K Number
K171238

Validate with FDA (Live)

Date Cleared
2017-05-25

(28 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.

The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.

Device Description

The KARL STORZ ICG Imaging System is used to provide real-time high definition (HD) endoscopic or telescope images of visible (VIS) and near-infrared (NIR) indocyanine green (ICG) dye fluorescence during minimally invasive surgery as well as plastic, micro- and reconstructive surgical procedures.

The overall system includes a 5mm &10mm HOPKINS ICG/NIR Endoscope for use in minimally invasive procedures and a VITOM II ICG/NIR Telescope for use in plastic, micro- and reconstructive surgical procedures for VIS and NIR illumination and imaging, a light source with foot switch for emission of VIS and NIR illumination, a color video camera head capable of capturing both VIS and NIR imaging, and KARL STORZ ICG Kit. Additional accessories used with the KARL STORZ ICG Imaging System include two standard fiber-optic light cables for transmission of VIS and NIR light and the Image1S Camera Control Unit (CCU).

The KARL STORZ ICG Imaging System can be used with any medical grade HD monitor with a DVI-D or 3G-SDI input.

AI/ML Overview

The provided document is a 510(k) summary for the KARL STORZ ICG Imaging System. It describes the device, its intended use, and argues for its substantial equivalence to a predicate device. However, this document does not contain information about acceptance criteria or a study that proves the device meets specific performance acceptance criteria in the context of diagnostic accuracy or a similar measure typically associated with AI/algorithm performance.

The document states:

  • "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence of the modifications."
  • "Additional bench testing was performed to ensure the device met its design specifications."
  • "The bench testing performed verified and validated that the KARL STORZ ICG Imaging System has met all its design specification and is substantially equivalent to its predicate devices."

This indicates that the device's approval was based on demonstrating safety, effectiveness, and substantial equivalence to existing devices through non-clinical (bench) testing, primarily focusing on aspects like electrical safety, electromagnetic compatibility, biocompatibility, and mechanical properties, not on diagnostic performance against ground truth in a clinical setting.

Therefore, I cannot provide the requested information for acceptance criteria and a study proving device performance in the context of diagnostic accuracy, as this type of information is not present in the provided text.

Based on the nature of the device (an imaging system for real-time visible and near-infrared fluorescence imaging) and the approval process described, it is highly probable that the "acceptance criteria" referred to in the document pertain to technical specifications, safety standards, and functional performance benchmarks rather than clinical diagnostic accuracy metrics (like sensitivity, specificity, or AUC).

If this were a typical AI/algorithm-based diagnostic device, the requested information would be crucial. However, for a medical imaging system like this, the focus is often on its ability to produce clear, reliable images and its safety for patient use, with appropriate regulatory standards and bench testing serving as the "proof."

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 25, 2017

KARL STORZ Endoscopy- America, Inc. Winkie Wong Senior Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

Re: K171238

Trade/Device Name: KARL STORZ ICG Imaging System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OWN Dated: April 26, 2017 Received: April 27, 2017

Dear Winkie Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171238

Device Name KARL STORZ ICG Imaging System

Indications for Use (Describe)

The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.

The KARL STORZ Endoscopic ICG System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

The KARL STORZ VITOM II ICG System is intended for capturing fluorescent mages for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive surgical procedures. The VITOM II ICG System is intended to provide a magnified view of the surgical field in standard white light.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for STORZ. The logo is in blue and consists of the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.

510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Submitter:KARL STORZ Endoscopy-America, Inc.
2151 E. Grand AvenueEI Segundo, CA 90245
Contact:Winkie WongSenior Regulatory Affairs SpecialistPhone: (424) 218-8379Fax: (424) 218-8519
Date ofPreparation:April 26, 2017
Type of 510(k)Submission:Special
DeviceIdentification:Trade Name: KARL STORZ ICG Imaging System
Classification Name: Confocal Optical Imaging
Regulatory Class:II
Product Code:OWN
Regulation:21 CFR part 876.1500 (Endoscope and Accessories)
PredicateDevice(s):Primary Predicate Device: KARL STORZ Endoscopic ICG System(K62882)
These predicate devices have not been subject to a design-relatedrecall.
Device Description:The KARL STORZ ICG Imaging System is used to provide real-timehigh definition (HD) endoscopic or telescope images of visible(VIS) and near-infrared (NIR) indocyanine green (ICG) dyefluorescence during minimally invasive surgery as well as plastic,micro- and reconstructive surgical procedures.
The overall system includes a 5mm &10mm HOPKINS ICG/NIREndoscope for use in minimally invasive procedures and a VITOMII ICG/NIR Telescope for use in plastic, micro- and reconstructivesurgical procedures for VIS and NIR illumination and imaging, alight source with foot switch for emission of VIS and NIRillumination, a color video camera head capable of capturing both
VIS and NIR imaging, and KARL STORZ ICG Kit. Additionalaccessories used with the KARL STORZ ICG Imaging Systeminclude two standard fiber-optic light cables for transmission ofVIS and NIR light and the Image1S Camera Control Unit (CCU).
The KARL STORZ ICG Imaging System can be used with anymedical grade HD monitor with a DVI-D or 3G-SDI input.
Intended Use:The KARL STORZ ICG Imaging System is intended to provide real-time visible and near-infrared fluorescence imaging.
Indications ForUse:The KARL STORZ ICG Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescence imaging.
The KARL STORZ Endoscopic ICG System enables surgeons toperform minimally invasive surgery using standard endoscopicvisible light as well as visual assessment of vessels, blood flow andrelated tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and commonhepatic duct), using near infrared imaging. Fluorescence imagingof biliary ducts with the KARL STORZ Endoscopic ICG System isintended for use with standard of care white light and, whenindicated, intraoperative cholangiography. The device is notintended for standalone use for biliary duct visualization.
The KARL STORZ VITOM II ICG System is intended for capturingand viewing fluorescent images for the visual assessment of bloodflow, as an adjunctive method for the evaluation of tissueperfusion, and related tissue-transfer circulation in tissue and freeflaps used in plastic, micro- and reconstructive surgicalprocedures. The VITOM II ICG System is intended to provide amagnified view of the surgical field in standard white light.
TechnologicalCharacteristics:The KARL STORZ ICG Imaging System is modification of andsubstantially equivalent to the KARL STORZ ICG Imaging System(K162882) in terms of its indications for use, design technology,and performance specifications.
The main difference between the subject and the primarypredicate device is the addition of a 5mm ICG Endoscope to theKARL STORZ Endoscopic ICG System to be used in minimalinvasive surgery.
Bench and comparative testing were used to demonstrate
substantial equivalence to the primary predicate device.
Therefore, the differences between the subject device and thepredicate device do not raise any new issues of safety andeffectiveness.
Non-ClinicalPerformance Data:There are no performance standards or special controlsdeveloped under Section 514 of the FD&C Act for endoscopes.However, the KARL STORZ ICG Imaging System follows the FDArecognized consensus standards and is tested according to thefollowing standard:• IEC 60601-1• IEC 60601-1-2• IEC 60601-2-18• ISO 8600-1• ISO 8600-3• ISO 8600-4• ISO 8600-5• ISO 8600-6• ISO 14971• ISO 10993Additional bench testing was performed to ensure the device metits design specifications.The bench testing performed verified and validated that the KARLSTORZ ICG Imaging System has met all its design specification andis substantially equivalent to its predicate devices.
ClinicalPerformance Data:Clinical testing was not required to demonstrate the substantialequivalence to the predicate devices. Non-clinical bench testingwas sufficient to assess safety and effectiveness and to establishthe substantial equivalence of the modifications.
Conclusion:The conclusions drawn from the non-clinical tests such as thecleaning and sterilization summary, the risk evaluation on themodification of the system and biological evaluation summarydemonstrated that the subject device is as safe as and as effectiveas the predicate device. As such, we concluded that the substantialequivalence of the subject and the predicate devices has been met,and the differences between the subject and the predicate devicesdo not raise new or different questions of safety and effectiveness.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and consists of the word "STORZ" in large, bold letters above the words "KARL STORZ - ENDOSKOPE" in smaller letters. The "O" in STORZ is stylized as a circle with a smaller circle inside it.

{5}------------------------------------------------

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.