K Number
K102520

Validate with FDA (Live)

Date Cleared
2011-01-06

(126 days)

Product Code
Regulation Number
880.5200
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

As indicated in 21 CFR Part 880.5200, The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made.

Device Description

The BD Nexiva™ Closed IV Catheter System consists of an over-the needle, peripheral intravascular catheter made from Vialon™ polyurethane, integrated extension tubing with a Y adapter and clamp, BD Q-Syte™ luer access port, and a passive needleshielding mechanism.

The design of the Nexiva™ IV catheter can be described as a closed system since it protects clinicians from blood exposure during the catheter insertion procedure. Since the needle is withdrawn through a septum that seals after the needle has been removed and both ports of the Y adapter are closed, blood is contained within the NexivaTM device during catheter insertion. The pressure exerted on the needle as it passes through the septum wipes blood from the needle, further reducing potential blood exposure. The slide clamp on the integrated extension tubing is provided to eliminate blood exposure when the vent plug is replaced with an infusion set connection of a BD Q-Syte™ luer access port.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the BD Nexiva™ Closed IV Catheter System, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The nonclinical tests compare the modified BD Nexiva™ Closed IV Catheter System to its predicate device (K032843). The reported device performance for all listed characteristics is "Pass," indicating that the device met the specified acceptance criteria.

Associated ModificationProduct Performance Characteristics/Verification TestingAcceptance Criteria for Product VerificationReported Device Performance
1Premature Decouple ForceDPPM < 280Pass
2Catheter Pull ForceDPPM < 60Pass
2Catheter Burst StrengthDPPM < 60Pass
2Catheter Penetration ForceDPPM <1350Pass
2Catheter DragDPPM <1350Pass
2Tip AdhesionDPPM <1350Pass
1System Drag -Cannula DragDPPM < 500Pass
1System Drag -Ferrule Retraction ForceDPPM < 500Pass
1Adapter Release ForceDPPM < 500Pass
3Needle Cover RemovalDPPM USL < 18000Pass
3(Needle Cover Removal)DPPM LSL < 200000Pass
3(Needle Cover Removal)-20%/+25%, DPPM < 5000Pass
4Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
5Pinch Clamp Fluid SealDPPM < 10000Pass
5Pinch Clamp Engagement Disengagement ForceDPPM < 2000Pass
8Wing deflection forceDPPM USL < 2000Pass
8(Wing deflection force)DPPM LSL < 500Pass
6Bond Strength -Tubing Bond to AdapterDPPM < 0.1Pass
4Extension Tube Burst PressureDPPM < 60Pass
10Unit Package Seal Peel Force1.33 CpkPass
10Seal Width1.33 CpkPass
10Unit Package Material Thickness - Post Formed (Bottom Wed)1.33 CpkPass
10Package Integrity (Water Leak)0 out of 2301Pass
10Ship Testing (Drop/Vibration)0 out of 2301Pass
7Y-Adapter Wishbone LoadingDPPM < 0.1Pass
7ISO Liquid LeakageDPPM < 500Pass
7ISO Air LeakageDPPM < 500Pass
7ISO Unscrewing TorqueDPPM < 500Pass
7ISO Resistance to OverridingDPPM < 500Pass
7ISO Thread SeparationDPPM < 500Pass
7Vent Plug Removal Force/TorqueDPPM USL < 6500Pass
7(Vent Plug Removal Force/Torque)DPPM LSL <10000Pass
7Q-Syte Removal TorqueDPPM USL < 6500Pass
7(Q-Syte Removal Torque)DPPM LSL <10000Pass
11ISO Thread Seperation -Both Y female luerDPPM < 500Pass
11ISO Thread Seperation -Straight female luerDPPM < 500Pass
11ISO Unscrewing Torque -Both Y female luerDPPM < 500Pass
11ISO Unscrewing Torque -Straight female luerDPPM < 500Pass
11ISO Resistance to Override -Both Y female luerDPPM < 500Pass
11ISO Resistance to Override -Straight female luerDPPM < 500Pass
11ISO Liquid LeakageDPPM < 500Pass
11ISO Air LeakageDPPM < 500Pass
12Vent Plug Torque RemovalDPPM USL < 10000, DPPM LSL< 20000Pass
11Bond Strength - Tubing Bond to AdapterDPPM < 60Pass
11Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
14, 15Proximal Re-exposureDPPM < 100Pass
17System Drag -Maximum Adhesion ForceDPPM < 1350Pass
17System Drag - Average Drag ForceDPPM < 10,000Pass
17System Drag - Offset Peak ForceDPPM < 2000Pass
17Flow Rate Label Claim-20%/+25%, DPPM < 5000, -10%/+15%, DPPM < 5000Pass
17Minimum FlowDPPM < 5000Pass
17FlashbackDPPM < 1000Pass
17Cannula PenetrationDPPM < 1350Pass
17Catheter PenetrationDPPM < 1350Pass
17, 19Premature DecoupleDPPM < 280Pass
20Needle Hub PullDPPM < 500Pass

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document specifies a sample size of 2301 units for "Package Integrity (Water Leak)" and "Ship Testing (Drop/Vibration)" tests. For other tests, specific sample sizes are not explicitly stated, but the acceptance criteria are generally expressed in "DPPM" (Defective Parts Per Million) or "Cpk" (Process Capability Index), suggesting statistical process control and quality assurance testing.
  • Data Provenance: The tests are described as "Nonclinical Tests" and "design verification activities," conducted by Becton Dickinson Infusion Therapy Systems, Inc. The data is retrospective in the sense that it was generated during device development and verification to support the 510(k) submission. There is no information on the country of origin of the data, but it can be assumed to be related to the manufacturer's testing facilities, potentially in the US (headquarters in Sandy, UT).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document primarily describes physical and mechanical performance testing of a medical device, not diagnostic or clinical performance involving expert interpretation of data (like radiology images). Therefore, the concept of "experts" establishing ground truth in this context doesn't directly apply as it would for a software device. The ground truth for these tests is based on established engineering standards (e.g., ANSI/AAMI/ISO 10993-1 2003 (E) for biocompatibility, and presumably other relevant engineering and ISO standards for mechanical performance), internal product specifications, and quality control methodologies.

4. Adjudication Method for the Test Set

Not applicable. As described above, these are physical and mechanical tests with quantitative acceptance criteria (e.g., DPPM, Cpk, force measurements). There is no "adjudication" in the sense of reconciling differing expert opinions or classifications. The results are objective measurements compared against predefined thresholds.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. This document does not describe an MRMC comparative effectiveness study. The submission relates to a physical medical device (intravascular catheter) and its modifications, not an AI or software device that would typically involve human readers interpreting cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No. This concept is not applicable as the submission is for a physical medical device, not an algorithm or AI. The tests are focused on the device's physical and mechanical properties.

7. The Type of Ground Truth Used

The ground truth used for these tests is based on:

  • Engineering Standards: Adherence to established industry and international standards (e.g., ANSI/AAMI/ISO 10993-1 2003 (E), ISO Liquid Leakage, etc.).
  • Product Specifications: Internal design requirements and performance targets defined by the manufacturer.
  • Predicate Device Performance: The modified device is benchmarked against its predicate device (K032843) to demonstrate "substantial equivalence."

8. The Sample Size for the Training Set

Not applicable. This is not a machine learning or AI device that requires a training set. The "training" for such physical devices involves design, prototyping, and iterative testing processes, not data-driven algorithmic training.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no "training set" in the context of this device.

{0}------------------------------------------------

Becton Dickinson Infusion Therapy Systems, Inc. 9450 South State Street Sandy UT 84070 tel: 801-565-2662 fax: 801-565-2749 iustice alder@hd.com

K102520/S2

Helping all people

live healthy lives

SPECIAL PREMARKET NOTIFICATION [510(K)] Summary- BD NexivaTM Closed IV Catheter System

Submitter:

Becton Dickinson Infusion Therapy Systems, Inc.

Address:

9450 South State Street Sandy, UT 84070

Contact Person:

Trade Names:

Common Name:

Classification Name:

Justice Alder Specialist, Regulatory Affairs

Telephone Number: Fax Number:

801-565-2749

801-565-2662

August 30, 2010

Date Summary Prepared:

BD Nexiva™ Closed IV Catheter System

Intravascular Catheter

Intravascular Catheter (FOZ) 880.5200

Unmodified/Predicate Device: (K032843) BD NexivaTM Closed IV Catheter System

Description of the device:

The BD Nexiva™ Closed IV Catheter System consists of an over-the needle, peripheral intravascular catheter made from Vialon™ polyurethane, integrated extension tubing with a Y adapter and clamp, BD Q-Syte™ luer access port, and a passive needleshielding mechanism.

The design of the Nexiva™ IV catheter can be described as a closed system since it protects clinicians from blood exposure during the catheter insertion procedure. Since the needle is withdrawn through a septum that seals after the needle has been removed and both ports of the Y adapter are closed, blood is contained within the NexivaTM device during catheter insertion. The pressure exerted on the needle as it passes through the septum wipes blood from the needle, further reducing potential blood exposure. The slide clamp on the integrated extension tubing is provided to eliminate blood exposure when the vent plug is replaced with an infusion set connection of a BD Q-Syte™ luer access port.

{1}------------------------------------------------

Intended Use of the BD Nexiva™ Closed IV Catheter System device: [21 CFR Part 880—GENERAL HOSPITAL AND PERSONAL USE DEVICES; General Hospital and Personal Use Therapeutic Devices Sec. 880.5200]

The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made.

Technological Characteristics Comparison:

The BD NexivaTM Closed IV Catheter products have the same technological characteristics with no changes in intended use. The subject device has similar design features and materials as the predicate device. The table below shows the design features and material comparisons between the predicate and subject device.

Subject DevicePredicate Device
1. Catheter adapter,tip shieldCatheter Adapter -Decreased adapter pocketwidth to 0.075" for tightercontrol of adapter retentionlock mechanismTip Shield -Decrease V-clippocket opening to 0.077" fortighter control of V-clipmovement when engagedwith adapterTip Shield -Scaling of InternalID through holes to cannulagauge to improve control ofcannula to tip shieldmovementCatheter Adapter -Adapterpocket width set at 0.088"Tip Shield -V-clip pocketwidth set at 0.084"Tip Shield -Internal ID holeswere generalized toaccommodate all cannulagauge sizes
2. Catheter material.Catheter material -VialonCatheter material - Vialon
Catheter adapterX40, Radio opaque materialCatheter Adapter-24Ga IDset to 0.0406" decreased by0.0004"X50, Radio opaque materialCatheter Adapter -24Ga ID0.0410"
3. Catheter adapter,needle coverCatheter Adapter -needlecover modified to removebumps, have a straightinterference fit and a solidstop featureNeedle Cover -Length set to2.26" (0.040" shortened)Catheter Adapter -needlecover interface had 2 bumpinterference features and atapered noseNeedle Cover -Length was2.30"
4. Extension tubingExt. Tubing -increaseddiameters for 18,20 GaproductExt. Tubing -all gaugesmicrobore tubing
5. ClampClamp -pinch clampClamp -slide clamp
6. Catheter adapter, LueradapterCatheter Adapter -tubingjunction increased ID formacrobore tubing,optimized for adhesiveaddition)Luer Adapter -tubingjunction increased ID formacrobore tubing,optimized for adhesiveaddition)Catheter Adapter -designed for microboretubing and solvent bondjointLuer Adapter -tubingjunction defined with aswedge design
7. Leur adapterLuer Adapter -PCTG "Y"adapter
8. Catheter adapterCatheter Adapter -smalleradapter wing structure(confined within overmoldedsoft wing)Catheter Adapter -largeadapter wing structure(confined within overmoldedsoft wing)
9. Tip shieldTip Shield -improvedergonomic gripping side walldesign to prevent userapplying force to catheteradapter directlyTip Shield -Straight side walldesign
10. PackagingPackage -Rigid designPackage -Blister package of
constrained unit package(dispenser, shipper scaled tonew unit package)assembly
11. Straight luer adapterStraight Luer Adapter -added new configurationStraight Luer Adapter -didnot exist
12. Vent plugVent Plug -placed in straightadapter for air ventingVent Plug -placed on "Y"adapter for air venting
13. Q-Syte, Q-Syte DustCover, and End CapQ-Syte, Q-Syte Dust Coverand End Cap -removed forstraight configurationQ-Syte, Q-Syte Dust Coverand End Cap -placed on "Y"configuration
14. CannulaCannula -bump indented intocannula for tip shield retentionCannula -ferrule bound tocannula for tip shield retention
15. Retaining featureRetaining Feature -washerretains cannula "bump" toprevent tip shield removalRetaining Feature -retentionplate retains cannula ferrule toprevent tip shield removal
16. V-ClipV-Clip -flag for cannula tipbarrierV-Clip -no cannula tip barrier(barrier provided by retentionplate)
17. Cannula, CatheterCannula -ID, OD gauge sizeincrease as ferrule removedCatheter -tip resized tocannula gauge increaseCannula -smaller gauge sizeto accommodate cannulaferruleCatheter -tip confirms tocannula gauge size
18. WedgeWedge -Improved reduced IDwedge design as ferrule wasremoved from cannulaWedge -larger ID wedge toaccommodate cannula ferrule
19. Tip shieldTip Shield -internal ID holesscaled to smaller cannula sizeTip Shield -internal IDincreased for larger cannulasize
20. GripGrip -internal ID increased forlarger cannula sizeGrip -internal ID holes scaledto smaller cannula size

{2}------------------------------------------------

·

.

{3}------------------------------------------------

{4}------------------------------------------------

Nonclinical Tests Support Substantial Equivalence:

The BD Nexiva™ Closed IV Catheter System is composed of material that have been tested in accordance with ANSI/AAMI/ISO 10993-1 2003 (E) standards. The design verification activities, acceptance criterion, and results have been summarized below which demonstrates safety and efficacy of the device.

.

AssociatedModificationProduct Performance Characteristics/Verification TestingAcceptance Criteria forProduct Verification
1Premature Decouple ForceDPPM < 280
2Catheter Pull ForceDPPM < 60
2Catheter Burst StrengthDPPM < 60
2Catheter Penetration ForceDPPM <1350
2Catheter DragDPPM <1350
2Tip AdhesionDPPM <1350
1System Drag -Cannula DragDPPM < 500
1System Drag -Ferrule Retraction ForceDPPM < 500
1Adapter Release ForceDPPM < 500
3Needle Cover RemovalDPPM USL < 18000
3DPPM LSL < 200000
3-20%/+25%, DPPM < 5000
4Flow Rate Label Claim-10%/+15%, DPPM < 5000
5Pinch Clamp Fluid SealDPPM < 10000
5Pinch Clamp Engagement Disengagement ForceDPPM < 2000
8Wing deflection forceDPPM USL < 2000
8DPPM LSL < 500
6Bond Strength -Tubing Bond to AdapterDPPM < 0.1
4Extension Tube Burst PressureDPPM < 60
10Unit Package Seal Peel Force1.33 Cpk
10Seal Width1.33 Cpk
10Unit PackageMaterial Thickness - Post Formed (Bottom Wed)1.33 Cpk
10Package Integrity (Water Leak)0 out of 2301
10Ship Testing (Drop/Vibration)0 out of 2301
7Y-Adapter Wishbone LoadingDPPM < 0.1
7ISO Liquid LeakageDPPM < 500
7ISO Air LeakageDPPM < 500
7ISO Unscrewing TorqueDPPM < 500
7ISO Resistance to OverridingDPPM < 500
7ISO Thread SeparationDPPM < 500
7Vent Plug Removal Force/TorqueDPPM USL < 6500
7DPPM LSL <10000
7Q-Syte Removal TorqueDPPM USL < 6500
7DPPM LSL <10000

{5}------------------------------------------------

AssociatedmodificationProduct Performance Characteristics/Verification TestingAcceptance Criteria forPerformance TestingResult
11ISO Thread Seperation -Both Y female luerDPPM < 500Pass
11ISO Thread Seperation -Straight female luerDPPM < 500Pass
11ISO Unscrewing Torque -Both Y female luerDPPM < 500Pass
11ISO Unscrewing Torque -Straight female luerDPPM < 500Pass
11ISO Resistance to Override -Both Y female luerDPPM < 500Pass
11ISO Resistance to Override -Straight female luerDPPM < 500Pass
11ISO Liquid LeakageDPPM < 500Pass
11ISO Air LeakageDPPM < 500Pass
12Vent Plug Torque RemovalDPPM USL < 10000DPPM LSL< 20000Pass
11Bond Strength - Tubing Bond to AdapterDPPM < 60Pass
11Flow Rate Label Claim-10%/+15%, DPPM < 5000Pass
14,15Proximal Re-exposureDPPM < 100Pass
17System Drag -Maximum Adhesion ForceDPPM < 1350Pass
17System Drag - Average Drag ForceDPPM < 10,000Pass
17System Drag - Offset Peak ForceDPPM < 2000Pass
17Flow Rate Label Claim-20%/+25%, DPPM < 5000-10%/+15%, DPPM < 5000Pass
17Minimum FlowDPPM < 5000Pass
17FlashbackDPPM < 1000Pass
17Cannula PenetrationDPPM < 1350Pass
17Catheter PenetrationDPPM < 1350Pass
17,19Premature DecoupleDPPM < 280Pass
20Needle Hub PullDPPM < 500Pass

Conclusions from Nonclinical Tests:

The BD Nexiva™ Closed IV Catheter System modifications presented in this submission are substantially equivalent to the currently marketed predicate device (K032843). The intended use and technological characteristics of the BD Nexiva™ Closed IV Catheter System product modifications described in this submission have not been altered, and therefore have remained the same since receiving their original 510(k) approval from the FDA.

The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without premarket approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to or in support of substantial equivalence herein shall be

{6}------------------------------------------------

construed as an admission against interest under the US Patent Laws or their application by the courts.


{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

C (a);

Ms. Justice Alder Specialist. Regulatory Affairs Becton Dickinson Infusion Therapy Systems, Incorporated 9450 South State Street Sandy, Utah 84070

Re: K102520

Trade/Device Name: BD Nexiva™ Closed IV Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: December 6, 2010 Received: December 10, 2010

Dear Ms. Alder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{8}------------------------------------------------

Page 2- Ms. Alder

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

James J. Roberts
for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital.\ Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{9}------------------------------------------------

K102520

INDICATIONS FOR USE

Device Proprietary Name:

BD Nexiva™ Closed IV Catheter System

Device Classification Name:

Intravascular Catheter (80 FOZ)

Indications for Use:

As indicated in 21 CFR Part 880.5200, The Nexiva™ intravascular catheter is inserted into a patient's vascular system for a short-term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature and luer access port, aid in the prevention of needle-stick injuries. Blood is contained within the device during the catheter insertion process aiding in the prevention of blood exposure. This catheter may be used for any patient population with consideration given to adequacy of vascular anatomy and appropriateness of procedure.

The 18-22 gauge Nexiva™ catheters are suitable for use with power injectors rated for a maximum of 300 psi when the luer access port(s) is removed and a direct connection is made.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use: X

OR Over-The-Counter Use: __

Acting for
RICHARD CHAPMAN

(Sign Off)

(Phusion of Anesthesiology, General Hospital infection Control, Dental Devices

Special 510(k): Device Modification Indications for Use

§ 880.5200 Intravascular catheter.

(a)
Identification. An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.(b)
Classification. Class II (performance standards).