Intended Use

The Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and reusable in infant, pediatric but not neonatal, and adult sizes.

Device Description

The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems, and wrap around the patient's limb. The cuff is non-sterile and reusable. The model difference of the proposed device includes bladder type, tube, color, joint model and size.

AI/ML Overview

The given text is a 510(k) Summary for a Reusable Blood Pressure Cuff, which is a submission to the FDA to demonstrate substantial equivalence to previously marketed devices. This document focuses on proving that the new device is as safe and effective as existing products, primarily through non-clinical testing and comparison to predicate devices, rather than an extensive study proving new performance criteria for a novel device or an AI algorithm.

Therefore, many of the requested details, such as "number of experts used to establish ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," and "sample size for training set," are not applicable to this type of regulatory submission and product. The device is a physical medical accessory, not an AI algorithm.

I will extract the available and relevant information.


Acceptance Criteria and Device Performance for Reusable Blood Pressure Cuff (K171889)

The acceptance criteria for the Reusable Blood Pressure Cuff (K171889) are established by demonstrating compliance with recognized international standards and substantial equivalence to predicate devices. The study conducted was primarily non-clinical testing to verify design specifications and biological compatibility.

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria / Performance AspectStandard/RequirementReported Device PerformanceComments on Performance
Basic Safety and Essential PerformanceIEC 80601-2-30: 2013Complies
Biocompatibility - CytotoxicityISO 10993-5: 2009Complies
Biocompatibility - Irritation & Skin SensitizationISO 10993-10: 2010Complies
Intended Use EquivalenceEquivalent to Predicate Device 1 (K133117) & Predicate Device 2 (K152468)Stated as same or slight differences that do not impact safety/effectivenessThe proposed devices include similar intended uses (accessory for noninvasive blood pressure measurement systems, non-sterile, reusable, various sizes). Differences like extended tube configurations or slight size variations are deemed not to affect safety or effectiveness.
Device ClassClass II (Regulatory)Class II
Product CodeDXQDXQ
Regulation Number21 CFR 870.112021 CFR 870.1120
Patient PopulationsAdult (for internal bladder), Infant/Pediatric/Adult (for integrated bladder)Matches Predicate Devices' populationsThe proposed device expands to include infant/pediatric populations for integrated bladder cuffs, which is aligned with Predicate Device 2.
Tube ConfigurationOne or two tubeTwo tube configuration as well as one tubeBroader than Predicate 1 (one tube only), but matched by Predicate 2 (one or two tubes). Not considered a safety/effectiveness issue.
Bladder TypeInternal Bladder / Integrated BladderBoth are offered by the proposed devicePredicate 1 uses Internal Bladder, Predicate 2 uses Integrated Bladder. The proposed device offers both variations.
ReusableYesYes
SterileNoNo
Pressure Range0-300mmHg0-300mmHgMatches Predicate Devices.
MaterialCompliant with Biocompatibility StandardsPolyester (internal bladder), TPU leather with nylon (integrated bladder)Different materials than predicate devices, but compliance with ISO 10993-5 and -10 confirmed.
Size Range (Internal Bladder)e.g., Small adult: 17-22cm; Large Adult: 30-42 cmStated as "included in the range of the predicate device 1"User manual indicates clear selection based on size. Not considered a safety/effectiveness issue.
Size Range (Integrated Bladder)e.g., Infant: 9-14.8cm; Child: 13.8-21.5cm; Thigh: 45-56.5cmStated as "different to that of predicate device 2," but "slight"User manual indicates clear selection based on size. Not considered a safety/effectiveness issue.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify sample sizes for the non-clinical tests (e.g., how many cuffs were tested for biocompatibility or performance). The data provenance is not detailed beyond the manufacturer being Suzhou Minhua Medical Apparatus Supplies Co., LTD in China, implying testing was likely conducted in China. The study refers to "non-clinical tests" and "test results" and "laboratory tests," which are inherently prospective for the specific devices being tested.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. The device is a physical medical accessory, and the "ground truth" for its performance is established by compliance with international engineering and biocompatibility standards, not by expert consensus on diagnostic interpretations.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. Adjudication methods are typically used in clinical studies involving interpretation (e.g., imaging reads) where there can be differing opinions among experts. This product's evaluation is based on objective measurements against engineering and biological standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. MRMC studies are relevant for diagnostic devices, particularly those involving human interpretation of data/images, often with AI assistance. This device is a blood pressure cuff and does not involve "readers" or "AI assistance" in that context.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical blood pressure cuff, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the Reusable Blood Pressure Cuff is compliance with:

  • Engineering Standards: IEC 80601-2-30: 2013 (Medical Electrical Equipment - Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers).
  • Biocompatibility Standards: ISO 10993-5: 2009 (Tests For In Vitro Cytotoxicity) and ISO 10993-10: 2010 (Tests For Irritation And Skin Sensitization).
  • Substantial Equivalence: Comparison to legally marketed predicate devices (K133117 and K152468) regarding intended use, principles of operation, materials, and performance specifications.

8. The sample size for the training set

This information is not applicable. "Training set" refers to data used to train algorithms, primarily in machine learning or AI. This device is a physical medical accessory.

9. How the ground truth for the training set was established

This information is not applicable as there is no "training set" for this physical device.

{0}------------------------------------------------

December 19, 2017

Suzhou Minhua Medical Apparatus Supplies Co., LTD % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CN

Re: K171889

Trade/Device Name: Reusable Blood Pressure Cuff, with list of 164 models attached Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: November 10, 2017 Received: November 15, 2017

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Diana Hong

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Wilhelm
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

List of 164 models

No.Product codeNameTubeArm size
1MH002RCP10-1Small adultOne17-22cm
2MH002RCP10-2Small adulttwo17-22cm
3MH002RCP11-1Small adultlongOne22-30 cm
4MH002RCP11-2Small adultlongtwo22-30 cm
5MH002RCP12-1AdultOne22-30 cm
6MH002RCP12-2Adulttwo22-30 cm
7MH002RCP14-1Large AdultOne30-42 cm
8MH002RCP14-2Large Adulttwo30-42 cm
9MH002RCP15-1Large AdultlongOne42-48 cm
10MH002RCP15-2Large Adultlongtwo42-48 cm

Table 1 Specifications of Reusable Blood Pressure Cuff – with Internal Bladder

No.Product codeNameTubeArm size
1MH002RCT06HP1InfantOne9-14.8cm
2MH002RCT06HP2Infanttwo9-14.8cm
3MH002RCT07HP1Infant longOne9-14.8cm
4MH002RCT07HP2Infant longtwo9-14.8cm
5MH002RCT08HP1ChildOne13.8-21.5cm
6MH002RCT08HP2Childtwo13.8-21.5cm
7MH002RCT09HP1Child longOne13.8-21.5cm
8MH002RCT09HP2Child longtwo13.8-21.5cm
9MH002RCT10HP1Small adultOne20.5-28.5cm
10MH002RCT10HP2Small adulttwo20.5-28.5cm
11MH002RCT11HP1Small adultlongOne20.5-28.5cm
12MH002RCT11HP2Small adultlongtwo20.5-28.5cm
13MH002RCT12HP1AdultOne27.5-36.5cm
14MH002RCT12HP2Adulttwo27.5-36.5cm
15MH002RCT13HP1Adult longOne27.5-36.5cm
16MH002RCT13HP2Adult longtwo27.5-36.5cm

{3}------------------------------------------------

17MH002RCT14HP1Big adultOne35.5-46cm
18MH002RCT14HP2Big adulttwo35.5-46cm
19MH002RCT15HP1Big adultlongOne35.5-46cm
20MH002RCT15HP2Big adultlongtwo35.5-46cm
21MH002RCT16HP1ThighOne45-56.5cm
22MH002RCT16HP2Thightwo45-56.5cm

Table 3 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (SL)

No.Product codeNameTubeArm size
1MH002RCT06SL1InfantOne9-14.8cm
2MH002RCT06SL2Infanttwo9-14.8cm
3MH002RCT07SL1Infant longOne9-14.8cm
4MH002RCT07SL2Infant longtwo9-14.8cm
5MH002RCT08SL1ChildOne13.8-21.5cm
6MH002RCT08SL2Childtwo13.8-21.5cm
7MH002RCT09SL1Child longOne13.8-21.5cm
8MH002RCT09SL2Child longtwo13.8-21.5cm
9MH002RCT10SL1Small adultOne20.5-28.5cm
10MH002RCT10SL2Small adulttwo20.5-28.5cm
11MH002RCT11SL1Small adultlongOne20.5-28.5cm
12MH002RCT11SL2Small adultlongtwo20.5-28.5cm
13MH002RCT12SL1AdultOne27.5-36.5cm
14MH002RCT12SL2Adulttwo27.5-36.5cm
15MH002RCT13SL1Adult longOne27.5-36.5cm
16MH002RCT13SL2Adult longtwo27.5-36.5cm
17MH002RCT14SL1Big adultOne35.5-46cm
18MH002RCT14SL2Big adulttwo35.5-46cm
19MH002RCT15SL1Big adultlongOne35.5-46cm
20MH002RCT15SL2Big adultlongtwo35.5-46cm
21MH002RCT16SL1ThighOne45-56.5cm
22MH002RCT16SL2Thightwo45-56.5cm

{4}------------------------------------------------

No.Product codeNameTubeArm size
1MH002RCT06MQ1InfantOne9-14.8cm
2MH002RCT06MQ2Infanttwo9-14.8cm
3MH002RCT07MQ1Infant longOne9-14.8cm
4MH002RCT07MQ2Infant longtwo9-14.8cm
5MH002RCT08MQ1ChildOne13.8-21.5cm
6MH002RCT08MQ2Childtwo13.8-21.5cm
7MH002RCT09MQ1Child longOne13.8-21.5cm
8MH002RCT09MQ2Child longtwo13.8-21.5cm
9MH002RCT10MQ1Small adultOne20.5-28.5cm
10MH002RCT10MQ2Small adulttwo20.5-28.5cm
11MH002RCT11MQ1Small adult longOne20.5-28.5cm
12MH002RCT11MQ2Small adult longtwo20.5-28.5cm
13MH002RCT12MQ1AdultOne27.5-36.5cm
14MH002RCT12MQ2Adulttwo27.5-36.5cm
15MH002RCT13MQ1Adult longOne27.5-36.5cm
16MH002RCT13MQ2Adult longtwo27.5-36.5cm
17MH002RCT14MQ1Big adultOne35.5-46cm
18MH002RCT14MQ2Big adulttwo35.5-46cm
19MH002RCT15MQ1Big adult longOne35.5-46cm
20MH002RCT15MQ2Big adult longtwo35.5-46cm
21MH002RCT16MQ1ThighOne45-56.5cm
22MH002RCT16MQ2Thightwo45-56.5cm

Table 4 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (MQ)

Table 5 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (DM)

No.Product codeNameTubeArm size
1MH002RCT06DM1InfantOne9-14.8cm
2MH002RCT06DM2Infanttwo9-14.8cm
3MH002RCT07DM1Infant longOne9-14.8cm
4MH002RCT07DM2Infant longtwo9-14.8cm
5MH002RCT08DM1ChildOne13.8-21.5cm
6MH002RCT08DM2Childtwo13.8-21.5cm
7MH002RCT09DM1Child longOne13.8-21.5cm
8MH002RCT09DM2Child longtwo13.8-21.5cm

{5}------------------------------------------------

9MH002RCT10DM1Small adultOne20.5-28.5cm
10MH002RCT10DM2Small adulttwo20.5-28.5cm
11Small adultOne20.5-28.5cm
MH002RCT11DM1long
12MH002RCT11DM2Small adulttwo20.5-28.5cm
long
13MH002RCT12DM1AdultOne27.5-36.5cm
14MH002RCT12DM2Adulttwo27.5-36.5cm
ા રMH002RCT13DM1Adult longOne27.5-36.5cm
16MH002RCT13DM2Adult longtwo27.5-36.5cm
17MH002RCT14DM1Big adultOne35.5-46cm
18MH002RCT14DM2Big adulttwo35.5-46cm
19MH002RCT15DM1Big adultOne35.5-46cm
long
20Big adulttwo35.5-46cm
MH002RCT15DM2long
21MH002RCT16DM1ThighOne45-56.5cm
22MH002RCT16DM2Thightwo45-56.5cm
Table 6 Specifications of Reusable Blood Pressure Cuff - with Integrated Bladder (NA)
---------------------------------------------------------------------------------------
No.Product codeNameTubeArm size
1MH002RCT06NA1InfantOne9-14.8cm
2MH002RCT06NA2Infanttwo9-14.8cm
3MH002RCT07NA1Infant longOne9-14.8cm
4MH002RCT07NA2Infant longtwo9-14.8cm
5MH002RCT08NA1ChildOne13.8-21.5cm
6MH002RCT08NA2Childtwo13.8-21.5cm
7MH002RCT09NA1Child longOne13.8-21.5cm
8MH002RCT09NA2Child longtwo13.8-21.5cm
9MH002RCT10NA1Small adultOne20.5-28.5cm
10MH002RCT10NA2Small adulttwo20.5-28.5cm
11MH002RCT11NA1Small adultlongOne20.5-28.5cm
12MH002RCT11NA2Small adultlongtwo20.5-28.5cm
13MH002RCT12NA1AdultOne27.5-36.5cm
14MH002RCT12NA2Adulttwo27.5-36.5cm
15MH002RCT13NA1Adult longOne27.5-36.5cm
16MH002RCT13NA2Adult longtwo27.5-36.5cm

{6}------------------------------------------------

17MH002RCT14NA1Big adultOne35.5-46cm
18MH002RCT14NA2Big adulttwo35.5-46cm
19MH002RCT15NA1Big adultOne35.5-46cm
long
20Big adulttwo35.5-46cm
MH002RCT15NA2long
21MH002RCT16NA1ThighOne45-56.5cm
22MH002RCT16NA2Thightwo45-56.5cm

Table 7 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (FL)

No.Product codeNameTubeArm size
1MH002RCT06FL1InfantOne9-14.8cm
2MH002RCT06FL2Infanttwo9-14.8cm
3MH002RCT07FL1Infant longOne9-14.8cm
4MH002RCT07FL2Infant longtwo9-14.8cm
5MH002RCT08FL1ChildOne13.8-21.5cm
6MH002RCT08FL2Childtwo13.8-21.5cm
7MH002RCT09FL1Child longOne13.8-21.5cm
8MH002RCT09FL2Child longtwo13.8-21.5cm
9MH002RCT10FL1Small adultOne20.5-28.5cm
10MH002RCT10FL2Small adulttwo20.5-28.5cm
11MH002RCT11FL1Small adult longOne20.5-28.5cm
12MH002RCT11FL2Small adult longtwo20.5-28.5cm
13MH002RCT12FL1AdultOne27.5-36.5cm
14MH002RCT12FL2Adulttwo27.5-36.5cm
15MH002RCT13FL1Adult longOne27.5-36.5cm
16MH002RCT13FL2Adult longtwo27.5-36.5cm
17MH002RCT14FL1Big adultOne35.5-46cm
18MH002RCT14FL2Big adulttwo35.5-46cm
19MH002RCT15FL1Big adult longOne35.5-46cm
20MH002RCT15FL2Big adult longtwo35.5-46cm
21MH002RCT16FL1ThighOne45-56.5cm
22MH002RCT16FL2Thightwo45-56.5cm

{7}------------------------------------------------

No.Product codeNameTubeArm size
1MH002RCT06MF1InfantOne9-14.8cm
2MH002RCT06MF2Infanttwo9-14.8cm
3MH002RCT07MF1Infant longOne9-14.8cm
4MH002RCT07MF2Infant longtwo9-14.8cm
5MH002RCT08MF1ChildOne13.8-21.5cm
6MH002RCT08MF2Childtwo13.8-21.5cm
7MH002RCT09MF1Child longOne13.8-21.5cm
8MH002RCT09MF2Child longtwo13.8-21.5cm
9MH002RCT10MF1Small adultOne20.5-28.5cm
10MH002RCT10MF2Small adulttwo20.5-28.5cm
11MH002RCT11MF1Small adult longOne20.5-28.5cm
12MH002RCT11MF2Small adult longtwo20.5-28.5cm
13MH002RCT12MF1AdultOne27.5-36.5cm
14MH002RCT12MF2Adulttwo27.5-36.5cm
15MH002RCT13MF1Adult longOne27.5-36.5cm
16MH002RCT13MF2Adult longtwo27.5-36.5cm
17MH002RCT14MF1Big adultOne35.5-46cm
18MH002RCT14MF2Big adulttwo35.5-46cm
19MH002RCT15MF1Big adult longOne35.5-46cm
20MH002RCT15MF2Big adult longtwo35.5-46cm
21MH002RCT16MF1ThighOne45-56.5cm
22MH002RCT16MF2Thightwo45-56.5cm

Table 8 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (MF)

{8}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K171889

Device Name Reusable Blood Pressure Cuff

Indications for Use (Describe)

The Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and reusable in infant, pediatric but not neonatal, and adult sizes.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{9}------------------------------------------------

510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: __K171889

    1. Date of Preparation: 12/18/2017
    1. Sponsor Identification

Suzhou Minhua Medical Apparatus Supplies Co., LTD You Yi Industrial Park, Songlin Town, Wujiang, 215222, CHINA

Establishment Registration Number: 3011529836

Contact Person: Yang Qidong Position: Vice general manager Tel: +86-21-59840221 Fax: +86-21-59841086 Email: QD@minhuayiliao.com

    1. Designated Submission Correspondent
      Ms. Diana Hong (Primary Contact Person) Mr. Betty Xiao (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

{10}------------------------------------------------

4. Identification of Proposed Device

Trade Name: Reusable Blood Pressure Cuff Common Name: Blood Pressure Cuff

Regulatory Information Classification Name: Blood Pressure Cuff Classification: 2 Product Code: DXQ Regulation Number:870.1120 Review Panel: Cardiovascular

Intended Use Statement:

The Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in infant, pediatric but not neonatal, and adult sizes.

Device Description

The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems, and wrap around the patient's limb. The cuff is non-sterile and reusable.

The model difference of the proposed device includes bladder type, tube, color, joint model and size.

ર. Identification of Predicate Device(s)

Predicate Device 1 510(k) Number: K133117 Product Name: Andon Blood Pressure Cuff

Predicate Device 2 510(k) Number: K152468 Product Name: Reusable Blood Pressure Cuff

6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

{11}------------------------------------------------

  • IEC 80601-2-30: 2013, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.

  • ISO 10993-5: 2009, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.

  • A ISO 10993-10: 2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
    1. Clinical Test Conclusion

No clinical study is included in this submission.

  • Substantially Equivalent (SE) Comparison 8.
ItemProposed DevicePredicate Device 1
Cuff with Internal BladderK133117
Device class22
Product codeDXQDXQ
Regulation Number21 CFR 870.112021 CFR 870.1120
Intended useThe Reusable Blood Pressure Cuff isan accessory used in conjunction withnoninvasive blood pressuremeasurement systems. The cuff isnon-sterile and reusable. It isavailable in infant, pediatric but notneonatal, and adult sizes.Blood Pressure Cuffs are intendedto be wrapped on the upper armand used with a non-invasive bloodpressure monitor system tocomplete the measurement ofblood parameters on adults.
Patient populationsadultadult
Tube configurationOne or two tubeOne tube
BladderInternal BladderInternal Bladder
Principles ofOperationBladder is wrapped around thepatient's limb and secured by hookand loop closure. Air hose isconnected to the noninvasive bloodpressure measurement systems.Bladder is wrapped around thepatient's limb and secured by hookand loop closure. Air hose isconnected to the noninvasive bloodpressure measurement systems.
SizeSmall adult: 17-22cmSmall adult long / Adult:22-30 cmLarge Adult: 30-42 cmLarge Adult long: 42-48 cmKD-525-P34: 15-24cmKD-525-P31: 20-34 cmKD-525-P32: 30-44 cmKD-525-P33: 40-48 cm

Table 1 Comparison for Cuff with Internal Bladder

{12}------------------------------------------------

ReusableyesYes
SterileNoNo
Pressure range0-300mmHg0-300mmHg
MaterialPolyesterNot available
BiocompatibilityComply with ISO 10993-5 and ISO10993-10.Comply with ISO 10993-5 andISO 10993-10.

There are four differences for the proposed cuff with internal bladder and predicate device 1.

For intended use.

The proposed devices include two types of bladder. The cuffs with integrated bladder include with sizes of infant, child and adult. But the cuffs with internal bladder only include adult size. The patient population of proposed cuffs with internal bladder include adult only. The intended use of proposed cuffs with internal bladder and predicate device 1 is same actually.

For tube configuration,

The proposed devices include one two tubes, while the predicate device only includes one tube. More tube configuration can be applicable for more types of noninvasive blood pressure measurement system. And the two tubes configuration is very common for blood pressure cuff. The predicate device 2, Reusable Blood Pressure Cuff, includes two tubes configuration. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

For size:

The cuff size range of proposed devices is included in the range of the predicate device 1. And the size is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

For material:

The material of proposed device and predicate device 1 may be different, but both of them comply with ISO 10993-5 and ISO 10993-10. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

ITEMProposed DevicePredicate Device 2
Cuff with Integrated BladderK152468
Device class22
Product codeDXQDXQ
RegulationNumber21 CFR 870.112021 CFR 870.1120
Intended useThe Reusable Blood Pressure CuffThe reusable blood pressure cuff is an
is an accessory used in conjunctionaccessory used in conjunction with
with noninvasive blood pressurebloodnoninvasivepressure
measurement systems. The cuff is
measurement systems. The cuff is
non-sterile and reusable. Itisnon-sterile and may be reused. It is
available in infant, pediatric but notavailable in infant, pediatric and adult
neonatal, and adult sizes.sizes
PatientInfant / pediatric / adultInfant / pediatric / adult
populations
TubeOne or two tubeOne or two tube
configuration
BladderIntegrated BladderIntegrated Bladder
PrinciplesofBladder is wrapped aroundtheBladder is wrapped aroundthe
Operationpatient's limb and secured by hookpatient's limb and secured by hook
and loop closure. Air hose isand loop closure. Air hose is
connected to the noninvasive bloodconnected to the noninvasive blood
pressure measurement systems.pressure measurement systems.
SizeInfant / infant long: 9-14.8cmInfant: 8-13 cm
Child /child long: 13.8-21.5cmChild: 12-19 cm
Small adult / small adult long:Small adult: 17-25cm
20.5-28.5cm
Small adult long / Adult:Adult: 23-33cm
27.5-36.5cm
Big adult /big adultlong:Large adult / largeadultlong:
35.5-46cm31-40cm
Thigh: 45-56.5cmThigh: 38-50cm
ReusableyesYes
SterileNoNo
Pressure range0-300mmHg0-300mmHg
MaterialTPU leather with nylon210D Nylon TPU and PVC
BiocompatibilityComply with ISO 10993-5 and ISOComply with ISO 10993-5 and ISO
10993-10.10993-10.

Table 2 Comparison for Cuff with Integrated Bladder

{13}------------------------------------------------

There are two differences for the proposed cuff with integrated bladder and predicate device 2.

For size,

The cuff size range of proposed devices is different to that of predicate device 2. But the different is slight. And the size is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

{14}------------------------------------------------

For material:

The material of proposed device and predicate device 2 is different, but both of them comply with ISO 10993-5 and ISO 10993-10. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

    1. Substantially Equivalent (SE) Conclusion
      Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).