K Number
K163528

Validate with FDA (Live)

Date Cleared
2017-03-03

(77 days)

Product Code
Regulation Number
892.1750
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Supria system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes.

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface.

The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Description

The Supria is a multi-slice computed tomography system designed to perform multi-slice CT scanning supported by 16-detector technology. The system allows optimum clinical applications ranging from routine exams in response to the diversified circumstances in imaging whole body regions.

The Supria system uses 16-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 16 slices of data simultaneously. The Xray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.

The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.

The Supria system consists of a Gantry, Operator's Workstation, Patient Table, High-Frequency X-ray Generator, and accessories.

AI/ML Overview

The provided text is a 510(k) Pre-market Notification for a medical device, specifically the HITACHI Supria Whole-body X-ray CT System Phase 3. It compares the new device to a predicate device (HITACHI SUPRIA w/guideShot Option K161748) to demonstrate substantial equivalence, rather than providing a detailed study proving the device meets acceptance criteria in the sense of a performance study with specific metrics and statistical analysis.

The document states that the new device has two new features: "Intelli IP Quick" (a faster noise reduction technique) and "HiMAR" (reduces artifacts caused by metal). The "Performance Testing - Clinical" section briefly mentions that "clinical images were collected and analyzed" for these new features to "ensure that images constructed by the Supria Whole-body X-ray CT System meet user needs." It also notes that "Hitachi has provided clinical images demonstrating the image quality of Intelli IP Quick and HiMAR features and validated by a physician."

Therefore, it is important to note that the provided text does NOT contain the detailed information typically found in a clinical study report that defines and proves specific acceptance criteria for performance metrics (like sensitivity, specificity, accuracy) using a rigorous methodology. Instead, it relies on demonstrating comparable performance to a predicate device and physician validation of image quality for new features.

Based on the provided text, here's what can be extracted and what remains unknown:


Acceptance Criteria and Device Performance (Based on provided text)

Since direct quantitative acceptance criteria and performance metrics are not explicitly stated in a table format for the new features or overall device, we can infer the "acceptance" is primarily based on substantial equivalence to the predicate device and subjective physician validation of image quality for the new features.

Acceptance Criterion (Inferred from 510(k) context)Reported Device Performance (From text)
Overall substantial equivalence to predicate device (SUPRIA w/guideShot Option K161748) in terms of:"The analysis confirms the performance characteristics of the Supria Whole-body X-ray CT System Phase 3 are comparable to the predicate device and support our conclusion that the Phase 3 system is substantially equivalent." "Hitachi believes that, based on the information included in the submission, Supria Whole-body X-ray CT System Phase 3 is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the SUPRIA w/guideShot Option (K161748)."
- HardwareNo differences identified (e.g., Gantry, Detector, X-ray Tube, X-ray Generator, Patient Table, Display, Image Storage parameters are identical to predicate).
- Base elements of softwareMost software features are identical to the predicate. Key differences (Intelli IP Quick, HiMAR, Orbit synchronization scan, Off-time mode, On-time standby, Shutter Scan Reduction) are described as improvements or additional functionalities.
- Safety"The technological characteristics do not impact safety and effectiveness." (Stated for the differences in Table 3). Compliance with applicable safety standards (e.g., IEC 60601 series, NEMA XR 25, NEMA XR26) is declared.
- Effectiveness"The technological characteristics do not impact safety and effectiveness." (Stated for the differences in Table 3). Clinical images demonstrating the image quality of Intelli IP Quick and HiMAR features were "collected and analyzed" and "validated by a physician" to "ensure that images constructed by the Supria Whole-body X-ray CT System meet user needs." The overall "Performance Comparison" section also states "A clinical evaluation comparison was conducted with the Supria Phase 3 system and the SUPRIA w/guideShot Option (K161748) and found to be substantially equivalent."
- FunctionalityDetailed comparison in Table 2 shows identical functionality for many parameters, with new features (Intelli IP Quick, HiMAR etc.) presented as enhancements that do not negatively impact core functionality.
Image Quality with new features (Intelli IP Quick, HiMAR)"clinical images demonstrating the image quality of Intelli IP Quick and HiMAR features and validated by a physician." The images were collected and analyzed "to ensure that images constructed by the Supria Whole-body X-ray CT System meet user needs."

Study Details:

  1. Sample sizes used for the test set and data provenance:

    • Test Set Sample Size: Not specified. The document only mentions that "clinical images were collected and analyzed" for the new features. No specific number of images or patients is provided.
    • Data Provenance: Not specified regarding country of origin. The study is described as "clinical images were collected," implying real patient data. It is a retrospective or prospective study is not specified, but the phrasing "were collected" could suggest a collection from existing or newly acquired cases.
  2. Number of experts used to establish the ground truth for the test set and their qualifications:

    • Number of Experts: Not specified. The document states the image quality was "validated by a physician." It does not specify if this was one physician or a panel.
    • Qualifications of Experts: The qualification is generally stated as "a physician." Specific specialties (e.g., Radiologist) or years of experience are not mentioned.
  3. Adjudication method for the test set:

    • Not specified. Given the limited detail ("validated by a physician"), it's unlikely a formal adjudication method (like 2+1 or 3+1) was employed and documented in this 510(k) submission.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • Not explicitly stated or described. The document refers to a "clinical evaluation comparison" but does not detail it as a formal MRMC study. It states the comparison "found to be substantially equivalent" and that clinical images were "validated by a physician." There is no mention of multiple human readers, a comparative effectiveness study design, or an effect size for human readers' improvement with AI assistance. The new features (Intelli IP Quick, HiMAR) are signal processing techniques within the CT system, not an AI assistance tool for human interpretation in the sense of a CAD system.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The "performance comparison" section mentions "Performance Testing - Bench" with a rationale analysis indicating "Hitachi judged that Supria Whole-body X-ray CT System Phase 3 is substantially equivalent to the predicate." It also notes that the device complies with "all applicable requirements for Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index." This sounds like standalone technical performance testing, but not necessarily a clinical "algorithm only" performance evaluation akin to an AI diagnostic tool. The device itself is a CT scanner, not an AI algorithm intended for standalone diagnosis.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the clinical evaluation of the new features (Intelli IP Quick, HiMAR), the ground truth appears to be based on "physician validation" of "image quality" and ensuring that the images "meet user needs." This is a subjective assessment, not pathology or outcomes data. For the comparison to the predicate, it's about comparable physical and performance characteristics.
  7. The sample size for the training set:

    • Not applicable / Not specified. The document describes a CT system and its new image processing features. It does not mention a "training set" in the context of machine learning. The "Intelli IP Quick" and "HiMAR" are described as "processing techniques" that reduce noise or reduce metal artifacts, implying algorithms that might have been developed and tuned, but the text does not detail any "training data" or "training set" as would be seen for a deep learning model.
  8. How the ground truth for the training set was established:

    • Not applicable / Not specified. As no training set or machine learning model is explicitly described, there is no information on how a ground truth for such a set would have been established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the heads.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 3, 2017

Hitachi Medical Systems America, Inc. % Mr. Doug Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087

Re: K163528

Trade/Device Name: HITACHI Supria Whole-body X-rav CT System Phase 3 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: December 9, 2016 Received: December 16, 2016

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K163528

Device Name

HITACHI Supria Whole-body X-ray CT System Phase 3

Indications for Use (Describe)

The Supria system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes.

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Section 5

510(k) Summary

{4}------------------------------------------------

Submitter:Hitachi Medical Systems America, Inc.1959 Summit Commerce ParkTwinsburg, Ohio 44087-2371
Contact:Douglas J. Thistlethwaite
Telephone number:330-425-1313
Telephone number:330-963-0749
E-mail:thistlethwaited@hitachimed.com
Date:November 1, 2016

Submitter Information

Subject Device Name

Trade/Proprietary Name:HITACHI Supria Whole-body X-ray CT System Phase 3
Regulation Number:21 CFR 892.1750
Regulation Name:Computed tomography x-ray system
Product CodeJAK, System, X-Ray, Tomography, Computed
ClassII
PanelRadiology

Predicate Device Name

Predicate Device(s):HITACHI SUPRIA w/guideShot Option (K161748)
Regulation Number:21 CFR 892.1750
Regulation Name:Computed tomography x-ray system
Product CodeJAK, System, X-Ray, Tomography, Computed
ClassII
PanelRadiology

Device Intended Use

The Supria system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dvnamic modes,

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface.

The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Description

Function

The Supria is a multi-slice computed tomography system designed to perform multi-slice CT scanning supported by 16-detector technology. The system allows optimum clinical applications ranging from routine exams in response to the diversified circumstances in imaging whole body regions.

{5}------------------------------------------------

Scientific Concepts

The Supria system uses 16-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 16 slices of data simultaneously. The Xray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.

The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.

Physical and Performance Characteristics

The Supria system consists of a Gantry, Operator's Workstation, Patient Table, High-Frequency X-ray Generator, and accessories. The system performance is similar to the predicate device.

Performance Comparison

A clinical evaluation comparison was conducted with the Supria Phase 3 system and the SUPRIA w/guideShot Option (K161748) and found to be substantially equivalent.

There is no change in the performance of the system in regards to Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index.

A rationale analysis was then conducted and the results are contained in Table 1.

Table 1 Performance Comparison Analysis
Testing TypeRationale Analysis
Performance Testing - BenchHitachi judged that Supria Whole-body X-ray CT System Phase 3is substantially equivalent to the predicate.
Performance Testing - ClinicalHitachi has provided clinical images demonstrating the image quality of Intelli IP Quick and HiMAR features and validated by a physician.

The analysis confirms the performance characteristics of the Supria Whole-body X-rav CT System Phase 3 are comparable to the predicate device and support our conclusion that the Phase 3 system is substantially equivalent.

Device Technological Characteristics

The technological characteristics of the Supria and the predicate device are listed in Table 2.

Table 2 Technological Characteristic Differences0
-------------------------------------------------------
ITEMHITACHI Supria Whole-body X-rayCT System Phase 3HITACHI SUPRIA w/guideShotOption (K161748)Difference
Gantry
GeometryRotate-rotate with offset detector system, slip ringRotate-rotate with offset detector system, slip ringNo
Scan Time0.75, 1.0, 1.5, 2.0 [s]0.75, 1.0, 1.5, 2.0 [s]No
X-ray Fan Beam Angle51 [deg]51 [deg]No
Gantry Tilt-30 to +30 [deg]-30 to +30 [deg]No
Gantry Aperture750 [mm]750 [mm]No
Gantry Dimensions1990 x 920 x 1842.5 [mm]1990 x 920 x 1842.5 [mm]No
Gantry Weight1600 [kg]1600 [kg]No
Scan LocalizerLaserLaserNo
ITEMHITACHI Supria Whole-body X-rayCT System Phase 3HITACHI SUPRIA w/guideShotOption (K161748)Difference
Detector
TypeSolid stateSolid stateNo
Number of Channels880 [ch] (8ch reference)880 [ch] (8ch reference)No
Number of Rows1616No
Number of Slices16 [slice/scan] (Axial)16 [slice/scan] (Axial)No
X-ray Tube
Anode Heat Storage5 [MHU]5 [MHU]No
Dissipation Rate470 [kHU/min]470 [kHU/min]No
Tube coolingOil/airOil/airNo
Tube focal spotDual 0.7 x 0.8, 1.2 x 1.4 [mm]Dual 0.7 x 0.8, 1.2 x 1.4 [mm]No
X-ray Generator
kW OutputSystem Maximum 48[kW] /Generator Maximum 51 [kW]System Maximum 48[kW] /Generator Maximum 51 [kW]No
Max. Power Input75 [kVA]75 [kVA]No
kVp Range80, 100, 120, 140 [kVp]80, 100, 120, 140 [kVp]No
mA Range10 to 400 [mA] @120kV, 48kW10 to 400 [mA] @120kV, 48kWNo
Patient Table
Range of Movement,Vertical450 to 1000 [mm] (CT-WT-21)450 to 1000 [mm] (CT-WT-21)No
Range of Movement,Longitudinal1910 [mm] (CT-WT-21)1910 [mm] (CT-WT-21)No
Range of Movement,LateralN/AN/ANo
Scannable Range155 cm155 cmNo
Maximum Load Capacity227 [kg]227 [kg]No
Display
Monitor Type24" LCD24" LCDNo
Matrices, Pixels1920 x 12001920 x 1200No
Image EnlargementsUp to 9.99xUp to 9.99xNo
Max. Slices Displayed at Once2525No
Image Storage
Hard Disk110 [GB] (images),200 [GB] (raw data)110 [GB] (images),200 [GB] (raw data)No
Storage Images200,000200,000No
Archival Storage (Media)DVD-R/RW, CD-R/RWDVD-R/RW, CD-R/RWNo
Scanning, Reconstruction
Localization ScanReal timeReal timeNo
Localization Scan Length150, 250, 350, 500, 750,1000, 1250, 1500, 1750 [mm]150, 250, 350, 500, 750,1000, 1250, 1500, 1750 [mm]No
Max. Scan Time100 [s]100 [s]No
Helical Beam Pitch0.56, 0.81, 1.06, 1.31, 1.560.56, 0.81, 1.06, 1.31, 1.56No
Collimation1.25, 5, 10 [mm]1.25, 5, 10 [mm]No
Reconstruction Matrix512 x 512 [pix]512 x 512 [pix]No
Reconstruction FOVs20 to 500 [mm]20 to 500 [mm]No
Slice Thickness0.625, 1.0, 1.25, 2.5, 3.75,5.0, 7.5, 10.0 [mm]0.625, 1.0, 1.25, 2.5, 3.75,5.0, 7.5, 10.0 [mm]No
ITEMHITACHI Supria Whole-body X-rayCT System Phase 3HITACHI SUPRIA w/guideShotOption (K161748)Difference
Range of CT numbers-2000 to +4000 (13bit)-32768 to +32767 (16bit)-2000 to +4000 (13bit)-32768 to +32767 (16bit)No
Reconstruction Time0.1 seconds per image or less0.1 seconds per image or lessNo
Performance
High-contrast spatialresolution0.35 [mm]0.35 [mm]No
Low-contrast resolution mm at% at ≤4 rads2.5 [mm] @ 0.25%2.5 [mm] @ 0.25%No
10% MTF14.7 [lp/cm]14.7 [lp/cm]No
50% MTF12.2 [lp/cm]12.2 [lp/cm]No
Dose Controls
Bow Tie FilterYes. NormalYes. NormalNo
Automatic Exposure ControlYes. IntelliECYes. IntelliECNo
Longitudinal ModulationYesYesNo
Angular ModulationYesYesNo
Maximum possible pitch withfull image quality1.561.56No
Dose Displays
CTDIvYesYesNo
DLPYesYesNo
Features
Axial ScanYesYesNo
Helical ScanYesYesNo
Dynamic ScanYesYesNo
Predict ScanYesYesNo
ECG Retrospective Scan(Helical)NoNoNo
ECG Prospective Scan(Axial)NoNoNo
guideShot Scan (option)YesYesNo
Automatic Exposure ControlYesYesNo
Automatic Exposure Controlusing Iterative ReconstructionNoNoNo
ECG Dose ModulationNoNoNo
Adaptive FilterNoNoNo
Iterative ReconstructionYes. Intelli IP QuickYes. Intelli IP AdvancedSee 01
Injector SynchronizationYesYesNo
Dose CheckYesYesNo
Access ControlYesYesNo
Automatic Cardiac PhaseSearchNoNoNo
Preview ScanNoNoNo
Double Slice at Axial ScanNoNoNo
Priority Recon.NoNoNo
Dose ReportYes. Simple Dose ReportYes. Simple Dose ReportNo
DICOMYesYesNo
ID ReaderYesYesNo
ITEMHITACHI Supria Whole-body X-rayCT System Phase 3HITACHI SUPRIA w/guideShotOption (K161748)Difference
Exam SplitYesYesNo
Multi-Planar Reconstruction(MPR)YesYesNo
Volume RenderingYesYesNo
CT Angiography (CTA)YesYesNo
SegmentationYesYesNo
RetouchYesYesNo
Quality ExamYesYesNo
Iterative ReconstructionYes. Intelli IP QuickYes. Intelli IP AdvancedSee 01
HiMARYesNoSee 02
Orbit synchronization scanYesNoSee 03
Off-time modeYesNoSee 04
On-time standbyYesNoSee 05
Shutter Scan ReductionYesNoSee 06

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

The differences from the predicate device to Supria Whole-body X-ray CT System Phase 3 are explained in Table 3.

Table 3 Analysis of Differences

Features
01Intelli IP Quick: Intelli IP Quick is a faster noise reduction processing technique as compared to Intelli IP.
02HIMAR: HIMAR reduces lack of photon caused by X-ray going through a highly absorptive body such as metals and ametal artifact caused by beam hardening.
03Orbit synchronization scan: By using the orbit synchronization, the orbit of the volume scan multiple sequences canbe combined and differences can be easily extracted by comparing each image together.
04Off-time mode: The power consumption of the detector system is saved and noise is reduced when the power of theCT system is off. According to the usage of the system, the starting time can be registered.
05On-time standby: During operation of the CT system, The power consumption under the standby condition is savedand noise is reduced.
06Shutter Scan Reduction: The number of shutter scans was reduced.

Therefore, based on a thorough analysis and comparison of the Supria Whole-body X-ray CT System Phase 3 and the predicate device, the technological characteristics do not impact safety and effectiveness.

{9}------------------------------------------------

Substantial Equivalence

A summary decision was based on analysis of Table 4.

Table 4 Rationale Analysis: Supria vs. Predicate
------------------------------------------------------
ITEMOverall Rationale Analysis
GantryBased on that there are no significant differences from the predicate device,
DetectorHitachi judges that the Supria Whole-body X-ray CT System Phase 3 device
X-ray Tubehas no additional issues with safety and effectiveness.
X-ray Generator
Patient Table
Display
Image Storage
Scanning,
Reconstruction
Performance
Dose Controls
Dose Displays
FeaturesBased on results of design inspection as confirmed in the Declaration ofConformity, Hitachi judges that the Supria Whole-body X-ray CT System Phase3 device has no additional issues with safety and effectiveness.

Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed Supria Whole-body X-ray CT System Phase 3 is considered substantially equivalent to the currently marketed predicate device (SUPRIA w/guideShot Option (K161748)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

{10}------------------------------------------------

Summary of Non-Clinical Testing

The Supria system is in conformance with the applicable parts of the following standards:

  • AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and ● A2:2010/(R)2012
    Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)

  • IEC 60601-1-2 Edition 3: 2007 . Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests

  • IEC 60601-1-3 Edition 2.0 2008-01 .

Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment

  • IEC 60601-2-44 Edition 3.0 2009-02 o Medical electrical equipment Part 2-44: particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography.
  • NEMA XR 25 Computed Tomography Dose Check
  • . NEMA XR26

Access Controls for Computer Tomography: Identification, Interlocks, and Logs

  • IEC 62304 First edition 2006-05, Medical device software Software life cycle processes 0
    In addition, the device complies with all applicable requirements for Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index.

Summary of Clinical Testing

Using the 2 new features, clinical images were collected and analyzed, to ensure that images constructed by the Supria Whole-body X-ray CT System meet user needs.

As a result of the analysis:

Testing TypeRationale Analysis
Performance Testing - ClinicalHitachi has provided clinical images demonstrating the image quality of Intelli IP Quick and HiMAR features and validated by a physician.

Conclusions

Hitachi believes that, based on the information included in the submission, Supria Whole-body X-ray CT System Phase 3 is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the SUPRIA w/quideShot Option (K161748).

§ 892.1750 Computed tomography x-ray system.

(a)
Identification. A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.