K Number
K193425

Validate with FDA (Live)

Device Name
Pre-Milled Blank
Date Cleared
2020-03-09

(90 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw.

ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:

ARUM Pre-Milled BlankImplant PlatformcompatibilityRestorative Platformdiameter [mm]Implant Bodydiameter [mm]AbutmentScrew
Ø10 mmØ14 mm
CIHE037CIHE038NobelActive NP3.53.5CSTO001
CIHE039CIHE040NobelActive RP3.94.3/5.0CSTO002
CIHE135CIHE136NobelActive WP5.15.5CSTO002

All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling center for manufacture.

Device Description

ARUM Dentistry Co., Ltd.'s Pre-Milled Blank abutments are a system of dental implant abutments which have an implant /abutment interface design compatible with the Reference device OEM Nobel Biocare NobelActive implant system. Each Subject device implant abutment has a pre-manufactured implant interface connection interface with a customizable cylindrical section above the implant/abutment interface. The compatible implant body diameters range from 3.5 to 5.5 mm, with restorative interface diameters ranging from 3.0 to 5.1 mm (NP, RP and WP diameters).

The Subject device abutments and corresponding abutment screws are all pre-manufactured from Ti-6Al-4V ELI (Grade 23) titanium conforming to ASTM F136 and provided non-sterile to the user.

The Subject device abutments are in two customizable-section diameters for each restorative interface. The cylindrical section of the Subject device abutments are intended to be customized by means of CAD/CAM technology to provide basis or support for single or multiple tooth prosthetic restorations. All digitally designed custom abutments from titanium blank abutments are to be sent to an ARUM Dentistry Co., Ltd.-validated milling center for manufacture.

AI/ML Overview

This document is an FDA 510(k) premarket notification for a medical device called "Pre-Milled Blank" (K193425). It details the device's administrative information, classification, predicate device, description, indications for use, and a comparison to the predicate device to demonstrate substantial equivalence.

While the document thoroughly addresses the substantial equivalence of the "Pre-Milled Blank" device to a predicate device based on indications for use, technological characteristics, and non-clinical performance data, it does not contain information related to an AI/ML-based medical device performance study.

Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, ground truth, or training set specifics, as these are evaluation criteria typically applied to AI/ML devices.

The "Pre-Milled Blank" is a physical dental implant abutment, and its approval is based on demonstrating equivalence to an existing device through non-clinical performance data (sterilization, biocompatibility, mechanical testing) rather than through AI/ML model performance metrics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 9, 2020

ARUM Dentistry Co. Ltd % Chris Brown Manager Aclivi. LLC 6455 Farley Road Pinckney, Michigan 48169

Re: K193425

Trade/Device Name: Pre-Milled Blank Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 27, 2019 Received: December 10, 2019

Dear Chris Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K193425

Device Name Pre-Milled Blank

Indications for Use (Describe)

ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw.

ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:

Compatibility Table
ARUM Pre-Milled BlankImplant PlatformcompatibilityRestorative Platformdiameter [mm]Implant Bodydiameter [mm]AbutmentScrew
Ø10 mmØ14 mm
CIHE037CIHE038NobelActive NP3.53.5CSTO001
CIHE039CIHE040NobelActive RP3.94.3/5.0CSTO002
CIHE135CIHE136NobelActive WP5.15.5

All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling center for manufacture.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K193425 - 510(k) Summary ARUM Dentistry Co., Ltd. Pre-Milled Blank 3/9/2020

5.1 ADMINISTRATIVE INFORMATION

Manufacturer Name ARUM Dentistry Co., Ltd. 1 Building, 44, Techno 8-ro, Yuseong-gu, Daejeon, 34028 Republic of Korea

Phone: +82-42-721-3644 Fax: +82-42-721-3645

Official ContactHeung-Yeon Hwang, Quality Manager
Email:hwanghy@arumdentistry.com

5.2 DEVICE NAME AND CLASSIFICATION

Trade/Proprietary Name:Pre-Milled Blank
Common Name:Abutment, Implant, Dental, Endosseous
Classification Name:Endosseous dental implant abutment
Classification Regulation:21 CFR 872.3630
Device Class:Class II
Product Code:NHA
Review Panel:Dental Products Panel
Reviewing Branch:Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)Dental Devices (DHT1B)

5.3 PREDICATE DEVICE INFORMATION

The devices within this submission are substantially equivalent in indications, intended use and design principles to the following primary predicate and reference devices:

Primary Predicate
510(k)Device NameCompany Name
K190299Elos Accurate Customized AbutmentElos Medtech Pinol A/S
Reference Device
510(k)Device NameCompany Name
K071370NobelActive Internal Connection ImplantNobel Biocare

Primary Predicate

5.4 DEVICE DESCRIPTION

ARUM Dentistry Co., Ltd.'s Pre-Milled Blank abutments are a system of dental implant abutments which have an implant /abutment interface design compatible with the Reference device OEM Nobel Biocare NobelActive implant system. Each Subject device implant abutment has a pre-manufactured implant interface connection interface with a customizable cylindrical section above the implant/abutment interface. The compatible implant body diameters range from 3.5 to 5.5 mm, with restorative interface diameters ranging from 3.0 to 5.1 mm (NP, RP and WP diameters).

The Subject device abutments and corresponding abutment screws are all pre-manufactured from Ti-6Al-4V ELI (Grade 23) titanium conforming to ASTM F136 and provided non-sterile to the user.

The Subject device abutments are in two customizable-section diameters for each restorative interface 510k Summary 5- 1

{4}------------------------------------------------

The cylindrical section of the Subject device abutments are intended to be customized by means of CAD/CAM technology to provide basis or support for single or multiple tooth prosthetic restorations. All digitally designed custom abutments from titanium blank abutments are to be sent to an ARUM Dentistry Co., Ltd.-validated milling center for manufacture.

5.5 INDICATIONS FOR USE

ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw.

ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:

ARUM Pre-Milled BlankImplant PlatformcompatibilityRestorative Platformdiameter [mm]Implant Bodydiameter [mm]AbutmentScrew
Ø10 mmØ14 mm
CIHE037CIHE038NobelActive NP3.53.5CSTO001
CIHE039CIHE040NobelActive RP3.94.3/5.0CSTO002
CIHE135CIHE136NobelActive WP5.15.5CSTO002

aatihility Tahla

All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling center for manufacture.

EQUIVALENCE TO MARKETED DEVICE 5.6

The Subject device is substantially equivalent to the predicate device with respect to Indications for Use and technological principles. The Comparison tables below compare the Indications for Use and Technological Characteristics of the Subject and Predicate devices.

DeviceIndications for Use Statement
Subject DevicePre-Milled BlankARUM DentistryCo., LTDARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to providesupport for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained singlerestorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will beattached to a dental implant using the included ARUM Dentistry prosthetic screw.
ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:Compatibility Table
ARUM Pre-Milled BlankImplant PlatformcompatibilityRestorative Platformdiameter [mm]Implant Bodydiameter [mm]Abutment Screw
Ø10 mmØ14 mm
CIHE037CIHE038NobelActive NP3.53.5CSTO001
CIHE039CIHE040NobelActive RP3.94.3/5.0CSTO002
CIHE135CIHE136NobelActive WP5.15.5
All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated millingcenter for manufacture.
PrimaryPredicateElos AccurateCustomizedAbutment(K190299)The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis forsingle or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dentalimplant using the included Elos Prosthetic screw.
The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1:Table 1.

Comparison of Indications for Use Statements

{5}------------------------------------------------

DeviceIndications for Use Statement
Elos MedtechPinol A/SElos AccurateCustomizedAbutment - Model TypePlatform compatibilityPlatformdiameter[mm]Implant Bodydiameter[mm]
AB-NBR35Nobel Replace NP3.53.5
AB-NBA30Nobel CC 3.03.03.0
AB-NBA43Nobel CC RP3.94.3 & 4.5
AB-NBA60Nobel CC WP5.15.5
AB-SBO33Straumann Bone Level NC3.33.3
AB-SBO41Straumann Bone Level RC4.1 & 4.84.1 & 4.8

Except for the device name, reference/model numbers, the list of specific compatible platforms, and company approving the milling facility the Subject and Predicate devices have nearly identical Indications for Use. These differences do not change the intended use of the Subject and Predicate devices to provide support for single or multi-unit prosthetic restorations. The slight differences in the wording "attachment to dental implants in order to provide support for" and "attaching to dental implants in order to provide basis for" are simply different ways to describe the same intended use. The inclusion of the language "screw-retained single restoration and cement-retained single or multi-unit restoration" specifies sub-sets of the types of single and multi-unit restorations included in the primary predicate. Identifying the specific types of single and multi-unit restorations does not change the intended use of the device.

Design ParameterSubject DevicePrimary PredicateEquivalence Discussion
Pre-Milled BlanksARUM Dentistry Co., LtdElos Accurate Customized Abutment(K190299)
FDA Classification &Product CodeEndosseous dental implant abutment,21 CFR 872.360, NHAEndosseous dental implant abutment,21 CFR 872.360, NHASubstantially Equivalent
MaterialsTitanium Ti-6Al-4V ELITitanium Ti-6Al-4V ELISubstantially Equivalent
Surface FinishNon-coatedNon-coatedOptional DLC coating on screwsSubstantially EquivalentThe optional coating does not changethe intended use of the device.
Implant InterfaceIndexedIndexedSubstantially Equivalent
Connection typeConicalFlat top and conicalSubstantially EquivalentDifferent connection interfaces donot change the intended use of thedevice.
Abutment Interfacediameter3.5-5.1 mm3.0-6.0 mmSubstantially EquivalentSlight differences in interfacediameters do not change theintended use of the device.
Prosthesis AttachmentCement-retainedCement-retainedSubstantially Equivalent
RestorationSingle-unit & Multi-UnitSingle-unit & Multi-UnitSubstantially Equivalent
Mode of operationProvide support for single and multi-unit prostheses.Provide support for single and multi-unit prostheses.Substantially Equivalent
SterilityProvided non-sterileProvided non-sterileSubstantially Equivalent
Sterilization methodSteam SterilizationSteam sterilizationSubstantially Equivalent
Abutment DesignParametersMinimum wall thickness: 0.5 mmMaximum post height: 13 mmMaximum angulation: 30°Maximum diameter: 14 mmMinimum post height: 4 mmMinimum gingiva height: 0.5 mmMaximum gingiva height: 4 mmMinimum wall thickness: 0.4 mmMaximum post height: 13 mmMaximum angulation: 30°Maximum diameter: 12 mmMinimum post height: 4 mmMinimum gingiva height: 0.5 mmMaximum gingiva height: 5mmSubstantially EquivalentA slightly larger minimum wallthickness, and slightly smallermaximum gingival height of theSubject device does not change theintended use of the device.

Comparison of Technological Characteristics

Minor differences in abutment interface geometry due to compatible implant systems or restorative interface diameters do not introduce new risk nor change the intended use of the device to provide support for single and multi-unit prosthetic restorations.

{6}------------------------------------------------

Overall, the Technological Characteristics, mode of operation and materials of the Subject device are substantially equivalent to that of the Predicate device.

5.7 NON-CLINICAL PERFORMANCE DATA

Non-clinical performance data submitted to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937; biocompatibility according to ISO 10993-5 and ANSI/AAMI ST72; reverse engineering of the OEM implant bodies, OEM abutments and OEM abutment screws to confirm compatibility; static and fatigue testing according to ISO 14801. No clinical data is included in this submission.

5.8 CONCLUSION

Slight variations in the wording of the Indications for the Subject and Predicate devices do not change the intended use of the devices to provide support for single or multi-unit prosthetic restorations.

Slight differences in the compatible implant system abutments available in the Subject and Predicate device systems does not change the intended use of the devices to provide support for single or multi-unit prosthetic restorations.

The slight differences in the Technological Characteristics of the Subject and Predicate devices do not change the intended use of the devices to provide support for single or multi-unit prosthetic restorations or introduce new risk or concerns. Differences in Subject device abutment design parameters were validated with respect to intended use through Performance testing.

Overall, the Pre-Milled Blank Subject device is substantially equivalent to the Predicate device.

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)