K Number
K171530

Validate with FDA (Live)

Manufacturer
Date Cleared
2017-12-21

(210 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Deltaven Closed I.V. Catheter systems are catheters for short-term peripheral venous access that allow the collection of blood samples and administration of fluids intravascularly.

Deltaven Closed I.V. Catheter systems are equipped with a passive system for the prevention of accidental needlestick injuries.

Blood is contained within the device during the catheter insertion process, aiding the prevention of blood exposure. The device can be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Deltaven Closed I.V. Catheter systems 16-24 gauge catheters are suitable for use with pressure injectors rated for a maximum of 330 psi when the access ports and stopcocks are removed and a direct connection is made with the proximal luer lock connector.

Deltaven Closed I.V. Catheter systems 26G are not suitable for the administration at high pressure.

Device Description

The devices consist of an over-the-needle, peripheral intravascular catheter made of polyurethane, integrated extension tubing with Luer lock adaptor and slide clamp. The devices are also equipped with a Luer lock final adaptor (single entry version) or a Y Luer lock final adapter (dual entry version).

Deltaven Closed I.V. Catheters are available in five versions, as follows:

  • Deltaven XiV Max
  • . Deltaven XiV Max Y
  • Deltaven XiV Max SC ●
  • Deltaven XiV Max Y-NL ●
  • Deltaven XiV Max SC-NL

The same versions are available with the needle provided with a notch along its surface that permits the early visualization of blood return inside the catheter tube. This version is generally named as Deltaven XV Max Fast Flash and is available in five versions, as follows:

  • . Deltaven XiV Max Fast Flash
  • Deltaven XiV Max Y Fast Flash
  • Deltaven XiV Max SC Fast Flash
  • Deltaven XiV Max Y- NL Fast Flash
  • . Deltaven XiV Max SC-NL Fast Flash

The devices provided with single entry Luer lock adaptors are also equipped with:

  • Luer lock white cap
  • 3 way stopcock
  • 3 way stopcock and needleless valve ●

The devices provided with dual entry Luer lock adaptors are also equipped with:

  • Luer lock white cap ●
  • Needleless valve connector
AI/ML Overview

The provided document describes the Deltaven Closed I.V. Catheter Systems and its substantial equivalence to a predicate device (BD Nexiva Closed IV Catheter System, K102520). The document outlines various bench/performance/non-clinical tests conducted to support this claim, referencing relevant ISO standards.

Here's an analysis of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states that in all testing, "the pre-determined acceptance criteria were met." However, it does not explicitly list the specific quantitative acceptance criteria or the numerical performance results for each test. Instead, it refers to compliance with ISO standards.

Here's a table based on the mentioned tests and the general statement of compliance:

Test CategorySpecific Test (Referenced Standard if applicable)Acceptance Criteria (Implied)Reported Device Performance
BiocompatibilityCytotoxicity, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogen, Bacterial-Mediated Pyrogen, Genotoxicity, Sub-Chronic Toxicity, Implantation, Hemolysis, Complement Activation, In Vivo Thrombogenicity, LAL testingCompliance with ISO 10993 series requirementsMet biocompatibility requirements
Physical/Mechanical (ISO 594-1, 594-2, 80369-7)GaugingCompliance with ISO 594-1 and 594-2Criteria met
Liquid leakageCompliance with ISO 594-2, 594-1, and 80369-7Criteria met
Air leakage during aspirationCompliance with ISO 594-2 and 594-1Criteria met
Sub-atmospheric pressure air leakageCompliance with ISO 80369-7Criteria met
Unscrewing torqueCompliance with ISO 594-2 and 80369-7Criteria met
Separation forceCompliance with ISO 594-2 and 80369-7Criteria met
Easy to assemblyCompliance with ISO 594-2Criteria met
Resistance of overridingCompliance with ISO 594-2 and 80369-7Criteria met
Stress crackingCompliance with ISO 594-2 and 80369-7Criteria met
Leakage by pressure decayCompliance with ISO 80369-7Criteria met
Catheter/Needle Properties (ISO 9626, 10555-1, 10555-5)Radio-detectabilityCompliance with ISO 10555-1Criteria met
Surface testingCompliance with ISO 9626, 10555-1, and 10555-5Criteria met
Corrosion testingCompliance with ISO 9626 and 10555-1Criteria met
Peak tensile forceCompliance with ISO 10555-1Criteria met
Freedom from leakage under pressureCompliance with ISO 10555-1Criteria met
Freedom from leakage during aspirationCompliance with ISO 10555-1Criteria met
Needle materialCompliance with ISO 9626Criteria met
Needle stiffnessCompliance with ISO 9626Criteria met
Resistance of tubing to breakageCompliance with ISO 9626Criteria met
Catheter unitCompliance with ISO 10555-5Criteria met
Strength union between needle hub and needle tubeCompliance with ISO 10555-5Criteria met
Vent fittingCompliance with ISO 10555-5Criteria met
Sharps Injury Protection (ISO 23908)Safety Closed IV Catheter system activation testCompliance with ISO 23908:2011Criteria met
Challenging safety device testCompliance with ISO 23908:2011Criteria met
Simulated Clinical Use TestCompliance with FDA "Guide for Industry and staff Medical Devices with Sharp injury Prevention Features" and ISO 23908:2011Criteria met
Additional Performance TestingPriming volumeVerified/ValidatedCriteria met
Clinical significant hydrationVerified/ValidatedCriteria met
Pressure resistance (up to 330 psi)Verified/Validated for 16-24G, 26G excludedCriteria met (for applicable gauges up to 330 psi)
Tensile force of 15 N for 15 seconds between tube and hubNo breakage or separationNo breakage or separation (demonstrated for further SE)

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample sizes for any of the individual tests. It states that "Performance testing was provide to verify/validate the sharps injury prevention feature" and "Performance data has been provided to verify pressure injection up to 330 psi". For the tensile force test, it mentions "to samples of each device type."

The data provenance is not explicitly stated as retrospective or prospective, nor does it mention the country of origin. However, given that these are bench and performance tests on a medical device for regulatory submission, they are typically prospective tests performed in a controlled laboratory environment.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This section is not applicable as the document describes bench and performance testing of a physical medical device, not a diagnostic or AI-driven device requiring expert interpretation for ground truth. The "ground truth" for these tests would be the physical properties and performance characteristics measured against established engineering and medical device standards.

4. Adjudication Method for the Test Set:

This section is not applicable for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are used for establishing consensus among human experts for diagnostic accuracy studies, not for physical performance testing.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

This section is not applicable. The document describes a medical device (intravascular catheter system) and its performance, not an AI-driven diagnostic tool that would involve human readers or an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

This section is not applicable. The device is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

The ground truth for this medical device's performance testing is established by objective measurements and reference to internationally recognized technical standards (e.g., ISO 594, ISO 9626, ISO 10555, ISO 23908, ISO 80369, ISO 10993 series). This includes:

  • Compliance with specified dimensions and tolerances.
  • Performance within defined ranges (e.g., pressure resistance, flow rates).
  • Absence of failures (e.g., leakage, breakage, separation).
  • Demonstrated activation of safety mechanisms.
  • Absence of biological adverse reactions (biocompatibility).

8. The Sample Size for the Training Set:

This section is not applicable. "Training set" refers to data used to train machine learning models. This document describes the testing of a physical medical device, which does not involve a training set in this context.

9. How the Ground Truth for the Training Set Was Established:

This section is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

December 21, 2017

Delta Med SpA % Roger Gray VP, Quality and Regulatory Donawa Lifescience Consulting Srl Piazza Albania, N 10 Rome, 00153 ITALY

Re: K171530

Trade/Device Name: Deltaven Closed I.V. Catheter Systems Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ. Dated: November 21, 2017 Received: November 24, 2017

Dear Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Tina
Kiang -S

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K171530

Device Name

Deltaven Closed I.V. Catheter Systems

Deltaven Closed I.V. Catheter systems are catheters for short-term peripheral venous access that allow the collection of blood samples and administration of fluids intravascularly.

Deltaven Closed I.V. Catheter systems are equipped with a passive system for the prevention of accidental needlestick injuries.

Blood is contained within the device during the catheter insertion process, aiding the prevention of blood exposure. The device can be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Deltaven Closed I.V. Catheter systems 16-24 gauge catheters are suitable for use with pressure injectors rated for a maximum of 330 psi when the access ports and stopcocks are removed and a direct connection is made with the proximal luer lock connector.

Deltaven Closed I.V. Catheter systems 26G are not suitable for the administration at high pressure.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

K171530 510(K) SUMMARY

Submitter:Delta Med S.p.AVia Guido Rossa, 20I-46019 Viadana, MantovaItaly
Phone: +39 03 757 85915Fax: +39 03 757 85201
Application Correspondent:Mr. Roger GrayVice President Quality and RegulatoryDonawa Lifescience ConsultingPiazza Albania 1000153 Rome Italy
Phone: +39 03 757 85915Email: rgray@donawa.com
Preparation Date:December 18, 2017
Trade Name:Deltaven Closed I.V. Catheter Systems
Common or Usual Name:Intravascular Catheter
Regulation Name:Intravascular Catheter
Regulation Number:21 CFR 880.5200
Product Code:FOZ
Device Class:Class II
Primary Predicate Device:K102520: BD Nexiva Closed IV Catheter System

Device Description:

The devices consist of an over-the-needle, peripheral intravascular catheter made of polyurethane, integrated extension tubing with Luer lock adaptor and slide clamp. The devices are also equipped with a Luer lock final adaptor (single entry version) or a Y Luer lock final adapter (dual entry version).

Deltaven Closed I.V. Catheters are available in five versions, as follows:

  • Deltaven XiV Max
  • . Deltaven XiV Max Y
  • Deltaven XiV Max SC ●
  • Deltaven XiV Max Y-NL ●
  • Deltaven XiV Max SC-NL

{4}------------------------------------------------

The same versions are available with the needle provided with a notch along its surface that permits the early visualization of blood return inside the catheter tube. This version is generally named as Deltaven XV Max Fast Flash and is available in five versions, as follows:

  • . Deltaven XiV Max Fast Flash
  • Deltaven XiV Max Y Fast Flash
  • Deltaven XiV Max SC Fast Flash
  • Deltaven XiV Max Y- NL Fast Flash
  • . Deltaven XiV Max SC-NL Fast Flash

The devices provided with single entry Luer lock adaptors are also equipped with:

  • Luer lock white cap
  • 3 way stopcock
  • 3 way stopcock and needleless valve ●

The devices provided with dual entry Luer lock adaptors are also equipped with:

  • Luer lock white cap ●
  • Needleless valve connector

Indications for Use:

Deltaven Closed I.V. Catheter systems are catheters for short-term peripheral venous access that allow the collection of blood samples and administration of fluids intravascularly.

Deltaven Closed I.V. Catheter systems are equipped with a passive system for the prevention of accidental needlestick injuries.

Blood is contained within the device during the catheter insertion process, aiding the prevention of blood exposure. The device can be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

Deltaven Closed I.V. Catheter systems 16-24 gauge catheters are suitable for use with pressure injectors rated for a maximum of 330 psi when the access ports and stopcocks are removed and a direct connection is made with the proximal luer lock connector.

Deltaven Closed I.V. Catheter systems 26G are not suitable for the administration at high pressure.

AttributeSubject Device - K171530 DeltavenClosed IV Catheter SystemPredicate Device - K102520 BDNexiva IV Closed Catheter SystemComparison
IV Catheter TypeSafety Closed IV Catheter systemSafety Closed IV Catheter systemSame
Intended useDeltaven Closed I.V. Cathetersystems are catheters forshort-term (less than 30 days)peripheral venous access thatallow the collection of bloodsamples and administration offluids intravascularly.The Nexiva intravascular catheteris inserted into a patient'svascular system for a short-termuse (less than 30 days) to sampleblood, monitor blood pressure, oradminister fluids intravascularly.Same
Single useYesYesSame
Sharps injuryprevention featureYesYesSame

{5}------------------------------------------------

AttributeSubject Device - K171530 DeltavenClosed IV Catheter SystemPredicate Device - K102520 BDNexiva IV Closed Catheter SystemComparison
Needle MaterialStainless SteelStainless SteelSame
Needle Gauge Sizes16 - 26 G18 - 24 GDifferent
Needle distal endconfigurationBack cut configurationBack cut configurationSame
Sterilization MethodEOEOSame
Pressure Resistance330 psi300 psiDifferent
Tubing proximal endconfigurationFemale 6 % Luer lock, inaccordance with ISO 594-2Female 6 % Luer lock, inaccordance with ISO 594-2Same
Catheter tubematerialPolyurethanePolyurethane (Vialon)Different
Tubing Extensionline dimension26 - 18 G; ID 1.2 mm16 G; ID 1.6 mmFor 24 - 22G ; ID 1.22mm20 - 18G; ID 1.65mmDifferent

Substantial Equivalence Discussion

The characteristics of the subject device, Delta Med Closed IV Catheter system, are compared with the selected predicate device cleared under K102520:

The indications for use statement and the intended use of the subject device are equivalent to the predicate device. The differences between the devices are:

    1. Sharps injury prevention feature:
      The subject device and the predicate device both have integral safety features to help prevent needlestick injuries. The needlestick protection engages passively, without any specific action from the user apart from withdrawal of the needle, which engages the safety mechanism to cover the needle point. Even though differences exist between the methods employed for protecting the needle tip, the two devices are equivalent in this respect. Performance testing was provide to verify/validate the sharps injury prevention feature.
    1. High pressure use:
      For the Nexiva range, gauge sizes from 22G to 18G are suitable for high pressure injection (300 psi), while the 24G model is excluded from high pressure usage. The subject device is suitable for high pressure usage from 24 G to 16 G, however, the 26G Deltaven device is excluded from high pressure use. Performance data has been provided to verify pressure injection up to 330 psi in these device configurations.
    1. Configurations:
      The predicate device is available in four proximal end configurations, two with a single entry connections, and two with dual entry connections. In comparison, the subject device has five basic proximal end configurations. Each of these five configurations is also available in a 'fast flash' version, which is substantially equivalent to the 'InstaflashTM' rapid flow visualization system integral to the predicate device range. These differences in proximal connection types do not affect the intended use of the devices

{6}------------------------------------------------

4. Material:

The predicate BD Nexiva catheters are made from a polyurethane material named Vialon whereas the subject Deltaven catheters are made from polyurethane named Pellethane®, which is the same material used for manufacture of the Neo Delta Self Safe Catheter, cleared under K121007.

5. Vent Plug:

Both the predicate and the subject catheters are provided with vent plugs on their final connections (single and dual entry). The presence of the vent plug prevents blood leakages from the final connections and visualization chamber. Testing was performed to verify the performance of the vent plug. The predicate and subject devices are therefore substantially equivalent in this respect.

6. Needle point:

The subject device and the reference device K121007 use stainless steel AISI 304 for the needles. The needle points are very similar and substantially equivalent between the predicate and the subject devices, as demonstrated photographically. Testing was performed to verify the performance and characteristics of the needle.

7. Extension Tubing:

The predicate BD Nexiva and subject Deltaven catheter systems are both provided with extension lines, but the dimensions are different. Both use microbore tubing and for the 18G to 22G catheters, the subject Deltaven device has a 0.45 mm smaller internal diameter. This difference has an impact on the flow rates, with the subject device range generally being lower, although substantial equivalence between the two ranges is not compromised by this difference. In order to further demonstrate substantial equivalence, a tensile force of 15 N was applied for 15 seconds between the tube and the hub to samples of each device type, demonstrating that there was no breakage or separation.

8. Catheter/Needle gauge size:

The predicate Nexiva range includes catheter/needle sizes between 24G and 18G, whereas the subject Deltaven range includes sizes from 26G to 16G. Performance testing was provided to verify use of the device at all gauge sizes.

Based on the above similarities and differences, the subject device does not raise different types of safety and effectiveness questions when compared to the predicate device.

Bench/Performance/Non-Clinical Testing

The Deltaven Closed IV Catheter Systems meet the relevant technical requirements of the following standards:

  • . ISO 594-1:1986 - Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements
  • ISO 594-2:1998 - Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
  • ISO 9626:2016 - Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
  • ISO 10555-1:2013 - Intravascular catheters -- Sterile and single-use intravascular catheters -- Part 1: General requirements
  • ISO 10555-5:2013 - Intravascular catheters -- Sterile and single-use catheters -- Part 5: Over-needle peripheral catheters

{7}------------------------------------------------

  • ISO 23908:2011 - Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
  • ISO 80369-7:2016 - Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications

The Deltaven Closed IV Catheter Systems are supplied sterile for single use, sterilized with ethylene oxide (ETO) gas, and meet the biocompatibility requirements of the applicable standards in the ISO 10993 series. The following biocompatibility testing were completed: cytotoxicity, irritation, acute systemic toxicity, material-mediated pyrogen, bacterial-mediated pyrogen, genotoxicity, sub-chronic toxicity, implantation, hemolysis, complement activation, in vivo thrombogenicity, and Limulus Amebocyte Lysate (LAL) testing.

The following testing was conducted in accordance with the applicable requirements of the above standards, to show safety and effectiveness as well as substantial equivalence to the predicate device:

  • Radio-detectability (ISO 10555-1)
  • Surface testing (ISO 9626, ISO 10555-1 and ISO 10555-5)
  • Corrosion testing (ISO 9626 and ISO 10555-1)
  • Peak tensile force (ISO 10555-1)
  • Freedom from leakage under pressure (ISO 10555-1)
  • Freedom from leakage during aspiration (ISO 10555-1)
  • Gauging (ISO 594-2 and ISO 594-1)
  • Liquid leakage (ISO 594-2, ISO 594-1 and ISO 80369-7)
  • Air leakage during aspiration (ISO 594-2-ISO 594-1)
  • Sub-atmospheric pressure air leakage (ISO 80369-7)
  • Unscrewing torque (ISO 594-2, ISO 80369-7)
  • . Separation force (ISO 594-2, ISO 80369-7)
  • Easy to assembly ( ISO 594-2)
  • Resistance of overriding (ISO 594-2, ISO 80369-7) ●
  • Stress cracking (ISO 594-2, ISO 80369-7)
  • Leakage by pressure decay (ISO 80369-7)
  • Needle material (ISO 9626)
  • Needle stiffness (ISO 9626)
  • Resistance of tubing to breakage (ISO 9626)
  • Catheter unit (ISO 10555-5)
  • Strength union between needle hub and needle tube (ISO 10555-5)
  • Safety Closed IV Catheter system activation test (ISO 23908)
  • Challenging safety device test (ISO 23908)
  • . Vent fitting (ISO 10555-5)

Additional testing was performed to verify:

  • Priming volume
  • . Clinical significant hydration
  • Pressure resistance

In addition, a Simulated Clinical Use Test was carried out to verify/validate use of the sharps protection device in accordance with:

{8}------------------------------------------------

  • FDA "Guide for Industry and staff Medical Devices with Sharp injury Prevention Features"
  • ISO 23908:2011 "Sharps injury protection. Requirements and Test method. Sharp protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling".

In all testing, the pre-determined acceptance criteria were met.

Conclusion

The subject device does not raise new or different questions of safety and effectiveness and are supported by non-clinical testing. The Deltaven Closed IV Catheter Systems is substantially equivalent to the BD Nexiva Closed IV Catheter System cleared under K102520.

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).