K Number
K171491

Validate with FDA (Live)

Date Cleared
2017-10-05

(136 days)

Product Code
Regulation Number
870.1340
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

R2PTM Destination Slender™ Guiding Sheath is indicated to be used for the introduction of interventional and diagnostic devices in the lower extremities of the peripheral vasculature through an access site, including tut not limited to the radial artery.

Device Description

The R2PTM Destination Slender™ Guiding Sheath is a low profile guiding sheath designed to perform as a guiding catheter and an introducer sheath. The sheath is coil reinforced, has a radiopaque tip, is hydrophilically coated and is available in 6Fr with a length of 119cm and 149cm. It comes packaged with a dilator and hemostatic valve.

AI/ML Overview

The provided FDA 510(k) summary for the Terumo R2P™ Destination Slender™ Guiding Sheath describes non-clinical performance testing to demonstrate substantial equivalence to predicate devices. It does not include clinical study data or information about AI/ML algorithms. Therefore, many of the requested categories related to clinical studies, expert-derived ground truth, and AI performance metrics cannot be populated from the provided text.

Here's a breakdown of the available information:

1. Table of acceptance criteria and the reported device performance

The document states that the R2P™ Destination Slender™ Guiding Sheath "met the predetermined acceptance criteria and results support a determination of substantial equivalence." However, the specific quantitative acceptance criteria for each test item are not explicitly listed in the provided tables, only the test items themselves and the reference standards used.

Test ItemReference Standard / Test TypeReported Device Performance
Three-Point BendIn-house standardMet predetermined acceptance criteria
Hub Joint StrengthISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013Met predetermined acceptance criteria
Tubing Tensile StrengthISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013Met predetermined acceptance criteria
Tip Tensile StrengthISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013Met predetermined acceptance criteria
OvalizationIn-house standardMet predetermined acceptance criteria
Lubricity and DurabilityIn-house standardMet predetermined acceptance criteria
Uncoated LengthIn-house standardMet predetermined acceptance criteria
Corrosion ResistanceISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013Met predetermined acceptance criteria
GaugingISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-1:1986Met predetermined acceptance criteria
Liquid Leakage of Luer under PressureISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-2:1998Met predetermined acceptance criteria
Air LeakageISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-2:1998Met predetermined acceptance criteria
Separation ForceISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-2:1998Met predetermined acceptance criteria
Unscrewing TorqueISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-2:1998Met predetermined acceptance criteria
Ease of AssemblyISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-2:1998Met predetermined acceptance criteria
Resistance to OverridingISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013Met predetermined acceptance criteria
Stress CrackingISO 594-2:1998, ISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013, ISO 594-2:1998Met predetermined acceptance criteria
Simulated Use and ParticulateFDA Guidance Doc 1608Met predetermined acceptance criteria
RadiodetectabilityISO 11070:2014, ISO 10555-1:2013Met predetermined acceptance criteria
Visual Inspection of External SurfaceISO 11070:2014, ISO 10555-1:2009, ISO 10555-1:2013Met predetermined acceptance criteria
Sheath Tip Inner DiameterIn-house standardMet predetermined acceptance criteria
Sheath Hub Inner DiameterIn-house standardMet predetermined acceptance criteria
Dilator Bump Inner DiameterIn-house standardMet predetermined acceptance criteria
Outer Diameter MeasurementIn-house standardMet predetermined acceptance criteria
Effective LengthIn-house standardMet predetermined acceptance criteria
Coating IntegrityFDA Guidance Doc 1608Met predetermined acceptance criteria
Liquid Leakage of the SheathISO 11070:2014Met predetermined acceptance criteria
Liquid Leakage Through Hemostasis ValveISO 11070:2014Met predetermined acceptance criteria
Torque StrengthFDA Guidance Doc 1608Met predetermined acceptance criteria
Dilator Tip Inner DiameterIn-house standardMet predetermined acceptance criteria
Dilator Tip PenetrationIn-house standardMet predetermined acceptance criteria
BiocompatibilityISO 10993-1 (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity (Acute), Pyrogenicity, Hemocompatibility, Thrombogenicity, Complement Activation, Physiochemical Testing)Demonstrated biocompatibility of the finished device
SterilizationISO 11135:2014 (Ethylene Oxide - SAL 10-6)Sterility assured, residuals meet EN ISO 10993-7:2008 limits

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified in the document.
  • Data Provenance: The document describes "non-clinical tests" and "performance testing" conducted. This refers to laboratory testing of the physical device, not patient data. Therefore, questions of country of origin or retrospective/prospective do not apply in the typical sense for clinical data. The manufacturer is Terumo Medical Corporation, located in the USA (Somerset, NJ and Elkton, MD).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This question is not applicable. The testing described is non-clinical performance and biocompatibility testing of a physical medical device. "Ground truth" in the context of expert review for diagnostic AI software is not relevant here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable. Adjudication methods are typically for resolving discrepancies in expert interpretations of clinical data, which is not part of this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This submission is for a physical medical device (guiding sheath), not AI software, and no clinical studies, especially MRMC, were conducted or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. This device is not an algorithm or AI software.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This question is not applicable in the context of clinical "ground truth." For the non-clinical performance tests, the "ground truth" would be established engineering specifications, industry standards (ISO), and in-house standards against which the device's physical properties are measured.

8. The sample size for the training set

This question is not applicable. This device does not involve a "training set" as it is a physical medical device, not an AI/ML algorithm.

9. How the ground truth for the training set was established

This question is not applicable. As explained above, there is no "training set" for this device.

{0}------------------------------------------------

October 5, 2017

Terumo Medical Corporation Liang Lu Senior Regulatory Affairs Specialist 950 Elkton Blvd Elkton, Maryland 21921

Re: K171491

Trade/Device Name: R2PTM Destination Slender™ Guiding Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: September 6, 2017 Received: September 7, 2017

Dear Liang Lu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171491

Device Name

R2PTM Destination Slender™ Guiding Sheath

Indications for Use (Describe)

R2P™ Destination Slender™ Guiding Sheath is indicated to be used for the introduction of interventional and diagnostic devices in the lower extremities of the peripheral vasculature through an access site, including tut not limited to the radial artery.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K171491 510(K) SUMMARY

A. SUBMITTER INFORMATION (807.92(a)(1))

Prepared by:

Liang Lu Senior Regulatory Affairs Specialist Terumo Medical Corporation Tel. (410) 392-7321 Fax (410) 398-6079

Prepared for: Owner/Operator Terumo Medical Corporation 2101 Cottontail Lane Somerset, NJ 08873, USA Registration Number: 2243441

Manufacturer (510(k) Applicant)

Terumo Medical Corporation 950 Elkton Boulevard, Elkton, MD 21921 Registration Number: 1118880

Sterilization Facility

Steris Isomedix Services Inc. 3459 South Clinton Ave South Plainfield, NJ 07080 Registration Number: 2246552

Contact Person: Liang Lu

Senior Regulatory Affairs Specialist Terumo Medical Corporation 950 Elkton Boulevard, Elkton, MD 21921 Tel. (410) 392-7321 Fax (410) 398-6079 E-mail: liang.lu@terumomedical.com

Date prepared: May 11, 2017

{4}------------------------------------------------

B. DEVICE NAME (807.92(a)(2))

Proprietary Name:R2PTM Destination Slender™ Guiding Sheath
Common Name:Guiding Sheath
Classification Name:Catheter Introducer
Classification Panel:Cardiovascular
Regulation:21 CFR 870.1340
Product Code:DYB
Classification:Class II

C. PREDICATE DEVICE (807.92(a)(3))

The legally marketed device(s) to which substantial equivalence is claimed are:

  • . Primary Predicate Device: K052185 - Destination Carotid Guiding Sheath, manufactured by Terumo Medical Corporation, USA
  • Reference Device: K161546 R2P SlenGuide, manufactured by Terumo . Corporation

D. REASON FOR 510(k) SUBMISSION

This premarket notification (510(k)) is being submitted for the R2P™ Destination Slender™ Guiding Sheath, manufactured by Terumo Medical Corporation, for the purposes of establishing substantial equivalence to a legally marketed predicate device.

E. DEVICE DESCRIPTION (807.92(a)(4))

The R2PTM Destination Slender™ Guiding Sheath is a low profile guiding sheath designed to perform as a guiding catheter and an introducer sheath. The sheath is coil reinforced, has a radiopaque tip, is hydrophilically coated and is available in 6Fr with a length of 119cm and 149cm. It comes packaged with a dilator and hemostatic valve.

F. INDICATIONS FOR USE (807.92(a)(5))

R2P™ Destination Slender™ Guiding Sheath is indicated to be used for the introduction of interventional and diagnostic devices in the lower extremities of the peripheral vasculature through an access site, including but not limited to the radial artery.

G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))

The R2PTM Destination Slender™ Guiding Sheath, subject of this 510(k), is substantially equivalent in its intended use/indications for use, technology/principal of operation, materials, and performance to the predicate device, manufactured by Terumo Medical Corporation.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green, block letters. The swoosh is positioned above and to the left of the word.

A comparison of the technological characteristics is summarized on the table below.

DeviceCharacteristicNew Device:R2PTM DestinationSlender™ GuidingSheathPrimary Predicate:Destination CarotidGuiding Sheath(K052185)Reference Device:R2P SlenGuide(K161546)
ManufacturerTerumo MedicalCorporationTerumo MedicalCorporationTerumo Corporation
Intended Use /Indications forUseR2PTM DestinationSlender™ GuidingSheath is indicated tobe used for theintroduction ofinterventional anddiagnostic devices inthe lower extremities ofthe peripheralvasculature through anaccess site, includingbut not limited to theradial artery.The Destination®Guiding Sheath isdesigned to be used forthe introduction ofinterventional anddiagnostic devices intothe human vasculatureincluding but notlimited to the carotidarteries.The R2P (radial toperipheral)SlenGuide isdesigned for theintroduction ofinterventional anddiagnostic devicesinto theperipheralvasculature ofthe lower extremities
OperationPrincipleOperated manually orby a manualprocess;SameSame
Design /ConstructionSheath, Dilator,Hemostatic Valve withside tube and three-waystopcockSameGuiding Catheter,Inner Guide

Table 12.1: Summary of Substantial Equivalence

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green. The font is sans-serif and bolded.

MaterialsSheath AssemblySheath AssemblyGuiding Catheter:
Tubing:Tubing:Outer layer:
Inner Layer: PTFEInner Layer: PTFEPolyamide elastomer(Nylon)
Middle Coil Layer:Middle Coil Layer:Middle Braid Layer:
Stainless SteelStainless SteelStainless steel
Outer Layer: NylonOuter Layer: NylonInner layer:
Polytetrafluoroethylene
Radiopaque Tip: Nylonwith TungstenTip: NylonRadiopaque Band:GoldRadiopaque Tip:Polyamide Elastomer(Nylon) with Tungsten
Hydrophilic Coating:Polyvinylpyrrolidone-based coatingHydrophilic Coating:Polyvinylpyrrolidone-based coatingHydrophilic Coating:Dimethyl acrylic amide- glycidyl methacrylate
Hub: NylonHub: NylonHub: Polyamide 12
Anti-kink protector:NylonAnti-kink protector:Polyester Elastomer
Dilator AssemblyTubing: PolypropyleneHub: Polypropylene//ThermoplasticElastomer BlendCoating: SiliconeCaulking Pin: StainlesssteelDilator AssemblySameInner Guide:Shaft: PolyesterElastomer withTungstenHub: PolyamideLock Adaptor:polycarbonate

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green. The font is sans-serif and bold.

Cross Cut ValveCross Cut ValveNone
Valve Assembly:Valve Assembly:
Housing:Same
Polypropylene
Cap: Polypropylene
Luer Lock Collar:
Polycarbonate
Valve: Silicone Rubber
Elastomer
Sidetube:
polybutadiene
Silicone: Non-reactive
silicone oil 1000cst
Side Tube Assembly:Side Tube Assembly:
Body: PolybutadieneSame
3Way (3WSC)3Way (3WSC)
Stopcock Assembly:Stopcock Assembly:
Body: PolycarbonateSame
Locking
Pin: Polyethylene
Cap: Polyethylene and
Colorant
Handle: Polyethylene
and ColorantTuohy-Borst Valve
Lock Adaptor:Similar material as
PolycarbonateCross Cut Valve
PackageUnit PouchUnit PouchUnit Pouch
Shelf BoxShelf BoxShelf Box
Shipping CartonShipping CartonShipping Carton

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red arc above the letters "TE". The font is bold and sans-serif. The logo is simple and modern.

SpecificationsSheath Size: 6 Fr.Sheath Size: 6-7 Fr.Catheter Size: 7Fr
Sheath ID/OD(nominal):Sheath ID/OD(nominal):I.D.: 2.20 mm (0.087")
6Fr.: 0.087"/0.100"(2.2mm /2.5mm)6Fr.: 0.087"/0.111"7Fr.: 0.100"/0.122"O.D.: 2.37 mm(0.093")
Sheath Length:119cm, 149cmSheath Length:80-110cmEffective lengths:120cm, 150cm
Hydrophilic Coating:full effective lengthHydrophilic Coating:Distal 15cm
Distal Shape: StraightDistal Shapes: Straight,Angled (40 degrees)
Dilator ID/OD(nominal):Dilator ID/OD(nominal):Inner GuideAccepts guide wirediameter: 0.035"
0.039"/0.086"6Fr.: 0.045”/0.084"7Fr.: 0.045”/0.097"
Dilator ExtendedLength: 5cmDilator ExtendedLength: 5cm
SterilizationEthylene Oxide(validated inaccordance with ANSI/ AAMI / ISO 11135-1to achieve SAL 10-6)SameSame
Shelf life30 monthsSame36 months
DisposableSingle UseYesSameSame

H. NON-CLINICAL TESTS (807.92(b)(1))

Performance

Performance testing was conducted to ensure that the R2P™ Destination Slender™ Guiding Sheath met the applicable design and performance requirements throughout the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device.

The following table provides a list of the performance tests that were performed on the proposed R2P™ Destination Slender™ Guiding Sheath.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red swoosh above the letters "TE". The font is bold and sans-serif. The logo is simple and modern.

Table 5.4: Summary of Performance Testing
Test ItemReferenceComponent(Sheath,Dilator, CCVAssembly)
Three-Point BendIn-house standardSheath
Hub Joint StrengthISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
Tubing Tensile StrengthISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
Tip Tensile StrengthISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
OvalizationIn-house standardSheath
Lubricity and DurabilityIn-house standardSheath
Uncoated LengthIn-house standardSheath
Corrosion ResistanceISO 11070:2014Sheath, Dilator
ISO 10555-1:2009
ISO 10555-1:2013
GaugingISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
ISO 594-1:1986
Liquid Leakage of Luerunder PressureISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
ISO 594-2:1998
Air LeakageISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
ISO 594-2:1998
Separation ForceISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
ISO 594-2:1998
Unscrewing TorqueISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
ISO 594-2:1998
Ease of AssemblyISO 11070:2014Sheath
ISO 10555-1:2009
ISO 10555-1:2013
Test ItemReferenceComponent(Sheath,Dilator, CCVAssembly)
ISO 594-2:1998
Resistance to OverridingISO 11070:2014ISO 10555-1:2009ISO 10555-1:2013Sheath
Stress CrackingISO 594-2:1998ISO 11070:2014ISO 10555-1:2009ISO 10555-1:2013ISO 594-2:1998Sheath
Simulated Use andParticulate*FDA Guidance Doc 1608Sheath, Dilator,CCV
RadiodetectabilityISO 11070:2014ISO 10555-1:2013Sheath, Dilator
Visual Inspection ofExternal SurfaceISO 11070:2014ISO 10555-1:2009ISO 10555-1:2013Sheath, Dilator
Sheath Tip Inner DiameterIn-house standardSheath
Sheath Hub Inner DiameterIn-house standardSheath
Dilator Bump InnerDiameterIn-house standardDilator
Outer DiameterMeasurementIn-house standardSheath, Dilator
Effective LengthIn-house standardSheath, Dilator
Coating IntegrityFDA Guidance Doc 1608Sheath
Liquid Leakage of theSheathISO 11070:2014Sheath
Liquid Leakage ThroughHemostasis ValveISO 11070:2014CCV
Torque StrengthFDA Guidance Doc 1608Sheath
Dilator Tip Inner DiameterIn-house standardDilator
Dilator Tip PenetrationIn-house standardDilator

Table 5.4: Summary of Performance Testing

{10}------------------------------------------------

The R2P™ Destination Slender™ Guiding Sheath tested met the predetermined acceptance criteria and results support a determination of substantial equivalence. Based on the results of the performance testing, the proposed R2PTM Destination Slender™ Guiding Sheath is substantially equivalent.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green. The word "TERUMO" is in a bold, sans-serif font.

Biocompatibility

Biocompatibility classification is based on the FDA Guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process."

The R2PTM Destination Slender™ Guiding Sheath is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours).

The following tests are recommended by FDA and ISO 10993-1 to be performed per this device classification:

    1. Cytotoxicity
    1. Sensitization
    1. Intracutaneous Reactivity
    1. Systemic Toxicity (Acute)
    1. Pyrogenicity
    1. Hemocompatibility
    1. Thrombogenicity
    1. Complement Activation (Immunology)
    1. Physiochemical Testing

Results of the testing demonstrate biocompatibility of the finished R2PTM Destination Slender™ Guiding Sheath.

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.

The sterilization process was validated utilizing the overkill half cycle approach to provide a Sterility Assurance Level (SAL) of 10-6.

R2P™ Destination Slender™ Guiding Sheath is a limited exposure device. After 24 hours of heated aeration, the level of residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) do not exceed an average daily dose of 4 mg and 9 mg respectively per EN ISO 10993-7:2008.

I. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the logo for Terumo. The logo consists of a red curved line above the word "TERUMO" in green, block letters. The red line appears to be a stylized representation of a blood vessel or a medical device.

J. CONCLUSION (807.92(b)(3))

In summary, the R2PTM Destination Slender™ Guiding Sheath, subject of this 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the predicate device(s):

  • Primary Predicate Device: K052185 Destination Carotid Guiding Sheath, ● manufactured by Terumo Medical Corporation, USA
  • Reference Device: K161546 R2P SlenGuide, manufactured by Terumo . Corporation

§ 870.1340 Catheter introducer.

(a)
Identification. A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.(b)
Classification. Class II (performance standards).