K Number
K162482

Validate with FDA (Live)

Date Cleared
2017-09-15

(374 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Two Stage Implants: I22, I5, I55, I10.

P4 and P14 angled abutments are to be used only with standard platform implants 3.5 mm in diameter or larger.

Device Description

A.B. DENTAL DEVICES® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants systems. Narrow platform implants are to be used with straight abutments only. The following components are being added:

ImplantPlatformDiameter, mmLength, mm
I22Standard3.758, 10, 11.5, 13, 16
4.28, 10, 11.5, 13, 16
4.58, 10, 11.5, 13, 16
58, 10, 11.5, 13, 16
I5Narrow3.310, 11.5, 13, 16
I55Narrow3.310, 11.5, 13, 16
3.58, 10, 11.5, 13, 16
Standard3.758, 10, 11.5, 13, 16
4.28, 10, 11.5, 13, 16
4.58, 10, 11.5, 13, 16
58, 10, 11.5, 13, 16
68, 10, 11.5, 13, 16
I10Standard4.510, 11.5, 13, 16

Abutments:

  • P0-14 Healing caps, 2.5 7 mm heights .
  • P3-5 straight abutments, 5 mm diameter, 7 9 mm length ●
  • P4 angled abutments, 15 - 25 degree angulation, standard platform, standard and long lengths
AI/ML Overview

This is a 510(k) summary for a dental implant system. The document focuses on demonstrating substantial equivalence to a predicate device rather than providing extensive performance data against specific acceptance criteria for a novel AI/ML device. Therefore, much of the requested information regarding AI/ML device performance studies is not available in this document.

Here's an analysis of the provided text based on your request, highlighting what is present and what is not:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria in the typical sense for an AI/ML device, as it describes a dental implant system. The performance data presented are primarily related to mechanical and biological safety, not diagnostic accuracy or effectiveness against quantifiable metrics for an AI system.

However, we can infer some "performance data" or testing performed:

Test PerformedReported Device Performance
Fatigue testing per ISO 14801:2007Performed on worst-case implant and abutment combination for standard platform implants and abutments. Engineering analysis to determine novel worst-cases. (Implication: Device withstands fatigue as per standard)
Cytotoxicity testing per ISO 10993-5Performed. (Implication: Device is biocompatible)
Biological risk assessmentPerformed. (Implication: Risks are assessed and managed)
Gamma sterilization validation per ISO 11137Validated. (Implication: Device is sterile)
Steam sterilization validation per ANSI/AAMI/ISO 17665-1:2006Validated. (Implication: Device is sterile)
Packaging validation and accelerated shelf-life testingPerformed. (Implication: Packaging maintains sterility, shelf life is established)
LAL testingPerformed. (Implication: Device is endotoxin-free)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. The performance data listed (fatigue, cytotoxicity, sterilization, etc.) are standard engineering and biological tests for medical devices, not clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and not provided. The "ground truth" for these tests relates to established standards and physical/chemical properties, not expert consensus on medical images or patient outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and not provided.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable and not provided. This is a conventional dental implant, not an AI-powered device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This information is not applicable and not provided.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the performance tests (fatigue, cytotoxicity, sterilization) would be the established international standards themselves (ISO 14801, ISO 10993-5, ISO 11137, ANSI/AAMI/ISO 17665-1) and their defined pass/fail criteria. For LAL testing, it would be the absence of endotoxins as per FDA guidance.

8. The sample size for the training set

This information is not applicable and not provided. This is not an AI/ML device.

9. How the ground truth for the training set was established

This information is not applicable and not provided. This is not an AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design. The eagle's head and neck are formed by the profiles, creating a visual representation of the department's focus on health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2017

A. B. Dental Device Ltd. % John Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 13th St. NW Washington, District of Columbia 20004

Re: K162482

Trade/Device Name: A.B. DENTAL DEVICES® Dental Implants System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: August 15, 2017 Received: August 15, 2017

Dear Dr. John Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Andrew I. Steen -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known)

K162482

Device Name

A.B. DENTAL DEVICES® Dental Implants System

Indications for Use (Describe)

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental lmplants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Two Stage Implants: 122, 15, 155, 110.

P4 and P14 angled abutments are to be used only with standard platform implants 3.5 mm in diameter or larger.

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

K162482 510(k) SUMMARY A.B. Dental Devices Dental Implant System

Submitter:

A.B. Dental Devices Ltd. 19 Havalomim St., Ashdod 7761117 lsrael

Phone: 972-8-8531388 Facsimile: 972-8-8522562 Contact Person: Gabi Krauss, RA Manager

Date Prepared: September 15, 2017

Name of Device: A.B. DENTAL DEVICES® Dental Implants System Common or Usual Name: Root-form Endosseous Dental Implants & Abutments Classification Name: Endosseous dental implant; 21 CFR §872.3640 Product Code: DZE, Additional Product Code NHA

Predicate Device

K132125 A.B. DENTAL DEVICES® Dental Implants System (primary) K051719 A.B. DENTAL DEVICES® Dental Implants System (reference) K112440 A.B. DENTAL DEVICES® Dental Implants System (reference) K071370 Nobel Biocare AB NobelActive (reference)

Indications for Use

A.B. DENTAL DEVICES® Dental Implants System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. A.B. DENTAL DEVICES® Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.

Two Stage Implants: 122, 15, 155, 110.

P4 and P14 angled abutments are to be used only with standard platform implants 3.5 mm in diameter or larger.

{4}------------------------------------------------

Device Description and Technological Characteristics

A.B. DENTAL DEVICES® Dental Implants System consists of one and two stage endosseous form dental implants, internal hexagonal and one piece implants systems. Narrow platform implants are to be used with straight abutments only. The following components are being added:

ImplantPlatformDiameter, mmLength, mm
I22Standard3.758, 10, 11.5, 13, 16
4.28, 10, 11.5, 13, 16
4.58, 10, 11.5, 13, 16
58, 10, 11.5, 13, 16
I5Narrow3.310, 11.5, 13, 16
I55Narrow3.310, 11.5, 13, 16
3.58, 10, 11.5, 13, 16
Standard3.758, 10, 11.5, 13, 16
4.28, 10, 11.5, 13, 16
4.58, 10, 11.5, 13, 16
58, 10, 11.5, 13, 16
68, 10, 11.5, 13, 16
I10Standard4.510, 11.5, 13, 16

Abutments:

  • P0-14 Healing caps, 2.5 7 mm heights .
  • P3-5 straight abutments, 5 mm diameter, 7 9 mm length ●
  • P4 angled abutments, 15 - 25 degree angulation, standard platform, standard and long lengths

Technological Characteristics Comparison

The purpose of this 510(k) is to expand the current product line to include two new endosseous implant product lines (122 and 155). Minor additions to existing implant lines have also been added. Comparison tables for each modification are provided below.

Subject I5Predicate I5Predicate I6bi
K-numberK162482K132125, K112440K132125
MaterialTi-6Al-4VTi-6Al-4VTi-6Al-4V
Diameter, mm3.33.5, 3.75, 4.2, 4.5, 5, 63
PlatformNarrowStandardNarrow
Length, mm10, 11.5, 13, 168, 10, 11.5, 13, 1610, 11.5, 13, 16
ProfileTaperedTaperedTapered
Surface TreatmentSand blasted with CaPSand blasted with CaPSand blasted with CaP
SterilitySterileSterileSterile
Subject I10Predicate I10
K-numberK162482K112440
MaterialTi-6Al-4VTi-6Al-4V
Diameter, mm4.53.75, 4.2, 5
PlatformStandardStandard
Length, mm10, 11.5, 13, 1610, 11.5, 13, 16
ProfileTaperedTapered

{5}------------------------------------------------

Surface TreatmentSand blasted with CaPSand blasted with CaP
SterilitySterileSterile
Subject 122Predicate 12
K-numberK162482K132125, K112440
MaterialTi-6Al-4VTi-6Al-4V
Diameter, mm3.75, 4.2, 4.5, 53.5, 3.75, 4.2, 4.5, 5, 6
PlatformStandardStandard
Length, mm8, 10, 11.5, 13, 168, 10, 11.5, 13, 16, 18, 20
ProfileTaperedTapered
Surface TreatmentSand blasted with CaPSand blasted with CaP
SterilitySterileSterile
Subject I55Predicate I5Predicate I6bi
K-numberK162482K132125, K112440K132125
MaterialTi-6Al-4VTi-6Al-4VTi-6Al-4V
Diameter, mm3.3,3.5, 3.75, 4.2, 4.5, 5, 63.5, 3.75, 4.2, 4.5, 5, 63
PlatformNarrow (≤3.5) and Standard (≥3.75)StandardNarrow
Length, mm8, 10, 11.5, 13, 168, 10, 11.5, 13, 1610, 11.5, 13, 16
ProfileTaperedTaperedTapered
Surface TreatmentSand blasted with CaPSand blasted with CaPSand blasted with CaP
SterilitySterileSterileSterile
Subject P0 Healing CapsPredicate P0 Caps
K-numberK162482K132125
MaterialTi-6Al-4VTi-6Al-4V
Connection MethodTaper fitScrew
PlatformP14 angled abutmentsNarrow, Standard
Length, mm2.5, 4, 5, 70.5, 2, 3, 4, 5, 6, 7
SterilityNon-SterileNon-Sterile
Subject P3-5Predicate P3
K-numberK162482K132125, K112440
MaterialTi-6Al-4VTi-6Al-4V
Diameter, mm53.75, 4.5, 5.5
PlatformStandardNarrow, Standard
Length, mm7,95, 7, 9, 11,12, 15
DesignAnti-rotationAnti-rotation
SterilityNon-SterileNon-Sterile
Subject P4Predicate P4
K-numberK162482K132125
MaterialTi-6Al-4VTi-6Al-4V
Diameter, mm3.75, 53.75
Angle, °15, 2515
PlatformStandardStandard

{6}------------------------------------------------

Length, mm8, 9, 15.515.7
DesignStraight, Standard, Narrow, LongLong, Anatomic (shoulder heights 1,3)
SterilityNon-SterileNon-Sterile

In addition, a comparison of the indications for use of the subject and primary predicate follows:

Subject A.B. DENTAL DEVICES®Dental Implants SystemPrimary Predicate A.B. DENTAL DEVICES®Dental Implants System
K-numberK162482K132125
Indications for UseA.B. DENTAL DEVICES® DentalImplants System is indicated foruse in surgical and restorativeapplications for placement in thebone of the upper or lower jaw toprovide support for prostheticdevices, such as artificial teeth, inorder to restore the patient'schewing function. A.B. DENTALDEVICES® Dental Implants Systemis indicated also for immediateloading when good primarystability is achieved and withappropriate occlusal loading.Two Stage Implants: I22, I5, I55,I10.P4 and P14 angled abutments are tobe used only with standard platformimplants 3.5 mm in diameter orlarger.A.B.DENTAL DEVICES® Dental ImplantsSystem is indicated for use in surgical andrestorative applications for placement inthe bone of the upper or lower jaw toprovide support for prosthetic devices,such as artificial teeth, in order to restorethe patient's chewing function. A.B.DENTAL DEVICES® Dental ImplantsSystem is indicated also for immediateloading when good primary stability isachieved and with appropriate occlusalloading.Two Stage Implants: I2, I5, I6BI.One Stage: I6, I6b, I6B.One Stage & One-Piece 3.0 mm diameterimplants: I6, I6B, I6BI, are intended forplacement at the mandibular central andlateral incisors and maxillary and lateralincisors. Indicated also for denturestabilization using multiple implants.One stage & One-Piece 2.4 mm diameterimplants for temporary use or long termuse: I6, I6b, permit immediate splintstability and long term fixation of new orexisting crown, bridge and prosthesis.P14 Angulated Abutment Adapter is to beused with implant diameter 4.2mm andhigher.

With regard to differences in the indications for use between the subject and primary predicate:

  • . The list of two stage implants was updated to include the new endosseous implant product lines.
  • Limitations for the 2.4 mm and 3.0 diameter implant bodies are not included because the subject ● devices do not contain any of that size.
  • One stage indications were removed because the subject devices are only indicated for two stage. ●
  • The indication for the P14 abutment was reduced to the standard platform implant diameter of 3.5 mm . or larger based on the fatigue testing provided.

Performance Data

Fatigue testing per ISO 14801:2007 of the worst-case implant and abutment combination for the standard platform implants and abutments Engineering analysis to determine if additional components constitute a new worst-case.

{7}------------------------------------------------

Cytotoxicity testing per ISO 10993-5 and biological risk assessment Gamma sterilization validation per ISO 11137 Steam sterilization validation per ANSI/AAMI/ISO 17665-1:2006 Packaging validation and accelerated shelf-life testing LAL testing in accordance with the FDA Guidance on Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile

Substantial Equivalence

The AB Dental Implant System has the same indications and similar technological characteristics and principles of operation as its predicate device. The minor technological differences between the AB Dental Implant Systems do not raise different issues. Performance data demonstrate that the AB Dental Implant System is substantially equivalent.

Conclusions

The AB Dental Implant System is substantially equivalent to the predicate device.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.