K Number
K171738

Validate with FDA (Live)

Device Name
Supria True64
Date Cleared
2017-08-18

(67 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes.

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface.

The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Description

The Supria True64 is a multi-slice computed tomography system designed to perform multi-slice CT scanning supported by 64-detector technology. The system allows optimum clinical applications ranging from routine exams in response to the diversified circumstances in imaging whole body regions.

The Supria True64 system uses 64-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.

The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.

The Supria True64 system consists of a Gantry, Operator's Workstation, Patient Table, High-Frequency X-ray Generator, and accessories.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Important Note: This submission is a 510(k) premarket notification for a new version of an existing device (Supria True64 CT system). The primary goal of a 510(k) is to demonstrate substantial equivalence to a previously legally marketed device (predicate device), not necessarily to establish novel clinical efficacy or new performance benchmarks against a specific disease. Therefore, the "acceptance criteria" here are largely focused on maintaining equivalent performance to the predicate and adhering to relevant standards for CT systems. The studies are primarily to confirm this equivalence.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a numerical or target-based fashion for all parameters. Instead, it relies heavily on demonstrating that the Supria True64's performance is "similar to" or "comparable to" the predicate device (HITACHI SUPRIA Whole-body X-ray CT System Phase 3, K163528) and complies with relevant industry standards.

Here's a table summarizing the implicit acceptance criteria and reported performance, derived from the "Performance Comparison" and "Technological Characteristic Differences" sections:

Testing Type / CharacteristicImplicit Acceptance CriteriaReported Device Performance (Supria True64)
Bench Validation TestingNo change / Substantial equivalence to predicate for specified parameters"No change about the following performance: Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, CT dose index."
Dose ProfileEquivalent to predicateEquivalent to predicate (implied, as "no change")
NoiseEquivalent to predicateEquivalent to predicate
Mean CT number & UniformityEquivalent to predicateEquivalent to predicate
Spatial ResolutionEquivalent to predicate; 0.35 mm (high-contrast)0.35 mm (high-contrast); Equivalent to predicate
Tomographic Section ThicknessEquivalent to predicateEquivalent to predicate
Sensitivity ProfileEquivalent to predicateEquivalent to predicate
Tomographic Plane LocationEquivalent to predicateEquivalent to predicate
CT dose indexEquivalent to predicateEquivalent to predicate
Low-contrast resolution2.5 mm @ 0.25% at <4 rads (match predicate)2.5 mm @ 0.25%
10% MTF14.7 lp/cm (match predicate)14.7 lp/cm
50% MTF12.2 lp/cm (match predicate)12.2 lp/cm
Clinical Validation TestingSufficient to judge clinical usability; low dose & high quality via IR"3 kinds of clinical image example which we judged to be sufficient to judge a clinical usability." and "images... realized both low dose and high quality through reduction of image noise and artifacts" (via FBP and Iterative Reconstruction comparison).
Technological CharacteristicsMaintain equivalence or minor differences that do not impact safety/effectivenessMostly "No" difference from the predicate, except for: - Detector: 64 rows added - Image Storage: Raw data disk for 64 slices added (larger capacity) - Helical Beam Pitch: Fine-tuned for 64 lines, new values provided
Standards ConformanceCompliance with specified IEC, AAMI, NEMA standardsConforms to AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-44, NEMA XR 25, NEMA XR26, IEC 62304.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document does not specify a numerical sample size for the "clinical image examples." It only states "3 kinds of clinical image example."
  • Data Provenance: The document does not explicitly state the country of origin or if the data was retrospective or prospective. Given the context of a 510(k) submission for a new device model, it's highly likely these "image examples" were selected from internal company testing or possibly from clinical partners, but specifics are not provided.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. The document only states, "...which we judged to be sufficient to judge a clinical usability." This implies internal judgment, but no details on who made this judgment or their qualifications are given.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not specified. The phrase "we judged to be sufficient" suggests an internal decision, but no formal adjudication process (like 2+1, 3+1 consensus) is outlined.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No, a formal MRMC comparative effectiveness study is not mentioned. The clinical evaluation focuses on qualitative judgment of "clinical usability" and a comparison of FBP vs. Iterative Reconstruction for image quality, not on human reader performance with or without AI assistance.

6. Standalone Performance Study (Algorithm Only)

  • Standalone Performance Study: Not explicitly. The device itself is an imaging system, not an AI algorithm performing a specific diagnostic task. The performance evaluation is for the imaging system's inherent capabilities (e.g., spatial resolution, noise, dose). The "Iterative Reconstruction" is an image processing algorithm, and its effect on image quality was evaluated, but not necessarily in a standalone study assessing its diagnostic accuracy independently.

7. Type of Ground Truth Used

  • Type of Ground Truth: For the "clinical image examples," the ground truth establishment method is not detailed beyond "we judged to be sufficient to judge a clinical usability." For the bench testing, the ground truth is based on physical phantom measurements and established engineering metrics for CT performance (e.g., measuring noise in a uniform phantom, spatial resolution using a line pair phantom).

8. Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This document describes a CT scanner (hardware and its integrated software for image acquisition and reconstruction), not a machine learning model that requires a "training set" in the conventional sense of AI/ML. The iterative reconstruction algorithms would have been developed and "trained" (in a broader sense of algorithm development and tuning) using various datasets, but these details are not provided in this 510(k) summary.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as explained above. The development of the CT system's underlying physics models and reconstruction algorithms would rely on established scientific principles and engineering validation, rather than a "ground truth" derived from expert consensus on medical images that is typical for AI models.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 18, 2017

Hitachi Healthcare Americas Corporation % Mr. Douglas J. Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087

Re: K171738

Trade/Device Name: Supria True64 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: July 26, 2017 Received: July 28, 2017

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171738

Device Name Supria True64 Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

Indications for Use (Describe)

The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes.

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

{4}------------------------------------------------

Submitter Information

Submitter:Hitachi Healthcare Americas1959 Summit Commerce ParkTwinsburg, Ohio 44087-2371
Contact:Douglas J. Thistlethwaite
Telephone number:330-425-1313
Telephone number:330-963-0749
E-mail:thistlethwaited@hitachihealthcare.com
Date:April 18, 2017

Subject Device Name

Trade/Proprietary Name:Supria True64
Regulation Number:21 CFR 892.1750
Regulation Name:Computed tomography x-ray system
Product CodeJAK, System, X-Ray, Tomography, Computed
ClassII
PanelRadiology

Predicate Device Name

Predicate Device(s):HITACHI SUPRIA Whole-body X-ray CT System Phase 3 (K163528)
Regulation Number:21 CFR 892.1750
Regulation Name:Computed tomography x-ray system
Product CodeJAK, System, X-Ray, Tomography, Computed
ClassII
PanelRadiology

Device Intended Use

The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dvnamic modes,

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface.

The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Description

Function

The Supria True64 is a multi-slice computed tomography system designed to perform multi-slice CT scanning supported by 64-detector technology. The system allows optimum clinical applications ranging from routine exams in response to the diversified circumstances in imaging whole body regions.

{5}------------------------------------------------

Scientific Concepts

The Supria True64 system uses 64-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.

The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.

Physical and Performance Characteristics

The Supria True64 system consists of a Gantry, Operator's Workstation, Patient Table, High-Frequency X-ray Generator, and accessories. The system performance is similar to the predicate device.

Performance Comparison

As part of our design validation performance comparison analysis was conducted to demonstrate continued conformance with a special control or recognized standard.

A rationale analysis was then conducted and the results are contained in Table 1:

Testing TypeRationale Analysis
Validation Testing - BenchThere is no change about the following performance.Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and SensitivityProfile, Tomographic Plane Location, CT dose indexTherefore, we judged that Supria is substantially equivalent to the predicate.
Validation Testing - ClinicalWe provide 3 kinds of clinical image example which we judged to be sufficient to judge a clinical usability.

Table 1 Performance Comparison Analysis

The analysis confirms the performance characteristics of the Supria True64 are comparable to the predicate device and support our conclusion that the subject system is substantially equivalent.

Device Technological Characteristics

The technological characteristics of the Supria True64 and the predicate device are listed in Table 2.

ITEMSupria True64HITACHI Supria PHASE 3 (K163528)Difference
Gantry
GeometryRotate-rotate with offset detector system,slip ringRotate-rotate with offset detector system,slip ringNo
Scan Time0.75, 1.0, 1.5, 2.0 [s]0.75, 1.0, 1.5, 2.0 [s]No
X-ray Fan Beam Angle51 [deg]51 [deg]No
Gantry Tilt-30 to +30 [deg]-30 to +30 [deg]No
Gantry Aperture750 [mm]750 [mm]No
Gantry Dimensions1990 x 920 x 1842.5 [mm]1990 x 920 x 1842.5 [mm]No
Gantry Weight1600 [kg]1600 [kg]No
Scan LocalizerLaserLaserNo
ITEMSupria True64HITACHI Supria PHASE 3 (K163528)Difference
Detector
TypeSolid stateSolid stateNo
Number of Channels880 [ch] (16 slice)888 [ch] (64 slice)880 [ch] (8ch reference)No
Number of Rows16 (16 slice)64 (64 slice)16See 01
Number of Slices16 [slice/scan] (Axial) (16 slice)64 [slice/scan] (Axial) (64 slice)16 [slice/scan] (Axial)No
X-ray Tube
Anode Heat Storage5 [MHU]5 [MHU]No
Dissipation Rate470 [kHU/min]470 [kHU/min]No
Tube coolingOil/airOil/airNo
Tube focal spotDual 0.7 x 0.8, 1.2 x 1.4 [mm]Dual 0.7 x 0.8, 1.2 x 1.4 [mm]No
X-ray Generator
kW OutputSystem Maximum 48[kW] /Generator Maximum 51 [kW]System Maximum 48[kW] /Generator Maximum 51 [kW]No
Max. Power Input75 [kVA]75 [kVA]No
kVp Range80, 100, 120, 140 [kVp]80, 100, 120, 140 [kVp]No
mA Range10 to 400 [mA] @120kV, 48kW10 to 400 [mA] @120kV, 48kWNo
Patient Table
Range of Movement,Vertical450 to 1000 [mm] (CT-WT-21)450 to 1000 [mm] (CT-WT-21)No
Range of Movement,Longitudinal1910 [mm] (CT-WT-21)1910 [mm] (CT-WT-21)No
Range of Movement,LateralN/AN/ANo
Scannable Range155 cm155 cmNo
Maximum Load Capacity227 [kg]227 [kg]No
Display
Monitor Type24" LCD24" LCDNo
Matrices, Pixels1920 x 12001920 x 1200No
Image EnlargementsUp to 9.99xUp to 9.99xNo
Max. Slices Displayed at Once2525No
Image Storage
Hard Disk110 [GB] (images),200 [GB] (raw data) (16 slice)800 [GB] (raw data) (64 slice)110 [GB] (images),200 [GB] (raw data)See 02
Storage Images200,000200,000No
Archival Storage (Media)DVD-R/RW, CD-R/RWDVD-R/RW, CD-R/RWNo
ITEMSupria True64HITACHI Supria PHASE 3 (K163528)Difference
Scanning, Reconstruction
Localization ScanReal timeReal timeNo
Localization Scan Length150, 250, 350, 500, 750,1000, 1250, 1500, 1750 [mm]150, 250, 350, 500, 750,1000, 1250, 1500, 1750 [mm]No
Max. Scan Time100 [s]100 [s]No
Helical Beam Pitch0.56, 0.81, 1.06, 1.31, 1.56 (16 slice)0.58, 0.83, 1.08, 1.33, 1.58 (64 slice)0.56, 0.81, 1.06, 1.31, 1.56See 03
Collimation1.25, 5, 10, 15, 20 [mm]1.25, 5, 10, 15, 20 [mm]No
Reconstruction Matrix512 x 512 [pix]512 x 512 [pix]No
Reconstruction FOVs20 to 500 [mm]20 to 500 [mm]No
Slice Thickness0.625, 1.0, 1.25, 2.5, 3.75,5.0, 7.5, 10.0 [mm]0.625, 1.0, 1.25, 2.5, 3.75,5.0, 7.5, 10.0 [mm]No
Range of CT numbers-2000 to +4000 (13bit)-32768 to +32767 (16bit)-2000 to +4000 (13bit)-32768 to +32767 (16bit)No
Reconstruction Time0.1 seconds per image or less0.1 seconds per image or lessNo
Performance
High-contrast spatial resolution0.35 [mm]0.35 [mm]No
Low-contrast resolution mm at% at <4 rads2.5 [mm] @ 0.25%2.5 [mm] @ 0.25%No
10% MTF14.7 [lp/cm]14.7 [lp/cm]No
50% MTF12.2 [lp/cm]12.2 [lp/cm]No
Dose Controls
Bow Tie FilterYes. NormalYes. NormalNo
Automatic Exposure ControlYes. IntelliECYes. IntelliECNo
Longitudinal ModulationYesYesNo
Angular ModulationYesYesNo
Iterative ReconstructionYes. Intelli IP Advanced, Intelli IP QuickYes. Intelli IP Advanced, Intelli IP QuickNo
Maximum possible pitch with fullimage quality1.56 (16 slice)1.58 (64 slice)1.56See 03
Dose Displays
CTDIvYesYesNo
DLPYesYesNo
Features
Axial ScanYesYesNo
Helical ScanYesYesNo
Dynamic ScanYesYesNo
Predict ScanYesYesNo
ECG Retrospective Scan(Helical)NoNoNo
ECG Prospective Scan(Axial)NoNoNo
guideShot ScanYesYesNo
ITEMSupria True64HITACHI Supria PHASE 3 (K163528)Difference
Automatic Exposure ControlYesYesNo
Automatic Exposure Controlusing Iterative ReconstructionNoNoNo
ECG Dose ModulationNoNoNo
Adaptive FilterNoNoNo
Iterative ReconstructionYes. Intelli IP Advanced, Intelli IP QuickYes. Intelli IP Advanced, Intelli IP QuickNo
Injector SynchronizationYesYesNo
Dose CheckYesYesNo
Access ControlYesYesNo
Automatic Cardiac PhaseSearchNoNoNo
Preview ScanNoNoNo
Double Slice at Axial ScanNoNoNo
Priority Recon.NoNoNo
Dose ReportYes. Simple Dose ReportYes. Simple Dose ReportNo
DICOMYesYesNo
ID ReaderYesYesNo
Exam SplitYesYesNo
Multi-Planar Reconstruction(MPR)YesYesNo
Volume RenderingYesYesNo
CT Angiography (CTA)YesYesNo
SegmentationYesYesNo
RetouchYesYesNo
Quality ExamYesYesNo
HiMARYesYesNo
Orbit synchronization scanYesYesNo
Off-time modeYesYesNo
On-time standbyYesYesNo
Shutter Scan ReductionYesYesNo

Table 2 Technological Characteristic Differences

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

The differences from the predicate device to Supria 64 are explained below table.

Detector
01The 64 lines detector is added.
Image Storage
02Raw data disk for 64 slices is added because the data amount for each raw data of 64 slices is .bigger than 16 slices.
Dose Controls
03Helical Beam Pitch is fine-tuned for 64 lines.

Therefore, based on a thorough analysis and comparison of Supra 64 and the predicate device, the technological characteristics do not impact safety and effectiveness.

{9}------------------------------------------------

Substantial Equivalence

A summary decision was based on analysis of Table 3.

ITEMOverall Rationale Analysis
GantryThere are no significant changes in technology characteristics, hardware, and software from the predicate device, HITACHI Supria PHASE 3 (K163528).
Detector
X-ray Tube
X-ray Generator
Patient Table
Display
Image Storage
Scanning,Reconstruction
Performance
Dose Controls
Dose Displays
Features

Table 3 Rationale Analysis: Supria vs. Predicate

Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed Supria True64 Whole-body X-ray CT System is considered substantially equivalent to the currently marketed predicate device (SUPRIA Whole-body X-ray CT System (K163528)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

{10}------------------------------------------------

Summary of Non-Clinical Testing

This device complies with all applicable requirements for Dose Profile. Noise. Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index.

Noise, Mean CT number, Uniformity and Spatial Resolution test results for the Supria True64 were compared to results from the Supria PHASE 3 (K163528. Our findings determined the systems to be equivalent. The summary report on these findings is provided in this submission.

In addition, the Supria True64 system is in conformance with the applicable parts of the following standards:

  • AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012
    Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)

  • IEC 60601-1-2 Edition 3: 2007 . Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

  • IEC 60601-1-3 Edition 2.0 2008-01

Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment

  • IEC 60601-2-44 Edition 3.0 2009-02 Medical electrical equipment Part 2-44: particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography.
  • NEMA XR 25 Computed Tomography Dose Check
  • NEMA XR26

Access Controls for Computer Tomography: Identification, Interlocks, and Logs

  • IEC 62304 First edition 2006-05. Medical device software Software life cvcle processes .

Summary of Clinical Testing

As part of design validation, Hitachi has provided 3 kinds of clinical image examples which we judged to be sufficient to judge a clinical usability.

In addition, a comparative evaluation was conducted of via FBP and Iterative Reconstruction. It resulted in finding the images from the Supria True64 realized both low dose and high quality through reduction of image noise and artifacts.

Conclusions

Hitachi believes that, based on the information included in the submission, Supria True64 Whole-body X-ray CT System is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the SUPRIA Whole-body Xray System Phase 3 (K163528).

N/A