K Number
K152552

Validate with FDA (Live)

Device Name
Patient Monitor
Date Cleared
2016-04-29

(234 days)

Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

iM20 Patient Monitor (hereinafter called monitor) is intended to be used for monitoring, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG. respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult only.

The monitor is additionally intended for use during patient transport inside of the hospital environment.

The monitor is not intended for airplane, helicopter transport, home use and MRI environments.

Device Description

The iM20 Patient Monitor System (hereinafter called iM20) can perform long-time continuous monitoring of multiple physiological parameters, including ECG, respiration (RESP), non-invasive blood pressure (NIBP), oxygen saturation of the blood (SpO2), pulse rate (PR), temperature (TEMP), invasive pressure (IBP), Carbon Dioxide (CO2). The system capabilities include storing, displaying, analyzing and controlling such data afterwards. When necessary, alarms will be produced so that doctors and nurses can manage patient care appropriately. The system is intended to be used during patient transport inside and outside of the hospital environment.
iM20 can be utilized in two ways, as an independent monitor and as a module of V series Patient Monitor (including models elite V5, elite V6 and elite V8). When used as an independent monitor, it can simultaneously monitor, store, review several parameters data. And transfer patient data to V series patient monitor only under the transport mode. As a highly portable monitor, its compact design makes it particularly appropriate inside hospital and vehicle ambulance transport environments.
When as a multi-measurement module, when the iM20 is directly connected to a V series patient monitor, it can provide the measurements, trends, and patient information. When connected, the V series Patient Monitor controls the connected iM20, including all alarm functionality. So no alarms are available on iM20 in such application and iM20 takes power from the V series Patient Monitor.

AI/ML Overview

This document focuses on the Edan Instruments Inc. Patient Monitor, model iM20 (K152552), and its equivalency to predicate devices. It lists performance data based on non-clinical and clinical studies to demonstrate substantial equivalence.

Here's the breakdown of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document provides a detailed comparison table between the proposed device (iM20) and a predicate device (Philips Medizin.Systeme Boeblingen GmbH IntelliVue MP2 cleared under K102562), focusing on various physiological parameters and technical specifications. A second comparison table is provided for the reference device Edan Instruments, Inc. iM70 (K131971), specifically for PR from NIBP and CO2 function.

Since the provided text does not present a separate "acceptance criteria" table and "reported device performance" table, but rather a comparison with performance metrics of a predicate device, the following table synthesizes the information from the "Predicate Device Comparison" section to represent acceptance criteria (from the predicate) and the reported device performance (for the iM20). Differences are explicitly noted as "Different." Where the proposed device's performance aligns with the predicate, it is marked "Same."

Acceptance Criteria (Based on Predicate Device) and Reported Device Performance (iM20)

Feature / ParameterAcceptance Criteria (Predicate: Philips MP2)Reported Device Performance (iM20)Match?
ECG Function - HR Calculation
HR Range (Adult/Pedi/Neo)Adult/pedi: 15 to 300 bpm, Neo: 15 to 350 bpmADU: 15 bpm to 300 bpm, PED/NEO: 15 bpm to 350 bpmSame
HR Accuracy±1% of range± 1% or 1 bpm, whichever is greaterDifferent
HR Resolution1 bpm1 bpmSame
Sensitivity≥ 200 µVpeak≥ 300 µVPPDifferent
PVC Rate Range0 to 300 bpmADU: 0 to 300 PVCs/ min, PED/NEO: 0 to 350 PVCs/ minDifferent
PVC Resolution1 bpm1 PVCs/minDifferent
ST Numeric Range-20 to +20 mm-2.0 mV to +2.0 mVDifferent
ST Accuracy± 0.5mm or 15%, whichever is greater-0.8 mV to +0.8 mV: ± 0.02 mV or 10%, whichever one is greater. Beyond this range: not specified.Different
ST Resolution0.1mm0.01 mVDifferent
Range of Sinus and SV Rhythms (Brady)Adult: 15 to 60 bpm, Pedi: 15 to 80 bpm, Neo: 15 to 90 bpmAdult: RR interval for 5 consecutive QRS complex ≥ 1.5 s. Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≥ 1 s.Different
Range of Sinus and SV Rhythms (Normal)Adult: 60 to 100 bpm, Pedi: 100 to 160 bpm, Neo: 90 to 180 bpmAdult: 0.5s < RR interval for 5 consecutive QRS complex < 1.5 s. Pedi/Neo: 0.375s < RR interval for 5 consecutive QRS complex < 1 s.Different
Range of Sinus and SV Rhythms (Tachy)Adult: >100 bpm, Pedi: >160 bpm, Neo: >180 bpmAdult: RR interval for 5 consecutive QRS complex ≤ 0.5 s. Pediatric/neonatal: RR interval for 5 consecutive QRS complex ≤ 0.375 s.Different
BandwidthDiagnosis: 0.05 to 150 Hz, Monitor: Adult: 0.5 to 40 Hz, Neo/pedi: 0.5 to 55 Hz, Filter Mode: 0.5 to 20 HzDiagnosis: 0.05Hz to 150Hz, Monitor: 0.5Hz to 40Hz, Surgery: 1Hz to 20HzDifferent
CMRRDiagnostic mode: >86 dB, Filter mode: >106 dBDiagnosis: >95dB (Notch filter off), Monitor: >105dB (Notch filter on), Surgery: >105dB (Notch filter on)Different
Differential Input Impedance>2 MΩ RA-LL leads (Resp), >5 MΩ at all other leads>5MΩDifferent
Input Signal Range±5mV (peak-to-peak value)±10mVPPDifferent
Electrode Offset Potential Tolerance±500mV±500mVSame
Auxiliary Current (detect)Active electrode: <100nA, Reference electrode: <900nAActive electrode: <100nA, Reference electrode: <900nASame
Pace Pulse Indicator±2 mV to ±700 mV and widths from 0.1 ms to 2.0 msAmplitude: ±2 mV to ±700 mV, Width: 0.1 ms to 2.0 ms, Ascending time: 10 us to 100 usDifferent
Pace Pulse Rejection±2 mV to ±700 mV and widths from 0.1 ms to 2.0 msAmplitude: ±2 mV to ±700 mV, Width: 0.1 ms to 2.0 ms, Ascending time: 10 us to 100 usDifferent
Max Start-up alarm time for Ventricular Tachycardia (1mV 206bpm)Gain 1.0 Range 6.1 to 6.9 seconds, Average 6.5 secondsGain 1.0: 10 sDifferent
Max Start-up alarm time for Ventricular Tachycardia (2mV 195bpm)Gain 1.0, Range 5.7 to 6.5 seconds, Average 6.1 secondsGain 1.0: 10 sDifferent
Tall T-wave RejectionExceeds ANSI/AAMI EC 13 Sect. 3.1.2.1(c) 1.2 mV T-Wave amplitudeComplied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17 minimum recommended 1.2 mV T-Wave amplitudeSame
HR Averaging MethodThree different methods used, typically 12 most recent RR intervalsMethod 1: excluding min/max values from 12 most recent RR intervals and averaging the residual 10. Method 2: 4 most recent RR intervals if 3 consecutive > 1200ms.Different
Accuracy of HR Meter and Response to Irregular RhythmVentricular bigeminy: 80 bpm, Slow alternating ventricular bigeminy: 60 bpm, Rapid alternating ventricular bigeminy: 120 bpm, Bidirectional systoles: 90 bpmVentricular bigeminy: 80 bpm±1 bpm, Slow alternating ventricular bigeminy: 60 bpm±1 bpm, Rapid alternating ventricular bigeminy: 120 bpm±1 bpm, Bidirectional systoles: 91 bpm±1 bpmDifferent
Response time of HR Meter to Change in HR (80 to 120 bpm)Range: [6.4 to 7.2 s], Average: 6.8 sWithin 11 sDifferent
Response time of HR Meter to Change in HR (80 to 40 bpm)Range: [5.6 to 6.4 s], Average: 6.0 sWithin 11 sDifferent
RESP Function
Respiration excitation waveformSinusoidal signal, 260 µA, 39 kHzSinusoid, 62.8 kHz (±10%), <500 µADifferent
RR Measuring RangeAdult/pedi: 0 to 120 rpm, Neo: 0 to 170 rpmAdult: 0 rpm to120rpm, Pediatric/neonate: 0 rpm to 150rpmDifferent
Accuracyat 0 to 120 rpm ±1 rpm, at 120 to 170 rpm ±2 rpmAdult: 6 to 120 rpm: ±2 rpm, 0 to 5 rpm: not specified. Neo/Ped: 6 to 150 rpm: ±2 rpm, 0 to 5 rpm: not specified.Different
Resolution1 rpm1 rpmSame
Waveform bandwidth0.3 to 2.5 Hz (-6 dB)0.2Hz to 2.5Hz (-3dB)Different
Apnea Alarm Time Setup10 to 40 seconds 5 second steps10s, 15s, 20s, 25s, 30s, 35s, 40sDifferent
NIBP Function
Principle of Operationoscillationoscillation (EDAN, SunTech NIBP)Same
Measurement Range (Adult Systolic)30 to 270 mmHgEDAN NIBP: 40 mmHg ~ 270 mmHg, SunTech NIBP: 60 mmHg ~ 250 mmHgDifferent
Measurement Range (Adult Diastolic)10 to 245 mmHgEDAN NIBP: 10 mmHg ~ 215 mmHg, SunTech NIBP: 30 mmHg ~ 190 mmHgDifferent
Measurement Range (Adult Mean)20 to 255 mmHgEDAN NIBP: 20 mmHg ~ 235 mmHg, SunTech NIBP: 40 mmHg ~ 210 mmHgDifferent
Measurement Range (Pediatric Systolic)30 to 180 mmHgEDAN NIBP: 40 mmHg ~ 230 mmHg, SunTech NIBP: 40 mmHg ~ 230 mmHgDifferent
Measurement Range (Pediatric Diastolic)10 to 150 mmHgEDAN NIBP: 10 mmHg ~ 180 mmHg, SunTech NIBP: 20 mmHg ~ 160 mmHgDifferent
Measurement Range (Pediatric Mean)20 to 160 mmHgEDAN NIBP: 20 mmHg ~ 195 mmHg, SunTech NIBP: 30 mmHg ~ 175 mmHgDifferent
Measurement Range (Neonate Systolic)30 to 130 mmHgEDAN NIBP: 40 mmHg ~135 mmHg, SunTech NIBP: 40 mmHg ~130 mmHgDifferent
Measurement Range (Neonate Diastolic)10 to 100 mmHgEDAN NIBP: 10 mmHg ~ 100 mmHg, SunTech NIBP: 20 mmHg ~ 90 mmHgSame/Different
Measurement Range (Neonate Mean)20 to 120 mmHgEDAN NIBP: 20 mmHg ~ 110 mmHg, SunTech NIBP: 30 mmHg ~ 100 mmHgDifferent
AccuracyMax. Std. Deviation: 8 mmHg, Max. Mean Error: ± 5 mmHgMax standard deviation: 8 mmHg (EDAN, SunTech); Max mean error: ± 5 mmHgSame
Pressure Resolution1mmHg1mmHg (EDAN, SunTech)Same
Maximum measuring period180 seconds (adult/pedi), 90 seconds (neonates)Adult/Pediatric 120s, Neonate 90sDifferent
Overpressure protection (Adult)>300 mmHg (40 kPa) >2 sec297 ± 3mmHg (EDAN), <300mmHg (SunTech)Different
Overpressure protection (Pediatric)>300 mmHg (40 kPa) >2 sec240 ± 3mmHg (EDAN), <300mmHg (SunTech)Different
Overpressure protection (Neonate)>150 mmHg (20 kPa) >2 sec147 ± 3mmHg (EDAN), <150mmHg (SunTech)Different
Measuring interval in AUTO Mode1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60 or 120 minutes1/2/3/4/5/10/15/30/60/90/120/240/480 min (EDAN), 1/2/3/4/5/10/15/30/60/90/120/240/480 min (SunTech)Different
Continuous5 minutes5min, interval is 5sSame
SpO2 Function (EDAN)
Measurement Range0 to 100 %0-100%Same
AccuracyPhilips Reusable Sensors: M1191A, etc. = 2 % (70 % to 100 %); M1193A, etc. = 3 % (70 % to 100 %)Adult/Pediatric: ± 2 % (70% to 100% SpO2), Undefined (0 to 69% SpO2). Neonate: ± 3 % (70% to 100% SpO2), Undefined (0 to 69% SpO2).Different
Resolution1 %1 %Same
Pulse Rate Measuring Range30 to 300 bpm25 bpm to 300 bpmDifferent
Pulse Rate Accuracy±2 % or 1 bpm, whichever is greater±2bpmDifferent
Pulse Rate Resolution1 bpm1 bpmSame
SpO2 Function (Nellcor)
Measurement Range1% to 100%1% to 100%Same
AccuracyPhilips Reusable Sensors: M1193A, etc. = 3 % (70 % to 100 %) (Adult); = 4 % (70 % to 100 %) (Neonate)DS-100A, OXI-A/N (Adult): ± 3% (70% to 100% SpO2). OXI-A/N (Neonate): ± 4% (70% to 100% SpO2).Same
Resolution1 %1 %Same
Pulse Rate Measuring Range30 to 300 bpm20bpm to 300bpmDifferent
Pulse Rate Accuracy±2 % or 1 bpm, whichever is greater± 3bpm (20bpm to 250bpm)Different
Pulse Rate Resolution1 bpm1 bpmSame
Temperature Function
Measurement Range-1 to +49°C (30 to 113°F)0 to 50°CDifferent
Accuracy±0.1°C (±0.2°F)±0.1 °CSame
Resolution0.1°C (±0.1°F)0.1 °CSame
Average Time ConstantLess than 10 seconds<30 sDifferent
IBP Function
Measurement Range-40 to 360 mmHg-50 to +300 mmHgDifferent
Accuracy4 % of reading or ±4 mmHg, whichever is greater± 2 % or ±1 mmHg, whichever is greaterDifferent
Resolution1 mmHg1 mmHgSame
Transducer Load/Output ImpedanceLoad Impedance: 200 to 2000 Ω, Output Impedance: 3000 Ω5 (μV/V/mmHg) 300 to 3000 ΩDifferent
Zero Range±200 mmHg±200 mmHgSame
General Specifications
Electrical Safety Anti-electroshock typeClass IIClass IISame
Working systemContinuous operation equipmentContinuous operation equipmentSame
Operating Temp Range0 to 40°C (32 to 104°F)0°C to +40°CSame
Storage Temp Range-20 to 60°C (-4 to 140°F)-30°C to +70°CDifferent
Operating Humidity Range15 % to 95 % RH15% to 95% (non-condensing)Same
Storage Humidity Range5 % to 95 % RH15% to 95% (non-condensing)Different
Operating Altitude Range-500 m to 3000 m (10000 ft)615hPa to 1060hPa (DC-DC power supply)(-400 m to 4000 m), 680hPa to 1060hPa (AC to DC power supply)(-400m to 3000m)Different
Storage Altitude Range-500 m to 4600 m (15000 ft)615hPa to 1060hPa (-400 m to 4000 m)Different
Weight1.25 kgiM20: <1.5kg, EFM: < 0.58 kgDifferent
Dimensions(W x H x D) 188 x 99 x 86 mmiM20: 185 mm (L) × 85.3 mm (W)× 116 mm (H), EFM: 207 mm (L) × 93.4 mm (W)× 116 mm (H)Different
Display Description72 x 54 mm (2.8 x 2.1 in)5 inches LCDDifferent
Display Resolution320 x 240800x480Different
Power consumption<12 W average, <30 W while battery is loading<25w (Max)Different
Power input36 to 60 V DC floating11.1V to 19.8VDCDifferent
Current0.7 to 0.4 A1.27~2.3ADifferent
AC-DC Power supply Line Voltage100 to 240 V ~100~240VacSame
AC-DC Power supply Output Voltage36 to 60 V DC floating15V±5%dc, 24VADifferent
AC-DC Power supply Frequency50/60 Hz ~50/60 Hz ~Same
DC-DC Power supply Line VoltageN/A10~16VdcDifferent
DC-DC Power supply Output VoltageN/A15V±5%dcDifferent
Standard Compliance (General)IEC 60601-1:1988 + A1:1991 + A2:1995IEC 60601-1:2005Different
Out-Of-Hospital Transport Shock TestsAccording to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-27 (peak acceleration up to 100g).According to IEC TR 60721-4-7, Class 7M3. IEC/EN60068-2-27 (peak acceleration up to 100g).Same
Out-Of-Hospital Transport Random VibrationAccording to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-64 (RMS acceleration 5g).According to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-64 (RMS acceleration 5g).Same
Out-Of-Hospital Transport Sinusoidal VibrationAccording to IEC TR 60721-4-7, Class 7M3. IEC/EN 60068-2-6 (acceleration up to amplitude 2g).According to EN1789:2007+A: 2010, 6.4.1 vibration test, IEC/EN 60068-2-6 (acceleration up to amplitude 2g).Different
Out-Of-Hospital Transport Bump-- (Not specified for predicate)According to EN1789:2007+A: 2010, 6.4.1 Bump test, IEC/EN 60068-2-6 (acceleration up to amplitude 15g).N/A
Out-Of-Hospital Transport Drop TestAccording to EN1789 (covers also IEC TR 60721-4-7 and Class 7M3). IEC/EN 60068-2-32 (height 0.75 m).according to EN1789:2007+A:2010, IEC 60068-2-32 (height 1.2 m).Different
Degrees of Protectionprovided by enclosures according to IEC/EN 60529: IP32provided by enclosures according to IEC/EN 60529: IP44Different
Medical Vehicles and EquipmentsEN 1789 +A1:2003 Medical vehicles and their equipment - Road ambulances (chapter 6 - Medical Devices).EN1789:2007+A: 2010 Medical vehicles and their equipment - Road ambulances (chapter 6 – Medical Devices).Same
Radiated Susceptibility20 V/m according to EN ISO 9919 (SpO2) and EN ISO 21647 (CO2).20 V/m according to EN ISO 9919 (SpO2) and EN ISO 21647 (CO2).Same
Alarm CategoriesPhysiological alarms, Technical alarms, Prompt informationPhysiological alarms, Technical alarms, Prompt informationSame
Alarm LevelsHigh level alarms, Medium level alarms, Low level alarmsHigh level alarms, Medium level alarms, Low level alarmsSame
Alarm ModesVisual alarms, Audible alarms, Alarm messages, Parameter flashesVisual alarms, Audible alarms, Alarm messages, Parameter flashesSame

Comparison to Reference Device iM70 (Specifically for PR from NIBP and CO2 Function)

Feature / ParameterAcceptance Criteria (Reference: iM70)Reported Device Performance (iM20)Match?
PR from NIBP
Measurement range40 to 240 bpm40 to 240 bpm (EDAN), 30 bpm ~220bpm (SunTech)Same/Different
Accuracy±3bpm or 3.5%, whichever is greater±3bpm or 3.5%, whichever is greater (EDAN); ±3bpm or ±2%, whichever is greater (SunTech)Same/Different
CO2 Function
EtCO2 Range0 to 150 mmHg0 mmHg ~ 150 mmHgSame
FiCO2 Range3 mmHg to 50 mmHg3 mmHg~50 mmHgSame
AwRR Range2 to 150 rpm2 rpm ~ 150 rpm (Sidestream)Same
EtCO2 Resolution1.0 mmHg1mmHgSame
FiCO2 Resolution1mmHg1mmHgSame
AwRR Resolution±1 rpm1 rpmSame
EtCO2 Accuracy (0-40 mmHg)±2.0 mmHg± 2 mmHgSame
EtCO2 Accuracy (41-70 mmHg)±5 % of reading± 5 %Same
EtCO2 Accuracy (71-100 mmHg)±8 % of reading± 8 %Same
EtCO2 Accuracy (101-150 mmHg)| ±10 % of reading± 10 %Same
RESP > 80rpm±12 % of actual± 12 %Same
AwRR Accuracy±1 rpm± 1 rpmSame
Sample Gas Flow rate50±10ml/min50±10 ml/minSame
O2 Compensation Range0 ~ 100%0~100%Same

2. Sample Sizes and Data Provenance for Test Set

  • Clinical Testing Sample Size: Not explicitly stated in terms of patient numbers or case count.
  • Data Provenance: The NIBP clinical testing was conducted as per ISO 81060-2: 2013, "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type." The document does not specify the country of origin for the clinical data or whether it was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth of Test Set

The document does not provide information on:

  • The number of experts used to establish ground truth for the test set.
  • The qualifications of those experts.

4. Adjudication Method for the Test Set

The document does not provide information on the adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No information about a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is provided. The study focuses on the device's technical and clinical performance against standards and a predicate device, not on human reader performance with or without AI assistance.

6. Standalone Performance Study (Algorithm Only)

The document primarily describes the performance of the entire "iM20 Patient Monitor device, consisting of all the modules and accessories in the system." While it details performance parameters of various physiological monitoring algorithms (ECG, NIBP, SpO2, CO2), it does not explicitly describe a standalone (algorithm only without human-in-the-loop performance) study in the detailed sense of a specific data set applied only to the algorithm for evaluation. However, the bench testing and NIBP clinical validation are evaluations of the device's inherent measurement capabilities.

7. Type of Ground Truth Used

  • For ECG, RESP, NIBP, SpO2, IBP, TEMP, CO2: The measurements of these physiological parameters are compared against the expected performance defined by international standards (e.g., ISO, IEC) or against the performance of a predicate device. For NIBP, the ground truth is established through clinical validation against "automated measurement type" as per ISO 81060-2:2013, which typically involves comparison to a reference standard measurement method (e.g., auscultatory method).
  • For Biocompatibility: Compliance with ISO 10993-1, (Cytotoxicity, Skin Sensitization, Skin Irritation).
  • For Electrical Safety and EMC: Compliance with IEC 60601-1:2005/A1:2012 and IEC 60601-1-2:2007.
  • For General Performance (Bench Testing): Compliance with various IEC/ISO standards specific to each parameter.

8. Sample Size for the Training Set

The document does not provide any information regarding a training set size, as it describes a medical device clearance submission based on equivalence to predicate devices and adherence to performance standards, not a machine learning model requiring a training set.

9. How Ground Truth for Training Set was Established

Not applicable, as no training set is mentioned in the context of this device's submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 29, 2016

Edan Instruments. Inc. % Doug Worth Sr. Dir. US RA/QA Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94086

Re: K152552

Trade/Device Name: Patient Monitor, model iM20 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, GXY, DPS, DRG Dated: March 22, 2016 Received: March 28, 2016

Dear Doug Worth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152552

Device Name Patient Monitor, model iM20

Indications for Use (Describe)

iM20 Patient Monitor (hereinafter called monitor) is intended to be used for monitoring, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.

The monitored physiological parameters include: ECG. respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult only.

The monitor is additionally intended for use during patient transport inside of the hospital environment.

The monitor is not intended for airplane, helicopter transport, home use and MRI environments.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (K) Summary

Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92

1. Submitter:Edan Instruments, Inc3/F - B, Nanshan MedicalEquipments Park, Nanhai Rd 1019#,Shekou, Nanshan Shenzhen,
518067 P.R. China
Tel: +86(0755) 26858736
Fax: +1 (408) 418-4059
Contact person:Queena Chen
Preparing date:March 22, 2015
2. Device name andclassification:Device Name: Patient Monitor
Model: iM20
Classification Name/ Product code:
870.1025 monitor, physiological, patient(with arrhythmiadetection or alarms)/ MHX
870.1025 Detector and Alarm, Arrhythmia/ DSI
870.1025 Monitor, ST Segment with Alarm/ MLD
870.2300 Cardiac monitor (including cardiotachometer and ratealarm)/ DRT
870.1130 Non-Invasive blood pressure/ DXN
870.1110 Blood pressure computer/ DSK
880.2910 Clinical Electronic Thermometers-TemperatureMonitor with Probe/ FLL
870.2700 Oximeter, Pulse/ DQA
868.1400 Carbon Dioxide Gas Analyzer/ CCK
882.1320 Cutaneous electrode/GXY
870.2340 Electrocardiograph/DPS
870.2910 Radiofrequency physiological signal transmitter andreceiver/ DRG
Regulatory Class: Class II
3.PremarketNotification Class IIICertification andSummaryNot applicable, the subject device is Class II.
4. Predicate Device(s):1) Philips Medizin.Systeme Boeblingen GmbH IntelliVue MP2 cleared under K102562.
2) Edan Instruments, Inc. iM70/ K131971
5. Reason forSubmissionIntroduce new device iM20
6. Pre-Submission,IDENot applicable, there is no prior submission.
7. Device Description:The iM20 Patient Monitor System (hereinafter called iM20) canperform long-time continuous monitoring of multiplephysiological parameters, including ECG, respiration (RESP),non-invasive blood pressure (NIBP), oxygen saturation of theblood (SpO2), pulse rate (PR), temperature (TEMP), invasivepressure (IBP), Carbon Dioxide (CO2). The system capabilitiesinclude storing, displaying, analyzing and controlling such dataafterwards. When necessary, alarms will be produced so thatdoctors and nurses can manage patient care appropriately. Thesystem is intended to be used during patient transport inside andoutside of the hospital environment.iM20 can be utilized in two ways, as an independent monitorand as a module of V series Patient Monitor (including modelselite V5, elite V6 and elite V8). When used as an independentmonitor, it can simultaneously monitor, store, review severalparameters data. And transfer patient data to V series patientmonitor only under the transport mode. As a highly portablemonitor, its compact design makes it particularly appropriateinside hospital and vehicle ambulance transport environments.When as a multi-measurement module, when the iM20 isdirectly connected to a V series patient monitor, it can providethe measurements, trends, and patient information.Whenconnected, the V series Patient Monitor controls the connectediM20, including all alarm functionality. So no alarms areavailable on iM20 in such application and iM20 takes powerfrom the V series Patient Monitor.
8. IntendedUse/Indications forUse:iM20 Patient Monitor (hereinafter called monitor) is intended tobe used for monitoring, storing, and reviewing of, and togenerate alarms for, multiple physiological parameters ofadults, pediatrics and neonates. The monitors are intended foruse by trained healthcare professionals in hospitalenvironments.The monitored physiological parameters include: ECG,respiration (RESP), temperature (TEMP), oxygen saturation of

{4}------------------------------------------------

{5}------------------------------------------------

pressure (NIBP), and invasive blood pressure (IBP), and carbon dioxide (CO2).

The arrhythmia detection and ST Segment analysis are intended for adult only.

The monitor is additionally intended for use during patient transport inside and outside of the hospital environment.

The monitor is not intended for airplane, helicopter transport, home use and MRI environments.

9. Predicate Device Comparison

The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters improvement, detailed substantial equivalence discussion are included in the following tables.

ItemProposed device: iM20Predicate device:IntelliVue MP2Comparison Result
510(k)NumberCurrent SubmissionK102562
Indications for Use
Intended UseiM20 Patient Monitor is intended to be used for monitoring, storing, and reviewing of, and to generate alarms for multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended to be used by trained healthcare professionals in a hospital environment. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), and carbon dioxide (CO2). The arrhythmia detection and ST-segment monitoring are intended for adult only. Neonatal and pediatricIndicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording of, and to generate alarms for multiple physiological parameters of adults, pediatrics and neonates in hospital environments. TheDifferent
patients are not clinically validated.The monitors are indicated for use byhealth care professionals or under theirdirection whenever there is a need formonitoring the physiological parametersof patients.The monitors are additionally intendedfor use during patient transport insideand outside of the hospital environment.The monitor is not intended for airplane,helicopter transport, home use or MRIenvironments.MP2, X2, MP5,MP5T, MP20,MP30, MP40, andMP50 areadditionallyintended for use intransport situationswithin hospitalenvironments. TheMP2, X2and MP5are also intendedfor use duringpatient transportoutside of ahospitalenvironment.
ECG Function
HR Calculation
RangeADU: 15 bpm to 300 bpmPED/NEO: 15 bpm to 350 bpmAdult/pedi: 15 to300 bpmNeo range: 15 to350 bpmDifferent
Accuracy$\pm$ 1% or 1 bpm, whichever is greater$\pm$ 1% of rangeDifferent
Resolution1 bpm1bpmSame
Sensitivity$\ge$ 300 $\mu$ VPP$\ge$ 200 $\mu$ VpeakDifferent
PVC Rate
RangeADU: 0 to 300 PVCs/ minPED/NEO: 0 to 350 PVCs/ min0 to 300 bpmDifferent
Resolution1 PVCs/min1 bpmDifferent
ST Numeric
Range-2.0 mV to +2.0 mV-20 to +20 mmDifferent
Accuracy-0.8 mV to +0.8 mV: $\pm$ 0.02 mV or 10%,which ever one is greater.Beyond this range: not specified.$\pm$ 0.5mm or 15%,which ever one isgreaterDifferent
Resolution0.01 mV0.1mmDifferent
Range of Sinus and SV Rhythm
BradyAdult: RR interval for 5 consecutiveQRS complex $\ge$ 1.5 s.Pediatric/neonatal: RR interval for 5consecutive QRS complex $\ge$ 1 s.Adult: 15 to 60bpmPedi: 15 to 80 bpmNeo: 15 to 90 bpmDifferent
NormalAdult: 0.5s < RR interval for 5consecutive QRS complex < 1.5 s.Adult: 60 to 100bpmDifferent
for 5 consecutive QRS complex < 1 s.bpm
Neo: 90 to 180 bpm
TachyAdult: RR interval for 5 consecutiveQRS complex ≤ 0.5 s.Pediatric/neonatal: RR interval for 5consecutive QRS complex ≤ 0.375 s.Adult: >100 bpmPedi: >160 bpmNeo: >180 bpmDifferent
Bandwidth(-3dB)Diagnosis: 0.05Hz to 150HzMonitor: 0.5Hz to 40HzSurgery: 1Hz to 20HzDiagnosis:Adult/neo/pedi:0.05 to 150 HzMonitor: Adult: 0.5to 40 HzNeo/pedi: 0.5 to 55 HzFilter Mode:Adult/neo/pedi: 0.5to 20 HzDifferent
CMRR(CommonModeRejectionRatio)Diagnosis: >95dB (the Notch filter isoff)Monitor: >105dB (the Notch filter is on)Surgery: >105dB (the Notch filter is on)Diagnosticmode: >86 dB(with a 51 k $/$ 47nF imbalance).Filter mode: >106dB (with a 51 k/47nF imbalance).Different
DifferentialInputImpedance>5MΩ>2 MΩ RA-LLleads (Resp)>5 MΩ at all otherleads (at 10 Hzincluding patientcable)Different
Input SignalRange±10mVPP±5mV(peak-to-peakvalue)Different
ElectrodeOffsetPotentialTolerance±500mV±500mVSame
AuxiliaryCurrent(Leads offdetection)Active electrode: <100nAReference electrode: <900nAActive electrode:<100nAReferenceelectrode: <900nASame
PacePulseIndicatorPulse is marked if the requirements ofIEC 60601-2-27: 2011, Sect.201.12.1.101.12 are met:±2 mV to ±700 mVand widths from0.1 ms to 2.0 msDifferent
Amplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 us to 100 us
Pulse is marked if the requirements of±2 mV to ±700 mV
IEC 60601-2-27: 2011, Sect.and widths from
Pulse201.12.1.101.12 are met:0.1 ms to 2.0 ms
RejectionAmplitude: ±2 mV to ±700 mV
Width: 0.1 ms to 2.0 ms
Ascending time: 10 us to 100 us
Maximum start-up alarm time for Tachycardia
VentricularGain 1.0: 10 sGain 1.0 Range 6.1Different
Tachycardiato 6.9 seconds,
1 mVAverage 6.5
206bpmseconds
VentricularGain 0.5: 10 sGain 0.5, Range 6.5
Rhythmto 8.4 seconds,
VentricularAverage 7.2
Bradycardiaseconds
Gain 2.0: 10 sGain 2.0, Range 5.9
to 6.7 seconds,
Average 6.3
seconds
Gain 1.0: 10 sGain 1.0, Range 5.7
to 6.5 seconds,
Average 6.1
seconds
VentricularGain 0.5: 10 sGain 0.5, Range 5.4
Tachycardiato 6.2 seconds,
2 mVAverage 5.8
195bpmseconds
Gain 2.0: 10 sGain 2.0, Range 5.3
to 6.1 seconds,
Average 5.7
seconds
Complied with IEC 60601-2-27: 2011,Exceeds
Sect. 201.12.1.101.17 minimumANSI/AAMI EC
recommended 1.2 mV T-Wave13 Sect. 3.1.2.1(c)
Tall T-waveamplitudeminimum
Rejectionrecommended 1.2
mV T-Wave
amplitude.
Heart RateMethod1.Heart rate is computed byThree differentDifferent
Averagingexcluding the minimum and maximummethods are used:
Methodvalues from the 12 most recent RRNormally, heart
intervals and averaging the residual 10rate is computed by
RR intervals.averaging the 12
Method 2.If each of three consecutivemost Recent RR
RR intervals is greater than 1200ms, thenintervals.
the four most recent RR intervals areFor runs of PVCs,
averaged to compute the HR.up to 8 RR
intervals are
averaged to
compute the HR. If
each of 3
consecutive RR
intervals is greater
than 1200 ms (that
is, rate less than 50
bpm), then the 4
most recent RR
intervals are
averaged to
compute the HR.
Complied with IEC 60601-2-27: 2011,VentricularDifferent
Sect. 201.7.9.2.9.101 b) 4), the HR valuebigeminy: 80 bpm
Accuracy ofafter 20 seconds of stabilization isSlow alternating
Heart Ratedisplayed as follows:ventricular
Meter andVentricular bigeminy: 80 bpm±1 bpmbigeminy: 60 bpm
Response toSlow alternating ventricular bigeminy:Rapid alternating
Irregular60 bpm±1 bpmventricular
RhythmRapid alternating ventricular bigeminy:bigeminy: 120 bpm
120 bpm±1 bpmBidirectional
Bidirectional systoles: 91 bpm±1 bpmsystoles: 90 bpm
ResponseHR range: 80 bpm to 120 bpmHR change from 80Different
time of HeartRange : Within 11 sto 120 bpm:
Rate MeterHR range: 80 bpm to 40 bpmRange: [6.4 to 7.2
to Change inRange : Within 11 sseconds] Average:
HR6.8 seconds
HR change from 80
to 40 bpm:
Range: [5.6 to 6.4
sec] Average: 6.0
seconds
RESP Function
Different
RespirationSinusoid, 62.8 kHz(= 10%), <500 μASinusoidal signal,
excitation260 µA, 39 kHz
waveform
RRAdult: 0 rpm to120rpmAdult/pedi: 0 toDifferent

Comparison to Primary Predicate Device MP2

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

MeasuringRangePediatric/neonate: 0 rpm to 150rpm120 rpmNeo: 0 to 170 rpm
AccuracyAdult: 6 to 120 rpm: ±2 rpm, 0 to 5 rpm:not specifiedNeo/Ped: 6 to 150 rpm: ±2 rpm0 to 5 rpm: not specifiedat 0 to 120 rpm ±1 rpmat 120 to 170 rpm ±2 rpmDifferent
Resolution1 rpm1 rpmSame
Waveformbandwidth0.2Hz to 2.5Hz (-3dB)0.3 to 2.5 Hz (-6 dB)Different
Apnea AlarmTime Setup10s, 15s, 20s, 25s, 30s, 35s, 40s;10 to 40 seconds 5 second stepsDifferent
NIBP Function
Principle ofOperationoscillation(EDAN, SunTech NIBP)oscillationSame
AdultDifferent
EDANNIBP40 mmHg ~ 270 mmHg30 to 270 mmHg
SystolicSunTechNIBP60 mmHg ~ 250 mmHg
DiastolicEDANNIBP10 mmHg ~ 215 mmHg10 to 245 mmHg
SunTechNIBP30 mmHg ~ 190 mmHg
MeanEDANNIBP20 mmHg ~ 235 mmHg20 to 255 mmHg
SunTechNIBP40 mmHg ~ 210 mmHg
Measurement RangePediatricEDANNIBP40 mmHg ~ 230 mmHg30 to 180 mmHg (4 to 24 kPa)
SystolicSunTechNIBP40 mmHg ~ 230 mmHg
DiastolicEDANNIBP10 mmHg ~ 180 mmHg10 to 150 mmHg
SunTechNIBP20 mmHg ~ 160 mmHg
MeanEDANNIBP20 mmHg ~ 195 mmHg20 to 160 mmHg
SunTechNIBP30 mmHg ~ 175 mmHg
Neonate
SystolicEDANNIBP40 mmHg ~135 mmHg30 to 130 mmHg (4 to 17 kPa)
SunTech NIBP40 mmHg ~130mmHg
EDANNIBP10 mmHg ~ 100mmHg10 to 100 mmHg
DiastolicSunTech NIBP20 mmHg ~ 90mmHg
EDANNIBP20 mmHg ~ 110mmHg20 to 120 mmHg
MeanSunTech NIBP30 mmHg ~ 100mmHg
Max. Std.Deviation: 8 mmHg(1.1 kPa)Same
AccuracyMax standard deviation: 8mmHg(EDAN, SunTech);Max mean error: $\pm$ 5 mmHgMax. Mean Error:$\pm$ 5 mmHg ( $\pm$ 0.7kPa)
PressureResolution1mmHg(EDAN, SunTech);1mmHgSame
Maximum time:180 seconds(adult/pediatric)90 seconds(neonates)Different
MaximummeasuringperiodAdult/Pediatric 120sNeonate 90s
Adult297 $\pm$ 3mmHg(EDAN)<300mmHg(SunTech)>300 mmHg (40kPa) >2 secDifferent
OverpressureprotectionPediatric240 $\pm$ 3mmHg(EDAN)<300mmHg(SunTech)>300 mmHg (40kPa) >2 sec
Neonate147 $\pm$ 3mmHg(EDAN)<150mmHg(SunTech)>150 mmHg (20kPa) >2 sec
Measuringinterval inAUTO Mode1/2/3/4/5/10/15/30/60/90/120/240/480min(EDAN)1/2/3/4/5/10/15/30/60/90/120/240/480min(SunTech)1, 2, 2.5, 3, 5, 10,15, 20, 30, 45, 60or 120 minutesDifferent
Continuous5min, interval is 5s5 minutesSame
SpO2 Function (EDAN)
Measurement Range0-100%0 to 100 %Same
AccuracyAdult /Pediatric : $\pm$ 2 % (70% to 100%SpO2), Undefined (0 to 69% SpO2)Neonate : $\pm$ 3 % (70% to 100% SpO2),Undefined (0 to 69% SpO2)Philips ReusableSensors:M1191A,M1191AL,M1191ANL,M1191B,Different

{11}------------------------------------------------

{12}------------------------------------------------

M1192A,M1192AN = 2 %(70 % to 100 %)M1193A,M1193AN,M1194A,M1194AN,M1195A,M1195AN,M1196A = 3 %(70 % to 100 %)M1191T, M1192T,M1193T (Adult),M1196T = 3%(70% to 100%)M1193T (Neonate)= 4 % (70 % to100 %)
Resolution1 %1 %Same
Pulse Rate
MeasuringRange25 bpm to 300 bpm30 to 300 bpmDifferent
Accuracy±2bpm±2 % or 1 bpm,whichever isgreaterDifferent
Resolution1 bpm1 bpmSame
SpO2 Function (Nellcor)
Measurement Range1% to 100%1% to 100%Same
AccuracyDS-100A, OXI-A/N(Adult): ± 3% (70%to 100% SpO2)OXI-A/N(Neonate): ± 4% (70% to 100%SpO2)Philips ReusableSensors:M1193A,M1193AN,M1194A,M1194AN,M1195A,M1195AN,M1196A = 3 %(70 % to 100 %)M1191T, M1192T,M1193T (Adult),M1196T = 3%(70% to 100%)M1193T (Neonate)Same
= 4 % (70 % to100 %)
Resolution1 %1 %Same
Pulse Rate
MeasuringRange20bpm to 300bpm30 to 300 bpmDifferent
Accuracy± 3bpm (20bpm to 250bpm)±2 % or 1 bpm,which eve one isgreaterDifferent
Resolution1 bpm1 bpmSame
Temperature Function
Measurement Range0 to 50°C-1 to +49°C (30 to113°F )Different
Accuracy(notincludingsensor)±0.1 °C±0.1°C (±0.2°F )Same
Resolution0.1 °C0.1°C (±0.1°F)Same
AverageTimeConstant<30 sLess than 10secondsDifferent
IBP Function
MeasurementRange-50 to +300 mmHg-40 to 360 mmHgDifferent
Accuracy (notincludingsensor)± 2 % or ±1 mmHg, whichever isgreater4 % of reading or±4 mmHg (±0.5kPa), whichever isgreaterDifferent
Resolution1 mmHg1 mmHgSame
Transducer5 (μV/V/mmHg)300 to 3000 ΩLoadImpedance:200 to2000 ^ (resistive)OutputImpedance:"3000^ (resistive)Different
ZeroRange: ±200 mmHg±200 mmHg (±26kPa)Same
Electronic Safety
Anti-electroshock typeClass IIClass IISame
WorkingsystemContinuous operation equipmentContinuousoperationequipmentSame
Environmental Specifications
Temperature Range
Operating0°C to +40°C0 to 40°C (32 to104°F)Same
Storageincludingtransportation-30°C to +70°C-20 to 60°C (-4 to140°F)Different
Humidity Range
Operating15% to 95% (non-condensing)15 % to 95 %Relative Humidity(RH)Same
Storageincludingtransportation15% to 95% (non-condensing)5 % to 95 %Relative Humidity(RH)Different
Altitude Range
Operating615hPa to 1060hPa (DC-DCpowersupply)(-400 m to 4000 m)680hPa to 1060hPa (AC to DC powersupply)(-400m to 3000m)-500 m to 3000 m(10000 ft)Different
Storageincludingtransportation615hPa to 1060hPa(-400 m to 4000 m)-500 m to 4600m(15000 ft)Different
Physical Characteristics
WeightiM20 : <1.5kgEFM : < 0.58 kg1.25 kgDifferent
DimensionsiM20:185 mm (L) × 85.3 mm (W)× 116mm (H)EFM: 207 mm (L) × 93.4 mm (W)× 116mm (H)(W x H x D)188 x99 x 86 mm
Display Specification
Description5 inches LCD72 x 54 mm (2.8 x2.1 in)Different
Resolution800x480320 x 240
Power Supply
Powerconsumption<25w(Max)<12 W average<30 W whilebattery is loadingDifferent
Power input11.1V to 19.8VDC36 to 60 V DCfloating
Current1.27~2.3A0.7 to 0.4 A
AC-DC Power supply
Line Voltage100~240Vac100 to 240 V ~Same
OutputVoltage15V±5%dc, 24VA36 to 60 V DC floatingDifferent
Frequency50/60 Hz ~50/60 Hz ~Same
DC-DC Power supply
Line Voltage10~16VdcN/ADifferent
OutputVoltage15V±5%dcN/A
Standard Compliance
GeneralrequirementIEC 60601-1:2005IEC 60601-1:1988 + A1:1991 +A2:1995Different
IEC 60601-2-25: 2011IEC60601-2-25:1993 +A1:1999
IEC 60601-2-27: 2011IEC60601-2-27:2005
IEC 60601-2-51: 2003IEC60601-2-51:2003
AAMI EC11EC13:1991/2002.AAMI EC11EC13:1991/2002
SpecialrequirementIEC 80601-2-30: 2009IEC60601-2-30:1999
IEC 60601-2-34: 2011IEC60601-2-34:2000
ISO 80601-2-61: 2011ISO 9919:2005
EN 12470-4: 2000+ A1: 2009EN 12470-4:2000
ISO 80601-2-55: 2011EN ISO21647:2004+Cor.1:2005
ISO 80601-2-56: 2009EN 12470-4:2000
ISO 81060-2: 2009--
IEC 60601-2-49: 2011IEC60601-2-49:2001).
EMCIEC 60601-1-2:2007IEC60601-1-2:2001 +A1:2004.
AlarmIEC 60601-1-8:2006IEC60601-1-8:2003
BiocompatibilityCompliance with ISO10993-1 ISO10993-5,ISO 10993-5-10Compliance withISO10993-1 ISO10993-5,ISO
Out-Of-Hospital Transport - Standards Compliance10993-5-10
Out-Of-Hospital Transport - Standards Compliance
Shock TestsAccording to IEC TR 60721-4-7, Class7M3. Test procedure according toIEC/EN60068-2-27 (peak accelerationup to 100g).According to IECTR 60721-4-7,Class 7M3. Testprocedureaccording toIEC/EN60068-2-27 (peakacceleration up to100g).Same
RandomVibrationAccording to IEC TR 60721-4-7, Class7M3. Test procedure according toIEC/EN 60068-2-64 (RMS acceleration5g).According to IECTR 60721-4-7,Class 7M3. Testprocedureaccording toIEC/EN60068-2-64 (RMSacceleration 5g).Same
SinusoidalVibrationAccording to EN1789:2007+A: 2010,6.4.1 vibration test, Test procedureaccording to IEC/EN 60068-2-6(acceleration up to amplitude 2g).According to IECTR 60721-4-7,Class 7M3. Testprocedureaccording toIEC/EN 60068-2-6(acceleration up toamplitude 2g).Different
BumpAccording to EN1789:2007+A: 2010,6.4.1 Bump test, Test procedureaccording to IEC/EN 60068-2-6(acceleration up to amplitude 15g).--Different
Drop Testaccording to EN1789:2007+A:2010,Test Procedure according to EN60068-2-32 (height 1.2 m).According toEN1789 (coversalso IEC TR60721-4-7 andClass 7M3). Testprocedureaccording to EN60068-2-32 (height0.75 m).Different
Degrees ofProtectionprovided by enclosures according toIEC/EN 60529: IP44provided byenclosuresaccording toDifferent
IEC/EN 60529:IP32
MedicalVehicles andEquipmentsEN1789:2007+A: 2010 Medicalvehicles and their equipment - Roadambulances (chapter 6 – MedicalDevices).EN 1789 +A1:2003Medical vehiclesand their equipment- Road ambulances(chapter 6 -Medical Devices).Same
RadiatedSusceptibility20 V/m according to EN ISO 9919(SpO2) and EN ISO 21647 (CO2).20 V/m accordingto EN ISO 9919(SpO2) and ENISO 21647 (CO2).Same
Alarm System
AlarmCategoriesPhysiological alarmsTechnical alarmsPrompt informationPhysiologicalalarmsTechnical alarmsPrompt informationSame
Alarm LevelsHigh level alarmsMedium level alarmsLow level alarmsHigh level alarmsMedium levelalarmsLow level alarmsSame
Alarm ModesVisual alarmsAudible alarmsAlarm messagesParameter flashesVisual alarmsAudible alarmsAlarm messagesParameter flashesSame

{13}------------------------------------------------

{14}------------------------------------------------

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

Comparison to Reference Device iM70

ItemProposed device, iM20Predicate device,Comparison
K#Current SubmissioniM70Result
K131971
PR from NIBP
Measurementrange40 to 240 bpm(EDAN)30 bpm ~220bpm(SunTech)40 to 240 bpmSameDifferent
Accuracy±3bpm or 3.5%, whichever isgreater(EDAN);±3bpm or ±2%, whichever isgreater(SunTech);±3bpm or 3.5%, whichever isgreaterSameDifferent
CO2 Function
Measuring Range
EtCO20 mmHg ~ 150 mmHg0 to 150 mmHg (0 to 20.0 kPa)Same
FiCO23 mmHg~50 mmHg3 mmHg to 50 mmHgSame
AwRR2 rpm ~ 150 rpm (Sidestream)2 to 150 rpmSame
Resolution
EtCO21mmHg1.0 mmHg (0.1 kPa)Same
FiCO21mmHg1mmHgSame
AwRR1 rpm±1 rpmSame
EtCO2 Accuracy
0 to 40 mmHg± 2 mmHg±2.0 mmHg (±0.29 kPa)Same
41 to 70 mmHg± 5 %±5 % of readingSame
71 to 100 mmHg± 8 %±8 % of readingSame
101 to 150 mmHg± 10 %±10 % of readingSame
RESP measurement value exceeds 80rpm (sidestream)± 12 %±12 % of actualSame
AwRR Accuracy± 1 rpm± 1 rpmSame
Sample Gas Flow rate (sidestream)50±10 ml/min50±10ml/minSame
O2 Compensation
Range0~100%0 ~ 100%Same
Resolution1%1%Same
Default16%16%Same
Barometric pressure compensationUser setupUser setupSame
Anesthetic Gas Compensation
Range0~20%0 ~ 20%Same
Resolution0.1%0.1%Same
Default0.0%0.0%Same
Balance Gas CompensationRoom air, N2O, heliumRoom air, N2O, heliumSame
Stability
Short TermDrift over 4 hours < 0.8 mmHgDrift over 4 hoursSame
Drift< 0.8 mmHg
Long Term Drift120 hours120 hoursSame
Total System Response Time4.7 s4.7 sSame
Alarm TypeEtCO2, FiCO2, AwRREtCO2, FiCO2, AwRRSame
Apnea Alarm Delay10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.Same
Data Sample Rate100 Hz100 HzSame
Sensor Response Time (sidestream)< 3 seconds, including transport time and rise time< 3 seconds, including transport time and rise timeSame
Interfering Gas and Vapor Effects on EtCO2 Measurement Values
Nitrous oxide 60%Dry and Saturated Gas0 – 40 mmHg: ± 1 mmHg additional errorDry and Saturated Gas0 – 40 mmHg: ± 1 mmHg additional errorSame
Halothane 4%41 – 70 mmHg: ± 2.5% additional error41 – 70 mmHg: ± 2.5% additional error
Desflurane 5%71 – 100 mmHg: ± 4% additional error71 – 100 mmHg: ± 4% additional error
Enflurane 5%101 – 150 mmHg: ± 5% additional error*Additional worst case error when compensation for PB, O2, N2O, anesthetic agents, or helium is correctly selected for the actual fractional gas constituents present.101 – 150 mmHg: ± 5% additional error*Additional worst case error when compensation for PB, O2, N2O, anesthetic agents, or helium is correctly selected for the actual fractional gas constituents
Isoflurane 5%
Sevoflurane 80%Desflurane:The presence of desflurane in the exhaled breath at concentrations greater than 5% will positively bias Carbon Dioxide values by up to an additional 3 mmHg at 38mmHg.
Xenon 50%Xenon:The presence of Xenon in the exhaled breath will negatively bias Carbon Dioxide values by up to an additional 5 mmHg at 38mmHg.
Helium 15%
Desflurane:The presence ofdesflurane in theexhaled breath atconcentrationsgreater than 5%will positively biasCarbon Dioxidevalues by up to anadditional 3mmHg at38mmHg.Xenon:The presence ofXenon in theexhaled breath willnegatively biasCarbon Dioxidevalues by up to anadditional 5mmHg at38mmHg.
Barometric Pressure on EtCO2 Measurement Values
0 – 40 mmHg$\pm$ 1 mmHg additional error$\pm$ 1 mmHgadditional errorSame
41 – 70mmHg$\pm$ 2.5% additional error$\pm$ 2.5% additionalerrorSame
71 – 100mmHg$\pm$ 4% additional error$\pm$ 4% additionalerrorSame
101 - 150mmHg$\pm$ 5% additional error$\pm$ 5% additionalerrorSame
Wi-Fi
IEEE802.11b/g/n802.11b/g/nSame
FrequencyBand2.4GHz ISM band2.4GHz ISM bandSame
ModulationOFDM with BPSK, QPSK, 16-QAM,and 64-QAM 802.11b with CCK andDSSSOFDM withBPSK, QPSK,16-QAM, and64-QAM802.11b with CCKand DSSSSame
TypicalTransmitPower ( $\pm$ 217 dBm for 802.11b DSSS17 dBm for 802.11b CCK15 dBm for 802.11g/n OFDM17 dBm for802.11b DSSS17 dBm forSame
dBm)802.11b CCK15 dBm for802.11g/n OFDM

{18}------------------------------------------------

{19}------------------------------------------------

{20}------------------------------------------------

{21}------------------------------------------------

As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The main technological differences between the subject and predicate devices are minor differences, and do not raise different questions of safety or effectiveness. As demonstrated in the non-clinical and clinical testing, the different technological characteristics do not affect the safety and effectiveness of the Edan iM20 system.

10. Performance Data:

Non-clinical data:

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the iM20 Patient Monitor were conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours. And the battery of testing included the following tests:

  • □ Cytotoxicity
  • □ Skin Sensitization
  • □ Skin Irritation

Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the iM20 Patient Monitor device, consisting of all the modules and accessories in the system complies with the IEC 60601-1:2005/A1: 2012 for safety and the IEC 60601-1-2:2007 standard for EMC.

Bench Testing

Bench testing was conducted on the iM20 Patient Monitor device, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1-8: 2006, IEC 60601-2-25: 2011, IEC 60601-2-27: 2011, IEC 80601-2-30: 2009, IEC 60601-2-34: 2011, IEC 60601-2-49: 2011, ISO 80601-2-55: 2011, ISO 80601-2-56: 2009 and ISO 80601-2-61: 2011 standards for performance effectiveness.

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance

{22}------------------------------------------------

for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

Clinical data:

Clinical testing required by FDA specific guidance on Non-Invasive Blood Pressure (NIBP) Monitor is conducted per ISO 81060-2: 2013 Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type.

Summary

Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device.

11. Conclusion

The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the iM20 Patient Monitor device should perform as intended in the specified use conditions. The clinical data demonstrate that the subject devices perform comparably to the predicate device that is currently marketed for the same intended use. In other words, the subject the iM20 Patient Monitor devices are substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.