K Number
K200910

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-05-02

(26 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

Device Description

The BOSS Balloon Guide Catheter system is a sterile, single-use intravascular catheter. The BOSS Balloon Guide Catheter is a 90 cm long, variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and a radiopaque marker band at the tip distal to the balloon. The catheter shaft has an annular inflation lumen and a coaxial central lumen with stainless steel coil and is braid reinforced. The hub central port leads to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port, connecting to the annular inflation lumen, and is used to facilitate inflating and deflating the balloon with a syringe. The BOSS Balloon Guide Catheter uses a distal hydrophilic coating to provide lubricity and reduce friction between the catheter shaft and the vessel wall. The BOSS Balloon Guide Catheter is packaged with the 6F dilator to facilitate the option of direct access into the blood vessel without the use of an introducer sheath.

AI/ML Overview

This document is a 510(k) Summary for the BOSS™ Balloon Guide Catheter, a medical device. It describes the non-clinical performance testing conducted to demonstrate substantial equivalence to legally marketed predicate devices, which means no clinical studies were required. Therefore, the information typically associated with AI/ML device studies (e.g., training set, test set, expert adjudication, MRMC studies, standalone performance) is not applicable here as this is a medical device, not an AI/ML diagnostic tool.

The "acceptance criteria" for this device are the pre-determined thresholds for various physical, functional, and biocompatibility tests, as detailed in the "NONCLINICAL PERFORMANCE TESTING SUMMARY" and "Biocompatibility Testing Summary" sections. The "study that proves the device meets the acceptance criteria" refers to the successful completion of all these non-clinical (bench and animal) tests.

Here's a breakdown of the requested information based on the provided text, reinterpreting "acceptance criteria" and "study" in the context of a medical device 510(k) submission:


1. Table of Acceptance Criteria and Reported Device Performance

Study NameDescriptionReference StandardAcceptance Criteria (Implied / Stated)Reported Device Performance
Visual Inspection and Dimensional VerificationTo demonstrate that the product meets the dimensional specificationsISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsMeeting pre-determined specificationsPass: All samples met the pre-determined acceptance criteria
Surface InspectionTo demonstrate the product satisfies the visual surface requirementsISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsMeeting pre-determined requirementsPass: All samples met the pre-determined acceptance criteria
Torque TestingTo demonstrate that the product is capable of 360 degrees of rotation about the central lumen axis without failure.FDA guidance PTCA Catheters:2010 §VIII.A.10360 degrees rotation without failure, meeting pre-determined criteriaPass: All samples met the pre-determined acceptance criteria
TensileTo demonstrate the product satisfies the tensile strength requirements for bonds and tip pull testISO 10555-1:2013, FDA guidance PTCA Catheters:2010 §VIII.A.7, 8Meeting tensile strength requirementsPass: All samples met the pre-determined acceptance criteria
Kink ResistanceTo demonstrate that the product has acceptable kink resistanceFDA Guidance PTCA:2010 §VIII.A.9 Kink TestAcceptable kink resistance, meeting pre-determined criteriaPass: All samples met the pre-determined acceptance criteria
Catheter LubricityPad friction test to compare coated to uncoated samplesCharacterization onlyCharacterization: Implied reduction in friction desiredResults show a 97% reduction in friction compared to uncoated samples
Particulates, Coating IntegrityThis study was conducted to determine the quantity and size of particles generated during simulated useAAMI TIR42:10 Evaluation of particulates associated with vascular medical devices, USP <788> Particulate Matter in InjectionsMeeting pre-determined acceptance criteria for particulate quantity and sizePass: All samples met the pre-determined acceptance criteria
RadiopacityTo determine the radiopaque characteristics of the device.ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsMeeting pre-determined acceptable radiopaque characteristicsPass: All samples met the pre-determined acceptance criteria
Push/Track, Device CompatibilityTo demonstrate that the device is deliverable through tortuous path model without kinking or buckling, is able inflate, deflate balloon at target, is compatible with treatment devices and removed without damage.ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsDeliverability, proper inflation/deflation, compatibility, removal without damage, meeting pre-determined criteriaPass: All samples met the pre-determined acceptance criteria
System Leak - Liquid LeakTo demonstrate that the product meets the liquid leakage under pressure requirementsISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsMeeting liquid leakage requirementsPass: All samples met the pre-determined acceptance criteria
System Leak - AspirationTo demonstrate that the product meets the hub aspiration air leakage requirementsISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsMeeting hub aspiration air leakage requirementsPass: All samples met the pre-determined acceptance criteria
Tip StiffnessTo demonstrate that the stiffness of the distal end of the product is similar to other marketed devices.N/A (Comparative, not a standard test)Stiffness similar to other marketed devices, meeting pre-determined criteriaPass: All samples met the pre-determined acceptance criteria
Corrosion ResistanceTo demonstrate the product satisfies the corrosion resistance requirementsISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsMeeting corrosion resistance requirementsPass: All samples met the pre-determined acceptance criteria
Catheter Burst Pressure Under Static ConditionsTo demonstrate the catheter does not leak or rupture up to 300 psi of internal pressure.ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirementsNo leak or rupture at up to 300 psiPass: All samples met the pre-determined acceptance criteria
Balloon Inflation / DeflationTo demonstrate balloon meets the inflation and deflation time specificationsISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation CathetersMeeting inflation and deflation time specificationsPass: All samples met the pre-determined acceptance criteria
Balloon FatigueTo demonstrate that there is no degradation of the balloon after 10 inflation cycles.ISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation CathetersNo degradation after 10 inflation cyclesPass: All samples met the pre-determined acceptance criteria
Balloon Diameter / VolumeTo demonstrate that the product meets the inflation volume vs balloon diameter specificationsISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation CathetersMeeting inflation volume vs. balloon diameter specificationsPass: All samples met the pre-determined acceptance criteria
Balloon Rated Burst VolumeTo demonstrate that the balloon is capable of withstanding an injection volume of 1.5 ml.ISO 10555-4:2013 Intravascular Catheters - Sterile and Single-Use Catheters - Part 4: Balloon Dilatation CathetersWithstanding 1.5 ml injection volumePass: All samples met the pre-determined acceptance criteria
Small Bore Connector Compliance with StandardTo demonstrate that the product meets the requirements for small bore connectorsISO 80369-7 2016 Small-bore connectors for liquids and gases in healthcare applications - Part 7, Connectors for intravascular or hypodermic applications.Meeting requirements for small bore connectorsPass: All samples met the pre-determined acceptance criteria
Usability Study / Simulated UseEvaluation of device performance to meet the user needs under simulated use conditions with accessories and treatment devicesFDA Guidance Human Factors and Usability Engineering 2016, IEC 62366:2015Meeting user needs under simulated use conditionsPass: All samples met the pre-determined acceptance criteria
Conditioning, Distribution, and Shelf life AgingTo demonstrate the device met all specifications at both baseline (T=0) and following accelerated aging to a 12-month shelf life equivalent (T=12)Evaluation of device following accelerated aging to a 12-month shelf life equivalent; ISTA procedure 3A (2018), ASTM D4169-16, ASTM F1980-16, ASTM F88-15, ASTM F2096-11 (for packaging)Meeting all specifications at T=0 and T=12-month AA equivalentPass: All samples met the pre-determined acceptance criteria
Visual Inspection Packaging and LabelingTo demonstrate that the product meets the packaging and labeling visual and adherence requirementsN/AMeeting visual and adherence requirements for packaging and labelingPass: All samples met the pre-determined acceptance criteria
Biocompatibility Testing
CytotoxicityTo demonstrate non-cytotoxicityISO 10993-5:2009Non-cytotoxicPass: Noncytotoxic according to the predetermined acceptance criteria
Intracutaneous IrritationTo demonstrate acceptable intracutaneous reactivityISO 10993-10:2010Meeting test requirements for intracutaneous reactivityPass: Test requirements for intracutaneous reactivity were met according to the predetermined acceptance criteria
SensitizationTo demonstrate no sensitization responseISO 10993-10:2010, Kligman Maximization TestNo sensitization responsePass: Did not elicit a sensitization response according to the predetermined acceptance criteria
Systemic ToxicityTo demonstrate acceptable systemic toxicityISO 10993-11:2017Meeting test requirements for systemic toxicityPass: Test requirements for systemic toxicity were met according to the predetermined acceptance criteria
Material Mediated PyrogenicityTo demonstrate non-pyrogenicityISO 10993-11:2017, USP 40 <151> Pyrogen TestNon-pyrogenicPass: Nonpyrogenic, met the predetermined acceptance criteria
HemolysisTo demonstrate non-hemolytic propertiesASTM F756-17, ISO 10993-4Non-hemolyticPass: Non-hemolytic, met the predetermined acceptance criteria
Complement ActivationTo demonstrate similar complement activation characteristics as control deviceISO 10993-4:2017, SC5b-9 Complement ActivationSimilar complement activation characteristics to control devicePass: Demonstrates similar complement activation characteristics as the control device, met the predetermined acceptance criteria
In vivo ThromboresistanceTo demonstrate similar thromboresistance characteristics as control deviceISO 10993-4:2017Similar thromboresistance characteristics to control devicePass: Demonstrates similar thromboresistance characteristics as the control device, met the predetermined acceptance criteria

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: The document repeatedly states "All samples met the pre-determined acceptance criteria" or similar for the non-clinical bench tests. While the exact numerical sample size for each test is not explicitly listed, it implies that a sufficient number of devices were tested to meet the requirements of the referenced standards (e.g., ISO 10555-1, FDA guidances). For the animal study, the sample size is not specified but it states "an acute porcine model."
  • Data Provenance:
    • Country of Origin: Not specified for the bench tests. The animal study used an "acute porcine model."
    • Retrospective or Prospective: All testing was prospective, designed specifically to evaluate the device against established standards and criteria for regulatory submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable. The "ground truth" for this medical device 510(k) submission is established by the performance standards (e.g., ISO, ASTM, FDA guidances) and the physical properties of the device itself (e.g., dimensions, strength, material characteristics). Experts were involved in setting these standards and conducting the tests, but not in the "ground truth" establishment in the sense of labeling data for an AI/ML model for diagnostic purposes.

4. Adjudication Method for the Test Set

Not applicable. This is not a study requiring human adjudication of results in the context of an AI/ML model. The "adjudication" is inherent in whether the device "passes" or "fails" the pre-defined engineering and performance criteria.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device (catheter) and the submission is based on non-clinical performance and substantial equivalence to predicate devices, not an AI/ML diagnostic tool requiring MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" in this context is based on:

  • Established international and national standards for medical devices (e.g., ISO 10555 series, ISO 10993 series, AAMI, ASTM).
  • FDA guidance documents for specific device types (e.g., PTCA Catheters).
  • Pre-determined engineering specifications and performance criteria for the device.
  • Results from a non-GLP (Good Laboratory Practice) animal study.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 2, 2020

Regulatory Technology Services LLC Prithul Bom Responsible Third Party Official 1000 Westgate Drive, Suite #510k Saint Paul, Minnesota 55114

Re: K200910

Device Name: BOSS™ Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: OJP, DOY Dated: April 4, 2020 Received: April 6, 2020

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K200910

Device Name BOSS Balloon Guide Catheter

Indications for Use (Describe)

The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

BOSS™ Balloon Guide Catheter 510(k) Summary

SUBMITTER [807.92(a)(1)]

Submitter's Name:Marblehead Medical, LLC
Address:11600 96th Ave N Maple Grove, MN 55369
Contact Person:Kristin Mortenson
Telephone:(612) 202-1142
emailMortenson.Kristin@MarbleheadMedical.com
Date Prepared:April 24, 2020
DEVICE [807.92(a)(2)]
Device Trade Name:BOSS™ Balloon Guide Catheter
Common Name:Balloon Guide Catheter
Classification Name:Percutaneous Catheter
Product Code:QJP, DQY
Regulatory Class:Class II
Regulation Number:21 CFR 870.1250
PREDICATE [807.92(a)(3)]
Primary Predicate Device:8F FlowGate2™ Balloon Guide Catheter, K153729
Secondary Predicate Device:Neuron™ MAX System, K111380

DEVICE DESCRIPTION [807.92(a)(4)]

The BOSS Balloon Guide Catheter system is a sterile, single-use intravascular catheter. The BOSS Balloon Guide Catheter is a 90 cm long, variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and a radiopaque marker band at the tip distal to the balloon. The catheter shaft has an annular inflation lumen and a coaxial central lumen with stainless steel coil and is braid reinforced. The hub central port leads to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port, connecting to the annular inflation lumen, and is used to facilitate inflating and deflating the balloon with a syringe. The BOSS Balloon Guide Catheter uses a distal hydrophilic coating to provide lubricity and reduce friction between the catheter shaft and the vessel wall. The BOSS Balloon Guide Catheter is packaged with the 6F dilator to facilitate the option of direct access into the blood vessel without the use of an introducer sheath.

INDICATION FOR USE [807.92(a)(5)]

The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular

{4}------------------------------------------------

systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

PREDICATE DEVICE COMPARISON [807.92(a)(6)]

The following table provides a comparison of the key characteristics of the BOSS Balloon Guide Catheter System to the predicate devices.

Primary Predicate deviceSecondary Predicate deviceSubject Device
FeatureFlowGate2TM Balloon Guide Catheter (K153720)NeuronTM MAX System (K111380)BOSS Balloon Guide Catheter System
FDA ClassificationClass IIClass IISame
Product Code(s)DQYDQYQJP, DQY
Regulation Number870.1250870.1250Same
Regulation NamePercutaneous CatheterPercutaneous CatheterSame
Indications for UseFlowGateTM Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.The NeuronTM MAX System is indicated for the introduction of interventional devices into the peripheral, coronary and neuro vasculature.The primary predicate and subject device have the same indications for use, except the primary predicate includes "and other angiographic" and the subject device does not include angiographic procedures. The primary predicate and subject device do not include coronary vasculature like the secondary predicate. The secondary predicate is included to support the direct vascular access and the hydrophilic coating. The secondary predicate does not include a balloon, therefore does not include temporary vascular occlusion. In addition, the secondary predicate does not indicate use as conduit for retrieval devices, a more specific use within the intended use.
Primary Predicate deviceSecondary Predicate deviceSubject Device
FeatureFlowGate2TM Balloon Guide Catheter (K153720)NeuronTM MAX System (K111380)BOSS Balloon Guide Catheter System
These differences in wording between the subject device and the predicates are not critical to the intended therapeutic or surgical use of the device, the introduction and guidance of intravascular devices into selected vasculature. The differences in wording do not affect the safety and effectiveness of the device when the device is used as labeled.
Anatomical LocationsPeripheral and neuro vasculaturePeripheral, coronary and neuro vascular systemsSame as primary predicate
Principle of OperationThe Balloon Guide Catheter is used to facilitate the selective placement of interventional devices. After needle access and insertion of up to a 0.038" guidewire into the femoral artery, the device is inserted into the artery through an introducer sheath over the guidewire. The lumen of device is used for insertion and guidance of an intravascular catheter and / or retrieval devices into a selected blood vessel for treatment.Same as primary predicate except includes the direct access use option without an introducer sheathSame as secondary predicate. Subject device includes a direct access use option whereas the primary predicate does not.
MaterialCommonly used medical grade plastics and stainless steelNot listedSame as primary predicate
Reinforced Catheter Shaft ReinforcementStainless Steel coil and braid reinforcedStainless steel braid reinforcedSame as primary predicate
Marker Band /0.02-in (0.5 mm) wideRadiopaque marker is0.08-in (2 mm) wide
Primary Predicate deviceSecondary PredicatedeviceSubject Device
FeatureFlowGate2™ BalloonGuide Catheter(K153720)Neuron™ MAX System(K111380)BOSS Balloon GuideCatheter System
Location0.03-in (0.75 mm) from thedistal tip edgeincluded on the distalend, width andlocation not listed0.02-in (0.5mm) from thedistal tip edge distal tothe balloon. Similar,difference does not raisenew questions regardingsafety and efficacy, alldevices utilizeradiopaque markerbands
CompliantBalloonYes, siliconeN/AYes, polyurethane.Similar to primarypredicate, materialdifferences do not raisenew questions of safetyand efficacy, bothmaterials are used forintravascular use.
Labeled ShaftOuterDiameter0.108-in (max)(2.7mm) 8Fr0.110-in (max)(2.8mm) 8F0.123-in (max)(3.1mm) 9.4FrSimilar, differences donot raise new questionsof safety and efficacy.
Labeled ShaftInnerDimension0.084-in (2.21mm) 6.4Fr0.088-in (2.24 mm)6.7FSame as secondarypredicate
Effectivelength85 and 95 cm80 and 90 cm90 cm, same assecondary predicate
Tip ShapeStraightStraightSame
MaximumBalloonVolume0.6mlN/ASame as primarypredicate
RadiopaqueDistal Tip has radiopaquemarker bands, stainlesssteel reinforcement in thecatheter shaft renders theshaft visible onfluoroscopyDistal Tip hasradiopaque zoneSame as primarypredicate
CoatingN/AHydrophilic Coating –distal shaftSame as the secondarypredicate. Coating is on
FeaturePrimary Predicate deviceSecondary Predicate deviceSubject Device
FlowGate2TM BalloonGuide Catheter(K153720)NeuronTM MAX System(K111380)BOSS Balloon GuideCatheter System
InternalConstructionCoaxial LumenSingle lumenthe distal portionproximal to the balloonSame as primarypredicate
Supplied items/ accessoriesDilator, RotatingHemostasis Valve, TuohyBorst Valve with Sideport,Peel Away Sheaths, Luer-Activated ValvesDilator, RotatingHemostasis Valve(RHV), ExtensiontubingDilator is provided, sameas both predicates. Theadditional accessoriesare commerciallyavailable devices.Differences do not raisenew questions of safetyand efficacy.
How SuppliedSterile, single useSterile, single useSame
SterilizationMethodEtOEtOSame
SterilityAssuranceLevel10-610-6Same

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

The differences between the subject device and predicates described in the comparison table above are not critical to the intended therapeutic or surgical use of the device, do not raise questions of safety and effectiveness, and as shown through testing and analysis, do not affect the safety and effectiveness of the device when used as labeled.

NONCLINICAL PERFORMANCE TESTING SUMMARY [807.92(b)]

Determination of substantial equivalence is based on an assessment of non-clinical performance bench test data. The BOSS Balloon Guide Catheter System has successfully completed the following relevant performance testing to demonstrate substantial equivalence. Testing was performed to evaluate physical integrity, functionality, and performance of the BOSS Balloon Guide Catheter. A summary of the tests performed is provided in the table below:

{8}------------------------------------------------

Performance Bench Testing Summary
Study NameDescriptionReference StandardResults
VisualInspection andDimensionalVerificationTo demonstrate that theproduct meets thedimensionalspecificationsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
SurfaceInspectionTo demonstrate theproduct satisfies thevisual surfacerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
Torque TestingTo demonstrate that theproduct is capable of360 degrees of rotationabout the central lumenaxis without failure.FDA guidance PTCACatheters:2010 §VIII.A.10PassAll samples metthe pre-determinedacceptance criteria
TensileTo demonstrate theproduct satisfies thetensile strengthrequirements for bondsand tip pull testISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: Generalrequirements, FDAguidance PTCACatheters:2010 §VIII.A.7,8PassAll samples met thepre-determinedacceptance criteria
Kink ResistanceTo demonstrate that theproduct has acceptablekink resistanceFDA Guidance PTCA:2010§VIII.A.9 Kink TestPassAll samples met thepre-determinedacceptance criteria
CatheterLubricityPad friction test tocompare coated touncoated samplesCharacterization onlyResults show a 97%reduction in frictioncompared touncoated samples
Particulates,CoatingIntegrityThis study wasconducted to determinethe quantity and size ofparticles generatedduring simulated useAAMI TIR42:10 Evaluationof particulates associatedwith vascular medicaldevices, USP <788>Particulate Matter inInjectionsPassAll samples met thepre-determinedacceptance criteria
RadiopacityTo determine theradiopaquecharacteristics of thedevice.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
Performance Bench Testing Summary
Study NameDescriptionReference StandardResults
Push/Track,DeviceCompatibilityTo demonstrate that thedevice is deliverablethrough tortuous pathmodel without kinkingor buckling, is ableinflate, deflate balloonat target, is compatiblewith treatment devicesand removed withoutdamage.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
System Leak -Liquid LeakTo demonstrate that theproduct meets the liquidleakage under pressurerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
System Leak -AspirationTo demonstrate that theproduct meets the hubaspiration air leakagerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
Tip StiffnessTo demonstrate that thestiffness of the distalend of the product issimilar to othermarketed devices.N/APassAll samples metthe pre-determinedacceptance criteria
CorrosionResistanceTo demonstrate theproduct satisfies thecorrosion resistancerequirementsISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
Catheter BurstPressure UnderStatic ConditionsTo demonstrate thecatheter does not leak orrupture up to 300 psi ofinternal pressure.ISO 10555-1:2013Intravascular catheters -Sterile and single-usecatheters - Part 1: GeneralrequirementsPassAll samples met thepre-determinedacceptance criteria
Balloon Inflation /DeflationTo demonstrate balloonmeets the inflation anddeflation timespecificationsISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation CathetersPassAll samples met thepre-determinedacceptance criteria
Performance Bench Testing Summary
Study NameDescriptionReference StandardResults
Balloon FatigueTo demonstrate thatthere is no degradationof the balloon after 10inflation cycles.ISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation CathetersPassAll samples met thepre-determinedacceptance criteria
BalloonDiameter /VolumeTo demonstrate that theproduct meets theinflation volume vsballoon diameterspecificationsISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation CathetersPassAll samples met thepre-determinedacceptance criteria
Balloon RatedBurst VolumeTo demonstrate that theballoon is capable ofwithstanding aninjection volume of 1.5ml.ISO 10555-4:2013Intravascular Catheters -Sterile and Single-UseCatheters - Part 4: BalloonDilatation CathetersPassAll samples met thepre-determinedacceptance criteria
Small BoreConnectorCompliancewith StandardTo demonstrate that theproduct meets therequirements for smallbore connectorsISO 80369-7 2016 Small-bore connectors forliquids and gases inhealthcare applications -Part 7, Connectors forintravascular orhypodermic applications.PassAll samples met thepre-determinedacceptance criteria
Usability Study /Simulated UseEvaluation of deviceperformance to meetthe user needs undersimulated useconditions withaccessories andtreatment devicesFDA Guidance HumanFactors and UsabilityEngineering 2016, IEC62366:2015PassAll samples met thepre-determinedacceptance criteria
Conditioning,Distribution,and Shelf lifeAgingTo demonstrate thedevice met allspecifications at bothbaseline (T=0) andfollowing acceleratedaging to a 12-monthshelf life equivalent(T=12)Evaluation of devicefollowing acceleratedaging to a 12-month shelflife equivalentPassAll samples met thepre-determinedacceptance criteria
Performance Bench Testing Summary
Study NameDescriptionReference StandardResults
Package Testing
Conditioning,Distribution,and Shelf LifeAgingEvaluation of packagingstrength and integrity atboth baseline (T=0) andfollowing acceleratedaging to a 12-monthshelf life equivalent(T=12)ISTA procedure 3A (2018),ASTM D4169-16, ASTMF1980-16,ASTM F88-15, ASTMF2096-11PassAll samples met thepre-determinedacceptance criteria
VisualInspectionPackaging andLabelingTo demonstrate that theproduct meets thepackaging and labelingvisual and adherencerequirementsN/APassAll samples met thepre-determinedacceptance criteria

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

The results of these tests provide reasonable assurance that the BOSS Balloon Guide Catheter has been designed and tested to assure conformance to the requirements for its intended use and indications for use. No new safety or performance issues were raised during the testing; therefore, this device is considered to be substantially equivalent to the predicate devices.

Biocompatibility Testing Summary

The device is categorized as Externally Communicating Device, Circulating Blood, Limited Contact (≤ 24 hours), per ISO 10993-1, the following testing was conducted:

Test NameTest MethodResults
CytotoxicityTested in accordance with ISO 10993-5:2009, Biological Evaluation of MedicalDevices – Part 5: Tests for in vitro toxicity,Neutral Red Uptake MethodPassNoncytotoxic according tothepredeterminedacceptance criteria
IntracutaneousIrritationTested in accordance with ISO 10993-10:2010, Biological Evaluation of MedicalDevices – Part 10: Tests for Irritation andSkin SensitizationPassTest requirements forintracutaneous reactivitywere met according to thepredeterminedacceptance criteria
SensitizationTested in accordance with ISO 10993-10,Biological Evaluation of Medical Devices –Part 10 Tests forIrritation and Skin Sensitization, KligmanMaximization TestPassDid not elicit asensitizationresponse according to thepredeterminedacceptance criteria
Test NameTest MethodResults
Systemic ToxicityTested in accordance with ISO 10993-11:2017, Biological Evaluation of MedicalDevices - Part 11: Tests for SystemicToxicityPassTest requirements forsystemic toxicity weremet according to thepredeterminedacceptance criteria
Material MediatedPyrogenicityTested in accordance with ISO 10993-11:2017, BiologicalEvaluation of Medical Devices - Part 11:Tests for Systemic Toxicity and USP 40<151> Pyrogen TestPassNonpyrogenic, met thepredeterminedacceptance criteria
HemolysisTested in accordance with ASTM F756-17,Standard Practice for Assessment ofHemolytic Properties of Materials and ISO10993-4, Biological Evaluation of MedicalDevices - Part 4: Selection of Tests forInteractions with Blood, Tests forHemolytic Properties, Direct and IndirectMethodsPassNon-hemolytic, met thepredeterminedacceptance criteria
ComplementActivationTested in accordance with ISO 10993-4,Biological Evaluation of Medical Devices -Part 4:2017: Selection of Tests forInteractions with Blood, SC5b-9Complement ActivationPassDemonstrates similarcomplement activationcharacteristics as thecontrol device, met thepredeterminedacceptance criteria
In vivoThromboresistanceTested in accordance with ISO 10993-4:2017, Biological Evaluation of MedicalDevices - Part 4: Selection of Tests forInteractions with BloodPassDemonstrates similarthromboresistancecharacteristics as thecontrol device, met thepredeterminedacceptance criteria

{12}------------------------------------------------

Sterilization and Shelf Life

The BOSS Balloon Guide Catheter sterilization process using 100% Ethylene Oxide (EO) has been validated in accordance with ISO 11135-1:2014 to achieve a SAL of 10 °. EO and ECH residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels (BET) were below the level of 2.15 EU/device in accordance with ANSI AAMI ST72:2011/(R)2016. Both baseline and accelerated shelf life testing were conducted (T = 0 and T = 12-month AA (accelerated aging)) demonstrating the device will perform as intended to support the proposed 12-month shelf-life.

{13}------------------------------------------------

Animal Testing

A non-GLP animal test was conducted to evaluate the in vivo performance of the device in an acute porcine model. Trackability and handling of the guide catheter, balloon inflation / deflation, temporary vascular occlusion, radiopacity, and catheter compatibility were assessed. The results for the subject device were comparable to the primary predicate device, demonstrating acceptable results.

Clinical Testing

No clinical studies were required to demonstrate substantial equivalence.

Conclusion

Marblehead Medical LLC concludes through a review of the benchtop assessments, the comparison of the device classification, intended use, operating principle, technological characteristics, sterility, and biocompatibility that the BOSS Balloon Guide Catheter System is substantially equivalent to the predicate devices.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).