K Number
K143551

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-04-28

(134 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Compex Wireless USA is an Over-The-Counter device intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is to be used by adults only.

The Compex Wireless USA is not intended for adjunctive therapy in the treatment of medical diseases and conditions of any kind. None of the Compex Wireless USA stimulation programs are designed for injured or disease afflicted muscles. Its use on such muscles is contraindicated. The work imposed on the muscles by the Compex Wireless USA programs is definitely not suitable for rehabilitation and physiotherapy. The Compex Wireless USA electrical impulses allow the triggering of action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

The Compex Wireless USA may therefore be considered a technique of muscle training.

Device Description

The Compex Wireless USA is a neuromuscular electrical stimulation (NMES) device, which stimulates nerve fibers by means of electrical impulses transmitted by electrodes. The electrical pulses generated by the Compex Wireless USA stimulate motor nerves to stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, and total session duration), different types of muscle work can be imposed on the stimulated muscles. The Compex Wireless USA may therefore be considered a technique of muscle training.

The device system is made up of a remote control, 4 stimulation modules, and electrodes, which are stored and shipped within the docking station which is used to recharge the remote and the modules. The docking station is powered by an AC-DC adapter.

The remote control is the interface between the stimulation modules and the user. It sends and receives information to and from the modules via a wireless network. The remote control allows the user to navigate through the user interface (UI), select stimulation program or objectives, set desired options and control the four (4) module intensities independently. The remote control is powered by a rechargeable battery.

The Compex Wireless USA stimulation module set is composed of 4 independent stimulation modules that are controlled via the remote control by a wireless connection. Each module is composed of two "pods" (1 battery "pod" and one stimulation "pod") linked by an electrical connection (cable). Two proprietary Compex standard snap gel electrodes are also needed to connect each "pod" to the body. The modules are powered by a Lithium Polymer (LiPo) rechargeable 3.7[V] /> 450 [mAh] battery.

The wireless protocol of the Compex Wireless USA Device is a proprietary design of a Radio-Frequency protocol operating the 2.4 GHz ISM band. It is used to 1) Send particular information from remote control to stimulation modules (stimulation settings) 2) Send particular information from stimulation modules to remote control, like current stimulation level and stimulation module subsystem status, 3) Transfer binary data to stimulation modules, and 4) Allow synchronization from stimulation modules to remote control clocks.

AI/ML Overview

The provided text describes the Compex Wireless USA device, its indications for use, and a comparison to a predicate device (Compex Sport Elite) for 510(k) clearance. However, it does not contain the acceptance criteria or results of a study that proves the device meets specific performance criteria in terms of accuracy, sensitivity, specificity, or other clinical efficacy metrics as would be expected for an AI/ML or diagnostic device.

The document focuses on demonstrating substantial equivalence to a predicate device through:

  • Identical Indications for Use.
  • Similar technological characteristics in terms of basic device attributes and output specifications.
  • Performance testing related to electrical safety, electromagnetic compatibility, FCC radio frequency, software verification, usability/human factors, and wireless coexistence, all against applicable standards.

Therefore, for the information requested in your prompt, I can only provide what is available in the text, which mainly pertains to regulatory compliance and safety/performance against engineering standards, not clinical performance or artificial intelligence specific metrics.

Here's an analysis of the provided text in response to your questions:

1. A table of acceptance criteria and the reported device performance

The document does not present clinical acceptance criteria (e.g., accuracy, sensitivity, specificity) for a specific diagnostic task or AI performance. Instead, the "acceptance criteria" are implied by compliance with various regulatory and engineering standards, and performance testing confirming functionality and safety.

Acceptance Criteria (Implied by Standards & Testing)Reported Device Performance (Summary from text)
Electrical Safety & Essential PerformanceComplied with AAMI/ANSI ES 60601-1:2005(R)2012 and C1:2009/(R)2012 and A2:2010/(R)2012 for medical electrical equipment.
Electromagnetic Compatibility (EMC)Complied with IEC 60601-1-2 Ed. 3:2007-03.
FCC Radio Frequency RequirementsComplied with 47 CFR 15.249.
Software VerificationSoftware was verified in accordance with FDA's guidance document "General Principles of Software Validation, January 11, 2002." Demonstrated that the software meets its design requirements. Compliance with IEC 62304.
Usability/Human FactorsUsability/Human Factors testing was performed, demonstrating that established requirements for usability were met and the design is appropriate for intended users and use environment. Results substantiate acceptability of use-related risks. Compliance with IEC 60601-1-6 and IEC 62366 and AAMI/ANSI HE75:2009.
Wireless CoexistenceEvaluated in environments with other Compex Wireless USA devices and other 2.4 GHz wireless devices (Bluetooth and Wi-Fi). Device met all specified requirements.
Risk ManagementCompliance with ISO 14971:2007.
Muscle Stimulation PerformanceThe device's electrical impulses allow triggering of action potentials on motoneurones of motor nerves, leading to mechanical muscle fiber responses. Parameters allow different types of muscle work. (This is a description of how NMES works and the intended physiological effect, not a specific performance metric against a benchmark). The 9 NMES programs are identical to the predicate device. Minor differences in maximum output voltage/current, pulse width, and phase duration are noted but not presented as failing or meeting a specific performance threshold.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document does not describe a clinical test set or data provenance in the context of typical AI/ML or diagnostic device studies (i.e., patient data, images, etc.). The mention of "Performance Testing" refers to engineering and software validation, not clinical validation on a patient cohort.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

This is not applicable as the document does not describe a clinical study involving experts establishing ground truth for a test set.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

This is not applicable as there is no described clinical test set or adjudication process.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A multi-reader, multi-case (MRMC) comparative effectiveness study was not done. This device is a muscle stimulator, not a diagnostic imaging device with human readers.

6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done

This is not applicable in the sense of an AI algorithm making a diagnosis or prediction. The device itself is "standalone" in its operation of muscle stimulation, controlled by the user.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the engineering and software validation tests performed, the "ground truth" would be the expected behavior and specifications defined by the design requirements and industry standards. For example, for "Software Verification," the ground truth is that the software performs its functions as designed. For "Electrical Safety," the ground truth is that the device passes specific tests outlined in the relevant IEC/AAMI standards. These are not clinical ground truths like pathology or outcomes data.

8. The sample size for the training set

This is not applicable. The device is not an AI/ML algorithm that requires a "training set" in the common sense of machine learning.

9. How the ground truth for the training set was established

This is not applicable.

In summary: The provided FDA 510(k) clearance document for the Compex Wireless USA is for a traditional medical device (a powered muscle stimulator). It demonstrates substantial equivalence to a predicate device and compliance with safety and performance standards through engineering tests. It does not present clinical performance data, AI/ML algorithm validation, or studies involving human readers, test sets, or ground truth establishment in the context of diagnostic or AI-powered devices.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 28, 2015

DJO, LLC Gina Flores Regulatory Specialist 1430 Decision Street Vista, CA 92081

Re: K143551

Trade/Device Name: Compex Wireless USA Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: March 25, 2015 Received: March 27, 2015

Dear Ms. Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143551

Device Name Compex Wireless USA

Indications for Use (Describe)

The Compex Wireless USA is an Over-The-Counter device intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is to be used by adults only.

The Compex Wireless USA is not intended for adjunctive therapy in the treatment of medical diseases and conditions of any kind. None of the Compex Wireless USA stimulation programs are designed for injured or disease afflicted muscles. Its use on such muscles is contraindicated. The work imposed on the muscles by the Compex Wireless USA programs is definitely not suitable for rehabilitation and physiotherapy. The Compex Wireless USA electrical impulses allow the triggering of action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.

The Compex Wireless USA may therefore be considered a technique of muscle training.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5. 510(k) Summary

This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92. The assigned 510(k) number is K143551.

Submitted by:DJO, LLC1430 Decision StreetVista, CA 92081
Contact Person:Gina FloresRegulatory Specialist760-734-3161
Date Summary Prepared:April 16, 2015
Trade Name:Compex® Wireless USA
Classification Name:Powered muscle stimulator (21 CFR 890.5850)
Product Code:NGX, Powered muscle stimulator, for muscleconditioning.
Regulatory Class:Class II
Predicate Device:Compex Sport Elite (K083140)

Device Description:

The Compex Wireless USA is a neuromuscular electrical stimulation (NMES) device, which stimulates nerve fibers by means of electrical impulses transmitted by electrodes. The electrical pulses generated by the Compex Wireless USA stimulate motor nerves to stimulate a muscular response. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, and total session duration), different types of muscle work can be imposed on the stimulated muscles. The Compex Wireless USA may therefore be considered a technique of muscle training.

The device system is made up of a remote control, 4 stimulation modules, and electrodes, which are stored and shipped within the docking station which is used to recharge the remote and the modules. The docking station is powered by an AC-DC adapter.

The remote control is the interface between the stimulation modules and the user. It sends and receives information to and from the modules via a wireless network. The remote control allows the user to navigate through the user interface (UI), select stimulation program or objectives, set desired options and control the four (4) module intensities independently. The remote control is powered by a rechargeable battery.

{4}------------------------------------------------

The Compex Wireless USA stimulation module set is composed of 4 independent stimulation modules that are controlled via the remote control by a wireless connection. Each module is composed of two "pods" (1 battery "pod" and one stimulation "pod") linked by an electrical connection (cable). Two proprietary Compex standard snap gel electrodes are also needed to connect each "pod" to the body. The modules are powered by a Lithium Polymer (LiPo) rechargeable 3.7[V] /> 450 [mAh] battery.

The wireless protocol of the Compex Wireless USA Device is a proprietary design of a Radio-Frequency protocol operating the 2.4 GHz ISM band. It is used to 1) Send particular information from remote control to stimulation modules (stimulation settings) 2) Send particular information from stimulation modules to remote control, like current stimulation level and stimulation module subsystem status, 3) Transfer binary data to stimulation modules, and 4) Allow synchronization from stimulation modules to remote control clocks.

Intended Use:

Indications for Use:

The Compex Wireless USA is an Over-The-Counter device intended to stimulate healthy muscles in order to improve or facilitate muscle performance. It is to be used by adults only.

The Compex Wireless USA is not intended for adjunctive therapy in the treatment of medical diseases and conditions of any kind. None of the Compex Wireless USA stimulation programs are designed for injured or disease afflicted muscles. Its use on such muscles is contraindicated. The work imposed on the muscles by the Compex Wireless USA programs is definitely not suitable for rehabilitation and physiotherapy. The Compex Wireless USA electrical impulses allow the triggering of action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The Compex Wireless USA may therefore be considered a technique of muscle training.

Programs:

Muscle Stimulation Training Programs: The Compex Wireless USA provides four muscle stimulation training programs: Endurance, Strength, and Explosive Strength. They correspond to the type of muscle performance the athlete wishes to improve or maintain. Each of the four training programs offers five different working levels that enable the amount of work to be gradually increased.

Specialized Muscle Training Programs: The Compex Wireless USA also offers five special muscle training programs: Potentiation, Training Recovery, Competition Recovery, Muscle Relaxation, and Pre-Warmup. Their objective is to prepare muscles

{5}------------------------------------------------

for explosive motions or to facilitate recovery after active muscle training and competition.

Comparison to the Predicate Device:

The indications for use for the Compex Wireless USA are identical to those of the predicate device, the Compex Sport Elite. In addition, the 9 NMES programs are identical in both devices.

The technological characteristics of the two devices are very similar, but there are a few minor differences. The following tables summarize the similarities and differences between the technological characteristics of the two devices.

CharacteristicNew DevicePredicate DeviceSimilar/Different
510(k) NumberK143551K083140Similar
Device Name,ModelCompex® Wireless USACompex Sport Plus(currently marketed asCompex Sport Elite)Similar
ManufacturerDJO LLCDJO LLC510(k) sponsor:Chattanooga GroupSimilar
Connection ofdevice to electrodesStimulation Module isdirectly connected to thecustom Compex femaleSNAP assembled in theelectrode. User Interface(LCD and buttons) isphysically separated (RemoteControl) and communicateswirelessly with up to four (4)Stimulation Modules.Stimulation safety remainsfully managed byStimulation Moduleelectronic circuit itself.With 6-pole cablesincluding female customSNAP plugged on thecustom Compex femaleSNAP assembled in theelectrode. Entireelectronic circuit forfour (4) StimulationChannels and UserInterface is combinedinto same casing,connected to theelectrodes with 6-polecables.Different
Power Source(s)Remote: : Lithium Polymer(LiPo)rechargeable 3.7[V] /≥1500[mAh]Stimulation Modules:Lithium Polymer (LiPo)rechargeable 3.7[V] /≥450[mAh]Rechargeable Ni-MhBattery 4.6V (4 cellsAA=R6)Different
Method of linecurrent isolationN/A (battery operateddevice)N/A (battery operateddevice)Similar
Patient Leakage CurrentN/A (battery operated device)N/A (battery operated device)Similar
• Normal conditionN/A (battery operated device)N/A (battery operated device)Similar
• Single fault conditionN/A (battery operated device)N/A (battery operated device)Similar
Number of Output ModesOne (NMES)One (NMES)Similar
Number of Output ChannelsFourFourSimilar
Synchronous or Alternating?Synchronous, but never 2 channels activated at the same timeSynchronous, but never 2 channels activated at the same timeSimilar
Method of Channel IsolationEach channel is the middle of a H-Bridge. Except when it is activated, each channel is always in high impedance state.Each channel is the middle of a H-Bridge. Except when it is activated, each channel is always in high impedance state.Similar
Regulated Current or Regulated Voltage?Regulated current (all channels)Regulated current (all channels)Similar
Software/Firmware/Microprocessor Control?YesYesSimilar
Automatic Overload Trip?YesYesSimilar
Automatic No-Load Trip?YesYesSimilar
Automatic Shut Off?"On/Off" switch"On/Off" switchSimilar
Patient Override Control?YesYesSimilar
Indicator Display - On/Off Status?YesYesSimilar
- Low Battery?YesYesSimilar
- Voltage/Current Level?Yes, unit = [Energy]Yes, unit = [Energy]Similar
Timer Range (minutes)Maximum = 48 minutes; Screen shows remaining time in minutes and displays image showing time remainingMaximum = 55 minute Screen shows remaining time in minutes and seconds and displays countdown timerDifferent
Compliance with 21 CFR 898?YesYesSimilar
Charging SystemAC/DC 5[v] 3.5[A]; Distributed through docking station to remoteAC/DC 9[v] 0.4[A] Distributed directly to the deviceDifferent
and 4 modules
Weight-Remote: 110 [g]- Stimulation Module:2x60 [g]- Docking Station 800 [g]300 [g]Different
Dimensions [W x H x D]-Remote 9x4.5x0.7[cm]- Stimulation Module: 6.5x2 [cm]- Docking Station: 25x25x2 [cm]99 x 142 x 36 [mm]3.9 x 5.6 x 1.4 [in]Different

Basic Device Characteristics – Comparison with Predicate Device

{6}------------------------------------------------

{7}------------------------------------------------

Output Specifications – Comparison with Predicate Device

CharacteristicNew DevicePredicate DeviceSimilar/Different
WaveformSymmetrical BiphasicSymmetrical BiphasicSimilar
ShapeRectangularRectangularSimilar
Maximum OutputVoltage (± 10%)60 V @ 500 Ω180 V @ 2 kΩ180 V @ 10 kΩ60 V @ 500 Ω165 V @ 2 kΩ165 V @ 10 kΩDifferent
Maximum OutputCurrent (± 10%)120 mA @ 500 Ω90 mA @ 2 kΩ18 mA @ 10 kΩ120 mA @ 500 Ω82 mA @ 2 kΩ16 mA @ 10 kΩDifferent
Pulse Width300 to 400 [μs](microseconds)200 to 400 [μs](microseconds)Different
Frequency1 to 120 Hz1 to 120 HzSimilar
For multiphasicwaveforms only:- Symmetricalphases?- Phase Duration(include units)(state range, ifapplicable) (bothphases, ifasymmetrical)Symmetrical, 300 - 400 μsSymmetrical, 200 - 400 μsDifferent
Net Charge[μC/pulse]0 [μC] @ 500ΩExcitation pulse fullycompensated0 [μC] @ 500ΩExcitation pulse fullycompensatedSimilar
Maximum PhaseCharge [μC]48 [μC] @ 500Ω48 [μC] @ 500ΩSimilar
Maximum Current(RMS) Density[mA/cm²]1.49 [mA/cm²] @ 500Ω1.49 [mA/cm²] @ 500ΩSimilar
Maximum PowerDensity [mW/cm²]27.6 [mW/cm²] @500Ω27.6 [mW/cm²] @500ΩSimilar

{8}------------------------------------------------

Although there are minor differences in the maximum output voltage and current, the physical dimensions, the timer range, the batteries used as a power source, and the battery charging specifications, the most significant difference between the two devices is the connection of the device to the electrodes. The Compex Wireless USA uses wireless technology, whereas the predicate device uses 6-pole cables. This type of wireless technology is used in other OTC devices, such as the WiTouch TENS device (k1120500). None of these differences, including the use of wireless technology, raise any new issues of safety or effectiveness.

Performance Testing:

Electrical Safety and Electromagnetic Compatibility: The Compex Wireless USA was tested and found to comply with recognized standards for electrical safety and electromagnetic compatibility.

FCC Radio Frequency Testing: The Compex Wireless USA was tested to FCC requirements and found to comply with the requirements of 47 CFR 15.249.

Software Verification: The device's software was verified in accordance with the requirements of FDA's guidance document: General Principles of Software Validation, January 11, 2002. The software testing demonstrated that the software meets its design requirements.

Usability/Human Factors Testing: Usability/Human Factors testing was performed, which demonstrated that the established requirements for usability were met, and the device's design is appropriate for the intended users and use environment. The result of this study substantiates the acceptability of the use-related risks identified during the risk assessment activities.

Wireless Coexistence Testing: The performance of Compex Wireless USA was evaluated in an environment with other Compex Wireless USA device and with other types of 2.4 GHz wireless devices (Bluetooth and Wi-Fi). The device met all specified requirements.

Standards:

DesignationNameFDA Recognition
AAMI/ANSI ES 60601-1:2005(R)2012 and C1:2009/(R)2012and A2:2010/(R)2012 (ConsolidatedText).Compex Wireless USA deviceMedical Electrical Equipment – Part 1:General Requirements for Basic Safetyand Essential Performance19-5
IEC 60950-1:2005 (2nd Ed.):Docking Station OnlyInformation Technology Equipment –Safety - Part 1: General requirementsN/A

The Compex Wireless USA conforms to the following standards.

{9}------------------------------------------------

IEC 60601-1-2 Ed. 3:2007-03:Compex Wireless USA deviceMedical Electrical Equipment - Part 1-2:General Requirements for Basic Safetyand Essential Performance - CollateralStandard: Electromagnetic Compatibility– Requirements and Tests19-1
EN 55022:2010/AC:2011Docking Station onlyInformation Technology Equipment –Radio Disturbance Characteristics –Limits and Methods of MeasurementN/A
EN 55024:2010Docking Station onlyInformation Technology Equipment –Immunity Characteristics – Limits andMethods of MeasurementN/A
EN 61000-3-2:2006 +A1:2009+A2:2009Docking Station onlyElectromagnetic Compatibility (EMC) –Part 3.2: Limits – Limits for HarmonicCurrent Emissions (Equipment InputCurrent Less Than or Equal To 16 A perPhase).N/A
EN 61000-3-3:2013Docking Station onlyElectromagnetic Compatibility (EMC) –Part 3.3: Limits - Limitation of VoltageChanges, Voltage Fluctuations andFlicker in Public Low-Voltage SupplySystems, for Equipment with RatedCurrent Less Than or Equal To 16 A perPhase and Not Subject to ConditionalConnectionN/A
IEC 60601-1-6 Ed. 3.0: 2010-01Medical Electrical Equipment - Part 1-6:General Requirements for Basic Safetyand Essential Performance - CollateralStandard: Usability5-85
IEC 60601-1-11 Ed. 1.0: 2010-04Medical Electrical Equipment – Part 1-11: General Requirements - CollateralStandard: Requirements for MedicalElectrical Equipment and MedicalElectrical Systems Used in the HomeHealthcare Environment19-6
IEC 60601-2-10 Ed. 2.0 2012-06Medical Electrical Equipment – Part 2-10: Particular Requirements for theBasic Safety and Essential Performanceof Nerve and Muscle Stimulators17-11
IEC 62366:2007Medical Devices – Application ofUsability Engineering to MedicalDevicesN/A
AAMI/ANSI HE75:2009Human Factors Engineering – Design ofMedical Devices5-57
ISO 14971:2007Medical Devices – Application of RiskManagement to Medical Devices5-40
IEC 62304 First Ed. 2006-05Medical Device Software - SoftwareLife Cycle Processes13-8

{10}------------------------------------------------

Conclusion:

Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the Compex Wireless USA is as safe and effective as, and substantially equivalent to, the predicate device.

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).