Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K150741
    Date Cleared
    2015-10-22

    (213 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OJP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TSolution One™ w/ACG Surgical System is intended for use as a device that uses diagnostic images of the patient acquired specifically to assist the physicial planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the surgeon, precisely implements the presurgical software plan.

    The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming/impacting techniques for femoral canal and acetabular cup preparation in primary cementless total hip arthroplasty (THA) using a posterior approach.

    The TSolution One™ w/ACG Surgical System is indicated for orthopedic procedures in which the broaching, reaming and impacting techniques used in primary cementless total hip arthroplasty (THA) using a posterior approach may be considered to be safe and effective and where references to rigid anatomical structures may be made.

    The TSolution One™ w/ACG Surgical System is also intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The TSolution One™ w/ACG Surgical System facilitates accurate positioning of THA implants, relative to these alignment axes.

    Device Description

    The TSolution One™ w/ACG Surgical System is a three-dimensional, graphical, preoperative planner and implementation tool for treatment of patients who require a total hip arthroplasty (THA) procedure. This device is intended as an alternative to manual template planning, broaching, reaming and impacting techniques for the preparation of femoral canal and acetabular cup for patients requiring a primary THA procedure. The system consists of the TPLAN™ Preoperative Planning Workstation and TCAT™, a robotic system composed of an electromechanical arm, arm base including control electronics and computer, display monitor, and miscellaneous accessories such as cutters, drapes, irrigation sets, probes, and markers. TPLANTM and TCAT™ when used according to the instructions for use, make precision femoral canal and acetabular cup preparation possible before and during THA surgical procedures.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the TSolution One™ w/ACG Surgical System, a robotic surgical system for total hip arthroplasty (THA). The document primarily focuses on demonstrating substantial equivalence to a predicate device through comparison of technological characteristics and performance testing.

    However, the document does not contain the level of detail required to fulfill all aspects of your request, particularly regarding specific acceptance criteria values, sample sizes for test sets where performance metrics are quantified, the number and qualifications of experts, adjudication methods, details of MRMC studies, or the provenance of detailed data. The "Performance Data" section lists "Benchtop Accuracy Testing" as verifying overall system accuracy against specified requirements, but it does not provide the specific numerical acceptance criteria or the reported device performance against those criteria. It only states "PASS".

    Based on the provided text, here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    Test/ActivityAcceptance CriteriaReported Device Performance
    Software TestingFunction as intended to successfully complete THA procedure including Acetabular Cup GuidancePASS
    Usability TestingUser interface functions as intended to successfully complete THA procedure including Acetabular Cup Guidance and meet user needsPASS
    Benchtop Accuracy TestingMeet specified requirements for overall system accuracyPASS
    Cadaver TestingWorkflow functions as intended to successfully complete a THA procedure including Acetabular Cup Guidance and meet customer requirements in a simulated use environmentPASS

    Important Note: The document states "Benchtop Accuracy Testing" verifies accuracy in "meeting specified requirements" but does not explicitly state what those numerical requirements are (e.g., a specific tolerance in mm or degrees) or the measured accuracy. It only indicates "PASS."

    2. Sample sizes used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated for any of the listed tests (Software, Usability, Benchtop Accuracy, Cadaver).
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective). The cadaver testing is a simulated use environment, which is a type of prospective lab-based testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication method for the test set

    • Adjudication Method: Not specified. Given the nature of these tests (software functionality, usability, benchtop accuracy, cadaver workflow), it's likely that a predefined protocol and objective measurements were used for "PASS/FAIL" rather than expert adjudication in the sense of consensus reading of medical images.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    • MRMC Study: No, an MRMC comparative effectiveness study is not mentioned. This type of study is typical for AI-assisted diagnostic devices to measure improvement in human reader performance. This device is a robotic surgical system, not a diagnostic AI.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: The "Benchtop Accuracy Testing" would be the closest to a standalone performance evaluation, as it aims to verify the "overall system (TPLAN presurgical planning, and TCAT surgical system) accuracy." However, the exact metrics and whether this is "algorithm only" are not detailed. The system, by its description, always involves a human surgeon "under the direction of the surgeon" and implementing a "presurgical software plan." Thus, a purely standalone algorithm evaluation in the sense of a diagnostic AI is not applicable.

    7. The type of ground truth used

    • Ground Truth Type:
      • For "Software Testing" and "Usability Testing": Implied ground truth is the predefined functional specifications and user requirements for the system.
      • For "Benchtop Accuracy Testing": Implied ground truth would be precise measurements from a known reference standard (e.g., phantom, jig) against which the system's output is compared. The "specified requirements" define this ground truth.
      • For "Cadaver Testing": Implied ground truth would be successful completion of the THA procedure workflow as per established surgical protocols and meeting "customer requirements" in a simulated environment.

    8. The sample size for the training set

    • Training Set Sample Size: Not applicable. This document describes a robotic surgical system, not a machine learning model that requires a distinct training set. The "software" mentioned refers to the system's control software and planning software, which are developed through traditional software engineering processes, not trained on a data set in the AI sense.

    9. How the ground truth for the training set was established

    • Training Set Ground Truth: Not applicable, as it's not a machine learning model that undergoes training on a data set with established ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140585
    Date Cleared
    2014-08-18

    (164 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OJP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TCAT™/TPLAN™ Surgical System is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.

    The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming techniques for femoral canal preparation in primary total hip arthroplasty (THA).

    The TCAT™/TPLAN™Surgical System is indicated for orthopedic procedures in which the broaching/reaming in primary total hip arthroplasty (THA) may be considered to be safe and effective and where references to rigid anatomical structures may be made.

    Device Description

    The TCAT™ TPLAN™ Surgical System is a three-dimensional, graphical, preoperative planner and implementation tool for treatment of patients who require a total hip arthroplasty (THA) procedure. This device is intended as an alternative to manual template planning, broaching, and reaming techniques for the preparation of bone for patients requiring a THA procedure. The system consists of the TPLAN™ Preoperative Planning Workstation and TCAT™ , a robotic system composed of an electromechanical arm, arm base including control electronics and computer, display monitor, and miscellaneous accessories such as cutters, drapes, irrigation sets, probes, and markers. TPLAN™ and TCAT™ when used according to the instructions for use, make precision bone preparation possible before and during THA surgical procedures.

    AI/ML Overview

    The provided document is a 510(k) summary for the TCAT™/TPLAN™ Surgical System. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with detailed acceptance criteria and performance statistics for the new device.

    Therefore, the information requested cannot be fully extracted from this document as it does not contain a study specifically designed to establish acceptance criteria and prove the device meets them in the way a clinical trial or a detailed performance validation study would for a novel device. The document emphasizes comparison to a predicate device.

    Here's an attempt to answer the questions based only on the provided text, highlighting what is present and what is missing:


    1. A table of acceptance criteria and the reported device performance

    This information is not explicitly provided in the document. The document states that "The TCAT™ Surgical System has been evaluated with non-clinical performance testing for the following modifications and or improvements" and lists various components. It also mentions "Bench and simulated use tests included functional software testing... and hardware, functional software, user interface, instrument/tool and sterile disposable accessory testing and simulated clinical use of new and changed instrument /tools for the TCAT™/TPLAN™ Surgical System." However, specific numerical acceptance criteria (e.g., accuracy, precision thresholds) and the corresponding reported performance values for the TCAT™/TPLAN™ Surgical System are not detailed.

    Table: Acceptance Criteria and Reported Device Performance (Information Not Available in Document)

    Acceptance CriteriaReported Device Performance
    Not specifiedNot specified

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not explicitly provided. The document mentions "non-clinical performance testing" and "Bench and simulated use tests" but does not detail the sample sizes (e.g., number of test cases, number of simulated surgeries) or the provenance of any data used for these tests. It is implied these are laboratory/bench tests, not patient data studies.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not explicitly provided. Since the document describes "non-clinical performance testing" and "simulated use tests," it's unlikely that experts were used in the context of establishing a clinical ground truth for a test set in the way a diagnostic AI would require. The ground truth for engineering or functional tests would typically be established by design specifications and measurement tools.


    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not explicitly provided. Given the nature of the described tests (bench and simulated use for functional software and hardware), a clinical adjudication method is not relevant or would not be described in this context.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned in the document. The TCAT™/TPLAN™ Surgical System is described as a robotic system that implements a presurgical plan, not an AI to assist human readers in, for instance, image interpretation. The comparison is between the robotic system and manual planning/broaching/reaming techniques, not between human readers with and without AI assistance for interpretation tasks.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is partially applicable but not fully detailed. The device consists of a planning software (TPLAN™) and a robotic surgical tool (TCAT™). The "algorithm only" performance would relate to the precision and accuracy of the robotic arm's movements in executing the presurgical plan. The document states that "the robotic surgical tool... precisely implements the presurgical software plan" and mentions "functional software testing" and "hardware... testing." While this suggests standalone performance evaluation of the system's components, specific metrics and studies are not detailed. The system is still "under the direction of the surgeon," implying continued human involvement.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not specify the type of ground truth used for performance validation. For a surgical robotic system, ground truth for bench tests would typically refer to highly accurate physical measurements of movement, accuracy, and precision against known targets or reference points. It would not typically involve expert consensus, pathology, or outcomes data in the context of this 510(k) submission's described testing.


    8. The sample size for the training set

    This information is not applicable or not provided. This document describes approval for a medical device (a surgical system), not an AI model trained on a dataset. Therefore, there is no "training set" in the context of machine learning. The term "training set" is generally used for machine learning algorithms, which is not the primary focus of this device's validation as presented in this 510(k).


    9. How the ground truth for the training set was established

    This question is not applicable as there is no mention of a "training set" for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140038
    Date Cleared
    2014-05-27

    (140 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OJP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DigiMatch™ ROBODOC®/ORTHODOC® Encore Surgical System is intended for use as a device which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intraoperative procedures. The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.

    The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming techniques for femoral canal preparation in primary total hip arthroplasty (THA).

    The DigiMatch™ ROBODOC®/ORTHODOC® Encore Surgical System is indicated for orthopedic procedures in which the broaching/reaming in primary total hip arthroplasty (THA) may be considered to be safe and effective and where references to rigid anatomical structures may be made.

    Device Description

    The DigiMatch™ ORTHODOC®/ROBODOC® Encore System is a three-dimensional, graphical, preoperative planner and implementation tool for treatment of patients who require a total hip arthroplasty (THA) procedure. This device is intended as an alternative to manual template planning, broaching, and reaming techniques for the preparation of bone for patients requiring a THA procedure. The system consists of the ORTHODOC® Preoperative Planning Workstation and ROBODOC® , a robotic system composed of an electromechanical arm, electronics control cabinet, computer, display monitor, and miscellaneous accessories such as cutters, drapes, irrigation sets, probes, and markers. ORTHODOC® and ROBODOC® when used according to the instructions for use, make precision bone preparation possible before and during THA surgical procedures.

    AI/ML Overview

    The provided text describes a 510(k) summary for the DigiMatch™ ORTHODOC®/ROBODOC® Encore Surgical System, a robotic system for Total Hip Arthroplasty (THA). This submission focuses on demonstrating substantial equivalence to a predicate device (DigiMatch™ ROBODOC® Surgical System, K072629) rather than establishing novel performance metrics or conducting a new clinical study. Therefore, some of the requested information, such as detailed quantitative acceptance criteria for a new device's performance, a standalone algorithm study, or a multi-reader, multi-case study, is not explicitly provided in the summary.

    However, based on the information available, here's a breakdown of the requested points:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are based on demonstrating substantial equivalence to the predicate device, meaning the new device performs at least as well as, and presents no new safety or effectiveness concerns compared to, the predicate. The "performance" is implicitly that the modified features function as intended and do not degrade the overall system's safety or effectiveness.

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance (Implied)
    Technological Characteristics & Principles of Operation- Similar intended use and indications.
    • Similar technological characteristics (e.g., CT scan input, presurgical planning, robotic arm driven by validated software, point-to-surface registration). | - The table clearly shows identical technological characteristics and principles of operation between the new device and the predicate.
    • "Any minor differences...raise no new questions of safety or effectiveness nor change the device's intended therapeutic effect." |
      | Performance Data | - Modifications and improvements do not adversely affect safety or effectiveness.
    • Functional verification of software and hardware.
    • Simulated clinical use performs as expected. | - Non-clinical performance testing for specific modifications (e.g., ORTHODOC® host computer, CT file checks, elimination of pin patient/robot registration, ROBODOC® electromechanical arm, Smart Bone Motion Monitor, etc.).
    • Bench and simulated use tests conducted, including functional software testing and simulated clinical use. |
      | Safety and Effectiveness | - No new questions of safety or effectiveness. | - Implicitly confirmed by the FDA's substantial equivalence determination. |

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the traditional sense of a patient-based clinical study for performance evaluation against a gold standard for a diagnostic or AI device. Instead, the testing described is primarily non-clinical performance testing and simulated use tests.

    • Sample Size: Not applicable in terms of patient numbers for a test set. The "samples" would be the modified components, software modules, and system configurations that underwent bench and simulated use tests. The number of such items or test runs is not specified.
    • Data Provenance: Not applicable in terms of country of origin or retrospective/prospective for patient data. The testing was laboratory-based, focusing on the functionality of the device's modifications.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the summary. For non-clinical, bench, and simulated use testing, "ground truth" would likely be defined by engineering specifications, software design documents, and expected performance under controlled conditions, rather than expert-established ground truth from medical data.

    4. Adjudication Method for the Test Set

    This information is not provided and is generally not applicable for non-clinical and simulated use testing as described. Adjudication methods (like 2+1 or 3+1) are typically reserved for clinical studies where multiple human readers assess medical images or findings to establish a consensus ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or reported in this summary. The submission is focused on demonstrating substantial equivalence of a modified surgical robotic system, not on assessing the improvement in human reader performance with or without AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    The device is a robotic surgical system that works with human intervention (under the direction of the surgeon). While the ORTHODOC® planning software and ROBODOC® robotic arm have algorithms, the performance evaluation described is for the integrated system, including its hardware, software, and simulated use. Therefore, a standalone "algorithm only" performance study in the context of, for example, a diagnostic AI device is not applicable or reported here. The "algorithm" here controls a physical robot and forms part of a surgical aid, not an independent diagnostic tool.

    7. The Type of Ground Truth Used

    For the non-clinical and simulated use testing, the "ground truth" would be based on:

    • Engineering specifications and design requirements: Ensuring that modified components meet their intended design outputs.
    • Functional correctness: Verifying that software and hardware perform as programmed and expected during various operational scenarios simulated in the lab.
    • Expected outcomes of simulated clinical scenarios: Ensuring the robotic system accurately plans and executes simulated bone preparations within specified tolerances.

    This is distinct from "expert consensus," "pathology," or "outcomes data" which are typically used for diagnostic or predictive AI devices involving patient data.

    8. The Sample Size for the Training Set

    This information is not provided. The document describes modifications to an existing system, rather than the development of a new AI model that requires a "training set" in the machine learning sense. The "training" for such a system would typically refer to the extensive development and validation cycles of the underlying control software and robotic components, not a patient data training set.

    9. How the Ground Truth for the Training Set was Established

    As no "training set" in the machine learning context is mentioned, the method for establishing its ground truth is not applicable or discussed. The "ground truth" for the development of the surgical system (both the predicate and the modified version) would be established through principles of mechanical engineering, software engineering, and surgical accuracy requirements, validated through extensive bench testing and cadaver studies (though not explicitly detailed in this summary).

    Ask a Question

    Ask a specific question about this device

    K Number
    K072629
    Date Cleared
    2008-08-06

    (324 days)

    Product Code
    Regulation Number
    882.4560
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    OJP

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DigiMatch™ ROBODOC® Surgical System is intended for use as a device, which uses diagnostic images of the patient acquired specifically to assist the physician with presurgical planning and to provide orientation and reference information during intra-operative procedures, The robotic surgical tool, under the direction of the surgeon, precisely implements the presurgical software plan.

    The preoperative planning software and robotic surgical tool is used as an alternative to manual planning and broaching/reaming techniques for femoral canal preparation in primary total hip arthroplasty (THA).

    The DigiMatch ROBODOC Surgical System is indicated for Orthopedic procedures in which the broaching/reaming in primary total hip arthroplasty (THA) may be considered to be safe and effective and where references to rigid anatomical structures may be made.

    Device Description

    The DigiMatch™ ROBODOC® Surgical System consists of the 510(k)-cleared (K960685) ORTHODOC® Preoperative Planning Workstation (ORTHODOC), and a Robot composed of an electromechanical arm, electronics control cabinet, display monitor and miscellaneous accessories (ROBODOC). The System (ORTHODOC and ROBODOC) is a threedimensional graphical preoperative planner and implementation tool and is indicated as an alternative to template planning and manual broaching for treatment of patients who require primary total hip arthroplasty (THA). The ORTHODOC component of the System allows a surgeon to preoperatively assess a patient's femoral anatomy and state of hip disease, select an optimally sized femoral stem implant from a library of prostheses, and determine where the implant should be positioned within the femur. Then ROBODOC component of the System, under direct control by the surgeon, will precisely implement the preoperative plan.

    AI/ML Overview

    The provided text describes the DigiMatch™ ROBODOC® Surgical System and its premarket notification (K072629) to the FDA. However, the document does not explicitly state specific quantitative acceptance criteria for performance metrics. Instead, it focuses on demonstrating substantial equivalence to predicate devices and proving safety and efficacy through various studies.

    The closest to "acceptance criteria" is the demonstration of "equivalence in Harris Hip Scores" in the clinical trials, implying that the device's performance should be at least as good as manual methods in this regard. The non-clinical studies focus on accuracy and biological response.

    Here's a breakdown of the requested information based on the provided text, with caveats where information is not explicitly stated:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Accuracy of Femur Cutting (Non-clinical)ROBODOC was more accurate than manual broaching in the cadaver study and animal study.
    Gaps between Implant and Bone (Non-clinical)Significant reduction in gaps between the implant and bone with ROBODOC compared to manual broaching in the cadaver study. In the animal study, ROBODOC showed better apposition of implants with cortical and cancellous bone.
    Bone Integrity/Damage (Non-clinical)Animal study revealed many instances of fracture or osteotomy in the control group (manual) compared to none in the ROBODOC group. ROBODOC group had more consistent cortical bone continuity.
    Clinical Efficacy (Harris Hip Scores)Equivalence in Harris Hip Scores was demonstrated between the ROBODOC group and the manual control group in both the First and Second U.S. Trials.
    Precise and Reliable Implantation of Preoperative PlanDemonstrated in the First U.S. Trial through fit and alignment of the femoral implant.
    SafetyConfirmed as safe and effective through the First and Second U.S. Trials, as well as non-clinical studies (e.g., absence of fractures/osteotomies in the animal study and histological findings related to bone integrity). The FDA's substantial equivalence determination implies it meets safety benchmarks comparable to predicate devices.
    Substantial Equivalence to Predicate DevicesThe device was determined to be substantially equivalent to Voyager/Tactile Guidance System - CT (K052851), Frameless Neuromate (K991081), and da Vinci Surgical System (K043153) based on technological characteristics and principles of operation (Table 1) and confirmed by the FDA's 510(k) clearance letter.

    Study Details:

    1. Sample Size used for the test set and the data provenance:

      • Cadaver Study: Sample size not specified. Data provenance: Not explicitly stated, but typically cadaver studies are conducted in a controlled lab environment.
      • Animal Study: 20 male greyhounds. Data provenance: Not explicitly stated, but typically animal studies are conducted in a controlled lab environment.
      • First U.S. Trial (Clinical): Sample size for ROBODOC group and manual control group not specified. Data provenance: U.S., prospective clinical trial.
      • Second U.S. Trial (Clinical): Sample size for ROBODOC group and manual control group not specified. Data provenance: U.S., prospective clinical trial (IDE # G000071).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      The document does not specify the number of experts or their qualifications for establishing ground truth for any of the studies (cadaver, animal, or clinical). For clinical trials, "Harris Hip Scores" are typically evaluated by orthopedic surgeons or trained clinicians. Histological examinations would involve pathologists.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
      The document does not specify any adjudication methods (e.g., 2+1, 3+1 consensus) for establishing ground truth in any of the studies.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done. If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      No, an MRMC study was NOT done. The studies described compare the robotic system (ROBODOC) to manual techniques for hip arthroplasty, not human readers with or without AI assistance in an imaging interpretation context. The device is a surgical system that implements a plan, not an AI diagnostic tool that assists human interpretation.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Yes, in essence, the "ROBODOC" component operates in a standalone manner in terms of implementing the presurgical plan. The document states: "The ROBODOC component of the System, under direct control by the surgeon, will precisely implement the preoperative plan." While under "direct control by the surgeon," the precision and execution of the plan by the robot itself is an algorithmic function. The performance of the robotic cutting was evaluated against manual cutting in cadaver and animal studies, and clinicallly, its implementation of the plan led to equivalent Harris Hip Scores.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Cadaver Study: "Accuracy of cutting the femur" and "measurement of the gap between the bone and implant." Implies objective measurements.
      • Animal Study: "Accuracy of cutting the femur," "histological examinations of the cavity/implant interface" (pathology), and observation of "fracture or osteotomy."
      • Clinical Studies (First & Second U.S. Trials): "Fit and alignment of the femoral implant" (likely based on imaging and surgical assessment) and "Harris Hip Scores" (patient-reported outcomes and clinical assessment of hip function).
    7. The sample size for the training set:
      The document does not specify any separate "training set" for the DigiMatch™ ROBODOC® Surgical System. The device's "training" and validation would have occurred during its development and prior testing, which is not detailed here. The studies mentioned (cadaver, animal, clinical) serve as validation/test sets to demonstrate safety and effectiveness for regulatory approval.

    8. How the ground truth for the training set was established:
      As no specific "training set" is described for the device's algorithmic development, this information is not available in the provided text. The ORTHODOC component is described as being "510(k)-cleared (K960685)," indicating prior regulatory review, but details of its development and ground truth establishment are not in this K072629 summary.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1