K Number
K190900

Validate with FDA (Live)

Device Name
Biograph Vision
Date Cleared
2019-05-06

(28 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Siemens Biograph Vision PET/CT systems are combined X-Ray Computed Tomography (CT) and Position Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray wansmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.

These systems are intended to be utilized by approxiately trained health care professionals to aid in detecting. diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

This CT system can be used for low dose lung cancer screening in high risk populations * * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lune Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Device Description

The Biograph Vision systems are combined multi-slice X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners. These systems are designed for whole body oncology, neurology and cardiology examinations.

The Biograph Vision systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

Biograph Vision software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

The Biograph Vision PET/CT, which is the subject of this application, is substantially equivalent to the commercially available Biograph Vision PET/CT (K180811). The key differences between the current Biograph Vision (predicate device) and the Biograph Vision PET/CT that is the subject of this application is the addition of two new models (Biograph Vision 450 and Biograph Vision 450 Edge) with a smaller axial PET Field of View, as well as additional software modifications to daily / weekly Quality Control that is applicable to all models of the Biograph Vision.

AI/ML Overview

The provided text describes the acceptance criteria and performance of the Siemens Biograph Vision PET/CT system, which is a combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanner. This is a 510(k) premarket notification for a medical device; therefore, the "study" proving acceptance is primarily a technical performance evaluation against established industry standards and comparison to a predicate device, rather than a clinical trial with human subjects for AI assistance.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The performance testing for the PET component was conducted in accordance with NEMA NU2:2018.

Device: Biograph Vision PET/CT (specifically models 450/450 Edge and 600/650 Edge are mentioned, performance results are generic for both where ranges are the same, or specific to each model where they differ).

Performance CriteriaAcceptance Criteria (New Models: 450/450 Edge)Reported Device Performance (New Models: 450/450 Edge)Acceptance Criteria (Predicate: 600/650 Edge)Reported Device Performance (Predicate: 600/650 Edge)
Resolution - Full Size
Transverse Resolution FWHM @ 1 cm≤ 4.0 mmPass≤ 4.0 mmPass
Transverse Resolution FWHM @ 10 cm≤ 4.8 mmPass≤ 4.8 mmPass
Transverse Resolution FWHM @ 20 cm≤ 5.2 mmPass≤ 5.2 mmPass
Axial Resolution FWHM @ 1 cm≤ 4.3 mmPass≤ 4.3 mmPass
Axial Resolution FWHM @ 10 cm≤ 5.4 mmPass≤ 5.4 mmPass
Axial Resolution FWHM @ 20 cm≤ 5.4 mmPass≤ 5.4 mmPass
Count Rate / Scatter / Sensitivity / etc.
Sensitivity @435 keV LLD≥ 8.0 cps/kBqPass≥ 15.0 cps/kBqPass
Count Rate peak NECR≥140 kcps @ ≤ 32 kBq/ccPass≥250 kcps @ ≤ 32 kBq/ccPass
Count Rate peak trues≥600 kcps @ ≤ 56 kBq/ccPass≥1100 kcps @ ≤ 56 kBq/ccPass
Scatter Fraction at peak NECR≤43%Pass≤43%Pass
Mean bias (%) at peak NEC≤ 6%Pass≤ 6%Pass
Co-Registration Accuracy≤ 5 mmPass≤ 5 mmPass
Time of Flight Resolution at peak NEC≤249 psPass≤249 psPass
Image Quality (4 to 1) - (% Contrast / Background Variability)
10mm sphere≥ 55% / ≤ 10%Pass≥ 55% / ≤ 10%Pass
13mm sphere≥ 60% / ≤ 9%Pass≥ 60% / ≤ 9%Pass
17mm sphere≥ 65% / ≤ 8%Pass≥ 65% / ≤ 8%Pass
22mm sphere≥ 70% / ≤ 7%Pass≥ 70% / ≤ 7%Pass
28mm sphere≥ 75% / ≤ 6%Pass≥ 75% / ≤ 6%Pass
37mm sphere> 80% / < 5%Pass> 80% / < 5%Pass

2. Sample Size Used for the Test Set and Data Provenance

The document describes performance testing in accordance with NEMA NU2:2018. This standard typically involves phantom-based measurements, not human subject data. Therefore, there's no "sample size" in the sense of patient data. The provenance is internal testing performed by Siemens Medical Solutions USA, Inc. The testing is retrospective in the sense that it evaluates the manufactured device against pre-defined standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Ground truth for these technical performance tests is established by physical measurements using phantoms and adherence to the NEMA NU2:2018 standard. Experts, in terms of human readers interpreting data, are not typically involved in establishing the "ground truth" for this type of technical performance testing. The "ground truth" is defined by the physical properties of the phantoms and the established physical parameters of the imaging system.

4. Adjudication Method for the Test Set

Not applicable. This is a technical performance test using phantoms and established measurement procedures, not a clinical study requiring adjudicated interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done for this submission. The submission focuses on the technical performance specifications of the PET/CT device itself and its substantial equivalence to a predicate device, not on the impact of AI assistance on human reader performance. The document mentions software modifications for Quality Control (QualityGuard and Bad Block Correction), but these are related to system maintenance and image reconstruction quality, not AI-assisted image interpretation.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study was Done

Not applicable in the context of what is typically considered a standalone AI algorithm study. The "device" here is a PET/CT scanner, and the performance testing described is for the imaging system as a whole, including image reconstruction capabilities. While the system has software, it's not described as a standalone diagnostic AI algorithm.

7. The Type of Ground Truth Used

The ground truth used for the performance testing is phantom-based measurements as specified by the NEMA NU2:2018 standard.

8. The Sample Size for the Training Set

Not applicable. This submission describes the performance of a physical imaging device, not a machine learning model that requires a training set of data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for an AI model in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Siemens Medical Solutions USA, Inc. % Alaine Medio PET Regulatory Projects Manager 810 Innovation Drive KNOXVILLE TN 37932

May 6, 2019

Re: K190900

Trade/Device Name: Biograph Vision PET/CT Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS, JAK Dated: April 5, 2019 Received: April 8, 2019

Dear Alaine Medio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K190900

Device Name Biograph Vision PET/CT

Indications for Use (Describe)

The Siemens Biograph Vision PET/CT systems are combined X-Ray Computed Tomography (CT) and Position Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray wansmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.

These systems are intended to be utilized by approxiately trained health care professionals to aid in detecting. diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

This CT system can be used for low dose lung cancer screening in high risk populations * * As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lune Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

Type of Use (Select one or both, as applicable) ☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

as required by 21 CFR Part 807.87(h)

Identification of the Submitter
Submitter:Alaine MedioRegulatory AffairsSiemens Medical Solutions USA, Inc.Molecular Imaging810 Innovation DriveKnoxville, TN 37932
Alternative Contact:Tabitha EstesRegulatory Affairs
Manufacturer:Siemens Medical Solutions USA, Inc.Molecular Imaging2501 North Barrington RoadHoffman Estates, IL 60192
Telephone Number:(865)206-0337
Fax Number:(865)218-3019
Date of Submission:April 5, 2019
Identification of the product
Device Proprietary Name:Biograph Vision PET/CT
Common Name:Positron Emission Tomography (PET) SystemComputed Tomography (CT) System
Classification Name:Emission Computed Tomography System per 21 CFR892.1200Computed Tomography X-Ray System per 21 CFR 892.1750
Product Code:90 KPS and 90 JAK
Classification Panel:Radiology
Device Class:Class II

{4}------------------------------------------------

Marketed Devices to which Equivalence is claimed

Predicate:

Device Proprietary Name:Biograph Vision PET/CT
Manufacturer:Siemens Medical Solutions USA, Inc
Product Code:90 KPS and 90 JAK
Device Class:Class II
510(k) Number:K180811
Reference Devices:
Device Name:Biograph mCT
510(k) Number:K173578

Device Description:

The Biograph Vision systems are combined multi-slice X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners. These systems are designed for whole body oncology, neurology and cardiology examinations.

The Biograph Vision systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

Biograph Vision software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

The Biograph Vision PET/CT, which is the subject of this application, is substantially equivalent to the commercially available Biograph Vision PET/CT (K180811). The key differences between the current Biograph Vision (predicate device) and the Biograph Vision PET/CT that is the subject of this application is the addition of two new models (Biograph Vision 450 and Biograph Vision 450 Edge) with a smaller axial PET Field of View, as well as additional software modifications to daily / weekly Quality Control that is applicable to all models of the Biograph Vision.

{5}------------------------------------------------

Intended Use:

The Biograph Vision PET/CT systems are radiological imaging systems that are a combination of a positron emission tomography (PET) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images. The nuclear medicine images and the x-ray images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (i.e., distribution of radiopharmaceuticals). The PET and CT portions of the system may be used independently or in combination. The PET and CT images may be transferred to other systems for radiation therapy planning or additional processing.

Indications for Use

The Siemens Biograph Vision PET/CT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

This CT system can be used for low dose lung cancer screening in high risk populations.*

*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

{6}------------------------------------------------

Summary of the Technological Characteristics of the Biograph Vision compared to the predicate device:

The Biograph Vision PET/CT, which is the subject of this application, is substantially equivalent to the commercially available Biograph Vision PET/CT (K180811). The key differences between the current Biograph Vision (predicate device) and the Biograph Vision PET/CT that is the subject of this application is the addition of two new models (Biograph Vision 450 and Biograph Vision 450 Edge) with a smaller axial PET Field of View:

FeatureBiograph Vision 450/ 450 EdgeBiograph Vision600 / 600 Edge(K180811)Comparison to PredicateDevice - K180811
General Specifications
Height203.6 cm (80.2 in)203.6 cm (80.2 in)No Change
Width234.4 cm (92.3in)234.4 cm (92.3in)No Change
Depth136 cm (53.6 in)136 cm (53.6 in)No Change
Weight3734 kg (8231 lbs)3759kg (8287 lbs)Reduction of detectors for newmodels.
Patient Port Diameter78 cm (30.7 in)78 cm (30.7 in)No Change
Power Requirements380-480V~, 50/60Hz,10 kVA Continuous,150kVA Intermittent380-480V~, 50/60Hz,10 kVA Continuous,150kVA IntermittentNo Change
PET System
DetectorsSiPM with LutetiumOxyorthosilicate (LSO)crystalsSiPM with LutetiumOxyorthosilicate(LSO) crystalsNo Change
PET Axial FOV200 mm263 mmReduction of detectors for newmodels with smaller PET AxialField of View.
PerformanceSpecificationsTested in Accordancewith NEMA NU2 2018Tested in Accordancewith NEMA NU2 2018See below for PET Performancespecifications. Changes inCount Rate peak NECR, CountRate peak Trues and Sensitivitydue to difference in Axial Fieldof View for new models.
CT System
CT subcomponentincorporatedSOMATOM DefinitionAS / AS+ andSOMATOM DefinitionEdgeSOMATOM DefinitionAS / AS+ andSOMATOM DefinitionEdgeNo Change

Additional changes include updates to the quality control software to implement:

  • . QualityGuard (a feature previously cleared with the Biograph mCT K173578) that provides for phantom-free daily calibration activities; and
  • . Bad Block Correction, where the system compensates for bad blocks in the reconstructed image without affecting image quality.

{7}------------------------------------------------

Both features are provided for all models of Biograph Vision, and are designed to improve quality control workflow while minimizing system downtime.

Performance Testing / Safety and Effectiveness:

PET Testing in accordance with NEMA NU2:2018 was conducted on the various Biograph Vision models.

Performance CriteriaBiograph Vision 450 / 450 EdgeBiograph Vision 600 / 650 Edge
Resolution - Full SizeResultsAcceptanceResultsAcceptance
Transverse ResolutionFWHM @ 1 cmPass≤ 4.0 mmPass≤ 4.0 mm
Transverse ResolutionFWHM @ 10 cmPass≤ 4.8 mmPass≤ 4.8 mm
Transverse ResolutionFWHM @ 20 cmPass≤ 5.2 mmPass≤ 5.2 mm
Axial ResolutionFWHM @ 1 cmPass≤ 4.3 mmPass≤ 4.3 mm
Axial ResolutionFWHM @ 10 cmPass≤ 5.4 mmPass≤ 5.4 mm
Axial ResolutionFWHM @ 20 cmPass≤ 5.4 mmPass≤ 5.4 mm
Count Rate / Scatter / Sensitivity / etc.
Sensitivity @435 keVLLDPass≥ 8.0 cps/kBqPass≥ 15.0 cps/kBq
Count Rate peak NECRPass≥140 kcps @ ≤ 32 kBq/ccPass≥250 kcps @ ≤ 32 kBq/cc
Count Rate peak truesPass≥600 kcps @ ≤ 56 kBq/ccPass≥1100 kcps @ ≤ 56 kBq/cc
Scatter Fraction atpeak NECRPass≤43%Pass≤43%
Mean bias (%) at peakNECPass≤ 6%Pass≤ 6%
Co-RegistrationAccuracyPass≤ 5 mmPass≤ 5 mm
Time of FlightResolution at peak NECPass≤249 psPass≤249 ps
Image Quality (4 to 1) - (% Contrast / Background Variability)
10mm spherePass≥ 55% / ≤ 10%Pass≥ 55% / ≤ 10%
13mm spherePass≥ 60% / ≤ 9%Pass≥ 60% / ≤ 9%
17mm spherePass≥ 65% / ≤ 8%Pass≥ 65% / ≤ 8%
22mm spherePass≥ 70% / ≤ 7%Pass≥ 70% / ≤ 7%
28mm spherePass≥ 75% / ≤ 6%Pass≥ 75% / ≤ 6%
37mm spherePass> 80% /<5%Pass> 80% /<5%

Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no changes affecting this testing.

{8}------------------------------------------------

Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.

Siemens claims compliance with the following product standards for the Biograph Vision:

  • IEC 60601-1: 2005+ A1:2012
  • IEC 60601-1-2: 2014
  • IEC 60601-1-3: 2013
  • IEC 60601-1-6:2010 +A1:2013
  • IEC 60601-2-28:2010 ●
  • IEC 60601-2-44: 2009 + A1:2012 + A2:2016
  • IEC 62366-1:2015
  • IEC 61223-2-6:2006
  • IEC 61223-3-5:2004
  • NEMA XR 25: 2010
  • NEMA XR 28: 2013
  • NEMA XR 29: 2013
  • NEMA PS3.1-3.20

Additionally, the Biograph Vision has been developed in accordance with the requirements of the following standards:

  • IEC 62304:2006 +A1:2015
  • ISO 14971:2012 (ISO 14971:2007)

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.

Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The Biograph Vision software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Biograph Vision and external devices.

Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.

{9}------------------------------------------------

All Performance testing met the predetermined acceptance values. This, coupled with the successful verification and validation testing demonstrates that the Biograph Vision functions as intended and that performance is comparable to the predicate devices.

Statement regarding Substantial Equivalence:

There have been no changes implemented in the Biograph Vision that impact either the fundamental scientific technology or the indications for use. The Biograph Vision described in this Premarket Notification is substantially equivalent to the currently commercially available Biograph Vision predicate device.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.