K Number
K153646

Validate with FDA (Live)

Date Cleared
2016-03-18

(88 days)

Product Code
Regulation Number
862.3250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SAFECARE Urine Test Amphetamine Cassette is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Amphetamine Cup is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Amphetamine DipCard is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine Cassette is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine Cup is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine DipCard is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana Cassette is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana Cup is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana DipCard is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

Device Description

SAFECARE Urine Test devices are immunochromatographic assays for cocaine, amphetamine and marijuana. Each assay test is a lateral flow, one step system for the qualitative detection of Benzoylecgonine, or D-amphetamine or 11-nor-A9-THC-9 COOH (target analyte) in human urine. The product is a single-use in vitro diagnostic device, which comes in the form of: DipCards, or Cups or Cassettes. It contains a Test Device (in one of the three formats), a package insert and a urine cup. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

The provided document describes the acceptance criteria and study results for the SAFECARE Urine Test Amphetamine, Cocaine, and Marijuana devices.

Here's an organized summary of the information requested:


Acceptance Criteria and Device Performance Study

The SAFECARE Urine Test devices are qualitative tests for the detection of Amphetamine, Cocaine (Benzoylecgonine), and Marijuana (Cannabinoids) in human urine at specific cutoff concentrations. The study aims to demonstrate the device's accuracy and reliability through analytical performance and comparison studies.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numerical targets (e.g., Sensitivity > X%, Specificity > Y%). However, the study design implicitly establishes acceptance criteria through its reported performance, which demonstrates consistent and accurate results around the cutoff concentrations. Given that this is a 510(k) submission and the conclusion states "acceptable performance characteristics" and "substantially equivalent to the predicate," the presented performance implicitly meets the FDA's requirements for substantial equivalence.

Interpreted Acceptance Criteria (Implicitly met by study results):

  • Precision: Consistent results across different lots and repeated testing at various concentrations relative to the cutoff. For concentrations well below the cutoff (-100%, -75%, -50% cutoff), results should be consistently negative. For concentrations well above the cutoff (+50%, +75%, +100% cutoff), results should be consistently positive. At concentrations near the cutoff (-25%, cutoff, +25% cutoff), there should be an expected distribution of positive and negative results, demonstrating appropriate cutoff sensitivity.
  • Cut-off Verification: All samples at and above +50% cut-off are positive, and all samples at and below -50% cut-off are negative.
  • Interference/Specificity: No interference from common physiological substances or cross-reactivity from specified compounds at tested concentrations should lead to false positives/negatives at 25% above cutoff.
  • Method Comparison: High agreement with GC/MS results for both negative and positive clinical samples. Low rates of discordant results, particularly for samples well below or well above the cutoff.
  • Lay-user Performance: High percentages of correct results when performed by lay users, indicating ease of understanding and use. This is especially critical for over-the-counter devices.

Reported Device Performance (Excerpt for Amphetamine Dip Card - Precision Study):

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 1Negative50505047241000
Positive00032649505050
Lot 2Negative50505048262000
Positive00022448505050
Lot 3Negative50505049263000
Positive00012447505050

Self-correction made: The table in the document uses "50-/0+" format where 50- means 50 negative results and 0+ means 0 positive results. This was separated into two rows for clarity as "Negative" and "Positive".

Lay-User Performance (Excerpt for Amphetamine Cup):

% of CutoffNumber of samplesd-Amphetamine Concentration by GC/MS (ng/mL)No. of PositiveNo. of NegativeThe percentage of correct results (%)
-100% Cutoff200020100
-75% Cutoff20250020100
-50% Cutoff20500020100
-25% Cutoff2075031785
+25% Cutoff20125018290
+50% Cutoff201500200100
+75% Cutoff201750200100

2. Sample Size for the Test Set and Data Provenance

All test sets (precision, cut-off, interference, specificity, method comparison, and lay-user studies) appear to use data that is retrospective in nature, as samples were "prepared by spiking drug in negative samples" or "unaltered clinical samples" were "blind labeled and compared to GC/MS results". The provenance of the data is not explicitly stated as a country of origin but implies laboratory-prepared or collected clinical samples, likely within China where the manufacturer is located, or a contract lab.

Sample Sizes for Test Sets:

  • Precision Study:
    • For each drug (Amphetamine, Cocaine, Marijuana) and each format (Dip Card, Cup, Cassette):
      • 9 concentration levels (-100% to +100% of cut-off).
      • Each concentration tested 50 times (2 runs per day for 25 days) for each of 3 lots.
      • Total observations per drug/format: 9 concentrations * 50 tests/concentration * 3 lots = 1350 observations.
  • Cut-off Verification Study:
    • For each drug (Amphetamine, Cocaine, Marijuana) and each format (Dip Card, Cup, Cassette):
      • 5 concentration levels (-50%, -25%, cut-off, +25%, +50% of cut-off).
      • Total samples: 150 samples (equally distributed across 5 concentrations, so 30 samples per concentration) tested using 3 different lots by 3 different operators.
  • Interference/Specificity Studies:
    • Urine samples (drug-free and spiked at +25% cut-off) tested for each interfering substance/cross-reactant compound. Number of tests per substance (e.g., 3 batches of each device for all formats) is not explicitly listed per substance but implied to be sufficient.
  • Method Comparison Studies:
    • For each drug (Amphetamine, Cocaine, Marijuana) and each format (Dip Card, Cup, Cassette):
      • 80 unaltered clinical samples (40 negative, 40 positive).
      • Total samples per drug/format tested by 3 viewers: 80 samples * 3 viewers = 240 observations.
  • Lay-user Study:
    • For Amphetamine devices: 420 lay persons.
    • For Cocaine devices: 420 lay persons.
    • For Marijuana devices: 420 lay persons.
    • Each person tested one blind labeled sample and a device.
    • Sample concentrations used: 7 levels (-100% to +75% of cut-off), 20 samples per level. Total samples per drug: 7 concentrations * 20 samples/concentration = 140 samples.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • For Analytical Performance (Precision, Cut-off, Interference, Specificity): The ground truth for spiked samples (negative samples spiked with known concentrations of drug) was established by GC/MS (Gas Chromatography/Mass Spectrometry), which is the "preferred confirmatory method" and considered the gold standard for drug detection and quantification in toxicology. No "experts" in the sense of clinicians or radiologists were used. The document doesn't specify the qualifications of the personnel performing GC/MS.
  • For Method Comparison Studies: The ground truth for the 80 unaltered clinical samples was established by GC/MS. These GC/MS results are the reference against which the device's performance was compared.
  • For Lay-user Study: The ground truth for the prepared urine samples was confirmed by GC/MS.

4. Adjudication Method for the Test Set

  • For Precision Studies: The testing was "performed two runs per day for 25 days" (total 50 tests per concentration per lot). The results are presented as counts of positive/negative. There is no mention of an adjudication method among multiple readers for these analytical studies.
  • For Cut-off Verification Study: Testing was performed by "three different operators." There is no explicit adjudication method stated for these observations. The results are presented as all positive or all negative for certain concentration ranges, which implies agreement or a predefined interpretation across operators.
  • For Method Comparison Studies: "Three different laboratory assistants" were the "operators" or "viewers" (referred to as Viewer A, B, C) who read the device results. The document lists discordant results individually for each viewer, indicating that results were compared viewer-by-viewer against GC/MS. There is no explicit multi-reader adjudication method (e.g., majority vote like 2+1 or 3+1) mentioned for reaching a consensus device result; rather, individual reader agreement with GC/MS is analyzed.
  • For Lay-user Study: Lay persons independently read the results. The comparison is between the lay person's reading and the GC/MS ground truth. No adjudication among lay persons is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was mentioned. The study performed a method comparison where individual "viewers" (laboratory assistants) read the devices and their results were compared to GC/MS. There is no comparison of human readers' performance with AI assistance versus without AI assistance, as these are rapid qualitative test devices, not AI-powered diagnostic algorithms.

6. Standalone (Algorithm Only) Performance

The SAFECARE Urine Test devices are immunochromatographic assays, not software algorithms or AI-driven systems. Their performance is inherently standalone (device-only, interpreted by a human user). The reported precision, cut-off, interference, specificity, and method comparison data represent the standalone performance of these physical tests.

7. Type of Ground Truth Used

  • GC/MS (Gas Chromatography/Mass Spectrometry): This is consistently stated as the ground truth method for confirming drug concentrations in all performance studies (precision, cut-off, interference, specificity, method comparison, and lay-user studies). GC/MS is a highly accurate analytical method considered the gold standard for drug testing.

8. Sample Size for the Training Set

These devices are immunochromatographic assays, not machine learning or AI models, therefore they do not have a "training set" in the computational sense. Their development typically involves chemical and biological optimization, with analytical and clinical validation serving as the equivalent of performance evaluation.

9. How the Ground Truth for the Training Set Was Established

As noted above, these devices do not involve a training set as they are not AI/ML-based. The development of such diagnostic tests relies on established immunochemistry principles and rigorous analytical validation to ensure sensitivity, specificity, and accuracy against a gold standard like GC/MS.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 18, 2016

SAFECARE BIOTECH C/O JOE SHIA LSI INTERNATIONAL 504 E DIAMOND AVE, SUITE I GAITHERSBURG, MD, 20877

Re: K153646

Trade/Device Name: SAFECARE Urine Test Amphetamine (Cassette, Cup, DipCard), SAFECARE Urine Test Cocaine (Cassette, Cup, DipCard), SAFECARE Urine Test Marijuana (Cassette, Cup, DipCard) Regulation Number: 21 CFR 862.3250 Regulation Name: Cocaine and Cocaine Metabolite Test System Regulatory Class: Class II Product Code: DIO, LAF, LDJ Dated: December 15th, 2015 Received: December 21st, 2015

Dear Mr. Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153646

Device Name

SAFECARE Urine Test Amphetamine (Cassette, Cup, DipCard) SAFECARE Urine Test Cocaine (Cassette, Cup, DipCard) SAFECARE Urine Test Marijuana (Cassette, Cup, DipCard)

Indications for Use (Describe)

SAFECARE Urine Test Amphetamine Cassette is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Amphetamine Cup is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Amphetamine DipCard is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine Cassette is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine Cup is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine DipCard is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana Cassette is a rapid test for the qualitative detection of Cannabinoids in human urine at

{3}------------------------------------------------

a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional iudgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana Cup is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana DipCard is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

Type of Use ( Select one or both, as applicable )

|X Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) SUMMARY

1. Date:March 17, 2016
2. Submitter:Safecare Biotech Co. Ltd.18 Haishu Road, Yuhang DistrictHangzhou, China
3. Contact person:Alex QiuSafecare Biotech18 Haishu Road, Yuhang DistrictHangzhou, ChinaTelephone: 86 571-89712897Fax: 86 571-80389223Email: alexqiu@safecare.com.cn
    1. Device Name: SAFECARE Urine Test Amphetamine (Cassette, Cup, DipCard) SAFECARE Urine Test Cocaine (Cassette, Cup, DipCard) SAFECARE Urine Test Marijuana (Cassette, Cup, DipCard)
Classification: Class II
ProductCodeCFR #Panel
DIO21 CFR, 862.3250 Cocaine Test SystemToxicology
DKZ21 CFR, 862.3100 Amphetamine Test SystemToxicology
LDJ21 CFR, 862.3870 Cannabinoid Test SystemToxicology

5. Predicate Devices:

K052115 First Check Multi Drug Cup 12

    1. Intended Use
      SAFECARE Urine Test Amphetamine Cassette is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

{5}------------------------------------------------

SAFECARE Urine Test Amphetamine Cup is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Amphetamine DipCard is a rapid test for the qualitative detection of Amphetamine in human urine at a cutoff concentration of 1000 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine Cassette is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine Cup is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Cocaine DipCard is a rapid test for the qualitative detection of Benzoylecgonine in human urine at a cutoff concentration of 300 ng/ml.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

{6}------------------------------------------------

SAFECARE Urine Test Marijuana Cassette is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana Cup is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

SAFECARE Urine Test Marijuana DipCard is a rapid test for the qualitative detection of Cannabinoids in human urine at a cutoff concentration of 50 ng/mL.

The tests provide only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. The tests are intended for over-the-counter and for prescription use.

    1. Device Description
      SAFECARE Urine Test devices are immunochromatographic assays for cocaine, amphetamine and marijuana. Each assay test is a lateral flow, one step system for the qualitative detection of Benzoylecgonine, or D-amphetamine or 11-nor-A9-THC-9 COOH (target analyte) in human urine. The product is a single-use in vitro diagnostic device, which comes in the form of: DipCards, or Cups or Cassettes. It contains a Test Device (in one of the three formats), a package insert and a urine cup. Each test device is sealed with a desiccant in an aluminum pouch.

8. Substantial Equivalence Information

A summary comparison of features of the SAFECARE Urine Test and the predicate device is provided in Table 1, Table 2 & Table 3.

Table 1: Features Comparison of SAFECARE Cocaine Test and the Predicate Device
----------------------------------------------------------------------------------------------
ItemDevicePredicate -K052115
--------------------------------------

{7}------------------------------------------------

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofBenzoylecgonine in human urine.Same
CalibratorBenzoylecgonineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values300 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Forover-the-counteruse.
ConfigurationsCup, Dip Card, CassetteCup

Table 2: Features Comparison of SAFECARE Amphetamine Test and the Predicate Device

ItemDevicePredicate - K052115
Indication(s)for UseFor the qualitative determination ofD-amphetamine in human urine.Same
CalibratorD-amphetamineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values1000 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.For over-the-counteruse.
ConfigurationsCup, Dip Card, CassetteCup

Table 3: Features Comparison of SAFECARE Marijuana Test and the Predicate Device

ItemDevicePredicate - K052115
Indication(s)For the qualitative determination ofSame

{8}------------------------------------------------

ItemDevicePredicate - K052115
for Use11-nor-Δ9-THC-9 COOH in human urine.
Calibrator11-nor-Δ9-THC-9 COOHSame
MethodologyCompetitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values50 ng/mLSame
IntendedPopulationFor over-the-counter and prescription uses.For over-the-counter use.
ConfigurationsCup, Dip Card, CassetteCup

9. Test Principle

SAFECARE Urine Tests are rapid tests for the qualitative detection of Benzoylecgonine, or D-amphetamine or 11-nor-A9-THC-9 COOH in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +75% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blinded labeled and randomized. For each concentration, tests were performed two runs per day for 25 days. The results obtained are summarized in the following tables:

Amphetamine (AMP) Dip Card Format

{9}------------------------------------------------

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+47-/3+24-/26+49+/1-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+26-/24+47+/3-50+/0-50+/0-50+/0-

Amphetamine (AMP) Cup Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot 150-/0+50-/0+50-/0+48-/2+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+47-/3+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+47-/3+27-/23+48+/2-50+/0-50+/0-50+/0-

Amphetamine (AMP) Cassette Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+47-/3+25-/25+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48/2+25-/25+47+/3-50+/0-50+/0-50+/0-

Cocaine (COC) Dip Card Format

ResultDrug-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot 150-/0+50-/0+50-/0+48-/2+24-/26+47+/3-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+47+/3-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+47-/3+26-/24+48+/2-50+/0-50+/0-50+/0-

COC Cup Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
-------------------------------------------------------------------------------------------------------------------------------------------------------------------

{10}------------------------------------------------

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+48-/2+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+25-/25+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+26-/24+47+/3-50+/0-50+/0-50+/0-

COC Cassette Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+47-/3+31-/29+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+28-/22+47+/3-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+47-/3+28-/22+49+/1-50+/0-50+/0-50+/0-

Marijuana (THC) Dip Card Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+49-/1+24-/26+47+/3-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+48+/2-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+26-/24+48+/2-50+/0-50+/0-50+/0-

THC Cup Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+48-/2+24-/26+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+48-/2+26-/24+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+48-/2+27-/23+47+/3-50+/0-50+/0-50+/0-

THC Cassette Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off

{11}------------------------------------------------

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot 150-/0+50-/0+50-/0+48-/2+25-/25+48+/2-50+/0-50+/0-50+/0-
Lot 250-/0+50-/0+50-/0+47-/3+26-/24+49+/1-50+/0-50+/0-50+/0-
Lot 350-/0+50-/0+50-/0+49-/1+25-/25+47+/3-50+/0-50+/0-50+/0-

b. Linearity

Not applicable, these are visually read devices.

c. Stability

The devices are stable at 4-30°C (39-86°F) for 24 months based on the accelerated stability study at 50°C. Control materials are not provided with the device. The labeling provides information on how to obtain control materials.

d. Cut-off

A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +50% cut-off and all negative at and below -50% cut-off for Amphetamine, Cocaine and Marijuana. The following cut-off values for the test devices have been verified.

TestCalibratorCut-off (ng/mL)
One Step Amphetamine TestD-amphetamine1000
One Step Cocaine TestBenzoylecogonine300
One Step Marijuana Test11-nor-Δ9-THC-9 COOH50

e. Interference

Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.

Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different formats.

{12}------------------------------------------------

4-AcetamidophenolL-EphedrineOxycodone
Acetophenetidin(-) Y EphedrineOxymetazoline
N-AcetylprocainamideErythromycinPapaverine
Acetylsalicylic acidẞ-EstradiolPenicillin-G
AminopyrineEstrone-3-sulfatePentazocaine
AmitryptylineEthyl-p-aminobenzoatePentobarbital
AmobarbitalFenfluraminePerphenazine
AmoxicillinFenoprofenPhencyclidine
AmpicillinFurosemidePhenelzine
Ascorbic acidGentisic acidPhenobarbital
AspartameHemoglobinPhetoin
AtropineHydralazineL-Phenylephrine
Benzilic acidHydrochlorothiazidePhenylpropanolamine
Benzoic acidHydrocodonePrednisolone
BenzoylecgonineHydrocortisonePrednisone
BilirubinO-Hydroxyhippuric acidProcaine
Brompheniramine3-HydroxytyraminePromazine
CaffeineIbuprofenPromethazine
CannabidiolImipramineD,L-Propanolol
Cannabinol(-) IsoproterenolD-Propoxyphene
ChloralhydrateIsoxsuprineQuinidine
ChloramphenicolKetamineQuinine
ChlordiazepoxideKetoprofenRanitidine
ChlorothiazideLabetalolSalicylic acid
(±) ChlorpheniramineLevorphanolSecobarbital
ChlorpromazineLoperamideSulfamethazine
ChlorquineMaprotilineSulindac
CholesterolMeperidineTemazepam
ClomipramineMeprobamateTetracycline
ClonidineMethadoneTetrahydrocortisone
Cocaine hydrochlorideMethylphenidateTetrahydrozoline
CodeineMorphine-3-Dglucuronide49-THC-COOH
CortisoneNalidixic acidThebaine
(-) CotinineNaloxoneThiamine
CreatinineNaltrexoneThioridazine
DeoxycorticosteroneNaproxenD,L-Thyroxine
DextromethorphanNiacinamideTolbutamine
DiazepamNifedipineTriamterene
DiclofenacNorcodeinTrifluoperazine

{13}------------------------------------------------

DiflunisalNorethindroneTrimethoprim
DigoxinD-NorpropoxypheneTrimipramine
DiphenhydramineNoscapineTryptamine
DoxylamineD,L-OctopamineD, L-Tyrosine
Ecgonine hydrochlorideOxalic acidUric acid
Ecgonine methylesterOxazepamVerapamil
(IR,2S)-(-)-EphedrineOxolinic acidZomepirac

COC

AcetominophenEstrone-3-sulfateOxymetazoline
AcetophenetidinEthyl-p-aminobenzoatePapaverine
N-AcetylprocainamideFenoprofenPenicillin-G
Acetylsalicylic acidFurosemidePentobarbital
AminopyrineGentisic acidPerphenazine
AmitryptylineHemoglobinPhencyclidine
AmobarbitalHydralazinePhenelzine
AmoxicillinHydrochlorothiazidePhenobarbital
AmpicillinHydrocodonePhentermine
L-Ascorbic acidHydrocortisoneL-Phenylephrine
DL-Amphetamine SulfateO-Hydroxyhippuric acidβ-Phenylethylamine
Apomorphinep-HydroxymethamphetaminePhenylpropanolamine
Aspartame3-HydroxytyraminePrednisolone
AtropineIbuprofenPrednisone
Benzilic acidImipramineProcaine
Benzoic acidIproniazidPromazine
Benzphetamine(±) - IsoproterenolPromethazine
(±) -BrompheniramineIsoxsuprineDL-Propranolol
CaffeineKetamineD-Propoxyphene
CannabidiolKetoprofenD-Pseudoephedrine
CannabinolLabetalolQuinidine
ChloralhydrateLevorphanolQuinine
ChloramphenicolLoperamideRanitidine
ChlordiazepoxideMaprotilineSalicylic acid
ChlorothiazideMeperidineSecobarbital
(±) -ChlorpheniramineMeprobamateSerotonin
ChlorpromazineMethadoneSulfamethazine
ChlorquineMethoxyphenamineSulindac
Cholesterol(±) -3,4-MethylenedioxyamphetamineTemazepam

{14}------------------------------------------------

Clomipraminehydrochloride(±)-3,4-Methylene- dioxymethamphetaminehydrochlorideTetracycline
ClonidineMorphine-3-B-D glucuronideTetrahydrocortisone3-(β-D glucuronide)
CodeineMorphine SulfateTetrahydrozoline
CortisoneNalidixic acidThebaine
(-) CotinineNaloxoneThiamine
CreatinineNaltrexoneThioridazine
DeoxycorticosteroneNaproxenDL-Tyrosine
DextromethorphanNiacinamideTolbutamide
DiazepamNifedipineTriamterene
DiclofenacNorcodeinTrifluoperazine
DiflunisalNorethindroneTrimethoprim
DigoxinD-NorpropoxypheneTrimipramine
DiphenhydramineNoscapineTryptamine
DoxylamineDL-OctopamineDL-Tryptophan
Ecgonine methylesterOxalic acidTyramine
(-) - Ψ-EphedrineOxazepamUric acid
ErythromycinOxolinic acidVerapamil
β-EstradiolOxycodoneZomepirac

THC

4-Acetamidophenolβ-EstradiolPapaverine
AcetophenetidinEstrone-3-sulfatePenicillin-G
N-AcetylprocainamideEthyl-p-aminobenzoatePentazocine
Acetylsalicylic acidFenoprofenPentobarbital
AminopyrineFurosemidePerphenazine
AmitryptylineGentisic acidPhencyclidine
AmobarbitalHemoglobinPhenelzine
AmoxicillinHydralazinePhenobarbital
AmpicillinHydrochlorothiazidePhentermine
Ascorbic acidHydrocodoneL-Phenylephrine
D,L-AmphetamineHydrocortisoneβ-Phenylethlamine
L-AmphetamineO-Hydroxyhippuric acidβ-Phenyllethylamine
Apomorphine3-HydroxytyraminePhenylpropanolamine
AspartameIbuprofenPrednisolone
AtropineImipraminePrednisone
Benzilic acidIproniazidProcaine
Benzoic acid(-) IsoproterenolPromazine
BenzoylecgonineIsoxsuprinePromethazine
BenzphetamineKetamineD,L-Propanolol
BilirubinLabetalolD-Propoxyphene
BrompheniramineLevorphanolD-Pseudoephedrine
CaffeineLoperamideQuinidine
ChloralhydrateMaprotilineQuinine
ChloramphenicolMeprobamateRanitidine
ChlordiazepoxideMethadoneSalicylic acid
ChlorothiazideMethoxyphenamineSecobarbital
(±) Chlorpheniramine(+)3,4-MethylenedioxyamphetamineSerotonin(5-Hydroxytyramine)
Chlorpromazine(+)3,4-MethylenedioxymethamphetamineSulfamethazine
ChlorquineMethylphenidateSulindac
CholesterolMethyprylonTemazepam
ClomipramineMorphine-3-β-DglucuronideTetracycline
ClonidineNalorphineTetrahydrocortisone3(5-Dglucuronide)
Cocaine hydrochlorideNaloxoneTetrahydrozoline
CodeineNalidixic acidThebaine
CortisoneNaltrexoneThiamine
(-) CotinineNaproxenThioridazine
CreatinineNiacinamideD, L-Thyroxine
DeoxycorticosteroneNifedipineTolbutamine
DextromethorphanNorcodeinTriamterene
DiazepamNorethindroneTrifluoperazine
DiclofenacD-NorpropoxypheneTrimethoprim
DiflunisalNoscapineTrimipramine
DigoxinD,L-OctopamineTryptamine
DiphenhydramineOxalic acidD, L-Tryptophan
DoxylamineOxazepamTyramine
Ecgonine hydrochlorideOxolinic acidD, L-Tyrosine
Ecgonine methylesterOxycodoneUric acid
(-) Y EphedrineOxymetazolineVerapamil
Erythromycinp-HydroxymethamphetamineZomepirac

{15}------------------------------------------------

f. Specificity

{16}------------------------------------------------

To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.

AMP(d-Amphetamine, Cut-off=1000 ng/mL)Result%Cross-Reactivity
d1-AmphetaminePositive at 3000 ng/mL33%
l-AmphetaminePositive at 20000 ng/mL5%
(+/-) 3,4-methylene-dioxyamphetamine (MDA)Positive at 20000 ng/mL5%
PhenterminePositive at 30000 ng/mL3%
HydroxyamphetaminePositive at 8000 ng/mL12.5%
d-Methamphetamine>100000<1%
l-Methamphetamine>100000<1%
ephedrine>100000<1%
3,4-Methylenedioxyethylamphetamine (MDE)>100000<1%
3,4-methylenedioxy-methamphetamine (MDMA)>100000<1%
COC(Benzoylecogonine, Cut-off=300 ng/mL)Result% Cross-Reactivity
Cocaine HClPositive at 750 ng/mL40%
CocaethylenePositive at 12500 ng/mL2.4%
EcgoninePositive at 32000 ng/mL0.9%
THCResult%
(11-Nor- $\Delta^9$ -Tetrahydrocannabinol-9-COOH,Cut-off=50 ng/mL)Positive at 50 ng/mLCross-Reactivity100%
11-Hydroxy- $\Delta^9$ -TetrahydrocannabinolPositive at 5000 ng/mL1%
11-Nor- $\Delta^8$ -Tetrahydrocannabinol-9-COOHPositive at 50100%

{17}------------------------------------------------

ng/mL
CannabinolPositive at20000 ng/mL0.3%
Δ8-TetrahydrocannabinolPositive at10000 ng/mL0.5%
Δ9-TetrahydrocannabinolPositive at10000 ng/mL0.5%
CannabidiolPositive at20000 ng/mL0.3%
11-Nor-Δ9-THC-carboxy glucuronidePositive at2500 ng/mL2%
(-)-11-nor-9-carboxy-Δ 9-THCPositive at2500 ng/mL2%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

    1. Comparison Studies
      The method comparison studies for the SAFECARE Urine Test (Amphetamine, Cocaine and Marijuana) were performed in-house with three different laboratory assistants for each format of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to GC/MS results are presented in the tables below:
CassetteformatAMI
NegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0011920
Negative10101910
Viewer BPositive0021920
Negative10101810

{18}------------------------------------------------

Viewer CPositive0021820
Negative10101820
Discordant Results of AMP Cassette
ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer A114130920Positive
Viewer B112663903Positive
Viewer B116513872Positive
Viewer C114130920Positive
Viewer C112663903Positive
Viewer A1160321086Negative
Viewer B1130011025Negative
Viewer C1130011025Negative
Viewer C1160321086Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Positive0011920
Viewer ANegative10101910
Viewer BPositive0021820
Negative10101820
Viewer CPositive0011820
Negative10101920

Discordant Results of AMP Dip Card

ViewerSample NumberGC/MS ResultDipcard FormatViewer Results
Viewer A114130920Positive
Viewer B114130920Positive
Viewer B112663903Positive
Viewer C112663903Positive
Viewer A1130011025Negative
Viewer B1130011025Negative
Viewer B1160321086Negative
Viewer C1160321086Negative

{19}------------------------------------------------

ViewerSample NumberGC/MS ResultDipcard FormatViewer Results
Viewer C1113051129Negative
CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0021820
Negative10101820
Viewer BPositive0021920
Negative10101810
Viewer CPositive0011920
Negative10101910

Discordant Results of AMP Cup

ViewerSample NumberGC/MS ResultCup Format Viewer Results
Viewer A112663903Positive
Viewer A116513872Positive
Viewer B114130920Positive
Viewer B112663903Positive
Viewer C114130920Positive
Viewer A1160321086Negative
Viewer A1113051129Negative
Viewer B1160321086Negative
Viewer C1130011025Negative

COC

CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50% )HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0021920

{20}------------------------------------------------

Negative10101810
Viewer BPositive0021920
Negative10101810
Viewer CPositive0011820
Negative10101920

Discordant Results of COC Cassette

Discordant Results of COC Cassette
ViewerSample NumberGC/MS ResultCassette Format Viewer Results
Viewer A126369294Positive
Viewer A125081293Positive
Viewer B126369294Positive
Viewer B125081293Positive
Viewer C126369294Positive
Viewer A123901325Negative
Viewer B124383319Negative
Viewer C123901325Negative
Viewer C124383319Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than -50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0021820
Negative10101820
Viewer BPositive0011820
Negative10101920
Viewer CPositive0021920
Negative10101810

Discordant Results of COC DipCard

ViewerSample NumberGC/MS ResultDipCard Format Viewer Results
Viewer A125081293Positive
Viewer A126518288Positive
Viewer B126369294Positive
Viewer C126369294Positive
Viewer C125081293Positive

{21}------------------------------------------------

ViewerSample NumberGC/MS ResultDipCard FormatViewer Results
Viewer A123901325Negative
Viewer A124383319Negative
Viewer B123901325Negative
Viewer B121307330Negative
Viewer C121307330Negative
CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0011820
Negative10101920
Viewer BPositive0011820
Negative10101920
Viewer CPositive0021920
Negative10101810

Discordant Results of COC Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer A126369294Positive
Viewer B125081293Positive
Viewer C126369294Positive
Viewer C125081293Positive
Viewer A124383319Negative
Viewer A121307330Negative
Viewer B123901325Negative
Viewer B121307330Negative
Viewer C123901325Negative

{22}------------------------------------------------

CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0021820
Negative10101820
Viewer BPositive0021820
Negative10101820
Viewer CPositive0021720
Negative10101830

Discordant Results of THC Cassette

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer A10393246Positive
Viewer A10631347Positive
Viewer B10358748Positive
Viewer B10631347Positive
Viewer C10358748Positive
Viewer C10393246Positive
Viewer A10320152Negative
Viewer A10239654Negative
Viewer B10320152Negative
Viewer B10465954Negative
Viewer C10320152Negative
Viewer C10465954Negative
Viewer C10239654Negative
DipCardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0021920
Negative10101810
Viewer BPositive0021820

{23}------------------------------------------------

Negative101010
Viewer CPositiveCl1800ﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ
Negative101018
ViewerSample NumberGC/MS ResultDipCard Format Viewer Results
Viewer A10358748Positive
Viewer A10393246Positive
Viewer B10358748Positive
Viewer B10413945Positive
Viewer C10393246Positive
Viewer C10631347Positive
Viewer A10465954Negative
Viewer B10320152Negative
Viewer B10465954Negative
Viewer C10320152Negative
Viewer C10239654Negative

Discordant Results of THC DipCard

CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0031720
Negative10101730
Viewer BPositive0021820
Negative10101820
Viewer CPositive0021820
Negative10101820

Discordant Results of THC Cup

ViewerSample NumberGC/MS ResultCup Format Viewer Results
Viewer A10358748Positive
Viewer A10393246Positive
Viewer A10631347Positive
Viewer B10358748Positive
Viewer B10393246Positive

{24}------------------------------------------------

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer C10358748Positive
Viewer C10340144Positive
Viewer A10320152Negative
Viewer A10236858Negative
Viewer A10465954Negative
Viewer B10320152Negative
Viewer B10239654Negative
Viewer C10320152Negative
Viewer C10465954Negative

Lay-user study

A lay user study was performed at three intended user sites with 420 lay persons testing the amphetamine devices, 420 lay persons testing the cocaine devices and 420 lay persons testing the marijuana devices. A total of 208 females and 212 males tested the amphetamine samples, 205 females and 215 males tested cocaine samples, and 212 females and 208 males tested the marijuana samples. They had diverse educational and professional backgrounds and ranged in age from 21 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

% of CutoffNumber of samplesd-AmphetamineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff20250020100
-50% Cutoff20500020100
-25% Cutoff2075021890
+25% Cutoff20125017385
+50% Cutoff201500200100
+75% Cutoff201750200100

Comparison between GC/MS and Lay Person Results AMP Cup

Comparison between GC/MS and Lay Person Results AMP Dip Card

% of CutoffNumberd-AmphetamineLay person resultsThe
-------------------------------------------------------------

{25}------------------------------------------------

ofsamplesConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100
-75%Cutoff20250020100
-50% Cutoff20500020100
-25% Cutoff2075031785
+25% Cutoff20125018290
+50% Cutoff201500200100
+75% Cutoff201750200100

Comparison between GC/MS and Lay Person Results AMP Cassette

% of CutoffNumber of samplesd-Amphetamine Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff200020100
-75% Cutoff20250020100
-50% Cutoff20500020100
-25% Cutoff2075021890
+25% Cutoff20125018290
+50% Cutoff201500200100
+75% Cutoff201750200100

Comparison between GC/MS and Lay Person Results COC Cup

% of CutoffNumberofsamplesBenzoylecgonineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022521890
+25% Cutoff2037518290
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results COC Dip Card

{26}------------------------------------------------

NumberofsamplesBenzoylecgonineConcentration by GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100
-75%Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022521890
+25% Cutoff2037517385
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results COC Cassette

% of CutoffNumber of samplesBenzoylecgonineConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100
-75% Cutoff2075020100
-50% Cutoff20150020100
-25% Cutoff2022531785
+25% Cutoff2037517385
+50% Cutoff20450200100
+75% Cutoff20525200100

Comparison between GC/MS and Lay Person Results THC Cup

% of CutoffNumber of samples11-nor-D9-THC-9-COOHConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.521890
+25% Cutoff2062.518290
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Comparison between GC/MS and Lay Person Results THC Dip Card

{27}------------------------------------------------

% of CutoffNumber of samples11-nor-D9-THC-9-COOHConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.531785
+25% Cutoff2062.518290
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Comparison between GC/MS and Lay Person Results THC Cassette

Numberofsamples11-nor-D9-THC-9-COOHConcentration by GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100
-75% Cutoff2012.5020100
-50% Cutoff2025020100
-25% Cutoff2037.531785
+25% Cutoff2062.517385
+50% Cutoff2075200100
+75% Cutoff2087.5200100

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.
  • 11.Conclusion
    Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the SAFECARE Urine Test devices are substantially equivalent to the predicate.

§ 862.3250 Cocaine and cocaine metabolite test system.

(a)
Identification. A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.(b)
Classification. Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).