(45 days)
The JuggerKnot™ Soft Anchor is intended to be used for soft tissue to bone fixation with indications for use in: Shoulder, Foot and Ankle, Elbow, Knee, Hand and Wrist, Hip.
The JuggerKnot™ Soft Anchors consist of a coreless sleeve structure and suture. The anchors are intended for use in soft tissue fixation by bunching against bone when deployed.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the JuggerKnot™ Soft Anchors:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Equivalent fixation strength to predicate devices | "The test results indicate that the Biomet Sports Medicine JuggerKnot™ Soft Anchors provide equivalent fixation strength to the predicate devices and would be functional within their intended use." |
Explanation of "Acceptance Criteria (Implicit)": The document doesn't explicitly state quantitative acceptance criteria (e.g., "must meet 90% of predicate device strength"). Instead, the acceptance criterion for the non-clinical testing is implied to be "equivalent fixation strength" when compared to the specified predicate devices.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated. The document refers to "mechanical pullout testing" and "test results" but does not provide details on the number of samples tested.
- Data Provenance: The testing was "Non-clinical laboratory testing" performed by Biomet Sports Medicine. The country of origin is not specified, but the company's address is in Warsaw, Indiana, USA, suggesting the testing likely occurred in the US or under US guidance. The data is entirely retrospective as it was conducted before the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable. The ground truth for this device's performance was established via "mechanical pullout testing," not expert opinion on diagnostic images or clinical outcomes.
- Qualifications of Experts: Not applicable for mechanical testing. The expertise would lie with engineers or technicians performing and analyzing the mechanical tests.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. As the testing was mechanical pullout testing, the "ground truth" was derived directly from the physical measurements and not through an adjudication process involving human reviewers.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done: No. The document explicitly states: "Clinical Testing: None provided as a basis for substantial equivalence." The evaluation was based solely on non-clinical mechanical testing comparisons to predicate devices.
- Effect Size of Human Readers Improvement: Not applicable, as no MRMC study was conducted.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Standalone Study Done: Yes, in a sense. The non-clinical mechanical testing represents a "standalone" evaluation of the device's physical performance without human intervention in its efficacy measurement. The device itself is a physical implant, not an algorithm. The "performance" being evaluated is its structural integrity and fixation strength.
7. Type of Ground Truth Used
- Ground Truth Type: Objective physical measurements from "mechanical pullout testing." The "truth" in this context is the quantitative force required to pull out the anchors, measured under controlled laboratory conditions, which formed the basis for comparison to predicate devices.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This device is a physical medical device, not an AI/ML algorithm. There is no concept of a "training set" in the context of its development and validation for regulatory submission.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text starts with "K110145" on the first line, followed by "p1/2" on the second line. The handwriting is somewhat uneven, and the characters are spaced irregularly.
Image /page/0/Picture/1 description: The image shows the logo for Biomet Sports Medicine. The logo is in black and white and features the word "BIOMET" in a stylized font. Below the word "BIOMET" is the phrase "SPORTS MEDICINE" in a smaller, sans-serif font. The logo is simple and professional, and it is likely used to represent the company's brand.
MAR - 4 2011
510(k) Summary
| Preparation Date: | January 17, 2011 |
|---|---|
| Applicant/Sponsor: | Biomet Sports Medicine |
| Contact Person: | Elizabeth Wray / Regulatory Project ManagerVictor Rodgers / Director of Quality, Clinical, & Regulatory Affairs |
| Proprietary Name: | JuggerKnot™ Soft Anchors |
| Common Name: | Soft Tissue Fixation Device |
| Classification Name: | Fastener, fixation, nondegradable, soft tissue(21CFR §888.3040) MBI |
Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
K071704 Sleeve and Button Soft Tissue Fixation Devices / Biomet Sports Medicine
Device Description:
The JuggerKnot™ Soft Anchors consist of a coreless sleeve structure and suture. The anchors are intended for use in soft tissue fixation by bunching against bone when deployed.
Intended Use / Indications for Use:
The JuggerKnot™ Soft Anchor is intended to be used for soft tissue to bone fixation with indications for use in:
Shoulder
Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps tenodesis
Foot and Ankle
Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair
Elbow
Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment
Knee
Extra-capsular repair: MCL, LCL, and posterior oblique ligament Iliotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure
Mailing Address:
P.O. Box 587
Warsaw, IN 46581-0587
Toll Free: 800.348.9500
Office: 574.267.6639
Main Fax: 574.267.8137
www.biomet.com
Shipping Address:
56 East Bell Drive
Warsaw, IN 46582
{1}------------------------------------------------
K110145
pH2
Hand and Wrist
Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction
Hip
Acetabular labral repair
Summary of Technologies:
The technological characteristics (materials, design, sizing and indications) of the JuggerKnot™ Soft Anchors are similar or identical to the predicate devices or other previously cleared devices.
Non-Clinical Testing:
Non-clinical laboratory testing was performed to verify the fixation strength of the JuggerKnot™ Soft Anchors in mechanical pullout testing as compared to the predicate devices for specific indications for use. The efficacy of the JuggerKnot™ Soft Anchors was compared to that of the Biomet Sports Medicine Sleeve and Button Soft Tissue Fixation Devices. The test results indicate that the Biomet Sports Medicine JuggerKnot™ Soft Anchors provide equivalent fixation strength to the predicate devices and would be functional within their intended use.
Clinical Testing:
None provided as a basis for substantial equivalence.
All trademarks are the property of Biomet, Inc.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the wings and body of the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet, Inc. % Ms. Elizabeth Wray Regulatory Project Manager 56 East Bell Drive - P.O. Box 587 Warsaw, Indiana 46581-0587
MAR - 4 2011
Re: K110145
Trade/Device Name: JuggerKnot" Soft Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: February 11, 2011 Received: February 14, 2011
Dear Ms. Wray:
We have reviewed your Section 510(k) premarket notification of-intent to-market the device referenced above ând have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Elizabeth Wray
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
/www.fda.gov/MedicalDevices/ResourcesforYouIndustry/ucm.htm
Sincerely yours,
Aliz B. nute
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
KI10145 510(k) Number (if known):
Device Name: JuggerKnot™ Soft Anchors
Indications For Use:
The JuggerKnot™ Soft Anchors are intended for soft tissue to bone fixation for the following indications:
Shoulder
Bankart lesion repair, SLAP lesion repair, Acromio-clavicular repair, Capsular shift/capsulolabral reconstruction, Deltoid repair, Rotator cuff tear repair, Biceps tenodesis
Foot and Ankle
Medial/lateral repair and reconstruction, Mid- and forefoot repair, Hallux valgus reconstruction, Metatarsal ligament/tendon repair or reconstruction, Achilles tendon repair
Elbow
Ulnar or radial collateral ligament reconstruction, Lateral epicondylitis repair, Biceps tendon reattachment
Knee
Extra-capsular repair: MCL, LCL, and posterior oblique ligament Iliotibial band tenodesis, Patellar tendon repair, VMO advancement, Joint capsule closure
Hand and Wrist
Collateral ligament repair, Scapholunate ligament reconstruction, Tendon transfers in phalanx, Volar plate reconstruction
Hip
Acetabular labral repair
Prescription Use_ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Melkerson
Page 1 of 1
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
K110145
510(k) Number
N/A