K Number
K161546

Validate with FDA (Live)

Device Name
R2P SlenGuide
Date Cleared
2016-10-14

(133 days)

Product Code
Regulation Number
870.1250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The R2P (radial to peripheral) SlenGuide is designed for the introduction of interventional and diagnostic devices into the peripheral vasculature of the lower extremities.

Device Description

The R2P SlenGuide is a single use, ethylene oxide sterilized device that is designed to perform as a guiding catheter for interventional procedures in the peripheral vasculature. It is packaged with a guiding catheter and an inner guide. The guiding catheter features a three-layer construction with a stainless steel mesh, polyamide elastomer, and polytetrafluoroethyelene. The distal end has a hydrophilic coating and a soft-tip visible under fluoroscopy. The inner guide is an accessory device made of polyester elastomer with a flexible distal portion containing tungsten, visible under fluoroscopy.

AI/ML Overview

The provided document describes the R2P SlenGuide, a percutaneous catheter, and its testing to demonstrate substantial equivalence to predicate devices. It does not contain information about a study proving the device meets acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy, or effect size for AI assistance. Instead, the "acceptance criteria" here refer to meeting standards for safety and performance characteristics through non-clinical testing.

Here's a breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The document lists various non-clinical tests performed and states that "Performance testing met the predetermined acceptance criteria." However, it does not explicitly state the numerical acceptance criteria for each test or the specific reported performance results in a detailed, quantitative table. It only confirms that the criteria were met.

For example, for "Peak tensile force," the test procedure is described (measure peak tensile strength), but the acceptable range (e.g., >X Newtons) and the actual measured value are not provided in this summary. The same applies to other tests.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document mentions that performance tests were performed on "non-aged and accelerated aged samples." However, it does not specify the number of samples used for each test.
  • Data Provenance: The tests are described as non-clinical performance testing conducted by the manufacturer, Terumo Corporation, or its Ashitaka Factory in Japan. The data is retrospective in the sense that it was collected as part of the device development and regulatory submission process.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable as the document describes non-clinical performance and biocompatibility testing of a physical medical device (a catheter), not an AI/software device that would require expert-established ground truth for a test set.

4. Adjudication Method

This information is not applicable for the same reason as point 3.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This information is not applicable as the device is a physical catheter, not an AI system. There is no mention of human readers or AI assistance in this context.

6. Standalone (Algorithm Only) Performance Study

This information is not applicable as the device is a physical catheter, not an algorithm.

7. Type of Ground Truth Used

This information is not applicable as the document describes non-clinical performance and biocompatibility testing of a physical medical device. The "ground truth" for these tests would be the established scientific and engineering principles and the specific requirements outlined in the referenced ISO and ASTM standards, as well as internal standards.

8. Sample Size for the Training Set

This information is not applicable as the device is a physical catheter, not an AI/machine learning device that would require a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.


Summary of Non-Clinical Tests Performed (as a proxy for "study")

The study to demonstrate that the device meets performance requirements involved a series of non-clinical, in-vitro tests based on established international standards (ISO and ASTM) and FDA guidance documents. These tests were categorized into performance testing and biocompatibility testing.

Performance Testing (Guiding Catheter & Inner Guide):

  • Standards Referenced: ISO 10555-1:2013, ASTM F640-12, USP <788>, FDA Guidance "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters."
  • Tests Included:
    • Radio-detectability
    • Surface quality
    • Peak tensile force
    • Freedom from leakage
    • Hub performance (for Guiding Catheter)
    • Distal tip appearance
    • Particulate evaluation
    • Torque Strength (for Guiding Catheter)
    • Coating lubricity (for Guiding Catheter)
    • Evaluation of flexibility and kink resistance (for Guiding Catheter)
    • Distal tip flexibility (for Guiding Catheter)
    • Kink condition (for Guiding Catheter)
    • Flexural rigidity
    • Distal tip strength (for Guiding Catheter)
    • Cleanliness
    • Product dimensions

Biocompatibility Testing:

  • Standards Referenced: ISO 10993-1, FDA General Program Memorandum #G95-1 (5/1/95), Draft Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993.
  • Classification: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours).
  • Biocompatibility Tests Performed (on non-aged and accelerated-aged samples for some tests):
    • Cytotoxicity
    • Sensitization
    • Intracutaneous Reactivity
    • Acute Systemic Toxicity
    • Pyrogenicity
    • Hemolysis
    • Thrombogenicity
    • Complement Activation (Immunology)
    • Physicochemical Profile (Physicochemical and FT-IR)

Conclusion of Testing: "Performance testing met the predetermined acceptance criteria and results support a determination of substantial equivalence." and "Results of the [biocompatibility] testing demonstrate that the device is biocompatible throughout the shelf life of the product."

Sterilization Validation:

  • Standard Referenced: ISO 11135:2014 ("Sterilization of Health Care Products - Ethylene Oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices").
  • Result: Achieved a Sterility Assurance Level (SAL) of 10-6.

In essence, the "study" demonstrating the device meets its "acceptance criteria" here refers to the comprehensive non-clinical testing outlined above, which confirmed adherence to recognized standards for safety, performance, and biocompatibility, thereby supporting the claim of substantial equivalence to predicate devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 14, 2016

Terumo Medical Corporation Ms. Monika McDole-Russell Senior Regulatory Affairs Specialist 265 Davidson Avenue, Suite 320 Somerset. NJ 08873

Re: K161546

Trade/Device Name: R2P SlenGuide Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: September 13, 2016 Received: September 14, 2016

Dear Ms. McDole-Russell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/7 description: The image shows the name "Brian D. Pullin -S" in a large, bold font. The text is black and is set against a light blue background. The letters are clear and easy to read.

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K161546

Device Name R2PTM SlenGuide™

Indications for Use (Describe)

The R2P (radial to peripheral) SlenGuide is designed for the introduction of interventional and diagnostic devices into the peripheral vasculature of the lower extremities.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

A. SUBMITTER INFORMATION (807.92(a)(1))

Prepared by:Monika McDole-Russell, MSRA, RACManager, Regulatory AffairsTerumo Medical CorporationTel. (732) 302-4900 ext.4273Fax (732) 302-4905
-------------------------------------------------------------------------------------------------------------------------------------------------------------------
Prepared for:Owner/Operator
Terumo Corporation
44-1, 2-Chome, Hatagaya
Shibuya-Ku, Tokyo
Japan 151-0072
Owner/Operator Number: 801 002 6

Manufacturer and Sterilization Facility (Applicant)

Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya, Shizuoka 418-0015, Japan Registration Number: 968 183 4

Contact Person:Monika McDole-Russell, MSRA, RAC
Manager, Regulatory Affairs
Terumo Medical Corporation
265 Davidson Avenue, Suite 320
Somerset, NJ 08873
Tel.(732) 302-4900 ext.4273
Fax(732) 302-4905
E-mail:monika.mcdolerussell@terumomedical.com

Date prepared: October 7, 2016

{4}------------------------------------------------

B. DEVICE NAME (807.92(a)(2))

Proprietary Name:R2PTM SlenGuideTM
Common Name:Guiding Catheter
Classification Name:Percutaneous Catheter
Classification Panel:Cardiovascular
Regulation:21 CFR 870.1250
Product Code:DQY
Classification:Class II

C. PREDICATE DEVICES (807.92(a)(3))

The legally marketed device(s) to which substantial equivalence is claimed are:

  • Predicate Device: K972978 Vista Brite Tip Guiding Catheters, manufactured by . Cordis Corporation.
  • Reference Device 1: K142819 Shuttle-SL Flexor Tuohy-Borst Side-Arm . Introducer Set, manufactured by Cook, Inc. (hereinafter referred to as "Flexor Introducer Set")*

*Flexor Introducer Set (K142819) is presented as a reference predicate for the Inner Guide.

  • Reference Device 2: K090040 Radifocus Glidecath/Glidecath XP, manufactured . by Terumo Corporation.
  • Radifocus Glidecath/Glidecath XP (K090040) is presented as a reference predicate for device length.

D. REASON FOR 510(k) SUBMISSION

This premarket notification (510(k)) is being submitted for the R2P SlenGuide, manufactured by Ashitaka Factory of Terumo Corporation, for the purposes of establishing substantial equivalence to a legally marketed predicate device.

E. DEVICE DESCRIPTION (807.92(a)(4))

Principle of Operation Technology

The R2P SlenGuide is operated by manual process.

Design/Construction

The R2P SlenGuide is a single use, ethylene oxide sterilized device that is designed to perform as a guiding catheter for interventional procedures in the peripheral vasculature. It is packaged with a guiding catheter and an inner guide; a description of

{5}------------------------------------------------

each component is provided below.

Guiding Catheter: features a three-layer construction, which consists of a stainless steel mesh sandwiched between an outer layer of polyamide elastomer and an inner layer of polytetrafluoroethyelene. The distal end has a hydrophilic coating and contains a "soft-tip," which is visible under fluoroscopy.

Inner Guide: is an accessory device for the guiding catheter. Its purpose is to help direct the guiding catheter to the lesion, and once achieved, the inner guide is removed, in order to proceed with the interventional procedure. It is comprised of polyester elastomer, and the distal portion is a flexible polyester elastomer, containing tungsten, which is visible under fluoroscopy.

Materials

The materials for the R2P SlenGuide are provided in the table below.

Name of ComponentRaw Material
GuidingCatheterShaftPolyamide elastomerPigment
Outer layer*DistalpartPolyamide elastomerPigment
Soft-tipPolyamide elastomerTungsten
Inner layer*Polytetrafluoroethylene
Braid†Stainless steel
Hub*Polyamide 12Pigment
Hydrophilic polymer coating*Dimethyl acrylic amide -glycidyl methacrylatecopolymer
Anti-kink protectorPolyester elastomerPigment
InnerGuideShaft*Distal partPolyester elastomerTungsten
Proximal partPolyester elastomer
Hub*Polyamide
Lock adaptorPolycarbonate
AdhesiveCyanoacrylate

Table 5.1: List of Materials

*Blood contacting material.

*Soft-tip is not braided.

{6}------------------------------------------------

Specifications

The specifications for the R2P SlenGuide are provided in the table below.

PartSpecification
Catheter Size7Fr.
Catheter ID/OD2.20 mm/2.37 mm
Catheter Effective Lengths*120 and 150 cm
Inner Guide ID/OD1.15 mm/2.10 mm
Inner Guide Extended Length†30 mm
Accepts Guide Wire Diameter0.035"

Table 5.2: R2P SlenGuide Specifications

*The length from the proximal anti-kink protector to the guiding catheter distal tip. *The length that the inner guide extends past the guiding catheter's tip.

F. INDICATIONS FOR USE (807.92(a)(5))

The R2P (radial to peripheral) SlenGuide is designed for the introduction of interventional and diagnostic devices into the peripheral vasculature of the lower extremities.

G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))

The R2P SlenGuide, subject of this Traditional 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to:

  • Predicate Device: K972978 Vista Brite Tip Guiding Catheters, manufactured by . Cordis Corporation.
  • Reference Device 1: K142819 Flexor Introducer Set*, manufactured by Cook, . Inc.

*Flexor Introducer Set (K142819) is presented as a reference predicate for the Inner Guide.

  • Reference Device 2: K090040 Radifocus Glidecath/Glidecath XP*, manufactured . by Terumo Corporation.
  • Radifocus Glidecath/Glidecath XP (K090040) is presented as a reference predicate for device length.

A comparison of the technological characteristics is summarized in the table below.

{7}------------------------------------------------

Device CharacteristicNew Device:Guiding Catheter of R2PSlenGuidePredicate Device:Vista Brite TipGuiding Catheters(K972978)*Reference Device 1:Flexor Introducer Set(K142819)*Reference Device 2:Radifocus Glidecathor Glidecath XP(K090040)
ManufacturerTerumo CorporationCordis CorporationCook, Inc.Terumo Corporation
IntendedUse/Indication forUseThe R2P (radial toperipheral) SlenGuide isdesigned for theintroduction ofinterventional anddiagnostic devices into theperipheral vasculature ofthe lower extremitiesThe guiding catheter isintended for use forintravascularintroduction ofinterventional /diagnostic devices intothe coronary orperipheral vascularsystemsFlexor Introducers andGuiding Sheaths areintended to introducetherapeutic ordiagnostic devices intothe vasculature,excluding coronary andneuro vasculatureThe Radifocus Glidecath(or Radifocus GlidecathXP) is intended for use inangiographic procedures.It delivers radiopaquemedia and therapeuticagents to selected sites inthe vascular system. It isalso used to lead a guidewire or catheter into thetarget site.
Operation PrincipleManualsamesamesame
Design/ConstructionGuidingCatheterSingle lumen catheter,stainless steel braid wire,catheter, radiopaque tip,and hubsameN/ASingle lumen catheter,stainless steel braidwire, cathetercontaining radiopaquematerial(tungsten orbarium sulfate) andhub
Inner GuideMade by plastic materials,featuring a distal taper thatallows for smoothtransition with theappropriately sized guidewireN/AsameN/A
Device CharacteristicNew Device:Guiding Catheter of R2PSlenGuidePredicate Device:Vista Brite TipGuiding Catheters(K972978)*Reference Device 1:Flexor Introducer Set(K142819)*Reference Device 2:Radifocus Glidecathor Glidecath XP(K090040)
MaterialsGuiding Catheter• Outer layer:Polyamide elastomer(Nylon)• Braid: Stainless steel• Inner layer:PolytetrafluoroethylenesameN/A
InnerGuide• Shaft: PolyesterelastomerN/A• Shaft: Polyamideelastomer (Nylon)or PolyethyleneN/A
Package• Individual package onwhich the product labeland thepeel-off labels areattached• 1 unit per packagesamesamesame
SpecificationsGuiding Catheter• Effective lengths: 120cm, 150 cm• French size: 7Fr• O.D.: 2.37 mm• I.D.: 2.20 mm (0.087")• Effective lengths:55 cm, 90 cm, 95cm• French size: 7Fr• O.D.: Soft tip:2.30 mm• I.D.: 2.0 mm(0.078")N/A• Effectivelengths: 30 - 150cm• French size:5Fr, 4Fr• O.D.:5Fr 1.70 mm4Fr 1.40mm• I.D.:5Fr 122mm(Doublebraided)1.12 mm(Singlebraided)4Fr 1.05 mm
Device CharacteristicNew Device:Guiding Catheter of R2PSlenGuidePredicate Device:Vista Brite TipGuiding Catheters(K972978)*Reference Device 1:Flexor Introducer Set(K142819)*Reference Device 2:Radifocus Glidecathor Glidecath XP(K090040)
ุการ อนเลAccepts guide wire●diameter: 0.035"N/AAccepts guide wire●diameter: 0.018",0.035", and 0.038"N/A
SterilizationEthylene oxidesamesamesame

Table 5.3: Summary of Comparative Information

{8}------------------------------------------------

{9}------------------------------------------------

*Based on publicly available information on the devices.

{10}------------------------------------------------

H. NON CLINICAL TESTS (807.92(b)(1))

Performance Testing

Performance testing was conducted to ensure that the R2P SlenGuide met the applicable design and performance requirements throughout its shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device. With the exception of the Radiodetectability] and Cleanliness tests2, the following performance tests were performed on non-aged and accelerated aged samples. The following tables provide a list of performance tests that were performed on the R2P SlenGuide.

ComponentTestStandardTest Method
GuidingCatheterRadio-detectabilityISO 10555-1:2013Section 4.2,ASTM F640-12Test samples using X-ray equipment,that are tested according to ASTMF640-12
Surface qualityISO 10555-1:2013Section 4.4Check external surfaces of samples witha magnifier (x 2.5 times or more)
Peak tensileforceISO 10555-1:2013Section 4.6VIII.A.7 of FDAGuidance3Using an autograph, measure peaktensile strength of samples
Freedom fromleakageISO 10555-1:2013Section 4.7Using a syringe filled with water, applyand maintain pressure to the catheter,and then inspect for liquid and airleakage
HubperformanceISO 10555-1:2013Section 4.8ISO 594-2: 1998Test hubs for gauging, liquid leakage,air leakage, separation force,unscrewing torque, ease of assembly,resistance to overriding, and stresscracking
Distal tipappearanceISO 10555-1:2013Section 4.12Check physical appearance of distal tipwith a microscope (x 10 times or more)
ParticulateevaluationVIII.A.13 of FDAGuidance3USP <788>In-house StandardSlide samples into an apparatus filledwith purified water, andmeasure/evaluate residual particles

Table 5.4: Summary of Performance Testing - Guiding Catheter

1 Only non-aged sample was tested since the amount of contrast media contained in the product would not change over time.

2 Only non-aged sample was tested since the particulates inside the catheter lumen would not increase over time.

3 Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary

Angioplasty (PTCA) Catheters, September 8, 2010.

{11}------------------------------------------------

ComponentTestStandardTest Method
Torque StrengthVIII.A.10 of FDAGuidanceIn-house Standard3Confirm R2P Slenguide has appropriatetorque strength by measuring therotation of the catheter, when twistingoff
CoatinglubricityVIII.A.12 of FDAGuidance3In-house StandardAfter the particulate evaluation test,visually evaluate coating undermagnification. Then, after pushing andpulling samples in an apparatus,confirm coating lubricity by measuringsliding forces
Evaluation offlexibility andkink resistanceVIII.A.9 of FDAGuidance3In-house StandardWind samples around mandrels, untilthe samples have kinked, and recordmandrel size that caused samples tokink
Distal tipflexibilityIn-house StandardMeasure tip force when samples arepushed by an apparatus
Kink conditionIn-house StandardKink the samples, and visually inspectwith a magnifying glass
Flexural rigidityIn-house StandardSamples are set on an apparatus, andmaximum force is applied, in order tomeasure shaft strength
Distal tipstrengthVIII.A.8 of FDAGuidance3In-house StandardSamples are clamped and pulled tomeasure tensile strength
CleanlinessIn-house StandardWater is injected into the samples and isvisually inspected for foreign matter
ProductdimensionsIn-house StandardMeasure by inner and outer diameter,effective length, and extended length

Table 5.5: Summary of Performance Testing – Inner Guide

ComponentTestStandardTest Method
Inner GuideRadio-detectabilityISO 10555-1:2013Section 4.2ASTM F640-12Test samples using X-ray equipment, when tested according to ASTM F640-12
Surface qualityISO 10555-1:2013Section 4.4Check external surfaces of samples with a magnifier (x 2.5 times or more)
Peak tensile forceISO 10555-1:2013Section 4.6VIII.A.7 of FDAGuidance3Using an autograph, measure peak tensile strength of samples
Freedom fromleakageISO 10555-1:2013Section 4.7Using a syringe filled with water, apply and maintain pressure to the catheter, and then inspect for liquid and air leakage
Distal tipappearanceISO 10555-1:2013Section 4.12Check physical appearance of distal tip with a microscope (x 10 times or more)

{12}------------------------------------------------

Flexural rigidityIn-house StandardSamples are set on an apparatus, andmaximum force is applied, in order tomeasure shaft strength
CleanlinessIn-house StandardWater is injected into the samples and isvisually inspected for foreign matter
Product dimensionsIn-house StandardMeasure by inner and outer diameter,effective length, and extended length

Performance testing met the predetermined acceptance criteria and results support a determination of substantial equivalence.

Biocompatibility

In accordance with ISO 10993-1, the R2P SlenGuide is classified as: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). The finished device's patient contacting parts were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and Draft Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." Screening tests were performed on accelerated aged devices to show that biocompatibility is maintained throughout the shelf life of the product. The table below provides a list of biocompatibility tests conducted on the R2P SlenGuide.

Non-aged, sterile, whole device
Cytotoxicity
Sensitization
Intracutaneous Reactivity
Acute Systemic Toxicity
Pyrogenicity
Hemoloysis
Thrombogencity
Complement Activation (Immunology)
Physicochemical Profile (Physicochemical and FT-IR)
Accelerated-aged (3 years), sterile, whole device
Cytotoxicity
Hemolysis
Physicochemical Profile (Physicochemical and FT-IR)

Table 5.6: Summary of ISO 10993 Biocompatibility Testing

{13}------------------------------------------------

Results of the testing demonstrate that the device is biocompatible throughout the shelf life of the product.

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10-6.

I. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

J. CONCLUSION (807.92(b)(3))

In summary, the R2P SlenGuide, subject of this 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to:

  • Predicate Device: K972978 Vista Brite Tip Guiding Catheters, manufactured by . Cordis Corporation.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).