AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ToogleLoc™ System devices, except the TogaleLoc XTender and TogaleLoc XL devices, are intended for soft tissue to bone fixation for the following indications:

Shoulder: Bankart lesion repair, SLAP lesion repairs Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps Tenodesis

Foot and Ankle: Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair, Ankle Syndesmosis fixation (Syndesmosis disruptions) and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures (only for ToggleLoc™ with ZipTight™)

Elbow: Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment

Knee: ACL/PCL repair / reconstruction, ACL/PCL patellar bone-tendon-bone grafts, Double-Tunnel ACL reconstruction, Extracapsular repair: MCL, LCL, and posterior oblique ligament, Iliotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure

Hand and Wrist: Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction

Hip: Acetabular labral repair

The ToggleLoc XTender and ToggleLoc XL devices are used for fixation of tendons and ligaments in cases of unanticipated intraoperative complications such as cortical breaching during orthopedic reconstruction procedures, such as Anterior Cruciate (ACL) or Posterior Cruciate (PCL) Reconstruction. The ToggleLoc™ XTender is for use in conjunction with a titanium alloy ToggleLoc™ device, excluding the ToggleLoc™ XL devices.

Device Description

This submission is for the ToggleLoc™ System in which a series of titanium and stainless steel ToggleLoc devices are being included. These items include the ToggleLoc™ XTender, ToggleLoc™ XL with ZipLoop™ Technology, ToggleLoc™ with ZipLoop™ Inline, expanded offering of components of the previously cleared ToggleLoc™ with ZipLoop™ System (K083070), and implants previously cleared in K083070 being included in this submission to add MRI language.

The ToggleLoc™ System is made up of various combinations of components, including: ToggleLoc™ buttons, ZipTight construct and buttons, Continuous Loops, and ZipLoop™ Technology.

AI/ML Overview

The acceptance criteria and study proving the device meets them are summarized below based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance:

Test TypeAcceptance CriteriaReported Device Performance
Static Load TestingNot explicitly stated, implied to meet predicate devices."met or exceeded the established acceptance criteria"
Cyclic Load TestingNot explicitly stated, implied to meet predicate devices."met or exceeded the established acceptance criteria"
MR Evaluation/SimulationsNot explicitly stated, implied to meet predicate devices."indicated that the devices are MR Conditional."

2. Sample size used for the test set and data provenance:

  • Sample Size: Not specified. The document states "The following tests were performed on the ToggleLoc™ System," implying testing on the new metallic ToggleLoc™ buttons and ZipLoop™ constructs.
  • Data Provenance: Not specified; however, the testing is non-clinical performance data, likely conducted in a laboratory setting by the manufacturer, Biomet Manufacturing Corp.

3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

This information is not applicable as the study is a non-clinical (mechanical and MR) performance study, not a study involving human interpretation of data where expert ground truth would be required.

4. Adjudication method for the test set:

This information is not applicable for a non-clinical performance study.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. This is a non-clinical performance study for a device (soft tissue fixation system), not an AI-powered diagnostic device.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable. This is a non-clinical performance study for a medical device (soft tissue fixation system), not an algorithm.

7. The type of ground truth used:

For the mechanical tests (Static and Cyclic Load Testing), the "ground truth" would be the established engineering specifications or performance characteristics of the predicate devices that the new devices aim to be substantially equivalent to.
For the MR Evaluation/Simulations, the "ground truth" would be the established industry standards and scientific principles for MR safety and compatibility.

8. The sample size for the training set:

This information is not applicable as this is a non-clinical performance study for a medical device, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established:

This information is not applicable for the same reason as above.

Study Description:

The study conducted was a series of non-clinical performance tests designed to demonstrate the substantial equivalence of the new ToggleLoc™ System components (additional metallic ToggleLoc™ buttons and ZipLoop™ constructs) to legally marketed predicate devices (K083070 ToggleLoc™ System, K102982 EndoButton CL, K112990 TightRope RT).

The tests performed included:

  • Static Load Testing: To evaluate the strength and structural integrity of the new line extensions under static conditions.
  • Cyclic Load Testing: To assess the fatigue resistance and durability of the devices under repeated loading, simulating physiological conditions.
  • MR Evaluation/Simulations: To determine the magnetic resonance (MR) compatibility of the devices, particularly to support the addition of MR compatibility language to the product labeling.

The document explicitly states that "All testing met or exceeded the established acceptance criteria," and "The results of mechanical testing indicated the devices performed mechanically equivalent to marketed devices and did not raise any new safety and efficacy issues. The results of the MR evaluation indicated that the devices are MR Conditional."

No clinical testing was necessary for the determination of substantial equivalence.

{0}------------------------------------------------

K130033

510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92

Submitter Information
NameBiomet Manufacturing Corp.
Address56 East Bell DriveWarsaw, IN 46581-0857
Phone number(574) 267-6639
Fax number(574) 371-1027
Establishment1825034
Registration Number
Name of contact personElizabeth Wray
Date preparedJune 21, 2013
Name of device
Trade or proprietarynameToggleLoc™ System
Common or usualnameSoft tissue fixation devices
Classification name- fastener, fixation, nondegradable, soft tissue- staple, fixation, bone
Classification panelOrthopedic
Regulation888.3040 and 888.3030
Product Code(s)MBI and JDR
Legally marketed device(s)to which equivalence isclaimedK083070 ToggleLoc™ System (Biomet)K102982 EndoButton CL (Smith & Nephew)K112990 TightRope RT (Arthrex)
Reason for 510(k)submissionAdditional metallic ToggleLoc™ buttons and ZipLoop™ constructs.Addition of magnetic resonance (MR) compatibility language toproduct labeling.
Device descriptionThis submission is for the ToggleLoc™ System in which a seriesof titanium and stainless steel ToggleLoc devices are beingincluded. These items include the ToggleLoc™ XTender,ToggleLoc™ XL with ZipLoop™ Technology, ToggleLoc™ withZipLoop™ Inline, expanded offering of components of thepreviously cleared ToggleLoc™ with ZipLoop™ System(K083070), and implants previously cleared in K083070 beingincluded in this submission to add MRI language.
Intended use of the deviceSoft tissue fixation
K130033
Indications for useThe ToggleLoc™ System devices, except the ToggleLoc XTenderand ToggleLoc XL devices, are intended for soft tissue to bonefixation for the following indications:
Shoulder: Bankart lesion repair, SLAP lesion repairsAcromio-clavicular repair, Capsular shift/capsulolabralreconstruction, Deltoid repair, Rotator cuff tear repair,Biceps Tenodesis
Foot and Ankle: Medial/lateral repair andreconstruction, Mid- and forefoot repair, Hallux valgusreconstruction, Metatarsal ligament/tendon repair orreconstruction, Achilles tendon repair, AnkleSyndesmosis fixation (Syndesmosis disruptions) and asan adjunct in connection with trauma hardware forWeber B and C ankle fractures ( only for ToggleLoc™with ZipTight™ )
Elbow: Ulnar or radial collateral ligamentreconstruction, Lateral epicondylitis repair, Bicepstendon reattachment
Knee: ACL/PCL repair / reconstruction, ACL/PCL patellarbone-tendon-bone grafts,Double-Tunnel ACL reconstruction, Extracapsular repair:MCL, LCL, and posterior oblique ligament, Iliotibial bandtenodesis, Patellar tendon repair, VMO advancement,Joint capsule closure
Hand and Wrist: Collateral ligament repair,Scapholunate ligament reconstruction, Tendon transfersin phalanx, Volar plate reconstruction
Hip: Acetabular labral repair
The ToggleLoc XTender and ToggleLoc XL devices are used forfixation of tendons and ligaments in cases of unanticipatedintraoperative complications such as cortical breaching duringorthopedic reconstruction procedures, such as Anterior Cruciate(ACL) or Posterior Cruciate (PCL) Reconstruction. TheToggleLoc™ XTender is for use in conjunction with a titaniumalloy ToggleLoc™ device, excluding the ToggleLoc™ XL devices.
Summary of the TechnologiesThe ToggleLoc™ System is made up of various combinations of components, including: ToggleLoc™

{1}------------------------------------------------

buttons, ZipTight construct and buttons, Continuous Loops, and ZipLoop™ Technology.

The technological characteristics of the ToggleLoc™ System are the same as those of predicate

{2}------------------------------------------------

K130033

devices (K083070) in terms of design, material, and principles of operation with the exception of slight modifications as described in this 510(k). The ToggleLoc™ System utilizes the identical manufacturing processes as the predicates (K083070).

Non-clinical testing was conducted to demonstrate that the differences did not adversely affect safety and efficacy, and to demonstrate substantial equivalence to the predicate components. All testing met or exceeded the established acceptance criteria. This information is detailed below in the Performance (Non-clinical) section.

PERFORMANCE DATA

SUMMARY OF NON-CLINICAL TESTS

Performance Test Summary-New Device

The following tests were performed on the ToggleLoc™ System:

Static Load Testing: ToggleLoc™ System Line Extensions Cyclic Load Testing: ToggleLoc™ System Line Extensions MR Evaluation/Simulations

SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION

Clinical Performance Data/Information: N/A

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

No clinical testing was necessary for a determination of substantial equivalence.

The results of mechanical testing indicated the devices performed mechanically equivalent to marketed devices and did not raise any new safety and efficacy issues. The results of the MR evaluation indicated that the devices are MR Conditional.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract shapes, possibly representing human figures or abstract forms, arranged in a dynamic, flowing manner.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Biomet Manufacturing Corporation % Ms. Elizabeth Wray Global Regulatory Project Manager 56 East Bell Drive Warsaw, Indiana 46581

August 1. 2013

Re: K130033

Trade/Device Name: Toggleloc™ System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, JDR Dated: June 28, 2013 Received: July 3, 2013

Dear Ms. Wray:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Elizabeth Wray

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, plcase note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincercly yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: Toggleloc™ System

Indications For Use:

The ToogleLoc™ System devices, except the TogaleLoc XTender and TogaleLoc XL devices, are intended for soft tissue to bone fixation for the following indications:

Shoulder

Bankart lesion repair SLAP lesion repairs Acromio-clavicular repair Capsular shift/capsulolabral reconstruction Deltoid repair Rotator cuff tear repair Biceps Tenodesis

Foot and Ankle

Medial/lateral repair and reconstruction Mid- and forefoot repair Hallux valqus reconstruction Metatarsal ligament/tendon repair or reconstruction Achilles tendon repair Ankle Syndesmosis fixation (Syndesmosis disruptions) and as an adjunct in connection with trauma hardware for Weber B and C ankle fractures (only for ToggleLoc™ with ZipTight™)

Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR -

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.
Division of Orthopedic Devices

Page 1 of 3

{6}------------------------------------------------

K130033 510(k) Number (if known):____

Device Name: Toggleloc™ System

Indications For Use:

Elbow

Ulnar or radial collateral ligament reconstruction Lateral epicondylitis repair Biceps tendon reattachment

Knee

ACL/PCL repair / reconstruction ACL/PCL patellar bone-tendon-bone grafts Double-Tunnel ACL reconstruction Extracapsular repair: MCL, LCL, and posterior oblique ligament Iliotibial band tenodesis Patellar tendon repair VMO advancement Joint capsule closure

Hand and Wrist

Collateral ligament repair Scapholunate ligament reconstruction Tendon transfers in phalanx Volar plate reconstruction

Hip

Acetabular labral repair

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.


Division of Orthopedic Devices

Page 2 of 3

510(k) Number (if known): _

{7}------------------------------------------------

K130033

Device Name: Toggleloc™ System

Indications For Use:

The ToggleLoc XTender device is used for fixation of tendons and ligaments in cases of unanticipated intraoperative complications such as cortical breaching during orthopedic reconstruction procedures, such as Anterior Cruciate (ACL) or Posterior Cruciate (PCL) Reconstruction. The ToggleLoc™ XTender is for use in conjunction with a titanium alloy ToggleLoc™ device, excluding the ToggleLoc™ XL devices.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey, L. Hanley, Ph.D.

Page 3 of 3

Division of Orthopedic Devices .. ..

N/A