K Number
K173578

Validate with FDA (Live)

Date Cleared
2017-12-20

(30 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Siemens mCT systems are combined X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-ray transmission data from either the same plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and/or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

This CT system can be used for low dose lung cancer screening in high risk populations. (High risk population has been defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.)

Device Description

The Biograph mCT family systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole body oncology, neurology and cardiology examinations. The Biograph mCT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

Biograph mCT software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

The software for the Biograph mCT systems which is the subject of this application is substantially equivalent to the commercially available Biograph mCT software. Modifications include:

  • Somaris Software (cleared in K152036)
    • o Upgrade to the latest revision of Somaris Software
  • PETsyngo software ●
    • Update of the software to add Low CT Does Protocols for PET AC O
    • Update to add Wholebody Scatter Correction O
    • QualityGuard O
    • FlowMotion Multi-Parametric PET O
    • O OncoFreeze
    • o CardioFreeze
    • Shuttle Mode PET Acquisition o
    • Parallel Reconstruction O
    • PET Dose Report O
    • TeamViewer O
    • Maximum Patient Clearance Mode o
    • Smart Mobile Connect o
    • O Update of the software corrections
    • Update of the TrueD software component O
      • Address anomalies
      • Update to improve usability and viewing of Multi-Parametric PET imaqes
    • Improvements to workflow o
    • Improvements to performance (faster reconstruction speeds) о
AI/ML Overview

This document describes the performance testing for the Siemens Biograph mCT and mCT Flow PET/CT Scanners (K173578). This is not an AI/ML device, therefore, some of the requested information (e.g., training set, expert adjudication, MRMC study, standalone performance) is not applicable.

1. Table of Acceptance Criteria and Reported Device Performance:

The performance testing was conducted in accordance with NEMA NU2:2012. The following table summarizes the acceptance criteria and the "Pass" results reported for two different configurations (3-ring and 4-ring versions) of the Biograph mCT systems.

Performance CriteriaAcceptanceReported Device Performance
Resolution - Full Size
Transverse Resolution FWHM @ 1 cm≤ 4.7 mmPass
Transverse Resolution FWHM @ 10 cm≤ 5.4 mmPass
Transverse Resolution FWHM @ 20 cm≤ 6.3 mmPass
Axial Resolution FWHM @ 1 cm≤ 4.9 mmPass
Axial Resolution FWHM @ 10 cm≤ 6.5 mmPass
Axial Resolution FWHM @ 20 cm≤ 8.8 mmPass
Resolution - 256 x 256
Transverse Resolution FWHM @ 1 cm≤ 7.3 mmPass
Transverse Resolution FWHM @ 10 cm≤ 7.5 mmPass
Transverse Resolution FWHM @ 20 cm≤ 7.7 mmPass
Axial Resolution FWHM @ 1 cm≤ 6.0 mmPass
Axial Resolution FWHM @ 10 cm≤ 6.6 mmPass
Axial Resolution FWHM @ 20 cm≤ 9.2 mmPass
Count Rate / Scatter / Sensitivity
Sensitivity @435 keV LLD≥ 5.0 cps/kBq ≥ 9.4 cps/kBq (TrueV)Pass
Count Rate peak NECR≥ 95 kcps @ ≤ 30 kBq/cc ≥ 165 kcps @ ≤ 28 kBq/cc (TrueV)Pass
Count Rate peak trues≥ 350 kcps @ ≤ 46 kBq/cc ≥ 575 kcps @ ≤ 40 kBq/cc (TrueV)Pass
Scatter Fraction at peak NECR≤ 40%Pass
Mean bias (%) at peak NEC+/- 6%Pass
Image Quality (4 to 1) - (% Contrast / Background Variability)
10mm sphere≥ 10% / ≤ 10%Pass
13mm sphere≥ 25% / ≤ 10%Pass
17mm sphere≥ 40% / ≤ 10%Pass
22mm sphere≥ 55% / ≤ 10%Pass
28mm sphere≥ 55% / ≤ 10%Pass
37mm sphere≥ 60% / ≤ 10%Pass

2. Sample size used for the test set and data provenance:

The document states that performance testing was conducted on two different configurations of the Biograph mCT systems: a 3-ring version and a 4-ring version. The testing was performed in accordance with NEMA NU2:2012, which typically involves standardized phantom studies rather than patient data. Therefore, the "sample size" refers to the number of phantom configurations tested. The data provenance is laboratory-based phantom testing, not patient data from a specific country or retrospective/prospective study.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the performance testing for a PET/CT scanner, according to NEMA NU2:2012 standards, relies on objective measurements from physical phantoms rather than human expert interpretation or ground truth establishment in a clinical context.

4. Adjudication method for the test set:

This information is not applicable as the performance testing for a PET/CT scanner, according to NEMA NU2:2012 standards, relies on objective measurements from physical phantoms rather than human expert interpretation or adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The device is a PET/CT scanner, a medical imaging hardware system, not an AI-powered diagnostic algorithm.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This is not applicable. The device is a PET/CT scanner, a hardware system with integrated software, not a standalone algorithm. Performance is assessed through physical measurements, not algorithmic output on its own.

7. The type of ground truth used:

The ground truth for this type of device is established by the physical properties of the phantoms used in the NEMA NU2:2012 testing. These phantoms have known dimensions, radioactive concentrations, and other physical characteristics that serve as the "true" values against which the scanner's measurements are compared.

8. The sample size for the training set:

This is not applicable. The device is a PET/CT scanner, which is a medical imaging hardware system. It does not employ machine learning that requires a "training set" in the conventional AI sense.

9. How the ground truth for the training set was established:

This is not applicable for the same reasons as #8.

{0}------------------------------------------------

December 20, 2017

Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Siemens Medical Solutions USA, Inc. % Ms. Tabitha Estes Regulatory Technical Specialist 810 Innovation Drive KNOXVILLE TN 37932

Re: K173578

Trade/Device Name: Biograph mCT and MCT Flow PET/CT Scanners Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS. JAK Dated: November 17, 2017 Received: November 21, 2017

Dear Ms. Estes:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Robert Oakes

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K173578

Device Name

Biograph mCT and mCT Flow PET/CT Scanners

Indications for Use (Describe)

Siemens mCT systems are combined X-ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-ray transmission data from either the same plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and/or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

This CT system can be used for low dose lung cancer screening in high risk populations. (High risk population has been defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.)

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

as required by 21 CFR Part 807.87(h)

Identification of the Submitter
Submitter:Tabitha EstesRegulatory Technical SpecialistSiemens Medical Solutions USA, Inc.Molecular Imaging810 Innovation DriveKnoxville, TN 37932
Manufacturer:Siemens Medical Solutions USA, Inc.Molecular Imaging2501 North Barrington RoadHoffman Estates, IL 60192
Telephone Number:(865)218-2421
Fax Number:(865)218-3019
Date of Submission:November 17th, 2017
Identification of the product
Device Proprietary Name:Biograph mCT Family of PET/CT Systems
Common Name:Positron Emission Tomography (PET) SystemComputed Tomography (CT) System
Classification Name:Emission Computed Tomography System per 21 CFR892.1200Computed Tomography X-Ray System per 21 CFR892.1750
Product Code:90 KPS and 90 JAK
Classification Panel:Radiology
Device Class:Class II

{4}------------------------------------------------

Marketed Devices to which Equivalence is claimed

Predicate:

Device Proprietary Name:Biograph mCT Family of PET/CT Systems
Manufacturer:Siemens Medical Solutions USA, Inc.
Product Code:90 KPS and 90 JAK
Device Class:Class II
510(k) Number:K151486
Reference Devices:
Device Name:Somatom Definition AS/AS+/Edge
510(k) Numbers:K152036 (Somatom Definition AS/AS+/Edge)
Device Name:Biograph Horizon
510(k) Number:K170904

{5}------------------------------------------------

Device Description:

The Biograph mCT family systems are combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. These systems are designed for whole body oncology, neurology and cardiology examinations. The Biograph mCT systems provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

Biograph mCT software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

The software for the Biograph mCT systems which is the subject of this application is substantially equivalent to the commercially available Biograph mCT software. Modifications include:

  • Somaris Software (cleared in K152036)
    • o Upgrade to the latest revision of Somaris Software
  • PETsyngo software ●
    • Update of the software to add Low CT Does Protocols for PET AC O
    • Update to add Wholebody Scatter Correction O
    • QualityGuard O
    • FlowMotion Multi-Parametric PET O
    • O OncoFreeze
    • o CardioFreeze
    • Shuttle Mode PET Acquisition o
    • Parallel Reconstruction O
    • PET Dose Report O
    • TeamViewer O
    • Maximum Patient Clearance Mode o
    • Smart Mobile Connect o
    • O Update of the software corrections
    • Update of the TrueD software component O
      • Address anomalies
      • Update to improve usability and viewing of Multi-Parametric PET imaqes
    • Improvements to workflow o
    • Improvements to performance (faster reconstruction speeds) о

{6}------------------------------------------------

Intended Use:

The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

This CT system can be used for low dose lung cancer screening in high risk populations. *

  • As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.

{7}------------------------------------------------

Performance Testing / Safety and Effectiveness:

Performance testing for the CT subsystem was included in the original premarket notification for the CT subsystems and there have been no changes affecting this testing.

PET Testing in accordance with NEMA NU2:2012 was conducted on two different configurations of the Biograph mCT systems, a 3 ring version and a 4 ring version.

Performance CriteriaResultsAcceptance
Resolution - Full Size
Transverse Resolution FWHM @ 1 cmPass≤ 4.7 mm
Transverse Resolution FWHM @ 10 cmPass≤ 5.4 mm
Transverse Resolution FWHM @ 20 cmPass≤ 6.3 mm
Axial Resolution FWHM @ 1 cmPass≤ 4.9 mm
Axial Resolution FWHM @ 10 cmPass≤ 6.5 mm
Axial Resolution FWHM @ 20 cmPass≤ 8.8 mm
Resolution - 256 x 256
Transverse Resolution FWHM @ 1 cmPass≤ 7.3 mm
Transverse Resolution FWHM @ 10 cmPass≤ 7.5 mm
Transverse Resolution FWHM @ 20 cmPass≤ 7.7 mm
Axial Resolution FWHM @ 1 cmPass≤ 6.0 mm
Axial Resolution FWHM @ 10 cmPass≤ 6.6 mm
Axial Resolution FWHM @ 20 cmPass≤ 9.2 mm
Count Rate / Scatter / Sensitivity
Sensitivity @435 keV LLDPass≥ 5.0 cps/kBq≥ 9.4 cps/kBq (TrueV)
Count Rate peak NECRPass≥ 95 kcps @ ≤ 30 kBq/cc≥ 165 kcps @ ≤ 28 kBq/cc (TrueV)
Count Rate peak truesPass≥350 kcps @ ≤ 46 kBq/cc≥ 575 kcps @ ≤ 40 kBq/cc (TrueV)
Scatter Fraction at peak NECRPass≤ 40%
Mean bias (%) at peak NECPass+/- 6%
Image Quality (4 to 1) - (% Contrast / Background Variability)
10mm spherePass≥ 10% / ≤ 10%
13mm spherePass≥ 25% / ≤ 10%
17mm spherePass≥ 40% / ≤ 10%
22mm spherePass≥ 55% / ≤ 10%
28mm spherePass≥ 55% / ≤ 10%
37mm spherePass≥ 60% / ≤ 10%

{8}------------------------------------------------

All Performance testing met the predetermined acceptance values.

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.

Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitiqation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.

Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.

Statement regarding Substantial Equivalence:

There have been no changes implemented in the modifications to the Biograph mCT that impact either the fundamental technology or the indications for use. The Biograph mCT with the modifications outlined in this Premarket Notification is substantially equivalent to the currently commercially available predicate device.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.