K Number
K163146

Validate with FDA (Live)

Date Cleared
2017-02-13

(96 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed device was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Description

The proposed device, POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed device is a Powder Free Nitrile Patient Examination Glove that is available in multiple sizes

The proposed device is provided non-sterile. The proposed device is made of Nitrile. The proposed device acts as a barrier.

AI/ML Overview

The document provided is a 510(k) premarket notification for "POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs." This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting a full clinical study with specific acceptance criteria, statistical analysis, and expert ground truth consensus typical for AI-powered diagnostic devices.

Therefore, the requested information about acceptance criteria, efficacy study design, sample sizes for test and training sets, expert qualifications, and adjudication methods (sections 1-9 of your prompt) does not directly apply in the context of this 510(k) submission for medical gloves.

The "study" in this context refers to bench testing against established industry standards for medical gloves, and the "acceptance criteria" are the performance specifications outlined in those standards.

Here's an analysis based on the available information:

1. Table of acceptance criteria and the reported device performance

The acceptance criteria are derived from various ASTM and ISO standards for medical gloves. The reported device performance indicates compliance with these standards.

Acceptance Criterion (from relevant standard)Reported Device Performance
Chemotherapy Drug Permeation (ASTM D6978-05)
Fluorouracil: Breakthrough Detection Time> 240 minutes
Etoposide (Toposar): Breakthrough Detection Time> 240 minutes
Cyclophosphamid (Cytoxan): Breakthrough Detection Time> 240 minutes
Carmustine (BCNU): Breakthrough Detection Time45.0 minutes
Thiotepa: Breakthrough Detection Time30.0 minutes
Paclitaxel (Taxol): Breakthrough Detection Time> 240 minutes
Doxorubicin Hydrochloride: Breakthrough Detection Time> 240 minutes
Dacarbazine (DTIC): Breakthrough Detection Time> 240 minutes
Cisplatin: Breakthrough Detection Time> 240 minutes
Carboplatin: Breakthrough Detection Time> 240 minutes
Docetaxel: Breakthrough Detection Time> 240 minutes
Ifosfamide: Breakthrough Detection Time> 240 minutes
Irinotecan: Breakthrough Detection Time> 240 minutes
Mechlorethamine HCL: Breakthrough Detection Time> 240 minutes
Methotrexate: Breakthrough Detection Time> 240 minutes
Mitomycin C: Breakthrough Detection Time> 240 minutes
Mitoxantrone: Breakthrough Detection Time> 240 minutes
Vincristine Sulfate: Breakthrough Detection Time> 240 minutes
Physical Properties - Before Aging (ASTM D6319-10)
Tensile Strength15 Mpa, min
Ultimate Elongation500% min
Physical Properties - After Aging (ASTM D6319-10)
Tensile Strength14 MPa, min
Ultimate Elongation400% min
Freedom from Holes (ASTM D5151-06)Free from holes, AQL 1.5
Powder Content (ASTM D6124-06)Max. 0.32 mg per glove
Biocompatibility - Irritation (ISO 10993-10)Not an irritant
Biocompatibility - Sensitization (ISO 10993-10)Not a sensitizer

2. Sample size used for the test set and the data provenance

The document does not specify exact sample sizes for each test within the standards (e.g., how many gloves were tested for permeation, holes, etc.). However, it refers to standards like ISO 2859-1:1999 ("Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection"), which would dictate the sample sizes used for quality control aspects like freedom from holes.

The data provenance is from bench testing conducted by the manufacturer, Hebei Hongsen Plastics Technology Co., Ltd. The country of origin for the manufacturing and testing data would be China, as indicated by the manufacturer's address. This is retrospective testing performed to demonstrate compliance for a new device submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. For medical gloves, "ground truth" is established by adherence to physical, chemical, and biological performance standards, not by expert consensus on clinical data interpretation. The tests are objective measurements (e.g., breakthrough time, tensile strength, presence of holes).

4. Adjudication method for the test set

Not applicable. The tests are objective measurements according to standardized protocols, eliminating the need for adjudication among human readers.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical glove, not an AI-powered diagnostic device. No human-in-the-loop performance or AI assistance considerations are relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical glove.

7. The type of ground truth used

The "ground truth" for the device's performance is established by objective measurements and adherence to performance specifications defined in international and national standards, such as:

  • Physical properties (tensile strength, elongation)
  • Dimensional properties (length, width, thickness)
  • Freedom from holes
  • Chemical permeation (chemotherapy drugs)
  • Biological safety (irritation, sensitization)

8. The sample size for the training set

Not applicable. There is no "training set" in the context of this device. The manufacturing process is designed to produce gloves meeting specifications, and the testing confirms this.

9. How the ground truth for the training set was established

Not applicable. There is no training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a stylized graphic. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion. The graphic element features a stylized depiction of three human profiles facing to the right, layered on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 13, 2017

Hebei Hongsen Plastics Technology Co, Ltd % Ray Wang Official Correspondent Beijing Believe Tech. Service Co., Ltd. 5-1206, Build 332, Dafangju, No.25 Banbidian Rd.,liyuan, Tongzhou District Beijing, 101121 CN

Re: K163146 Trade/Device Name: POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: January 3. 2017 Received: January 9, 2017

Dear Ray Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, Ph.D.

Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K163146

Device Name

POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs

Indications for Use (Describe)

The POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed device was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Chemotherapy Drug Permeation

The following chemicals have been tested with proposed device.
Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Fluorouracil50.0 mg/ml (50,000 ppm)> 240
Etoposide (Toposar)20.0 mg/ml (20,000 ppm)> 240
Cyclophosphamid (Cytoxan)20.0 mg/ml (20,000 ppm)> 240
*Carmustine (BCNU)3.3 mg/ml (3,300 ppm)45.0
*Thiotepa10.0 mg/ml (10,000 ppm)30.0
Paclitaxel (Taxol)6.0 mg/ml (6,000 ppm)> 240
Doxorubicin Hydrochloride2.0 mg/ml (2,000 ppm)> 240
Dacarbazine (DTIC)10.0 mg/ml (10,000 ppm)> 240
Cisplatin1.0 mg/ml (1,000 ppm)> 240
Carboplatin10.0 mg/ml (10,000 ppm)> 240
Docetaxel10.0 mg/ml (10,000 ppm)> 240
Ifosfamide50.0 mg/ml (50,000 ppm)> 240
Irinotecan20.0 mg/ml (20,000 ppm)> 240
Mechlorethamine HCL1.0 mg/ml (1,000 ppm)> 240
Methotrexate25.0 mg/ml (25,000 ppm)> 240
Mitomycin C0.5 mg/ml (500 ppm)> 240
Mitoxantrone2.0 mg/ml (2,000 ppm)> 240
Vincristine Sulfate1.0 mg/ml (1,000 ppm)> 240
  • Please note that the following drugs have low permeation times: Carmustine (BCNU): 45 minutes and Thiotepa: 30 minutes

Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K163146

    1. Date of Preparation: 02/08/2017
    1. Sponsor Identification

HEBEI HONGSEN PLASTICS TECHNOLOGY CO., LTD EASTERN INDUSTRIAL ZONE, NANGONG CITY, HEBEI PROVINCE, CHINA

Contact Person: Mr. ShaoZhang Nan Tel: +86-0319-7295820 Fax: +86-0319-7295801 Email: nanshaozhang@163.com

    1. Designated Submission Correspondent
      Mr. Ray Wang

Beijing Believe Technology Service Co., Ltd. 5-1206, Build 332, DaFangJu, No.25 BanBiDian Rd., LiYuan Town, TongZhou District, Beijing, 101121, China

Tel: +86-21-59120817 Fax: +86-21-59120817 Email: Ray.Wang@believe-med.com

{5}------------------------------------------------

4. Proposed Device Identification

Trade Name: POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs Device Name: NITRILE Patient Examination Gloves (Powder Free) Common Name: Patient Examination Gloves

Regulatory Information

Classification: I Product Code: LZA, LZC Regulation Number: 21 CFR 880.6250 Review Panel: General Hospital

Indication for Use:

The POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed device was tested for use with Chemotherapy Drugs as per ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:

Chemotherapy Drug Permeation

The following chemicals have been tested with proposed device.

Chemotherapy DrugConcentrationBreakthrough Detection Time in Minutes
Fluorouracil50.0 mg/ml (50,000 ppm)> 240
Etoposide (Toposar)20.0 mg/ml (20,000 ppm)> 240
Cyclophosphamid (Cytoxan)20.0 mg/ml (20,000 ppm)> 240
*Carmustine (BCNU)3.3 mg/ml (3,300 ppm)45.0
*Thiotepa10.0 mg/ml (10,000 ppm)30.0
Paclitaxel (Taxol)6.0 mg/ml (6,000 ppm)> 240
Doxorubicin Hydrochloride2.0 mg/ml (2,000 ppm)> 240
Dacarbazine (DTIC)10.0 mg/ml (10,000 ppm)> 240
Cisplatin1.0 mg/ml (1,000 ppm)> 240
Carboplatin10.0 mg/ml (10,000 ppm)> 240
Docetaxel10.0 mg/ml (10,000 ppm)> 240
Ifosfamide50.0 mg/ml (50,000 ppm)> 240
Irinotecan20.0 mg/ml (20,000 ppm)> 240
Mechlorethamine HCL1.0 mg/ml (1,000 ppm)> 240
Methotrexate25.0 mg/ml (25,000 ppm)> 240
Mitomycin C0.5 mg/ml (500 ppm)> 240
Mitoxantrone2.0 mg/ml (2,000 ppm)> 240
Vincristine Sulfate1.0 mg/ml (1,000 ppm)> 240

*Please note that the following drugs have low permeation times:

Carmustine (BCNU): 45 minutes and Thiotepa: 30 minutes

{6}------------------------------------------------

    1. Predicate Device Identification
      510(k) Number: K141982

Product Name: Dermagrip Powder Free Blue Nitrile Patient Examination Gloves, Non-Sterile, Tested for use with Chemotherapy Drugs

Manufacturer: WRP Asia Pacific Sdn Bhd.

  • Device Description 6.
    The proposed device, POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

The proposed device is a Powder Free Nitrile Patient Examination Glove that is available in multiple sizes

The proposed device is provided non-sterile. The proposed device is made of Nitrile. The proposed device acts as a barrier.

The proposed device was tested according to the following standards: ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, and ASTM D6978-05. These standards are identified in the following section "Non-clinical test conclusion.

    1. Non-Clinical Test Conclusion
      Bench tests were conducted to verify that the proposed device met all specifications and the proposed device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.

ASTM D5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.

ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs

ISO 2859-1:1999, "Sampling Procedures for Inspection by Attributes - Part I: Sampling Plans Indexed by Acceptable Quality Level (AQL) for Lot-by-Lot Inspection.

ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

{7}------------------------------------------------

ItemProposed DevicePOWDER FREE Blue NitrileGLOVES, Tested for Use withChemotherapy Drugs(K163146)Predicate DeviceDermagrip Powder Free Blue NitrilePatient Examination Gloves Tested forUse with Chemotherapy Drugs(K141982)RemarkSize
Product CodeLZA, LZCLZA, LZCSameProposed DeviceDesignationXSSMLXLTolerance
Regulation Number21 CFR 880.625021 CFR 880.6250SamePOWDER FREE Blue Nitrile GLOVES, Tested for Usewith Chemotherapy Drugs(K163146)Length, mm230230230230230min
ClassIISameWidth, mm708095110120±10
Indication for useThe POWDER FREE Blue NitrileGLOVES, Tested for Use withChemotherapy Drugs is a disposabledevice intended for medical purposesthat is worn on the examiner's handsto prevent contamination betweenpatient and examiner.A patient examination glove is adisposable device intended for medicalpurposes that is worn on theexaminer's hand to preventcontamination between patient andexaminer.SameThickness, mm:
Powdered orPowered freePowdered freePowdered freeSameFinger0.10±0.03
Design FeatureambidextrousambidextrousSamePalm0.08±0.03
LabelingInformationSingle-use indication, powder free,device name, glove size and quantity,Nitrile Examination Gloves,Non-SterileSingle-use indication, powder free,device name, glove size and quantity,Nitrile Examination Gloves,Non-SterileSameCuff0.06±0.03
Chemotherapy DrugPermeation ClaimFluorouracil, Etoposide (Toposar),Cyclophosphamid (Cytoxan),Carmustine (BCNU), Thiotepa,Paclitaxel (Taxol), DoxorubicinHydrochloride, Dacarbazine (DTIC),Cisplatin, Carboplatin, Docetaxel,Ifosfamide, Irinotecan,Mechlorethamine HCL,Methotrexate, Mitomycin C,Mitoxantrone, Vincristine SulfateFluorouracil, Etoposide (Toposar),Cyclophosphamid (Cytoxan),Carmustine (BCNU), Thiotepa,Paclitaxel (Taxol), DoxorubicinHydrochloride, Dacarbazine (DTIC),Cisplatin, Ifosfamide, Mitoxantrone,Vincristine SulfateAnalysis 1Predicate DeviceSize: Min. 240 mm
Dermagrip Powder Free Blue Nitrile PatientExamination Gloves Tested for Use with ChemotherapyDrugs(K141982)Thickness: Finger (0.07-0.10); Palm(0.07-0.09); Cuff (0.06-0.08)
RemarkAnalysis 2

8. Substantially Equivalent (SE) Comparison Conclusion

Table 1 General Comparison

Analysis 1:

The proposed and predicate devices both have a tested for use with chemotherapy drugs claim. However, different chemotherapy drugs have been tested for the proposed device and the results meet the specifications of ASTM D6978

{8}------------------------------------------------

Table 2 Device Dimensions Comparison

Analysis 2:

The proposed device has different size specification as compared to the predicate device, but the proposed device meets the specifications of ASTM D6319.

{9}------------------------------------------------

ItemProposed DevicePOWDER FREE Blue Nitrile GLOVES,Tested for Use with Chemotherapy Drugs(K163146)Predicate DeviceDermagrip Powder FreeBlue Nitrile PatientExamination Gloves Testedfor Use with ChemotherapyDrugs(K141982)Remark
ColorantBlueBlueSame
PhysicalpropertiesBeforeAgingTensileStrength15 Mpa, minMeet the Requirements ofASTM D 6319Analysis 3
UltimateElongation500% minMeet the Requirements ofASTM D 6319
AfterAgingTensileStrength14 MPa, minMeet the Requirements ofASTM D 6319
UltimateElongation400% minMeet the Requirements ofASTM D 6319
Comply with ASTM D6319Comply with ASTM D631914 MPa. Min/500% min.before aging;14 MPa. Min./400% min.After agingSame
Freedom from HolesBe free from holes when tested inaccordance with ASTM D5151AQL 1.5Be free from holes whentested in accordance withASTM D5151under AQL 2.5/InspectionLevel G-ISame
Powder ContentMax. 0.32 mg per gloveMeet the requirements ofASTM D6319Less than 2mg per gloveSame

Table 3 Performance Comparison

Analysis 3:

The proposed device has a different Ultimate Elongation after aging as compared to the predicate device, but the proposed device meets the specifications of ASTM D6319.

{10}------------------------------------------------

ItemProposed DevicePOWDER FREE BlueNitrile GLOVES, Tested forUse with ChemotherapyDrugs(K163146)Predicate DeviceDermagrip Powder Free Blue NitrilePatient Examination Gloves Tested for Usewith Chemotherapy Drugs(K141982)Remark
MaterialNitrileNitrileSame
BiocompatibilityIrritationUnder the conditions of thestudy, not an irritantComply with ISO 10993-10Same
SensitizationUnder the conditions of thestudy, not a sensitizer
Label and LabelingMeet FDA'sRecommendationsMeet FDA's RecommendationsSame

Table 4 Safety Comparison

9. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device, POWDER FREE Blue Nitrile GLOVES, Tested for Use with Chemotherapy Drugs, is as safe, as effective, and performs as well as or better than the legally marketed predicate device, Dermagrip Powder Free Blue Nitrile Patient Examination Gloves Tested for Use with Chemotherapy Drugs.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.