K Number
K152583

Validate with FDA (Live)

Date Cleared
2016-05-31

(264 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KARL STORZ Endoscopic ICG Imaging System is intended to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The KARL STORZ ICG Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Device Description

The KARL STORZ Endoscopic ICG Imaging System is used to provide real-time high definition (HD) endoscopic video images of visible (VIS) and near-infrared (NIR) indocyanine green (ICG) dye fluorescence during minimally invasive surgery. The system components are rigid ICG Endoscopes for VIS and NIR illumination and imaging, a light source with foot switch for emission of VIS and NIR illumination, a 3 CCD (charge coupled device) color video camera head capable of capturing both VIS and NIR imaging, and KARL STORZ Indocyanine Green (ICG Kit). Additional accessories used with the KARL STORZ Endoscopic ICG Imaging System include a standard fiber-optic light cable for transmission of VIS and NIR light and the Image 1 SPIES Camera Control Unit (CCU) cleared in K131953. The KARL STORZ Endoscopic ICG Imaging System can be used with any medical grade HD monitor with a DVI-D or 3G-SDI input. The KARL STORZ Endoscopic ICG Imaging System is a Class II device under 21 CFR 876.1500.

AI/ML Overview

This document is a 510(k) summary for the KARL STORZ Endoscopic ICG Imaging System (K152583). It explicitly states that clinical performance data was not required to demonstrate substantial equivalence to the predicate device. Therefore, no studies were conducted to prove the device meets specific acceptance criteria based on human performance. The following information is based on the provided text, and many sections will indicate that the information is not available as no clinical study was performed.

1. A table of acceptance criteria and the reported device performance

Since no clinical study was performed to establish performance against acceptance criteria in a clinical setting, this table cannot be populated with device performance. The non-clinical performance data focused on functional and safety aspects.

Acceptance Criteria (Clinical)Reported Device Performance (Clinical)
Not established/applicable (No clinical study performed)Not applicable (No clinical study performed)

The document mentions non-clinical performance data:

  • "The KARL STORZ Endoscopic ICG Imaging System has been successfully tested for its functions and performance, including verification that the spectral characteristics of the ICG system illumination light source, light transmission system and endoscope enable a selective visualization of the ICG fluorescence signal as detected by the camera system."
  • "Furthermore, a GLP animal study was successfully performed by the NAMSA testing facility to evaluate the performance of the KARL STORZ Endoscopic ICG Imaging System in the porcine minimally-invasive laparoscopy model."
  • Safety testing: IEC 60601-1 and 60601-2-18 (electrical safety), IEC 60601-1-2 (electromagnetic compatibility), ISO 10993 (biocompatibility).
  • Additional validations for system software, manual cleaning method, and sterilization process.

These are functional and safety-related performance measures, not clinical efficacy or diagnostic accuracy metrics typically found in acceptance criteria for AI/diagnostic devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable, as no clinical test set was used for human performance evaluation. A GLP animal study was performed in a porcine model. The specific sample size for the animal study is not provided, nor is the country of origin.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable, as no clinical test set requiring expert ground truth was used.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as no clinical test set was used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. No MRMC study was done, as clinical testing was not required and the device is an imaging system, not explicitly an AI/diagnostic algorithm that enhances human reader performance in a comparative study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device itself is an imaging system. It performs a standalone function of capturing and displaying images. However, the FDA states, "The device is not intended for standalone use for biliary duct visualization," meaning it requires a human surgeon's interpretation in conjunction with white light imaging and potentially intraoperative cholangiography. No 'algorithm-only' performance was assessed independently of the visual output for a human.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical functional tests, the ground truth would have been established by engineering specifications and physical measurements. For the GLP animal study, the "performance" evaluation would likely involve visual assessment by qualified personnel regarding the visualization of ICG fluorescence in the porcine model. There is no mention of pathology or outcomes data as ground truth.

8. The sample size for the training set

Not applicable. This device is an imaging system, not an AI/ML algorithm that requires a training set in the conventional sense.

9. How the ground truth for the training set was established

Not applicable, as there was no training set for an AI/ML algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 31, 2016

Karl Storz Endoscopy-America, Inc. Winkie Wong Senior Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

Re: K152583 Trade/Device Name: Karl Storz Endoscopic ICG Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN Dated: April 13, 2016 Received: April 15, 2016

Dear Winkie Wong,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152583

Device Name

KARL STORZ Endoscopic ICG Imaging System

Indications for Use (Describe)

The KARL STORZ Endoscopic ICG Imaging System is intended to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The KARL STORZ ICG Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are in smaller, blue letters.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Applicant:Karl Storz Endoscopy-America, Inc2151 E. Grand AvenueEl Segundo, CA 90245
Contact:Winkie WongSenior Regulatory Affairs Specialist
Phone: (424) 218-8379Fax: (424) 218-8519
Date of Preparation:May 18th, 2016
Device Identification:Trade Name: KARL STORZ Endoscopic ICG Imaging SystemCommon Name: Endoscopic Video Imaging SystemClassification Name: Endoscope and accessories
Product Code:OWN
Regulation:21 CFR part 876.1500 (Endoscope and Accessories)
Predicate Device(s):The KARL STORZ Endoscopic ICG Imaging System issubstantially equivalent to the Intuitive Surgical® da Vinci®Fluorescence Imaging Vision System (K124031), marketedby Intuitive Surgical, Inc.
Indications For Use:The KARL STORZ Endoscopic ICG Imaging System isintended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The KARL STORZ ICGImaging System enables surgeons to perform minimallyinvasive surgery using standard endoscopic visible light aswell as visual assessment of vessels, blood flow and relatedtissue perfusion, and at least one of the major extra-hepaticbile ducts (cystic duct, common bile duct and commonhepatic duct), using near-infrared imaging.
Fluorescence imaging of biliary ducts with the KARL STORZEndoscopic ICG Imaging System is intended for use withstandard of care white light and, when indicated,intraoperative cholangiography. The device is not intendedfor standalone use for biliary duct visualization.

{4}------------------------------------------------

Device Description:The KARL STORZ Endoscopic ICG Imaging System is used toprovide real-time high definition (HD) endoscopic videoimages of visible (VIS) and near-infrared (NIR) indocyaninegreen (ICG) dye fluorescence during minimally invasivesurgery. The system components are rigid ICG Endoscopesfor VIS and NIR illumination and imaging, a light source withfoot switch for emission of VIS and NIR illumination, a 3 CCD(charge coupled device) color video camera head capable ofcapturing both VIS and NIR imaging, and KARL STORZIndocyanine Green (ICG Kit). Additional accessories usedwith the KARL STORZ Endoscopic ICG Imaging Systeminclude a standard fiber-optic light cable for transmission ofVIS and NIR light and the Image 1 SPIES Camera ControlUnit (CCU) cleared in K131953. The KARL STORZ EndoscopicICG Imaging System can be used with any medical grade HDmonitor with a DVI-D or 3G-SDI input. The KARL STORZEndoscopic ICG Imaging System is a Class II device under21 CFR 876.1500.
Technological Comparison Summary:The KARL STORZ Endoscopic ICG Imaging System issubstantially equivalent to the Intuitive Surgical® da Vinci®Fluorescence Imaging Vision System (K124031) in terms ofits indications for use, design technology and performancespecification.The differences between the subject and predicate deviceare:The subject device uses a Filtered Xenon Light sourcewhereas the Predicate (da Vinci) uses a Laser Diodeduring NIR Illumination Mode. The subject device projects 2D fluorescence image ona dichromatic background whereas the predicatedevice (da Vinci) projects 3D fluorescence image on amonochromatic background in NIR Illumination Mode. Bench and comparative testing were used to demonstratesubstantial equivalence to the predicate Intuitive Surgical®da Vinci® Fluorescence Imaging Vision System, K124031.Therefore, the differences between the subject device andthe predicate device do not raise any new issues of safetyand effectiveness.
Non-ClinicalPerformance Data:The KARL STORZ Endoscopic ICG Imaging System has beensuccessfully tested for its functions and performance,including verification that the spectral characteristics of theICG system illumination light source, light transmissionsystem and endoscope enable a selective visualization of theICG fluorescence signal as detected by the camera system.Furthermore, a GLP animal study was successfully performedby the NAMSA testing facility to evaluate the performance ofthe KARL STORZ Endoscopic ICG Imaging System in the

{5}------------------------------------------------

porcine minimally-invasive laparoscopy model. Safety testing was performed including electrical safety IEC 60601-1 and 60601-2-18, electromagnetic compatibility per IEC 60601-1-2, and biocompatibility of the patient contacting materials per ISO 10993. Additional validations were conducted for system software, the manual cleaning method, and the sterilization process.
Clinical Performance Data:Clinical testing was not required to demonstrate substantial equivalence to the predicate device.
Conclusion:The KARL STORZ Endoscopic ICG Imaging System is substantially equivalent to the predicate Intuitive Surgical® da Vinci® Fluorescence Imaging Vision System (K124031) with regards to device design, materials, operational principles, and performance characteristics. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as or better than the legally marketed device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.