K Number
K181037

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-12-21

(246 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

Patient specific abutment is intended for use with the UF Implant Systems provided in the chart. All digitally designed abutments for use with DIO CAD/CAM Abutments are intended to be manufactured at a DIO Corporation validated milling center.

Device Description

The DIO CAD/CAM Abutment includes two CAD/CAM abutment designs, Hybrid Link Abutment and Patient-Specific Abutment.

  1. Hybrid Link abutment: Hybrid Link abutment is intended to provide support for customized prosthetic restorations such as crowns and bridges. The hybrid link abutment is composed of two-piece abutment that is a hybrid link at the bottom and a coping (CAD/CAM patient specific superstructure) at the top. The hybrid link abutments are pre-manufactured (stock) abutments, made from a titanium alloy conforming to ASTM F136. The diameters of Hybrid Link Abutment are 4.0/4.5/5.5mm. Hybrid Link abutment is provided non-sterile therefore must be sterilized after the cementation of the patient-specific superstructure on the Hybrid Link Abutment.

  2. Patient-Specific Abutment: Patient-specific abutment is made from titanium alloy conforming to ASTM F136 titanium abutment to be used in fabricating patient-specific abutments. The subject abutments are indicated for cemented or "Screw-and Cement-Retained Prosthesis(SCRP)" restorations. Each patient-specific abutment is individually prescribed by the clinician. The diameters of patient-specific Abutment are 3.0/3.3/3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 and two connection designs (Hex, Non-hex). Patient-specific abutments are supplied with an abutment screw previous cleared device as K122519 and K161987 and provided non-sterile.

AI/ML Overview

The document provided is a 510(k) Premarket Notification from the FDA for a dental implant abutment. It does not describe a study involving an AI/ML device, nor does it provide acceptance criteria or performance metrics in the format requested. The document focuses on demonstrating substantial equivalence to predicate devices through material properties, design specifications, and non-clinical testing for dental implant abutments.

Therefore, I cannot provide the requested information regarding acceptance criteria and device performance from this document. The sections you asked for, such as sample size, expert qualifications, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth establishment, are not applicable to the content of this 510(k) summary for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2018

DIO Corporation % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620

Re: K181037

Trade/Device Name: DIO CAD/CAM Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 19, 2018 Received: November 26, 2018

Dear April Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Andrew I. Steen -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181037

Device Name DIO CAD/CAM Abutment

Indications for Use (Describe)

Indication for Use

DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

Implant System CompatibilityImplant Diameter (mm)Platform Diameter (mm)
UF(II) Narrow Implant System3.0/3.33.0/3.3
UF Sub merged Implant System3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0
UF(II) Implant System3.8/4.0/4.5/5.0/5.53.8/4.0/4.5/5.0/5.5

Patient specific abutment is intended for use with the UF Implant Systems provided in the chart. All digitally designed abutments for use with DIO CAD/CAM Abutments are intended to be manufactured at a DIO Corporation validated milling center.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter

DIO Corporation Jiae, Park 66, Centum Seo-ro, Haeundae-gu Busan, 48058 Republic of Korea Email: japark@dio.co.kr Phone: +82-51-745-7836

Official Correspondent

Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122

Device Information

  • Trade Name: DIO CAD/CAM Abutment
  • Common Name: Endosseous dental implant abutment
  • Classification Name: Abutment, Implant, Dental, Endosseous .
  • Product Code: NHA
  • . Panel: Dental
  • Regulation Number: 21 CFR 872.3630 ●
  • . Device Class: Class II
  • Date prepared: 12/21/2018 ●

Predicate Devices:

The subject device is substantially equivalent to the following predicate device:

Primary Predicate

  • Medentika CAD/CAM Abutment (K150203)
    Reference Devices

  • Sirona Dental CAD/CAM System (K111421)

  • Zimmer® patient Specific Abutment, Internal Hex, Titanium(K143505) ●

  • Encode Patient Specific Abutment for Straumann Bone Level Connection(K112730) ●

  • . UF(II) Narrow Implant System (K161987)

  • UF Sub Merged Implant System (K122519) ●

  • UF(II) Implant System (K170608)

  • Self-Adhesive Resin Cement (K073173)

  • . ZIRMON SERIES (K131117)

{4}------------------------------------------------

Device Description

The DIO CAD/CAM Abutment includes two CAD/CAM abutment designs, Hybrid Link Abutment and Patient-Specific Abutment.

1) Hybrid Link abutment

Hybrid Link Abutment

Hybrid Link abutment is intended to provide support for customized prosthetic restorations such as crowns and bridges. The hybrid link abutment is composed of two-piece abutment that is a hybrid link at the bottom and a coping (CAD/CAM patient specific superstructure) at the top. The hybrid link abutments are pre-manufactured (stock) abutments, made from a titanium alloy conforming to ASTM F136. The diameters of Hybrid Link Abutment are 4.0/4.5/5.5mm.

Hybrid Link abutment is provided non-sterile therefore must be sterilized after the cementation of the patient-specific superstructure on the Hybrid Link Abutment.

The proposed devices are compatible with the following device.

Dental Implants

Proprietary NameUF(II) NarrowImplant SystemUF Sub merged ImplantSystemUF(II) Implant System
Compatible Implants(Knumber)K161987K122519K170608
Implant diameter size3.0/3.33.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.8/4.0/4.5/5.0/5.5
Implant InterfaceConnectionType/Size(mm)Internal connectiontype/2.3Internal connectiontype/3.35Internal connectiontype/3.35
Type of Implant-Abutment ConnectionHex/Non HexHex/Non HexHex/Non Hex

Raw material blanks

  • K131117 ZIRMON SERIES manufactured by KUWOTECH CO., LTD .

Raw material cement

  • . K073173 Self-Adhesive Resin Cement manufactureed by Densply is the cement
    The coping that composes the final abutment should be designed and milled through the CAD/CAM software, according to the prosthetic planning and patient clinical situation. The coping and crowns designed using these or more recent versions of the CAD/CAM System, within the design limits as defined within the design software, are compatible with the Hybrid link abutment.

The coping would be manufactured by DIO Corporation only with design input using CAD/CAM Software from and by DIO Corporation milling center.

Design Limitation for Superstructure:

Super structure design Limitation (Unit: mm)
Range (Diameter)Range (Height)Range (Wallthickness)Range (Angle)

{5}------------------------------------------------

Min .4.0mmland of the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of thMin. OmmU.Smmا ا 1 1 111 C ( .रे
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

2) Patient-Specific Abutment

Patient-specific abutment is made from titanium alloy conforming to ASTM F136 titanium abutment to be used in fabricating patient-specific abutments. The subject abutments are indicated for cemented or "Screw-and Cement-Retained Prosthesis(SCRP)" restorations. Each patient-specific abutment is individually prescribed by the clinician.

The diameters of patient-specific Abutment are 3.0/3.3/3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0 and two connection designs (Hex, Non-hex).

Patient-Specific Abutment is compatible with following Implant Systems.

Proprietary NameUF(II) Narrow ImplantSystemUF Sub mergedImplant SystemUF(II) ImplantSystem
Compatible Implants(Knumber)K161987K122519K170608
Implant diameter size3.0/3.33.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.8/4.0/4.5/5.0/5.5
Implant InterfaceConnectionType/Size(mm)Internal connectiontype/2.3Internal connectiontype/3.35Internal connectiontype/3.35
Type of Implant-Abutment ConnectionHex/Non HexHex/Non HexHex/Non Hex

Patient-specific abutments are supplied with an abutment screw previous cleared device as K122519 and K161987 and provided non-sterile.

Model NameRange (Diameter)Range (Height)Range (Angle)
CASSHCASSN3.9~12.03.1~14.90~29.8°
CAUNHCAUNN3.9~12.03.1~14.90~29.3°

Indication for Use

DIO CAD/CAM Abutment is intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

Implant System CompatibilityImplant Diameter (mm)Platform Diameter (mm)
UF(II) Narrow Implant System3.0/ 3.33.0/ 3.3
UF Sub merged Implant System3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0
UF(II) Implant System3.8/4.0/4.5/5.0/5.53.8/4.0/4.5/5.0/5.5

Patient specific abutment is intended for use with the UF Implant Systems provided in the chart. All digitally designed abutments for use with DIO CAD/CAM Abutments are intended to be manufactured at a DIO Corporation validated milling center.

{6}------------------------------------------------

Summaries of Technological Characteristics

The subject device is substantially equivalent to the currently cleared devices. They are substantially equivalent in intended use, material and connection interfaces to the implants are identical for each individual diameter and connection type. Comparison demonstrating Substantial Equivalence follows:

1) Hybrid Link Abutment

Subject DevicePrimary Predicate DeviceReference Device
ApplicantDIO CorporationMedentika GmbHSirona Dental Systems GmbH
Trade NameDIO CAD/CAM AbutmentMedentika CAD/CAM AbutmentSirona Dental CAD/CAM SystemTiBase Abutment
510(K) No.K181037K150203K111421
Classification NameEndosseous Dental Implant, Abutment(872.3630)Endosseous Dental Implant, Abutment(872.3630)Endosseous Dental Implant, Abutment(872.3630)
Product CodeNHANHANHA
ClassIIIIII
MaterialTi-6Al-4V ELI (ASTM F136)Zirconia OxideTitanium AlloyTi-6Al-4V ELI (ASTM F136)Zirconia Oxide
DesignTitanium BaseTitanium BaseTitanium Base
Diameters (mm)4.0/4.5/5.53.25-7.03.5~6.5
SuperstructureAngulation (°)0~15Up to 30Up to 20
SuperstructurematerialZirconiaZirconiaZirconia
SterileSteam Sterilization by user(Delivered non sterile)Steam Sterilization by user(Provided Non-Sterile)Steam Sterilization by user(Delivered non sterile)
Type of RetentionScrew-retained or cement retainedCement-retainedScrew-retained or cement retained
Indications For UseDIO CAD/CAM Abutment is intended for usewith dental implants as a support for single ormultiple tooth prostheses in the maxilla ormandible of a partially or fully edentulous patient.Medentika Preface CAD/CAM Abutments are intended foruse with dental implants as a support for single or multipletooth prostheses in the maxilla or mandible of a partially orfully edentulous patient.The Sirona Dental CAD/CAM System isintended for use in partially or fullyedentulous mandibles and maxillae insupport of single or multiple-unit cementretained restorations. For the titaniumbases SSO 3.5L and SBL 3.3L, theindication is restricted for replacement ofsingle lateral incisors in the maxilla and

{7}------------------------------------------------

NarrowImplantSystemNobel BiocareReplace™ SelectE3.5, 4.3,5.0, 6.03.5, 4.3,5.0, 6.0lateral and central incisors in the mandible.The system consists of three major parts:TiBase, inCoris mesostructured, andCAD/CAM software. Specifically, theinCoris mesostructured and TiBasecomponents make up a two-pieceabutment which is used in conjunctionwith endosseoous dental implants torestore the function and aesthetics in theoral cavity. The inCoris mesostructuredmay also be used in conjunction with theCamlog Titanium base CAD/CAM(typesK2244.xxxx)(K083496) in the CamlogImplant System. The CAD/CAM softwareis intended to design and fabricate theinCoris mesostructure. The inCorismesostructured and TiBase two-pieceabutment is compatible with the followingimplants systems:
UF SubmergedImplantSystem3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0Nobel BiocareNobel Active™F3.5, 4.3,5.03.5, 3.9(4.3), 3.9(5.0)
UF(II)ImplantSystem3.8/4.0/4.5/5.0/5.53.8/4.0/4.5/5.0/5.5Biomet 3iOsseotite®H3.25, 4.0,5.03.4, 4.1,5.0Nobel Biocare Replace(K020646)
CertainBiomet 3iOsseotite®I3.25, 3.75,4.0, 5.03.4, 4.1,5.0Nobel Biocare Branemark(K022562)
Nobel BiocareBranemarkK3.3, 3.75,4.0, 5.03.5, 4.1,4.1, 5.1Friadent Xive(K013867)
Patient specific abutment is intended for use withthe UF Implant Systems provided in the chart. Alldigitally designed abutments for use with DIOCAD/CAM Abutments are intended to bemanufactured at a DIO Corporation validatedmilling center.Straumann BoneLevelL3.3, 4.1,4.83.3, 4.1,4.8Biomet 3i Osseotite(K980549)
StraumannStnadardN3.3, 4.1,4.83.5(NNC),4.8, 6.5Astra Tech Osseospeed(K091239)
Zimmer TaperedScrew-vent®R3.3, 3.7,4.1, 4.7,6.03.5, 4.5,5.7Zimmer Tapered Screw-Vent(K061410)
Austra TechOsseoSpped™S3.5, 4.0,4.5, 5.03.5, 4.0,4.5, 5.0Straumann SynOcta(K061176)
DentsplyFriadent®Frialit/XiVE®T3.4, 3.8,4.5, 5.53.4, 3.8,4.5, 5.5Straumann Bone Level(K053088,K062129, K060958)
DentsplyFriadent®Ankylos®Y3.5, 4.5,5.5, 7.03.5, 4.5,5.5, 7.0Biomet 3i Certain(K014235, K061629)
Medentika PreFace is intended for use with theStraumann® CARES® System. All digitally designedabutments for use with Medentika CAD/CAM Abutmentsare intended to be manufactured at a Straumann®CARES® validated milling center.Nobel Biocare Active(K071370)
SubstantialEquivalenceComparisonThe subject Hybrid Link Abutment is substantially equivalent in designs, dimensions, material, indications, and technological characteristicswith the identified primary predicate device. The DIO CAD/CAM Abutment is similar in fundamental scientific technology to thepredicate device in that they all have been designed, manufactured and tested in compliance with FDA's Class II special controlsguidance document root-food endosseous dental implants and endosseous dental implant abutments.The diameters of the subject device are slightly different from the predicate devices. However, the subject diameters are in the range ofdiameters of predicates and this dimensional difference doesn't affect device safety and effectiveness.The Indications for Use of the subject and primary predicate device are identical other than the compatible implant bodies. This difference ismitigated by fatigue testing, reverse engineering, dimensional analysis and identification of reference device for compatible implant bodies

{8}------------------------------------------------

Both the predicate and subject devices are intended to be milled into patient specific abutments using CAD/CAM technology under the manufacturing control of the sponsor. Any differences in technology characteristics are accompanied by information that demonstrated the device is substantially equivalent as the predicate and do not raise different questions of safety and effectiveness than the predicate.

2) Patient-Specific Abutment

Subject DevicePrimaryPredicate DeviceReference Device
ApplicantDIO CorporationMedentika GmbHZimmer Dental, Inc.BIOMET 3i
Trade NameDIO CAD/CAM AbutmentMedentika CAD/CAM AbutmentZimmer® patientSpecific Abutment,Internal Hex, TitaniumEncode PatientSpecific Abutment forStraumann Bone LevelConnection
510(K) No.Not yet assignedK150203K143505K112730
ClassificationNameEndosseous Dental Implant, Abutment(872.3630)Endosseous Dental Implant, Abutment(872.3630)Endosseous DentalImplant, Abutment(872.3630)Endosseous DentalImplant, Abutment(872.3630)
Product CodeNHANHANHANHA
ClassIIIIIIII
MaterialTi-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTM F136)Ti-6Al-4V ELI (ASTMF136)Ti-6Al-4V ELI(ASTM F136)
DesignCAD/CAM Patient-Specific AbutmentCAD/CAM BlankCAD/CAM Patient-Specific AbutmentCAD/CAM Patient-Specific Abutment
Diameters (mm)CAD/CAM Patient-Specific Abutment: 3.0/3.3/3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.0-7.03.5/ 4.5/ 5.73.3/ 4.1/ 4.8
SterileSteam Sterilization by user(Provided Non-Sterile)Steam Sterilization by user(Provided Non-Sterile)Steam Sterilization byuser(Provided Non-Sterile)Steam Sterilization byuser(Provided Non-Sterile)
Type of RetentionScrew-retained or cement retainedCement-retainedScrew-retained orcement retainedScrew-retained orcement retained
Abutment SeatSits on TaperSits on TaperSits on TaperSits on Taper
Anatomical SiteOral CavityOral CavityOral CavityOral Cavity
ConstructionsMachinedMachinedMachinedMachined
Indications ForUse/Intended UseDIO CAD/CAM Abutment is intended for usewith dental implants as a support for single ormultiple tooth prostheses in the maxilla orMedentika Preface CAD/CAM Abutments are intendedfor use with dental implants as a support for single ormultiple tooth prostheses in the maxilla or mandible ofThe Zimmer® PatientSpecific Abutment isused as a terminal orThe 3i Patient-SpecificDental Abutment isintended for use as an

{9}------------------------------------------------

mandible of a partially or fully edentulous patient.a partially or fully edentulous patient.intermediate abutmentfor a cementedprosthesis. Theabutment can be usedfor a single of multiple-unit restoration.accessory to anendosseous dentalimplant to support aprosthetic device in apartially or edentulouspatient. It is intendedfor use to supportsingle and multipletooth prostheses, in themandible or maxilla.The prostheses can bescrew or cementretained to theabutment.
ImplantSystemCompatibilityImplantDiameter(mm)PlatformDiameter(mm)Implant SystemCompatibilitySeriesImplantDiameter(mm)PlatformDiameter(mm)
UF(II)NarrowImplantSystem3.0/ 3.33.0/ 3.3Nobel BiocareReplaceTM SelectE3.5, 4.3, 5.0,6.03.5, 4.3,5.0, 6.0
UF SubmergedImplantSystem3.8/4.0/4.5/5.0/5.5/6.0/6.5/7.03.8/4.0/4.5/5.0/5.5/6.0/6.5/7.0Nobel BiocareNobel ActiveTMF3.5, 4.3, 5.03.5, 3.9(4.3), 3.9(5.0)
Biomet 3iOsseotite®CertainH3.25, 4.0, 5.03.4, 4.1,5.0
Biomet 3iOsseotite®I3.25, 3.75,4.0, 5.03.4, 4.1,5.0
Nobel BiocareBranemarkK3.3, 3.75,4.0, 5.03.5, 4.1,4.1, 5.1
UF(II)ImplantSystem3.8/4.0/4.5/5.0/5.53.8/4.0/4.5/5.0/5.5Straumann BoneLevelL3.3, 4.1, 4.83.3, 4.1,4.8
StraumannStnadardN3.3, 4.1, 4.83.5(NNC),4.8, 6.5
Zimmer TaperedScrew-vent®R3.3, 3.7, 4.1,4.7, 6.03.5, 4.5,5.7
Austra TechOsseoSppedTMS3.5, 4.0, 4.5,5.03.5, 4.0,4.5, 5.0
DentsplyFriadent®Frialit/XiVE®T3.4, 3.8, 4.5,5.53.4, 3.8,4.5, 5.5
DentsplyFriadent®Ankylos®Medentika PreFace is intended for use with theStraumann® CARES® System. All digitally designedabutments for use with Medentika CAD/CAMAbutments are intended to be manufactured at aStraumann® CARES® validated milling center.Y3.5, 4.5, 5.5,7.03.5, 4.5,5.5, 7.0
Patient specific abutment is intended for use with the UF implant systems provided in the chart.All digitally designed abutments for use with DIO CAD/CAM Abutments are intended to bemanufactured at a DIO Corporation validated milling center.
The subject Patient-Specific Abutment is substantially equivalent in designs, dimensions, material, indications, abutment seat, screw seat,anatomical site, connection, and technological characteristics with the identified primary predicate device. The DIO CAD/CAM Abutment issimilar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested incompliance with FDA's Class II special controls guidance document root-food endosseous dental implants and endosseous dentalimplant abutments.
SubstantialEquivalenceComparisonCompared to the primary predicate device, the subject device different with predicate device design, the predicate device design is CAD/CAM
Blank and the subject device is patient specific device. To support this discrepancy of different component inclusion, K143505 and K112730
were added as both predicates have the CAD/CAM patient specific abutment with equivalent design, dimensions, materials and indications.

{10}------------------------------------------------

The Indications for Use of the subject and primary predicate device than the compatible implant bodies. This difference is mitigated by fatigue testing, reverse engineering, dimensional analysis, and identification of reference device for compatible implant bodies. Both the predicate and subject devices are intended to be milled into patient specific abutments using CAD/CAM technology under the manufacturing control of the sponsor. Any differences in technology characteristics are accompanied by information that device is substantially equivalent

as the predicate and do not raise different questions of safety and effectiveness than the predicate.

{11}------------------------------------------------

K181037

Non-clinical Testing

The subject device was tested to evaluate its substantial equivalence according to the following standards.

  • . Fatigue Tests according to ISO 14801:2016
  • End User Steam Sterilization Tests according to ISO 17665-1:2006, 17665-2:2009 and ANSI/AAMI ST79:2010
  • Cytotoxicity testing according to ISO 10993-5:2009

Below tests were performed for predicate devices. K122519 and leveraged for the subject device:

  • Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-● 10:2010.
    Biocompatibility test data was used to evaluate the proposed device's substantial equivalence compared to the predicate device which is owned by the applicant. The results of the above tests have met the criteria of the standard and demonstrated the substantial equivalence with the predicate device.

Non-clinical testing was conducted in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", and it consisted of testing finished assembled implant/abutment systems of the worst-case scenario, (smallest diameter with maximum angulation) through fatigue testing.

Clinical testing was not necessary to establish substantial equivalency of the device.

Conclusion

The DIO CAD/CAM Abutment constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate device. Therefore, DIO CAD/CAM Abutment sand its predicates are substantially equivalent

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)