cutoff20125918290
+50% Cutoff401508400100
+75% Cutoff201759200100

COC:

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2074020100
-50% Cutoff1601520160100
-25% Cutoff2022411995
+25% Cutoff2037318290
+50% Cutoff40452400100
+75% Cutoff20522200100

THC:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentage
No. ofPositiveNo. ofNegativeof correct results(%)
-100% Cutoff200020100
-75% Cutoff2013020100
-50% Cutoff160250160100
-25% Cutoff203721890
+25% Cutoff206219195
+50% Cutoff4075400100
+75% Cutoff2087200100

{23}------------------------------------------------

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2076020100
-50% Cutoff1601510160100
-25% Cutoff2022631785
+25% Cutoff2037219195
+50% Cutoff40448400100
+75% Cutoff20529200100

BAR:

BZO:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff1601490160100
-25% Cutoff2022411995
+25% Cutoff2037618290
+50% Cutoff40454400100
+75% Cutoff20520200100
% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20251020100
-50% Cutoff1605020160100
-25% Cutoff2075831785
+25% Cutoff20125919195
+50% Cutoff401512400100
+75% Cutoff201759200100

MTD:

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2074020100
-50% Cutoff1601510160100
-25% Cutoff2022521890
+25% Cutoff2037619195
+50% Cutoff40447400100
+75% Cutoff20522200100

{24}------------------------------------------------

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentageof correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff20505020100
-50% Cutoff16010090160100
-25% Cutoff20151031785
+25% Cutoff20251119195
+50% Cutoff403031400100
+75% Cutoff203521200100

OPI:

MDMA:

% of CutoffNumber of samplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
-100% Cutoff200020100
-75% Cutoff20123020100
-50% Cutoff1602480160100
-25% Cutoff2037621890
+25% Cutoff2062217385
+50% Cutoff40741400100
+75% Cutoff20871200100

OXY:

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff2025020100
-50% Cutoff160490160100
-25% Cutoff207511995
+25% Cutoff2012419195
+50% Cutoff40149400100
+75% Cutoff20175200100

BUP:

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff203020100
-50% Cutoff16050160100
-25% Cutoff20821890
+25% Cutoff201319195
+50% Cutoff4015400100
+75% Cutoff2018200100

{25}------------------------------------------------

% of CutoffNumber of samplesDrug Concentration by LC/MS(ng/mL)Lay person ResultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100
-75% Cutoff206020100
-50% Cutoff160130160100
-25% Cutoff201921890
+25% Cutoff203118290
+50% Cutoff4038400100
+75% Cutoff2044200100

PCP:

TCA:

% of CutoffNumber of samplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20252020100
-50% Cutoff1604990160100
-25% Cutoff2075921890
+25% Cutoff20125819195
+50% Cutoff401503400100
+75% Cutoff201751200100

PPX:

% of CutoffNumber ofsamplesDrugConcentration byLC/MS(ng/mL)Lay person ResultsThe percentageof correct results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2077020100
-50% Cutoff1601490160100
-25% Cutoff2022521890
+25% Cutoff2037419195
+50% Cutoff40448400100
+75% Cutoff20523200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11. Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off,

{26}------------------------------------------------

interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the SAFECARE Dip Card Tests and SAFECARE Cup Tests are substantially equivalent to the predicate.

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).