(67 days)
The Biomet JuggerLoc™ Bone system is intended for repair in the foot and ankle including indications for: Midfoot repair including but not limited to Lisfranc repair, ankle syndesmosis fixation (syndesmosis disruptions), and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures.
The new JuggerLoc Bone to Bone system is a suture based construct intended to stabilize fractures. The construct consists of a JuggerKnot Soft Anchor with ZipLoop Technology, featuring combinations of suture and metal button fixation.
The provided text describes a medical device called the "JuggerLoc Bone to Bone System," a suture-based construct intended to stabilize fractures. The submission is a 510(k) for a new device construct, claiming substantial equivalence to predicate devices (K130033 - Biomet ToggleLoc System, K110145 - Biomet JuggerKnot Soft Anchors, K133275 - Arthrex Mini-Tightrope).
Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative or numerical format (e.g., a specific tensile strength in Newtons, or a particular displacement under load). Instead, it refers to "Preclinical performance tests are provided to address the subject construct's strength." The general acceptance is that "Results indicate that the subject construct is substantially equivalent to legally marketed devices."
Therefore, a table cannot be fully constructed with explicit numerical criteria. However, based on the text, the performance goal is implicitly to meet or exceed the performance of the predicate device(s) in terms of strength.
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Demonstrated strength and performance similar to legally marketed predicate devices. | Preclinical performance tests were conducted to address the subject construct's strength. Results indicate that the subject construct is substantially equivalent to legally marketed devices. The technological characteristics (design, dimensions, function, materials) are similar. |
| Biocompatibility | A biocompatibility rationale was provided. |
| No new concerns of safety or efficacy compared to predicate devices. | "No new concerns of safety or efficacy have been raised." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "Preclinical performance tests" but does not specify the sample size used for these tests. It also does not specify the data provenance beyond stating they are "Preclinical performance tests," implying laboratory testing rather than human subject data. There is no information on country of origin or whether it was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable as the study described is a preclinical performance test (likely mechanical and material testing), not a study requiring human expert assessment of a test set, like in imaging or diagnostic device evaluations. There is no "ground truth" to be established by experts in the context of this submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable for the same reasons as point 3. No human expert adjudication was involved in these preclinical performance tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices involving human interpretation, not for mechanical fixation systems like the JuggerLoc Bone to Bone System. There is no mention of AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. The device is a physical medical implant, not an algorithm or software. Therefore, an algorithm-only standalone performance evaluation is not relevant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the preclinical performance tests, the "ground truth" would be objective physical measurements of the device's mechanical properties (e.g., tensile strength, displacement, fatigue resistance), compared against established industry standards or the performance of predicate devices. The document does not specify the exact parameters measured, but implies mechanical strength testing as the basis for evaluation.
8. The sample size for the training set
This is not applicable. There is no mention of a "training set" as this device is a physical construct, not a learning algorithm or AI.
9. How the ground truth for the training set was established
This is not applicable for the same reason as point 8.
{0}------------------------------------------------
JUL 1 8 2014
Section 5 - 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter Information | |
|---|---|
| Name | Biomet Manufacturing Corp. |
| Address | 56 East Bell DriveP.O. Box 587Warsaw, IN 46581 |
| Phone number | (574) 372-1941 |
| Fax number | (574) 371-1027 |
| Establishment | 1825034 |
| Registration Number | |
| Name of contact person | Jared CooperRegulatory Affairs Specialist |
| Date prepared | 29 April 2014 |
| Name of device | |
| Trade or proprietary name | JuggerLoc Bone to Bone System |
| Common or usual name | Bone Fixation Anchor |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories. |
| Classification panel | Orthopedics |
| Regulation | 21 CFR § 888.3030 |
| Product Code(s) | HTN - washer, bolt nut |
| Legally marketed device(s) to which equivalence is claimed | K130033 - Biomet ToggleLoc SystemK110145 - Biomet JuggerKnot Soft AnchorsK133275 - Arthrex Mini-Tightrope |
| Reason for 510(k) submission | New Device Construct |
| Device description | The new JuggerLoc Bone to Bone system is a suture based construct intended to stabilize fractures. The construct consists of a JuggerKnot Soft Anchor with ZipLoop Technology, featuring combinations of suture and metal button fixation. |
| Indications for use | The JuggerLoc Bone to Bone System is intended for repair in the foot and ankle including indications for:Midfoot repair including but not limited to Lisfranc repair, ankle syndesmosis fixation (syndesmosis disruptions), and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures. |
| Summary of the Technologies |
The technological characteristics of the JuggerLoc Bone to Bone System are similar to the predicate devices including design, dimensions, function, and materials. The JuggerLoc design differs from the predicate devices by having one soft anchor and one metal button to achieve fixation rather than two metal buttons.
{1}------------------------------------------------
PERFORMANCE DATA
SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Performance Test Summary - New Device
The JuggerLoc Bone to Bone System components have the same basic technological and material characteristics as the predicate devices except for slight modifications to the general design as described in this 510(k) notification. Preclinical performance tests are provided to address the subject construct's strength. A biocompatibility rationale was also provided. Results indicate that the subject construct is substantially equivalent to legally marketed devices.
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical Performance Data/Information: None provided as a basis for substantial equivalence.
CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The results of preclinical tests and the similarities with legally marketed devices indicate the JuggerLoc Bone to Bone System is substantially equivalent to currently marketed devices. No new concerns of safety or efficacy have been raised.
{2}------------------------------------------------
Public Health Service
Food and Drue Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2014
Biomet Manufacturing Corp. Dr. Jared Cooper Regulatory Affairs Specialist 56 East Bell Drive. PO Box 587 Warsaw, Indiana 46581
Re: K141219
Trade/Device Name: JuggerLoc TM Bone to Bone System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN Dated: May 23, 2014 Received: May 27, 2014
Dear Dr. Cooper:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{3}------------------------------------------------
Page 2 - Dr. Jared Cooper
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.hun for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Lori A. Wiaqins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Section 4 - Indications for Use Statement
K141219 510(k) Number:
JuggerLoc™ Bone to Bone System Device Name:
Indications For Use:
The Biomet JuggerLoc™ Bone system is intended for repair in the foot and ankle including indications for:
Midfoot repair including but not limited to Lisfranc repair, ankle syndesmosis fixation (syndesmosis disruptions), and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures.
× Prescription Use_ (Per 21 CFR 801 Subpart D) AND/OR Over-the-Counter_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth 臨歐ank -S
Division of Orthopedic Devices
N/A