K Number
K153696

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-04-14

(113 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chattanooga Revolution Wireless is a clinical electrotherapy device intended for use under the supervision of a Healthcare Professional.

Indications for Use:

As an NMES device, indications are for the following conditions:

  • Retarding or preventing disuse atrophy
  • Maintaining or increasing range of motion
  • Re-educating muscles
  • Relaxation of muscle spasms
  • Increasing local blood circulation
  • Prevention of venous thrombosis of the calf muscles immediately after surgery

As a TENS device, indications are for the following conditions:

  • Symptomatic relief and management of chronic, intractable pain
  • Post-surgical and post-trauma acute pain

As a pulsed mode device, indications are for the following conditions:

  • Relaxation of muscle spasm
  • Increasing local blood circulation
  • Retardation or prevention of disuse atrophy
  • Maintenance or increase of range of motion
Device Description

The Chattanooga Revolution Wireless is a neuromuscular electrical stimulation (NMES) device, which stimulates nerve fibers by means of electrical impulses transmitted by electrodes. The electrical pulses generated by the Chattanooga "Revolution Wireless stimulate motor nerves to stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, and total session duration), different types of muscle work can be imposed on the stimulated muscles.

The device system is made up of a remote control, 4 stimulation modules, and electrodes, which are stored and shipped within the docking station which is used to recharge the remote and the modules. The docking station is powered by an AC-DC adapter.

The remote control is the interface between the stimulation modules and the user. It sends and receives information to and from the modules via a wireless network. The remote control allows the user to navigate through the user interface (UI), select stimulation program or objectives, set desired options and control the four (4) module intensities independently. The remote control is powered by a rechargeable battery.

The Chattanooga® Revolution Wireless stimulation module set is composed of 4 independent stimulation modules that are controlled via the remote control by a wireless connection. Each module is composed of two "pods" (1 battery "pod" and one stimulation "pod") linked by an electrical connection (cable). Two proprietary standard snap gel electrodes are also needed to connect each "pod" to the body. The modules are powered by a Lithium Polymer (LiPo) rechargeable 3.7[V] / ≥ 450 [mAh] battery.

The docking station is the part on which all the other components (modules, remote and accessories) are stored when the Chattanooga Revolution Wireless system is not in use. The docking is continuously connected to an AC wall socket and it allows charging simultaneously the 4 stimulation modules and the remote control. The docking main part is intended to be fixed and not for transportation purposes, but the tablet may be removed and used as a portable charging station for the remote and modules. The docking station includes a removable tray where the user can store electrodes, quick start guide, the remote and modules for transportation.

The wireless protocol of the Chattanooga® Revolution Wireless Device is a proprietary design of a Radio-Frequency protocol operating the 2.4 GHz ISM band. It is used to

  1. Send particular information from remote control to stimulation modules (stimulation settings)

  2. Send particular information from stimulation modules to remote control, like current stimulation level and stimulation module subsystem status,

  3. Transfer binary data to stimulation modules, and

  4. Allow synchronization from stimulation modules to remote control clocks.

AI/ML Overview

The provided documentation describes the Chattanooga Revolution Wireless, a powered muscle stimulator and transcutaneous electrical nerve stimulator (TENS) device. The information details the device's characteristics, indications for use, comparison to a predicate device (Vectra Neo), and performance testing conducted for its 510(k) premarket notification.

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated in a quantitative table format with pass/fail values. Instead, the document discusses compliance with recognized standards and successful completion of various tests. The reported device performance is presented as a confirmation of compliance.

Acceptance Criteria CategoryReported Device Performance
Electrical SafetyThe device was tested and found to comply with AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012, a recognized standard for medical electrical equipment.
Electromagnetic Compatibility (EMC)The device was tested and found to comply with IEC 60601-1-2 Edition 3:2007-03, a recognized standard for electromagnetic compatibility in medical electrical equipment.
Nerve and Muscle Stimulator Specific SafetyThe device was tested and found to comply with IEC 60601-2-10 Edition 2.0 2012-06, a recognized standard for the basic safety and essential performance of nerve and muscle stimulators.
FCC Radio Frequency ComplianceThe device was tested to FCC requirements and found to comply with the requirements of 47 CFR 15.249.
Software VerificationThe device's software was verified in accordance with FDA's guidance document: General Principles of Software Validation, January 11, 2002. The testing demonstrated that the software meets its design requirements. Also complies with IEC 62304 First Edition 2006-05 (Medical Device Software - Software Life Cycle Processes).
Usability/Human FactorsUsability/Human Factors testing was performed, demonstrating that established requirements for usability were met, and the design is appropriate for intended users and use environment. The study "substantiates the acceptability of the use-related risks." Complies with IEC 60601-1-6 Edition 3.0 2010-01 (Usability) and IEC 62366:2007, Ed. 1 (Application of Usability Engineering).
Wireless CoexistenceThe performance was evaluated in an environment with other Chattanooga Revolution Wireless devices and other 2.4 GHz wireless devices (Bluetooth and Wi-Fi). The device met all specified requirements.
Risk ManagementComplies with ISO 14971:2007 (Medical Devices - Application of Risk Management).
Electrode Lead Wires & Patient CablesComplies with 21 CFR 898 (Performance standard for electrode lead wires and patient cables).

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample sizes used for any of the performance tests (e.g., electrical safety, software verification, usability). There is also no mention of data provenance (e.g., country of origin, retrospective/prospective). The tests described are primarily engineering verification and validation studies rather than clinical trials with human participant data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Given that the performance testing primarily involves engineering compliance with standards and functional verification, the concept of "ground truth" established by experts in the clinical sense (e.g., radiologists interpreting images) is not applicable here. The "ground truth" for these tests would be the specifications outlined in the recognized standards themselves, and the expertise would lie with the engineers and testers performing the measurements and assessments against those standards. No specific number or qualifications of such experts are provided beyond implied expertise in regulatory and engineering compliance.

4. Adjudication Method for the Test Set

As the studies described are compliance-based engineering tests and not clinical studies requiring expert consensus on complex outcomes, an explicit adjudication method like "2+1" or "3+1" is not mentioned or relevant. Compliance is typically determined by whether the device's performance meets the defined parameters of the relevant standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study is mentioned. The device is a physical therapy device (stimulator), and the evaluation focuses on its safety and performance against established standards and a predicate device, not on assessing the improvement of human readers (e.g., doctors interpreting scans) with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

The device itself is a physical medical device, not an AI algorithm. Therefore, a "standalone algorithm only" performance study is not applicable. The performance studies are for the integrated device system.

7. Type of Ground Truth Used

The "ground truth" for the performance testing is implicitly defined by:

  • Recognized Standards: Such as AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-6, IEC 62366, IEC 62304, ISO 14971.
  • Design Requirements: For the software verification and usability testing, the device's own design requirements serve as the ground truth against which performance is measured.
  • FCC Regulations: For radio frequency compliance.

There is no mention of pathology, clinical outcomes data, or expert consensus (in a diagnostic context) as ground truth.

8. Sample Size for the Training Set

This information is not provided. The document describes premarket notification for a physical medical device, not an AI/ML product that typically involves a "training set." The software verification mentioned refers to traditional software engineering verification, not machine learning model training.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a "training set" in the context of an AI/ML model, this question is not applicable to the provided document. The ground truth for the device's design and functionality is established through engineering specifications, regulatory standards, and intended use definitions.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 14, 2016

DJO, LLC Lorri Trotter Regulatory Affairs Manager 1430 Decision Street Vista. California 92081

Re: K153696

Trade/Device Name: Chattanooga Revolution Wireless Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF; GZJ Dated: March 11, 2016 Received: March 14, 2016

Dear Ms. Trotter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael J.Hoffmann -A

  • for Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153696

Device Name Chattanooga Revolution Wireless

Indications for Use (Describe)

The Chattanooga Revolution Wireless is a clinical electrotherapy device intended for use under the supervision of a Healthcare Professional.

Indications for Use:

As an NMES device, indications are for the following conditions:

  • Retarding or preventing disuse atrophy
  • Maintaining or increasing range of motion
  • Re-educating muscles
  • Relaxation of muscle spasms
  • Increasing local blood circulation
  • Prevention of venous thrombosis of the calf muscles immediately after surgery

As a TENS device, indications are for the following conditions:

  • Symptomatic relief and management of chronic, intractable pain
  • Post-surgical and post-trauma acute pain

As a pulsed mode device, indications are for the following conditions:

  • Relaxation of muscle spasm
  • Increasing local blood circulation
  • Retardation or prevention of disuse atrophy
  • Maintenance or increase of range of motion
Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitted by:DJO, LLC
1430 Decision Street
Vista, CA 92081
Contact Person:Lorri Trotter
Regulatory Affairs Manager
(760)734-3049
Date Summary Prepared:April 11, 2016
Trade Name:Chattanooga® Revolution Wireless
Classification Name:Powered muscle stimulator (21 CFR 890.5850);Transcutaneous electrical nerve stimulator for pain relief(21 CFR 5890)
Product Code:IPF, GZJ
Regulatory Class:Class II
Predicate Device:Vectra Neo (K132284)

Device Description:

The Chattanooga Revolution Wireless is a neuromuscular electrical stimulation (NMES) device, which stimulates nerve fibers by means of electrical impulses transmitted by electrodes. The electrical pulses generated by the Chattanooga "Revolution Wireless stimulate motor nerves to stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, and total session duration), different types of muscle work can be imposed on the stimulated muscles.

The device system is made up of a remote control, 4 stimulation modules, and electrodes, which are stored and shipped within the docking station which is used to recharge the remote and the modules. The docking station is powered by an AC-DC adapter.

The remote control is the interface between the stimulation modules and the user. It sends and receives information to and from the modules via a wireless network. The remote control allows the user to navigate through the user interface (UI), select stimulation program or objectives, set desired options and control the four (4) module intensities independently. The remote control is powered by a rechargeable battery.

The Chattanooga® Revolution Wireless stimulation module set is composed of 4 independent stimulation modules that are controlled via the remote control by a wireless connection. Each module is composed of two "pods" (1 battery "pod" and one stimulation "pod") linked by an

{4}------------------------------------------------

electrical connection (cable). Two proprietary standard snap gel electrodes are also needed to connect each "pod" to the body. The modules are powered by a Lithium Polymer (LiPo) rechargeable 3.7[V] / ≥ 450 [mAh] battery.

The docking station is the part on which all the other components (modules, remote and accessories) are stored when the Chattanooga Revolution Wireless system is not in use. The docking is continuously connected to an AC wall socket and it allows charging simultaneously the 4 stimulation modules and the remote control. The docking main part is intended to be fixed and not for transportation purposes, but the tablet may be removed and used as a portable charging station for the remote and modules. The docking station includes a removable tray where the user can store electrodes, quick start guide, the remote and modules for transportation.

The wireless protocol of the Chattanooga® Revolution Wireless Device is a proprietary design of a Radio-Frequency protocol operating the 2.4 GHz ISM band. It is used to

  1. Send particular information from remote control to stimulation modules (stimulation settings)

  2. Send particular information from stimulation modules to remote control, like current stimulation level and stimulation module subsystem status,

  3. Transfer binary data to stimulation modules, and

  4. Allow synchronization from stimulation modules to remote control clocks.

Indications for Use:

The Chattanooga Revolution Wireless is a clinical electrotherapy device intended for use under the supervision of a Healthcare Professional.

Indications for Use:

As an NMES device, indications are for the following conditions:

  • Retarding or preventing disuse atrophy -
  • -Maintaining or increasing range of motion
  • Re-educating muscles -
  • Relaxation of muscle spasms -
  • -Increasing local blood circulation
  • Prevention of venous thrombosis of the calf muscles immediately after surgery -

As a TENS device, indications are for the following conditions:

  • Symptomatic relief and management of chronic, intractable pain -
  • Post-surgical and post-trauma acute pain -
  • As a pulsed mode device, indications are for the following conditions:
    • Relaxation of muscle spasm -
    • Increasing local blood circulation -
    • -Retardation or prevention of disuse atrophy
    • Maintenance or increase of range of motion -

Programs:

{5}------------------------------------------------

The Chattanooga Revolution Wireless provides 14 programs: Trigger Point, Acute Pain, Chronic Pain, Muscle Pump Cycle, Muscle Pump Continuous, Increase ROM (Range of Motion), Decrease Muscle Tone, Fast Twitch Function, Muscle Atrophy, Slow Twitch Function, VMS-FR Dynamic 2 Channel, VMS-FR Dynamic 4 Channel, VMS-FR Static 2 Channel, and VMS-FR Static 4 Channel.

Comparison to the Predicate Device: Includes Table

The indications for use for the Revolution Wireless USA are a subset of those for the predicate device the Vectra Neo Clinical Therapy System. The predicate device has additional features, functions, and applications that are not included in the Revolution Wireless. Therefore, we are comparing the Revolution Wireless only to those functions of the predicate that are applicable.

The technological characteristics of the two devices are very similar for the comparable functions, but there are a few differences. The following table summarizes the similarities and differences between the technological characteristics of the two devices. The power sources of the two devices differ (Revolution is battery powered) and output specifications differ slightly between the two devices, but the Chattanooga Revolution Wireless conforms to required standards, and small differences in output do not raise any issues of safety or effectiveness. The Revolution Wireless differs from the predicate in that it communicates wirelessly with the stimulation modules, but the design of the wireless feature is identical to that of the Compex Wireless USA (K143551).

Basic Device Characteristics – Comparison with Predicate Device
CharacteristicNew DevicePredicate DeviceSimilar/ Different
510(K) NumberDevice Name, ModelChattanoogaWireless PROK132284Vectra Neo ClinicalTherapy SystemDifferent
ManufacturerDJO, LLCDJO, LLCSimilar
Prescription/OTCPrescriptionPrescriptionSimilar
Where usedPhysician office,physical therapyClinic, Hospital,Nursing Home, PostAcute Care,Chiropractic ClinicPhysician office,physical therapyClinic, Hospital,Nursing Home, PostAcute Care,Chiropractic ClinicSimilar
Target populationAdult PopulationAdult PopulationSimilar
Anatomical libraryYesYesSimilar
Audio IndicatorYesYesSimilar
Connection of device to electrodesStimulation Moduleis directly connectedto the custom maleSNAP assembled inthe electrode. UserInterface (LCD andbuttons) is physicallyseparated (RemoteWith cablesincluding pins toconnect to electrodespins. There is 1 cableper channel with amaximum of 4channelsDifferent
CharacteristicNew DevicePredicate DeviceSimilar/Different
Control) andcommunicateswirelessly with up tofour (4) stimulationmodules.Stimulation safetyremains fullymanaged byStimulation Moduleelectronic circuititself.
Power Source (s)Rechargeable battery100-240V~, 2.6-1.0A, 47-63HzDifferent
Method of Line Current IsolationNABattery operateddeviceIndependenttransformer isolatedDifferent
Electrical TypeNABattery operateddeviceType BFDifferent
Patient Leakage Current - NormalCondition (μΑ)NABattery operateddevice<100uA patientleakageDifferent
Patient Leakage Current - SingleFault Condition (μΑ)NABattery operateddevice<300uA line leakageDifferent
Number of Output ModesMuscle stimulator:ElectrodesMuscle stimulator:ElectrodesUltrasound:applicatorsEMG:electrodesLaser: applicatorDifferent
Number of Output Channels40, 2, or 4Similar
Synchronous or AlternatingSee OutputSpecifications BelowSee OutputSpecifications Below
Method of Channel IsolationEach channel is themiddle of a H Bridge.Except when it isactivated, eachchannel is always inhigh impedance stateThroughtransformers andopto-couplersDifferent
Regulated Current or RegulatedVoltage (output signals only)Regulated current onall channelsConfigurable foreither constantvoltage or constantcurrent, see OutputSpecifications below.Similar
Software/Firmware/MicroprocessorControlYesYesSimilar
CharacteristicNew DevicePredicate DeviceSimilar/Different
Automatic No-Load TripYesYesSimilar
Automatic Shut OffYes, On/off switchYesSimilar
Patient Override ControlYesYesSimilar
Indicator DisplaySimilar
- On/Off StatusYesYes
- Low BatteryYesNo battery
- Voltage/Current level?Yes, unit [miliAmps]Yes, unit [mA]
Timer Range (Minutes)Yes, unit [minutes]Max 30 [minutes]0-60 [minutes]Different
Compliance with 21 CFR 898?Yes21CFR 898-Performance standardfor electrode leadwires and patientcables.21CFR 898-Performancestandard forelectrode lead wiresand patient cables.Similar
WeightRemote: 0.24 lbswith batteryDevice: 20.7 lbsDifferent
Stimulation Module:0.11 lbs per moduleModule: 1 lbsDifferent
Docking Station: 3lbsDevice + CartAccessory: 48.9 lbsDifferent
Dimension (in.) [W x H x D]Remote: 11 x 7 x 2cmDevice: [15.89 x20.05 x 15.89]Different
Stimulation Module:2x(6 x 2) cmModule: [11.12 x1.43 x 6.34]Different
Docking Stationtablet: 14x27x3 [cm]Device + Cart:[23.94 x 52.85 x26.19]Different
OUTPUT SPECIFICATIONS
VMS Symmetrical Biphasic Waveform
ShapeRectangularRectangularSimilar
Maximum Output Voltage (± 10%)60 V @ 500 Ω180V @ 2 kΩ180 V @ 10 kΩ57 V @ 500 Ω200 V @ 2 kΩ100V @ 10 kΩ(open lead detectedabove 10[mA])Different
Maximum Output Current (± 10%)120 mA @ 500 Ω90 mA @ 2 kΩ18 mA @ 10 kΩ114 mA @ 500 Ω100 mA @ 2 kΩ10 mA @ 10 kΩ(open lead detectedabove 10[mA])Different
Pulse Width300 to 400 [μs](microseconds)20 to 400 [μs](microseconds)Different
Frequency1 to 100 Hz1 - 200 [Hz]Different
Net Charge [μC/pulse]0 [μC] @ 500ΩExcitation pulse fullycompensated0 [μC] @ 500ΩSimilar
CharacteristicNew DevicePredicate DeviceSimilar/Different
Maximum Phase Charge [µC]48 [µC] @ 500Ω45.6 [µC] @ 500ΩDifferent
Maximum Current (RMS) Density(mA/cm2)2.1mA/cm22.4 mA/cm2Different
Maximum Power Density[mW/cm²]36 [mW/cm²] @ 500Ω55 [mW/cm²] @ 500ΩDifferent
ON Time (seconds)1 - 10 [sec]1 - 60 [sec]Different
OFF Time (seconds)3 - 50 [sec]1 - 60 [sec]Different
Additional Features (if applicable)100 [µs](microseconds)100 [µs](microseconds)Similar
VMS-FR Waveform
ShapeRectangularRectangularSimilar
Maximum Output Voltage (± 10%)60 V @ 500 Ω180V @ 2 kΩ180 V @ 10 kΩ57 V @ 500 Ω200 V @ 2 kΩ100V @ 10 kΩ(open lead detected above 10[mA])Different
Maximum Output Current (± 10%)120 mA @ 500 Ω90 mA @ 2 kΩ18 mA @ 10 kΩ114 mA @ 500 Ω100 mA @ 2 kΩ10 mA @ 10 kΩ(open lead detected above 10[mA])Different
Pulse Width200 to 300 [µs](microseconds)20 to 400 [µs](microseconds)Different
Frequency35 - 80 Hz20 - 80 [Hz]Different
Net Charge [µC/pulse]0 [µC] @ 500ΩExcitation pulse fullycompensated0 [µC] @ 500ΩSimilar
Maximum Phase Charge [µC]36 [µC] @ 500Ω45.6 [µC] @ 500ΩDifferent
Maximum Current (RMS) Density(mA/cm2)1.34mA/cm21.52 mA/cm2Different
Maximum Power Density[mW/cm²]14.4 [mW/cm²] @ 500Ω22 [mW/cm²] @ 500ΩDifferent
Burst Mode (i.e. pulse trains)N/A, no burst modea) Pulses per burstb) Bursts per secondc) Burst duration (seconds)d) Duty Cycle[Line (b) x Line (c)]VMS burstDifferent
ON Time (seconds)1 - 10 [sec]1 - 60 [sec]Different
OFF Time (seconds)3 - 50 [sec]1 - 60 [sec]Different
Additional Features (if applicable)100 [µs](microseconds)100 [µs](microseconds)Similar
Interphase interval(microseconds)(microseconds)

Rasic Device Characteristics - Comparison with Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Performance Testing:

Electrical Safety and Electromagnetic Compatibility: The Chattanooga® Revolution Wireless was tested and found to comply with recognized standards for electrical safety and electromagnetic compatibility.

FCC Radio Frequency Testing: The Chattanooga® Revolution Wireless was tested to FCC requirements and found to comply with the requirements of 47 CFR 15.249.

Software Verification: The device's software was verified in accordance with the requirements of FDA's guidance document: General Principles of Software Validation, January 11, 2002. The software testing demonstrated that the software meets its design requirements.

Usability/Human Factors Testing: Usability/Human Factors testing was performed, which demonstrated that the established requirements for usability were met, and the device's design is appropriate for the intended users and use environment. The result of this study substantiates the acceptability of the use-related risks identified during the risk assessment activities.

Wireless Coexistence Testing: The performance of Chattanooga® Revolution Wireless was evaluated in an environment with other Chattanooga® Revolution Wireless device and with other types of 2.4 GHz wireless devices (Bluetooth and Wi-Fi). The device met all specified requirements.

Standards:

RecognitionNo.StandardDescription
19-4AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 AndA2:2010/(R)2012 (ConsolidatedText)Medical Electrical Equipment - Part1: General Requirements for BasicSafety and Essential Performance(IEC 60601-1, Mod.)
19-1IEC 60601-1-2 Edition 3:2007-03Medical Electrical Equipment - Part1-2: General Requirements for BasicSafety and Essential Performance –Collateral Standard: ElectromagneticCompatibility – Requirements andTests
17-11IEC 60601-2-10 Edition 2.0 2012-06Medical Electrical Equipment -- Part2-10: Particular Requirements forthe Basic Safety and EssentialPerformance of Nerve and MuscleStimulators
5-85IEC 60601-1-6 Edition 3.0 2010-01Medical Electrical Equipment - Part

The Chattanooga Revolution Wireless conforms to the following standards.

{10}------------------------------------------------

1-6: General Requirements for BasicSafety and Essential Performance –Collateral Standard: Usability
N/A*IEC 62366:2007, Ed. 1* Ed. 1 was used because it is citedin IEC 60601-1-6 Ed. 3.0.Medical Devices - Application ofUsability Engineering to MedicalDevices
13-8IEC 62304 First Edition 2006-05Medical Device Software - SoftwareLife Cycle Processes
5-40ISO 14971:2007Medical Devices - Application ofRisk Management to MedicalDevices

Conclusion:

Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the Chattanooga Revolution Wireless is as safe and effective as, and substantially equivalent to, the predicate device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).