K Number
K133117

Validate with FDA (Live)

Date Cleared
2014-03-04

(155 days)

Product Code
Regulation Number
870.1120
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Blood Pressure Cuffs are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.

Device Description

The proposed device, Blood Pressure Cuff, is a rectangle soft inelastic sleeve reusable with a bladder. There is a single-tube connected to the bladder and the Non Invasive Blood Pressure Monitor for inflating and deflating. The device should connected to a Non Invasive Blood Pressure Monitor to complete the function. There is various sizes for different arm range as follows:

ModelArm range
KD-525-P3120cm-34cm
KD-525-P3230cm-44cm
KD-525-P3340cm-48cm
KD-525-P3415cm-24cm
KD-5917-P4222cm-42cm
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Andon Blood Pressure Cuff:

Acceptance Criteria and Device Performance

The provided text for the Andon Blood Pressure Cuff (K133117) states that "the device met all applicable requirements of the standard" for its clinical tests. However, it does not explicitly provide a table of specific numerical acceptance criteria and the reported device performance against those criteria. Instead, it refers to compliance with standards.

Here's a breakdown of the information available:

Acceptance Criteria (Standard Compliance)Reported Device Performance
IEC 80601-2-30 (Clinical Test)Met all applicable requirements of the standard.
IEC 80601-2-30 (Non-clinical Test)Safety and performance characteristics tested. No new questions of safety/effectiveness raised.
ISO 10993-1, ISO 10993-5, ISO 10993-10Biocompatibility tests performed and conform to standards. No new questions of safety/effectiveness raised.

Study Details

2. Sample size used for the test set and data provenance

  • Sample Size for Test Set: Not explicitly stated. The document mentions "Clinical test has been done in accordance with IEC 80601-2-30...". IEC 80601-2-30 Section 201.12.1.101 (Clinical investigation of AUTOMATED SPHYGMOMANOMETERS) typically specifies a minimum of 85 subjects for accuracy evaluation. Without the full standard report, the exact number used in this specific study cannot be determined from the provided text.
  • Data Provenance: Not explicitly stated. It's not mentioned whether the data was from a specific country or if it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and qualifications of those experts

  • This information is not provided in the given text. For blood pressure cuff validation according to standards like IEC 80601-2-30, ground truth is typically established by trained observers (auscultation method) using a reference sphygmomanometer, but the number and qualifications of such observers are not detailed here.

4. Adjudication method for the test set

  • This information is not provided in the given text.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC comparative effectiveness study was done or mentioned. This device is a blood pressure cuff, a passive medical device, not an AI-powered diagnostic tool that human readers would interact with.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, a standalone performance assessment was conducted implicitly. The clinical and non-clinical tests, particularly the clinical test in accordance with IEC 80601-2-30, evaluate the device's performance (i.e., its ability to measure blood pressure) as a standalone component when used with an NIBP monitor. The "algorithm" here refers to the cuff's physical characteristics and how it interfaces with the NIBP monitor to facilitate the measurement.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Based on the reference to IEC 80601-2-30, the ground truth for blood pressure measurement studies typically involves:
    • Simultaneous auscultation: Trained observers (experts) using a mercury sphygmomanometer or another validated reference device on the opposite arm or using a Y-connector with a reference device. This method provides the reference systolic and diastolic blood pressure values.

8. The sample size for the training set

  • Not applicable/Not provided. This device is a blood pressure cuff, not an AI/machine learning model that would require a distinct "training set." The tests performed are for validation of the physical device's performance against established standards.

9. How the ground truth for the training set was established

  • Not applicable/Not provided. As explained above, a training set associated with ground truth is not relevant for this type of device and its validation.

{0}------------------------------------------------

f

510(k) Summary

K133117 page 1 of 3

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information

Name:Andon Health Co., Ltd.
Address:No 3, Jinping Street Ya An Road, Nankai District, Tianjin,
P.R. China
Phone number:86-22-6052 6161
Fax number:86-22-6052 6162
Contact:Liu Yi
Date of Application: 09/25/2013

2.0 Device information

Trade name: Andon Blood Pressure Cuff Blood Pressure Cuff Device name: Classification name: Blood pressure Cuff

3.0 Classification

Production code: DXQ Regulation number: 870.1120 Classification: II Panel: Cardiovascular

4.0 Predicate device information

1Manufacturer:APK Technology Co, Ltd
Device:Blood Pressure Cuff
510(k) number:K102825

{1}------------------------------------------------

5.0 Intended use

Blood Pressure Cuffs are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.

6.0 Device description

The proposed device, Blood Pressure Cuff, is a rectangle soft inelastic sleeve reusable with a bladder. There is a single-tube connected to the bladder and the Non Invasive Blood Pressure Monitor for inflating and deflating. The device should connected to a Non Invasive Blood Pressure Monitor to complete the function. There is various sizes for different arm range as follows:

ModelArm range
KD-525-P3120cm-34cm
KD-525-P3230cm-44cm
KD-525-P3340cm-48cm
KD-525-P3415cm-24cm
KD-5917-P4222cm-42cm

7.0 Summary comparing technological characteristics with predicate device

Technological CharacteristicsComparison result
Design principleIdentical
AppearanceSimilar
Patients contact MaterialsIdentical
PerformanceSimilar
BiocompatibilityIdentical
FunctionIdentical

8.0 Discussion of non-clinical and clinical test performed

Non-clinical Tests have been done as follows:

a. Safety and performance characteristics of the test according to IEC 80601-2-30

b. Biocompatibility test has been performed according to ISO 10993-5 and ISO 10993-10

None of the test demonstrates that the Blood Pressure Cuffs bring new questions of safety and effectiveness.

{2}------------------------------------------------

Clinical Test Concerning the Compliance of ANSI/AAMI SP10

Clinical test has been done in accordance with IEC 80601-2-30-Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers, and the test result shows, the device met all applicable requirements of the standard.

9.0 Performance summary

The new cuffs conform to the following standards:

  • · IEC 80601-2-30: Particular requirements for the basic safety and performance of of of automated non-invasive essential sphygmomanometers
  • · ISO 10993-1: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk manangement process
  • · ISO 10993-5: Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytoto xicity.
  • · ISO 10993-10: Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization

10.0 Comparison to the predicate device and the conclusion

Compared to the predicate device, the new cuffs are very similar in the intended use, the design principle, the material, the performance and the applicable standards.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

So we claims the proposed cuffs to be Substantially Equivalent (SE) to the predicate devices Blood Pressure Cuff(K102825).

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

DEPARTMENT OF HEALTH & HUMAN SERVICES - USA

[Image of the Department of Health and Human Services logo]

March 4, 2014

Andon Health Co., Ltd Liu Yi No. 3 Jinping Street, Ya An Road, Nankai District Tianjin, 300190 CN

Re: K133117

Trade/Device Name: Andon Blood Pressure Cuff (Model: KD-525-P31, KD-525-P32, KD-525-P33, KD-525-P34, KD-5917-P42) Regulation Number: 21 CFR 870.1120 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXQ Dated: January 2, 2014 Received: January 6, 2014

Dear Liu Yi,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems

{4}------------------------------------------------

Page 2 - Liu Yi

(OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Andon Blood Pressure Cuff FDA 510(k) Files

Statement of Indications for Use

K133117 510(k) Number :

Andon Blood Pressure Cuff (Model: KD-525-P31, Device name: KD-525-P33, KD-525-P34, KD-525-P32, KD-5917-P42)

Indications for use:

Blood Pressure Cuffs are intended to be wrapped on the upper arm and used with a non-invasive blood pressure monitor system to complete the measurement of blood parameters on adults.

AND/OR Over-The-Counter Use YES Prescription use (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Date:
FDA 2024.03.04
14:44:16 -05'00'
for Bram Zuckerman

Page 1 of 1

§ 870.1120 Blood pressure cuff.

(a)
Identification. A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.(b)
Classification. Class II (performance standards).