Intended Use

The VISERA ELITE II Infrared Imaging System is intended to provide real-time endoscopic visible and near infrared fluorescence imaging. The VISERA ELITE II Infrared Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels. blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the VISERA ELITE II Infrared Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

[VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S200]
This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation.

[VISERA ELITE II XENON LIGHT SOURCE OLYMPUS CLV-S200-IR]
The light source has been designed to be used with Olympus endoscopes, video system centers, light guide cables, and other ancillary equipment for endoscopic observation through visible and near-infrared fluorescence imaging with fluorescence agent.

[HD 3CMOS AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S200-XZ-EA]
The camera head has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

[HD 3CMOS CAMERA HEAD OLYMPUS CH-S200-XZ-EB]
The camera head has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

[TELESCOPE IR/ULTRA WAIR500A,WAIR530A,WAIR100A,WAIR130A]
These endoscopes are intended to be used for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indication of transanal and transvaginal natural orifice surgery. In combination with a compatible infrared imaging system, the telescope allows for fluorescence imaging.

Device Description

The subject devices OTV-S200, CH-S200-XZ-EA, and CH-S200-XZ-EB were cleared in 510(k) K190449. After clearances, no technological modification for the subject devices were made. Therefore, the above devices have been omitted from the device description in this pre-market notification.

CLV-S200-IR: This device consists of the source circuit, the control circuit, the illumination lamp, and the optical filter. The control circuit connects to the diaphragm to regulate the light intensity, the source circuit supplies the power to the illumination lamp, the operation panel and the rear panel. By switching on the illumination lamp, this device provides the light for endoscopic observation. This device regulates the endoscopic image brightness constantly from the video system center. The observation mode can be switched by the optical filter extracting the specific wavelengths.

WAIR100A, WAIR130A, WAIR500A, WAIR530A: The "IR" Telescopes are rigid endoscopes. An image relay system of rod lenses transmits the endoscopic image. A bundle of optical fibers transmits light from an external light source to illuminate the endoscopic image. The "IR" Telescopes are delivered non-sterile. They are reusable and fully autoclavable. Before the first and each subsequent use of the device, it must be inspected and reprocessed according to defined reprocessing methods in the Instructions for Use. The "IR" Telescopes are available with different directions of view to allow use for various applications in accordance with the intended use as submitted with this 510(k).

AI/ML Overview

This document describes the VISERA ELITE II Infrared Imaging System components (Xenon Light Source, Telescopes, Video System Center, and Camera Heads). There is no explicit "device" (like an AI algorithm) for which acceptance criteria and a study proving the device meets those criteria are directly provided in the format you described.

The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than proving a novel device meets specific performance acceptance criteria for a new function. Therefore, much of the requested information regarding an AI study or specific performance metrics linked to acceptance criteria is not present.

However, I can extract information related to the device's validation and compliance:

1. Table of Acceptance Criteria and Reported Device Performance:

The document mentions several types of performance testing and compliance with standards. It does not provide a table with quantitative acceptance criteria and corresponding reported performance values for a specific device feature that would typically be seen in an AI performance study (e.g., sensitivity, specificity for disease detection). Instead, it lists validation activities.

Category of TestingAcceptance Criteria (Implied by Standards/Guidance)Reported Device Performance Summary
Reprocessing ValidationAdherence to "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling" guidance.Reprocessing instruction and method validation testing for WAIR100A/130A/500A/530A was conducted and documentation provided.
Biocompatibility TestingAdherence to ISO 10993-1 guidance.Biocompatibility testing for WAIR100A/130A/530A included Cytotoxicity, Intracutaneous Study, Guinea Pig Maximization Sensitization, Systemic Toxicity, and Material mediated pyrogenicity tests.
Software Verification & ValidationAdherence to FDA guidances for software in medical devices and cybersecurity.Software V&V for OTV-S200 and CLV-S200-IR was conducted and documentation provided.
Electrical Safety & EMCCompliance with ANSI/AAMI ES 60601-1 and IEC 60601-1-2, IEC 60601-2-18 standards.The subject devices comply with ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012, IEC 60601-2-18:2009, and IEC 60601-1-2:2014 standards.
Performance Testing - BenchMeets design specifications and intended performance (e.g., optical properties, photobiological safety).Verification for photobiological safety of illumination light. For WAIR100A/130A/500A/530A: DOV (Direction of View), FOV (Field of View), MTF (Modulation Transfer Function), Distortion, Ghost Image, Internal reflections, Spectral transmission of imaging system, Expected Service Life, Transport Drop, Design Validation / Usability, Illumination System performance Data, Comparison of Optical Properties of Subject Device and predicate Device were assessed.
Performance Testing - AnimalValidation of device performance in simulated body environment, demonstration of substantial equivalence in IR observation.An animal study (canines and swines) was performed, and videos were evaluated by an independent Healthcare Professional to demonstrate substantial equivalence in terms of IR observation.
Risk AnalysisCompliance with ISO 14971:2007, and establishment of in-house acceptance criteria.A Risk analysis for CLV-S200-IR and WAIR100A/130A/500A/530A was conducted, and design verification tests and acceptance criteria were identified and performed based on this analysis.

The subsequent points (2-9) are generally applicable to studies for AI/CADe devices and are not fully addressed by this document, which pertains to a general surgical imaging system rather than a specific AI-driven diagnostic or assistive tool.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not applicable. The document describes an animal study but does not specify a "test set" in the context of an AI algorithm evaluation. For the animal study, it mentions "canines and swines" but does not give a specific number of animals.
  • Data Provenance: The animal study involved "canines and swines." It's a prospective study as it involved taking videos to validate performance. Country of origin is not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • For the animal study videos, evaluation was done by an "independent Health Care Professional." The number of professionals and their specific qualifications (e.g., radiologist with X years of experience) are not detailed.

4. Adjudication method for the test set:

  • Not specified. It only mentions evaluation by "an independent Health Care Professional."

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC or comparative effectiveness study involving human readers with and without AI assistance was described. This document is not for an AI-assisted device in that context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is an imaging system, not an AI algorithm. Its performance is demonstrated as an imaging tool used by surgeons.

7. The type of ground truth used:

  • For the animal study, the "ground truth" for demonstrating substantial equivalence in IR observation was based on the performance observed and evaluated by a healthcare professional in a simulated body environment. This is closer to expert assessment/observation in a controlled animal setting, rather than pathology or outcomes data.

8. The sample size for the training set:

  • Not applicable. This is not an AI algorithm.

9. How the ground truth for the training set was established:

  • Not applicable. This is not an AI algorithm.

In summary:

This 510(k) pertains to an endoscopic imaging system. It demonstrates substantial equivalence to predicate devices through various engineering, safety, and limited performance tests (bench and animal study). It does not involve a software-based AI algorithm for image interpretation or diagnosis that would typically require validation against a test set with established ground truth by multiple experts. Therefore, many of the requested points related to AI device evaluation are not applicable or cannot be extracted from this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in a sans-serif font, with "FDA" in a blue square and the rest of the text in blue.

Olympus Medical Systems Corp. % Lisa Boyle Regulatory Affairs Specialist II Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, Pennsylvania 18034-0610

July 17, 2020

Re: K200542

Trade/Device Name: Visera Elite II Xenon Light Source, Telescope Ultra. Visera Elite II Video System Center, HD 3CMOS Autoclavable Camera Head, HD 3CMOS Camera Head Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN, NWB, GCJ, HET, FET Dated: June 19, 2020 Received: June 19, 2020

Dear Lisa Boyle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Neil R.P. Ogden, MS Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K200542

Device Name VISERA ELITE II Infrared Imaging System

Indications for Use (Describe)

[VISERA ELITE II Infrared Imaging System]

The VISERA ELITE II Infrared Imaging System is intended to provide real-time endoscopic visible and near infrared fluorescence imaging. The VISERA ELITE II Infrared Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels. blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the VISERA ELITE II Infrared Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

[VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S200]

This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation.

[VISERA ELITE II XENON LIGHT SOURCE OLYMPUS CLV-S200-IR]

The light source has been designed to be used with Olympus endoscopes, video system centers, light guide cables, and other ancillary equipment for endoscopic observation through visible and near-infrared fluorescence imaging with fluorescence agent.

[HD 3CMOS AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S200-XZ-EA]

The camera head has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

[HD 3CMOS CAMERA HEAD OLYMPUS CH-S200-XZ-EB]

The camera head has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

[TELESCOPE IR/ULTRA WAIR500A,WAIR530A,WAIR100A,WAIR130A]

These endoscopes are intended to be used for endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indication of transanal and transvaginal natural orifice surgery. In combination with a compatible infrared imaging system, the telescope allows for fluorescence imaging.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. A thin, horizontal, yellow line is located beneath the word. The registered trademark symbol is located to the right of the letter "S".

K200542

Date Prepared: July 16, 2020

510(k) Summary

5.1 GENERAL INFORMATION

  • 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
  • Contact Person: Lisa M. Boyle Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3676 Fax: 484-896-7128 Email: lisa.boyle@olympus.com

5.2 DEVICE IDENTIFICATION

5.2.1 VISERA ELITE II Infrared Imaging System

  • Device Name VISERA ELITE II Infrared Imaging System

  • Model Name [System components]

    • VISERA ELITE II XENON LIGHT SOURCE OLYMPUS CLV-S200-IR
    • TELESCOPE IR WAIR100A, WAIR130A
    • TELESCOPE ULTRA WAIR500A,WAIR530A
    • VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S200
    • HD 3CMOS AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S200-XZ-EA
    • HD 3CMOS CAMERA HEAD OLYMPUS CH-S200-XZ-EB
    • Indocyanine Green
  • Common Name

  • 876.1500 ■ Regulation Number

  • Endoscope and accessories ■ Regulation Name

{5}------------------------------------------------

OLYMPUS®

■ Regulatory ClassII
■ Product Code-
■ Classification PanelGeneral & Plastic Surgery
■ Manufacturing site:-
5.2.2 CLV-S200-IR■ Device NameVISERA ELITE II XENON LIGHT SOURCE
■ Model Name■ Common NameCLV-S200-IREndoscopic Video Imaging System/Component
■ Regulation Number876.1500
■ Regulation NameEndoscope and accessories
■ Regulatory ClassII
■ Product CodeOWN; Confocal optical imagingNWB; Endoscope, Accessories, Narrow Band Spectrum
■ Classification PanelGeneral & Plastic Surgery
■ Manufacturing site:Shirakawa Olympus Co., Ltd.3-1 Okamiyama, Odakura, Nishigo-mura, Nishishirakawa-gun,Fukushima 961-8061, Japan
5.2.3 WAIR100A, WAIR130A, WAIR500A, WAIR530A
■ Device NameTELESCOPE "ULTRA"TELESCOPE "IR"
■ Model NameWAIR100A, WAIR130A,WAIR500A,WAIR530A
■ Common NameLaparoscopes
■ Regulation Number876.1500884.1720
■ Regulation NameEndoscope and accessoriesGynecologic laparoscope and accessories
■ Regulatory ClassII
■ Product CodeGCJ; Laparoscope, general & plastic surgeryHET; Laparoscope, gynecologic (and accessories)
■ Classification PanelGeneral & Plastic Surgery
■ Manufacturing site:Olympus Winter & Ibe GmbHKuehnstr.61, 22045 Hamburg, Germany
5.2.4 OTV-S200
■ Device NameVISERA ELITE II VIDEO SYSTEM CENTER
■ Model NameOTV-S200
■ Common NameEndoscopic Video Imaging System/Component
■ Regulation Number876.1500
■ Regulation NameEndoscope and accessories
■ Regulatory ClassII
■ Product CodeOWN; Confocal optical imagingFET; Endoscopic Video Imaging System/Component,Gastroenterology-Urology
■ Classification PanelNWB; Endoscope, Accessories, Narrow Band SpectrumGeneral & Plastic Surgery
■ Manufacturing site:Shirakawa Olympus Co., Ltd.3-1 Okamiyama, Odakura, Nishigo-mura, Nishishirakawa-gunFukushima 961-8061, Japan
5.2.5 CH-S200-XZ-EA/ CH-S200-XZ-EB
■ Device NameHD 3CMOS AUTOCLAVABLE CAMERA HEADHD 3CMOS CAMERA HEAD
■ Model NameCH-S200-XZ-EACH-S200-XZ-EB
■ Common NameEndoscopic Video Imaging System/Component
■ Regulation Number876.1500
■ Regulation NameEndoscope and accessories
■ Regulatory ClassII
■ Product CodeOWN; Confocal optical imaging
FET; Endoscopic Video Imaging System/Component,Gastroenterology-UrologyNWB; Endoscope, Accessories, Narrow Band Spectrum
■ Classification PanelGeneral & Plastic Surgery
■ Manufacturing site:Shirakawa Olympus Co., Ltd.3-1 Okamiyama, Odakura, Nishigo-mura, Nishishirakawa-gunFukushima 961-8061, Japan

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. There is a registered trademark symbol to the right of the word.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. A registered trademark symbol is located to the right of the word.

5.3 PREDICATE DEVICE

5.3.1 CLV-S200-IR

Subject Device(Part of this submission)Predicate DevicePredicate Device510(k) No.
VISERA ELITE II XENONLIGHT SOURCE OLYMPUSCLV-S200-IRKarl Storz Endoscopic ICGImaging SystemK152583

5.3.2 WAIR100A, WAIR130A, WAIR500A, WAIR530A

Subject Device(Part of this submission)Predicate DevicePredicate Device510(k) No.
TELESCOPE IRWAIR100A, WAIR130AWA4KL100/WA4KL130K150633
TELESCOPE ULTRAWAIR500A,WAIR530AWA4KL500/WA4KL530

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image features the word "OLYMPUS" in bold, blue letters. The letters are large and fill most of the frame. A registered trademark symbol is located to the right of the word.

5.3.3 OTV-S200

Subject Device(Part of this submission)Predicate DevicePredicate Device510(k) No.
VISERA ELITE II VIDEOSYSTEM CENTEROLYMPUS OTV-S200VISERA ELITE II VIDEOSYSTEM CENTER OLYMPUSOTV-S200K190449

5.3.4 CH-S200-XZ-EA/ CH-S200-XZ-EB

Subject Device(Part of this submission)Predicate DevicePredicate Device510(k) No.
HD 3CMOSAUTOCLAVABLE CAMERAHEAD OLYMPUSCH-S200-XZ-EAHD 3CMOSAUTOCLAVABLE CAMERAHEAD OLYMPUSCH-S200-XZ-EAK190449
HD 3CMOS CAMERA HEADOLYMPUS CH-S200-XZ-EBHD 3CMOS CAMERA HEADOLYMPUS CH-S200-XZ-EBK190449

5.4 REFERENCE DEVICE 5.4.1 CLV-S200-IR

Reference DevicePredicate Device510(k) No.
PINPOINT Endoscopic Fluorescence Imaging SystemK161792
daVinci Firefly Imaging SystemK141077

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced and appear to be a sans-serif font. A registered trademark symbol is located to the right of the word.

5.4.2 WAIR100A, WAIR130A, WAIR500A, WAIR530A

Reference DeviceReference Device510(k) No.
KARL STORZ Endoscopic ICG Imaging SystemK152583
KARL STORZ Endoscopic ICG Imaging SystemK162882
KARL STORZ Endoscopic ICG Imaging SystemK171238

5.5 DEVICE DESCRIPTION

The subject devices OTV-S200, CH-S200-XZ-EA, and CH-S200-XZ-EB were cleared in 510(k) K190449. After clearances, no technological modification for the subject devices were made. Therefore, the above devices have been omitted from the device description in this pre-market notification.

5.5.1 CLV-S200-IR

This device consists of the source circuit, the control circuit, the illumination lamp, and the optical filter. The control circuit connects to the diaphragm to regulate the light intensity, the source circuit supplies the power to the illumination lamp, the operation panel and the rear panel. By switching on the illumination lamp, this device provides the light for endoscopic observation. This device regulates the endoscopic image brightness constantly from the video system center. The observation mode can be switched by the optical filter extracting the specific wavelengths.

Subject DeviceCLV-S200-IRPrimary PredicateDevice 1Reference Device 1Reference Device 2
510(k)NumberK200542K152583K161792K141077
Regulationnumber21 CFR part876.150021 CFR part876.150021 CFR part876.150021 CFR part876.1500
RegulatoryClassIIIIIIII
Product codeOWN, NWBOWNGCJ; IZINAY, GCJ, IZI

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is centered in the image. There is a registered trademark symbol to the right of the letter "S". The background is white.

Subject DeviceCLV-S200-IRPrimary PredicateDevice 1Reference Device 1Reference Device 2
ManufacturerOLYMPUSMEDICALSYSTEMS CORPKarl StorzEndoscopy-America,Inc.NovadaqTechnologies Inc.Intuitive SurgicalIncorporated
ICG agentCo packageCo packageCo packageCo package
Light sourceXenonXenonLaserLaser
3D imageNot availableNot availableNot availableAvailable
WLI*AvailableAvailableAvailableAvailable
NBI*AvailableNot availableNot availableNot available
IR*AvailableAvailableAvailableAvailable
  • White Light Imaging (WLI), Narrow band imaging (NBI), Infrared Imaging (IR)

5.5.2 WAIR100A, WAIR130A, WAIR500A, WAIR530A

The "IR" Telescopes are rigid endoscopes. An image relay system of rod lenses transmits the endoscopic image. A bundle of optical fibers transmits light from an external light source to illuminate the endoscopic image.

The "IR" Telescopes are delivered non-sterile. They are reusable and fully autoclavable. Before the first and each subsequent use of the device, it must be inspected and reprocessed according to defined reprocessing methods in the Instructions for Use.

The "IR" Telescopes are available with different directions of view to allow use for various applications in accordance with the intended use as submitted with this 510(k).

Subject devicePredicate device
WAIR100A, WAIR130A,WAIR500A, WAIR530AWA4KL100/WA4KL130WA4KL500/WA4KL530
510(k) NumberK200542K150633
Regulation number21 CFR 876.150021 CFR 884.172021 CFR 876.150021 CFR 884.1720
Regulatory ClassIIII
Product codeGCJ / HETGCJ / NMH / HET

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced. A registered trademark symbol is located to the right of the letter "S".

Subject devicePredicate device
Subject deviceWAIR100A, WAIR130A,WAIR500A, WAIR530AWA4KL100/WA4KL130WA4KL500/WA4KL530
ManufacturerOLYMPUS WINTER &IBE GMBHOLYMPUS WINTER &IBE GMBH
Laparoscoperigid endoscoperigid endoscope
Ocular coupling (y/n)yesyes
Integrated filtersIR filternone
Lens coatingNear-infrared (NIR)optimized anti-reflectivecoatingStandard anti-reflectivecoating
Autoclavabilityautoclavableautoclavable
Reprocessingmanual and automatedmanual and automated
PackagingCardboard boxNon-sterile, re-usableCardboard boxNon-sterile, re-usable

5.6 INDICATIONS FOR USE VISERA ELITE II Infrared Imaging System

The VISERA ELITE II Infrared Imaging System is intended to provide real-time endoscopic visible and near infrared fluorescence imaging. The VISERA ELITE II Infrared Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the VISERA ELITE II Infrared Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization

VISERA ELITE II XENON LIGHT SOURCE OLYMPUS CLV-S200-IR

The light source has been designed to be used with Olympus endoscopes, video system centers, light guide cables, and other ancillary equipment for endoscopic observation through visible and near-infrared fluorescence imaging with fluorescence agent.

{12}------------------------------------------------

TELESCOPE IR WAIR100A, WAIR130A TELESCOPE ULTRA WAIR500A,WAIR530A

These endoscopes are intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The device is also indicated for visualization of transanal and transvaginal natural orifice surgery. In combination with a compatible infrared imaging system, the telescope allows for fluorescence imaging.

VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S200

This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis.treatment, and video observation.

HD 3CMOS AUTOCLAVABLE CAMERA HEAD OLYMPUS CH-S200-XZ-EA HD 3CMOS CAMERA HEAD OLYMPUS CH-S200-XZ-EB

The camera head has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.

5.7 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE

5.7.1 CLV-S200-IR

The CLV-S200-IR has the same technological characteristics and design as the predicate device except for the following new features:

Specified IR color modes –

5.7.2 WAIR100A, WAIR130A, WAIR500A, WAIR530A

The WAIR100A, WAIR130A, WAIR500A, and WAIR530A have the same technological characteristics and design as the predicate device except for the following new features:

An additional filter optimized for fluorescence imaging has been added within the ocular

  • । Coating of internal lenses for optimization of infrared imaging
    All other technological characteristics of both the subject and predicate devices are identical. Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

{13}------------------------------------------------

5.8 PERFORMANCE DATA

The following performance data is provided in support of the substantial equivalence determination.

1) Reprocessing validation testing

Reprocessing instruction and reprocessing method validation testing for the WAIR100A/130A/500A/530A was conducted and documentation is provided as recommended by Guidance for Industry and Food and Drug Administration Staff. "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".

2) Biocompatibility testing

Biocompatibility testing for the WAIR100A/130A/530A was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

  • Cytotoxicity Study Using the Colony Assay ।
  • Intracutaneous Study in Rabbits ।
  • Guinea Pig Maximization Sensitization Test ।
  • Systemic Toxicity Study in Mice i
  • Material mediated pyrogenicity ।

3) Software verification and validation testing

Software verification and validation testing for the OTV-S200 and CLV-S200-IR was conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

4) Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing was conducted. The subject devices comply with the ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.

5) Performance testing - Bench

Bench testing as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

[VISERA ELITE II Infrared Imaging System]

  • Verification for photobiological safety of the illumination light

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. A registered trademark symbol is located to the right of the word. The background is white, providing a strong contrast to the blue letters.

  • [WAIR100A/130A/500A/530A] -DOV -FOV -MTF -Distortion -Ghost Image -Internal reflections -Spectral transmission of imaging system -Expected Service Life -Transport Drop -Design Validation / Usability -Illumination System performance Data -Comparison of Optical Properties of Subject Device and predicate Device

6) Performance testing - Animal

The animal study using canines and swines was performed by taking videos that validate the performance of subject device in simulated body environment. The videos were evaluated by an independent Health Care Professional to demonstrate substantial equivalence in terms of IR observation.

7) Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

8) Risk analysis

A Risk analysis for the CLV-S200-IR and WAIR100A/130A/500A/530A was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

The following voluntary standards have been applied to the CLV-S200-IR and WAIR100A/130A/500A/530A respectively;

CLV-S200-IR

  • AAMI / ANSI ES 60601-1:2005 and A1:2012, C1:2009 and A2:2010 -
  • IEC 60601-1-2:2014 -
  • IEC 60601-2-18:2009 -
  • -ISO 14971:2007
  • -AAMI/ANSI/IEC 62304:2006

WAIR100A/130A/500A/530A

  • AAMI / ANSI ES 60601-1:2005 and A1:2012, C1:2009 and A2:2010 -
  • IEC 60601-2-18:2009 -
  • IEC 60601-1-6:2013 -

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly condensed and have a sans-serif typeface. A thin, gold line is visible beneath the word, adding a subtle accent to the logo.

  • ISO 8600-1:2015 -
  • ISO 8600-3:1997 + AM1(2003) -
  • ISO 8600-4:2014 -
  • ISO 8600-5:2005 -
  • ISO 8600-6:2005 -
  • ISO 10993-1:2018 -
  • AAMI/ANSI/ISO 10993-5:2009 -
  • ISO 10993-12:2012 -
  • ISO 14971:2007 -
  • AAMI/ANSI/ISO 14937:2009 -
  • -ISO 17665:2006

5.9 CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the OTV-S200, CH-S200-XZ-EA, CH-S200-XZ-EB, CLV-S200-IR and WAIR100A/130A/500A/530A raise no new issue of safety and effectiveness. They are substantially equivalent to the predicate devices in safety, efficacy, and performance.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.