(225 days)
As prescribed by a physician for the following:
TENS- Transcutaneous Nerve Stimulation
- Symptomatic relief of chronic intractable pain .
- . Post traumatic and post surgical pain relief
EMS- Electrical Muscle Stimulation
- . Relaxation of muscle spasm
- . Increasing local blood circulation
- . Muscle re-education
- . Prevention or retardation of disuse atrophy
- Prevention of venous thrombosis of the calf muscles immediately after surgery
- . Maintaining or increase range of motion
The Flex-MT + is a combination TENS and EMS device which delivers nerve or muscle stimulation by applying an electrical current to electrodes, which are attached on the patient's skin. The output and waveform is adjustable according to the intended treatment of patient. The stimulator has 2 output channels, accessed through jacks at the top of the housing, so that it may stimulate either 2 or 4 patient electrodes. The device is powered by 700 mAh 4.8V Ni MH rechargeable battery pack. A patient compliance timer can memorize 60 sets of operation records; the total recordable time is 999 hours.
The provided text is a 510(k) summary for the Flex-MT+ device. It outlines the device's technical specifications and compares them to predicate devices to establish substantial equivalence.
Here's an analysis of the acceptance criteria and the "study" (in this case, performance testing) that proves the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for performance, but rather demonstrates substantial equivalence by comparing the Flex-MT+ device's technical characteristics and performance to two predicate devices (K083030 EMSI TENS-EMS-14 (Flex MT) and K021100 EMPI 300 PV). The "Comment" column in the provided tables effectively acts as the "acceptance criteria" by stating whether the subject device's performance is comparable to or within acceptable limits of the predicate devices and relevant standards.
Acceptance Criteria (Derived) and Reported Device Performance (Flex-MT+)
| Characteristic | Acceptance Criteria (Derived from Predicates & Standards) | Reported Device Performance (Flex-MT+) | Fulfilled? |
|---|---|---|---|
| Indications for Use | Same as predicate TENS and NMES modes. | Same as K083030 EMSI TENS-EMS-14 (Flex MT) and K021100 EMPI 300 PV | Yes |
| Power Source | Battery operated; battery charger available (consistent with EMPI predicate). | 700 mAh 4.8V Ni-MH, rechargeable battery pack and charger | Yes |
| Patient Leakage Current | < 1 microamp (consistent with EMSI predicate). | < 1 microamp | Yes |
| Number of Output Modes | 2 (consistent with EMSI predicate). | 2 | Yes |
| Number of Output Channels | 2 (consistent with both predicates). | 2 | Yes |
| Method of Channel Isolation | Transformer (consistent with both predicates). | Transformer | Yes |
| Regulated Current or Voltage? | Regulated Voltage (consistent with EMSI predicate). | Regulated Voltage | Yes |
| Software/Firmware/Microprocessor | Yes (consistent with both predicates). | Yes | Yes |
| Automatic Overload Trip? | Provides additional safety (improved over EMSI predicate). | Yes | Yes |
| Automatic Shut Off? | Yes (consistent with both predicates). | Yes | Yes |
| Patient Override Control? | Yes (consistent with both predicates). | Yes | Yes |
| Indicator Display (On/Off, Low Bat, Voltage/Current) | Yes (consistent with both predicates). | Yes (1-10 bars displayed for Voltage/Current Level) | Yes |
| Timer Range (minutes) | Minor difference from EMPI predicate is acceptable and does not affect safety/effectiveness. | 5-90 minutes, or continuous | Yes |
| Compliance with Voluntary Standards | ANSI/AAMI ES1-1993, IEC 60601-1-2 as applicable (consistent with EMSI predicate). | ANSI/AAMI ES1-1993, IEC 60601-1-2 as applicable | Yes |
| Compliance with 21 CFR 898? | Yes (consistent with both predicates). | Yes | Yes |
| Weight & Dimensions | Minor differences do not change device performance. | 156 g (with battery), 12cm x 5.4cm x 3.3cm | Yes |
| Housing Materials | Plastic (consistent with both predicates). | Plastic | Yes |
| Waveform (TENS/EMS) | Asymmetrical biphasic / Biphasic symmetrical (consistent with both predicates). | Asymmetrical biphasic / Biphasic symmetrical | Yes |
| Waveform Shape | Rectangular (consistent with both predicates). | Rectangular | Yes |
| Maximum Output Voltage | Within the range of predicates and acceptable limits of IEC 60601-1. | 43.2V @ 500Ω | Yes |
| Maximum Output Current | Within the range of predicates and acceptable limits of IEC 60601-1. | 86.4mA @ 500Ω | Yes |
| Pulse Width per phase | Same or increased selection as predicates. | 50-400µsec (consistent with EMPI predicate) | Yes |
| Pulse Frequency | Same available frequencies as predicates. | 2-150 Hz | Yes |
| Net Charge ($\mu$C per pulse) | Within acceptable limits of IEC 60601-1. | 14.6 @ 500Ω (TENS), 0 (EMS) | Yes |
| Maximum Current Density (mA/cm²) | Within acceptable limits of IEC 60601-1. | 1.83 @500Ω (TENS), 2.82 @500Ω (EMS) | Yes |
| Maximum Power Density (W/cm²) | Within acceptable limits to avoid thermal burn. | 0.000858W/cm2 @500Ω | Yes |
| Additional Features | Patient compliance timer, battery charger (consistent with predicates). | Patient compliance timer, battery charger | Yes |
2. Sample size used for the test set and the data provenance
The document explicitly states: "No clinical study data is provided in support of this submission."
Therefore, there is no "test set" in the sense of clinical data from patients. The evaluation is based on technical, electrical, and mechanical performance testing, likely conducted on a single or a few units of the device to verify its specifications. The document doesn't specify the sample size for these technical performance tests. The data provenance would be laboratory testing by the manufacturer (EMSI).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Since no clinical study or test set involving human interpretation of medical data was conducted, no experts were used to establish ground truth in that context. The "ground truth" for the technical performance is derived from recognized industry standards (IEC, ANSI/AAMI) and the specifications of the predicate devices.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as there was no test set requiring human adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a powered muscle stimulator/transcutaneous nerve stimulator, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a stimulator, not an algorithm, and its performance is assessed based on its physical and electrical outputs, not an algorithmic standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for evaluating the Flex-MT+ device's technical characteristics and performance is based on:
- Industry Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11, EN 60601-1-4, EN 62304, and ANSI/AAMI ES1-1993. The device's compliance with these standards serves as a form of "ground truth" for safety and basic performance.
- Predicate Device Specifications: The technical specifications (e.g., maximum output voltage, current, pulse width, frequency) of the legally marketed predicate devices (K083030 EMSI TENS-EMS-14 (Flex MT) and K021100 EMPI 300 PV) serve as the benchmark for substantial equivalence. The argument is that if the new device's specifications are within acceptable ranges of previously cleared devices, it is substantially equivalent.
8. The sample size for the training set
Not applicable. This device is not an AI/machine learning model, so there is no training set.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for an AI/ML model, there is no ground truth established for a training set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked slightly offset from each other. The profiles are connected by a flowing line that forms the base of the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 7, 2014
EMSI % Cherita James M Squared Associates, Inc. 815 King St. Suite 206 Alexandria, VA 22314
Re: K140467
Trade/Device Name: Flex-MT+ Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: August 20, 2014 Received: August 22, 2014
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Felipe Aquel -S
- Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K140467
Device Name Flex-MT +
Indications for Use (Describe)
TENS - Transcutaneous Nerve Stimulation
- · Symptomatic relief of chronic intractable pain
· Post traumatic and post surgical pain relief
EMS - Electrical Muscle Stimulation
- · Relaxation of muscle spasm
- Increasing local blood circulation
- · Muscle re-education
- · Prevention or retardation of disuse atrophy
- · Prevention of venous thrombosis of the calf muscles immediately after surgery
- · Maintaining or increase range of motion
Type of Use (Select one or both, as applicable)
| Registration Use (Part 21 CFR 201.Subpart D) | Over-The-Counter Use (21 CFR 201.Subpart C) |
|---|---|
| --------------------------------------------------------------------------- | -------------------------------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) SUMMARY
The following information is provided as required by 21 CFR § 807.92 for EMSI's Flex-MT + 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
- Sponsor: EMSI 3504 Cragmont Dr. Suite#100 Tampa, Florida 33619 Ph: 813-471-0129 Fax: 813-471-0130 Registration Number: 3003573572
M Squared Associates, Inc. Contact: Cherita James 815 King Street, Suite 206 Alexandria, VA 22314 Ph. 703-562-9800 Ext 257 Fax. 703-562-9797
| Date of Submission: | September 30, 2014 |
|---|---|
| Proprietary Name: | Flex-MT + |
| Common Name: | Powered Muscle Stimulator, Transcutaneous Nerve Stimulator |
| Regulatory Class: | II |
| Regulation: | 21 CFR 890.5850, 21 CFR 882.5890 |
| Panel: | Physical Medicine |
| Product Codes: | IPF, GZJ |
| Predicate Device(s): | K083030 EMSI TENS-EMS-14 (Flex MT), K021100 EMPI 300 PV |
The Flex-MT + is a combination TENS and EMS device which delivers Device Description: nerve or muscle stimulation by applying an electrical current to electrodes, which are attached on the patient's skin. The output and waveform is adjustable according to the intended treatment of patient. The stimulator has 2 output channels, accessed through jacks at the top of the housing, so that it may stimulate either 2 or 4 patient electrodes. The device is powered by 700 mAh 4.8V Ni
{4}------------------------------------------------
MH rechargeable battery pack. A patient compliance timer can memorize 60 sets of operation records; the total recordable time is 999 hours.
Intended Use: As prescribed by a physician for the following:
TENS- Transcutaneous Nerve Stimulation
- Symptomatic relief of chronic intractable pain .
- . Post traumatic and post surgical pain relief
EMS- Electrical Muscle Stimulation
- . Relaxation of muscle spasm
- . Increasing local blood circulation
- . Muscle re-education
- . Prevention or retardation of disuse atrophy
- Prevention of venous thrombosis of the calf muscles immediately after surgery
- . Maintaining or increase range of motion
Performance Testing
The Flex-MT + is compliant with the following standards and has outputs which are within the same range as the predicate devices.
- . IEC 60601-1:2005 = Corr 1:2006, Corr 2:2007 Medical Electrical Equipment - Part 1: General Requirements For Safety
- . IEC 60601-1-2:2007/AC:2010 Medical Electrical Equipment - Part 1-2: General Requirements For Safety - Collateral Standard: Electromagnetic Compatibility -Requirements And Test
- IEC 60601-2-10 Edition 2.0 2012-06 Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
- IEC 60601-1-11:2010 Medical electrical equipment -- Part 1-11: General requirements . for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- EN 60601-1-4:1996+A1:1999 Medical device Programmable Electrical Medical Systems (PEMS);
- . EN 62304 : 2006/AC:2008 Medical device software-Software life cycle processes.
{5}------------------------------------------------
Technical and Performance Comparison
Both the subject device and the predicate device are battery powered, handheld devices with similar unit characteristics including device controls, power supply, output modes and channels, waveforms, output voltage, current and pulse durations, as well as device material.
| Subject DeviceFlex-MT + | Predicate K08303EMSI TENS-EMS-14(Flex MT) | PredicateK021100EMPI 300PV(TENS/NMESmodes) | Comment | |
|---|---|---|---|---|
| Indication for Use | TENS-TranscutaneousNerve Stimulation•Symptomaticrelief of chronicintractable pain•Post traumaticand post surgicalpain reliefEMS- ElectricalMuscleStimulation•Relaxation ofmuscle spasm•Increasing localblood circulation•Muscle re-education•Prevention orretardation ofdisuse atrophy•Prevention ofvenousthrombosis of thecalf musclesimmediately aftersurgery•Maintaining orincrease range ofmotion | TENS-TranscutaneousNerve Stimulation•Symptomaticrelief of chronicintractable pain•Post traumatic andpost surgical painreliefEMS- ElectricalMuscle Stimulation•Relaxation ofmuscle spasm•Increasing localblood circulation•Muscle re-education•Prevention orretardation ofdisuse atrophy•Prevention ofvenous thrombosisof the calf musclesimmediately aftersurgery•Maintaining orincrease range ofmotion | TENS-TranscutaneousNerve Stimulation•Symptomaticrelief of chronicintractable pain•Adjunctivetreatment for postsurgical and posttrauma painEMS- ElectricalMuscle Stimulation•Relaxation ofmuscle spasm•Increasing localblood circulation• Re-educationmuscles• Retarding orpreventing disuseatrophy•Prevention ofvenous thrombosisof the calf musclesimmediately aftersurgery•Maintaining orincrease range ofmotion | Same indicationsin the subject andpredicates TENSand NMES modes. |
| Subject DeviceFlex-MT + | Predicate K08303EMSI TENS-EMS-14(FlexMT) | PredicateK021100EMPI 300PV | Comment | |
|---|---|---|---|---|
| 4. Power Source | 700 mAh 4.8VNi-MH,rechargeable | four batteries, sizeAA, alkaline | Two batteries, sizeAA rechargeable,and charger | All devices arebattery operated.The subject and |
| Subject DeviceFlex-MT + | Predicate K08303EMSI TENS-EMS-14(FlexMT) | PredicateK021100EMPI 300PV | Comment | |
| battery pack andcharger | EMPI devicesoffer a batterycharger. | |||
| – Methodof Line CurrentIsolation | n/a | n/a | n/a | NA |
| - PatientLeakage Current- Normal condition- Singlefault condition | not measurable< 1 microamp | not measurable< 1 microamp | Unknown | The subject andEMSI predicatehave the samePatient Leakagelevel |
| 5. Number ofOutput Modes | 2 | 2 | 4 | The subject andEMSI predicatehave the samenumber ofchannels. TheEMPI predicatesoffers the same 2channels, as wellas 2 additionalchannels for theirIFS and FESoutputs. |
| 6. Number ofOutput Channels | 2 | 2 | 2 | Same number andtype of outputchannels andchannel isolation. |
| - Synchronousor Alternating | Synchronous orAlternating | Synchronous orAlternating | Synchronous orAlternating | |
| – Method ofChannel Isolation | Transformer | Transformer | Transformer | |
| 7. RegulatedCurrent orRegulated Voltage? | RegulatedVoltage | Regulated voltage | Unknown | Same regulatedvoltage method asthe EMSI predicatedevice. |
| 8.Software/Firmware/MicroprocessorControl? | Yes | Yes | Yes | All devices aresoftware controlled |
| 9. AutomaticOverload Trip? | Yes | No | Unknown | The automaticoverloadprotectionprovides additionalsafety whenoperating thesubject device |
| Subject DeviceFlex-MT + | Predicate K08303EMSI TENS-EMS-14(FlexMT) | PredicateK021100EMPI 300PV | Comment | |
| 11. Automatic ShutOff? | Yes | Yes | Yes | Same features |
| 12. PatientOverride Control? | Yes | Yes | Yes | Same features |
| 13. IndicatorDisplay:On/Off Status?Low Battery?Voltage/CurrentLevel? | YesYesYes (1-10 barsdisplayed) | YesYesYes (1-10 barsdisplayed) | YesYesYes | Same features |
| 4. Timer Range(minutes) | 5-90 minutes, orcontinuous | 5-90 minutes, orcontinuous | 5-99 minutes, orcontinuous | Minor timerdifference whencompared to theEMPI predicate donot affect safety oreffectiveness. |
| 15. Compliancewith VoluntaryStandards? | ANSI/AAMIES1-1993IEC 60601-1-2 asapplicable | ANSI/AAMI ES1-1993IEC 60601-1-2 asapplicable | Unknown | NA |
| 16. Compliancewith 21 CFR 898? | Yes | Yes | Yes | Same features |
| 17. Weight | 156 g (includingbattery) | 140 g | 226g | Minor differencesin weight and |
| 18. Dimensions [Wx H x D] | 12cm x 5.4cm x3.3cm | 12 cm x 5.5cm x2.5cm | unknown | dimensions do notchange device |
| 19. HousingMaterials andConstruction | plastic | plastic | plastic | performance.Same materials. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Subject Device- Flex-MT + | Predicate K083030EMSI TENS-EMS-14(Flex MT) | Comments | |||||||
|---|---|---|---|---|---|---|---|---|---|
| TENS mode | EMS mode | TENS mode | EMS mode | ||||||
| Waveform | Asymmetricalbiphasic | Biphasicsymmetrical | Asymmetricalbiphasic | Biphasicsymmetrical | Asymmetricalbiphasic | Biphasicsymmetrical | Asymmetricalbiphasic | Biphasicsymmetrical | Samewaveforms |
| Shape | Rectangular | Rectangular | Rectangular | Rectangular | Sameshape | ||||
| MaximumOutputVoltage | 43.2V @ 500Ω | 43.2V @ 500Ω | 41V@500 Ω | 28V@500 Ω | 41V @ 500 Ω | 29.5V@ 500 Ω | Themaximumoutputvoltage @500Ω ofthe subjectdevice iswithin therange ofthe EMSIand EMPIpredicates.All withinacceptablelimits ofIEC60601-1 | ||
| MaximumOutputCurrent | 82mA @ 500Ω | 56mA @ 500Ω | 82 mA @500Ω | 59 mA@500Ω | Subjectdevice hasincreaseoutputcurrentavailablewhencomparedto theEMSIpredicate,but withinrange ofthe EMPIpredicate.All withinacceptablelimits ofIEC60601-1 | ||||
| 86.4mA @ 500Ω | 86.4mA @500Ω | 86.4mA @ 500Ω | 86.4mA @ 500Ω | ||||||
| PulseWidth perphase | 50-400µsec | 50-400µsec | 50-300 µsec | 50-300 µsec | Subjectdevice hasincreasedselectionof pulsewidthoptionwhencomparedto EMSIpredicate,but sameoption asEMPIpredicate. | ||||
| MaxPhaseDuration(PositivePhase) | $400\mu s$=0.4ms | $400\mu s$=0.4ms | $400\mu s$=0.4ms | $400\mu s$=0.4ms | N/A | 50-300 $\mu s$ | N/A | 50-300 $\mu s$ | |
| MaxPhaseDuration(NegativePhase) | 2.6ms | $400\mu s$=0.4ms | 2.6ms | $400\mu s$=0.4ms | - | - | |||
| PulseFrequencyMax Dutyfactor | 2-150 Hz0.060 | 2-150 Hz0.120 | 2-150 Hz0.060 | 2-150 Hz0.120 | 2-150Hz0.045 | 2-150 Hz0.090 | 2-150Hz0.045 | 2-150 Hz0.090 | Samepulsefrequencies available |
| Subject Device- Flex-MT + | Predicate K083030EMSI TENS-EMS-14(Flex MT) | ||||||||
| Multi-phasicwaveforms | Yes | yes | Yes | yes | yes | yes | yes | yes | |
| NetCharge( $\mu$ C perpulse) | 14.6 @ 500 $\Omega$ | 0(symmetricalphases resultin 0) | 14.6 @ 500 $\Omega$ | 0(symmetricalphases) | 24.5 @ 500 $\Omega$ | 0 (symmetricalphases) | 24.6 $\mu$ C @500 $\Omega$ | 0(symmetrical phases) | All withinacceptablelimits ofIEC 60601-1 |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| MaximumCurrentDensity,(mA/cm²) | 1.83 @500Ω | 2.82 @500Ω | 1.83 @500Ω | 2.82 @500Ω | 0.26 @500 Ω | 0.35 @ 500 Ω | 0.26 @ 500 Ω | 0.37 @ 500 Ω | All withinacceptablelimits ofIEC60601-1 |
|---|---|---|---|---|---|---|---|---|---|
| MaximumPowerDensity,(W/cm²) | 0.000858@500Ω | 0.000858@500Ω | 0.000858@500Ω | 0.000858@500 Ω | 0.0105@ 500 Ω | 0.0098 @ 500 Ω | 0.0105 @500 Ω | 0.0109 @500 Ω | |
| MaximumPulseDuration | 400µs+2.6ms=0.4ms+2.6ms=3.0ms | 400µs+400µs=800µs=0.8ms | 400µs+2.6ms=0.4ms+2.6ms=3.0ms | 400µs+400µs=800µs=0.8ms | - | - | - | - | |
| AdditionalFeatures(ifapplicable) | Patient compliance timer, battery charger | Patient compliance timer |
{12}------------------------------------------------
| Subject Device- Flex-MT + | Predicate K021100EMPI 300 PV (TENS and EMSmodes only) | Comments | |||||
|---|---|---|---|---|---|---|---|
| TENS mode | EMS mode | TENS and EMS mode | Same modes | ||||
| Waveform | Asymmetricalbiphasic | Biphasicsymmetrical | AsymmetricalBiphasicbiphasicsymmetrical | Asymmetrical biphasic, Biphasicsymmetrical | Same waveforms | ||
| Shape | Rectangular | Rectangular | Rectangular | Same shape | |||
| MaximumOutputVoltage | 43.2V @ 500Ω | 43.2V @ 500Ω | 43.2V @ 500Ω | 43.2V @ 500Ω | 50V@500 Ώ | The maximum output voltage of thesubject device is within the range ofthe EMSI and EMPI predicates. Allwithin acceptable limits of IEC60601-1 | |
| MaximumOutputCurrent | 86.4mA @ 500Ω | 86.4mA @500Ω | 86.4mA @ 500Ω | 86.4mA @500Ω | 100mA @, 500 Ώ | The maximum output current of thesubject device is within the range ofthe EMSI and EMPI predicates. Allwithin acceptable limits of IEC60601-1 | |
| PulseWidth perphase | 50-400 usec | 50-400 usec | 50-400 usec | Same range | |||
| Max PhaseDuration(PositivePhase) | 400us=0.4ms | 400us=0.4ms | 400us=0.4ms | 400us=0.4ms | |||
| Max PhaseDuration(NegativePhase) | 2.6ms | 400us=0.4ms | 2.6ms | 400us=0.4ms | |||
| Phaseduration | 50-400 µsec | 50-400 µsec | 50-400 µsec | 50-400 µsec | Unknown | --- | |
| NetCharge(µC perpulse) | 14.6 @500Ω | 0(symmetricalphases resultin 0) | 14.6 @500Ω | 0(symmetricalphases) | Unknown | --- | |
| MaximumPhaseCharge,( μ C) | 14.6uC@ 500Ω | 0uC @ 500Ω | 14.6 uC @500Ω | 0 uC @500Ω | 40 uC @ 500 Ώ | All within acceptable limits of IEC60601-1 | |
| MaximumCurrentDensity,(mA/cm²) | 14.6 @500Ω | 0@500Ω | 14.6 @500Ω | 0@500Ω | 0.84 @ 500 Ω2"square electrode | All within acceptable limits of IEC60601-1 | |
| MaximumPowerDensity,(W/cm²) | 0.000858W/cm2@500Ω | 0.000858W/cm2@ 500Ω | 0.000858W/cm2@500Ω | 0.000858W/cm2@500Ω | 0.0088W/cm2 @ 500 Ώ2"square electrode | All within acceptable limits to avoidthermal burn. | |
| MaximumPulseDuration | 400µs+2.6ms=0.4ms+2.6ms=3.0ms | 400µs+400µs=800µs=0.8ms | 400µs+2.6ms=0.4ms+2.6ms=3.0ms | 400µs+400µs=800µs=0.8ms | --- | ||
| AdditionalFeatures(ifapplicable) | Patient compliance timer, battery charger | Battery charger | Same feature |
{13}------------------------------------------------
{14}------------------------------------------------
Clinical Data: No clinical study data is provided in support of this submission.
Substantial Equivalence
Based on the Flex-MT+ and the predicate device technical characteristics, performance, and indications for use, the substantially equivalent to the EMSI Flex MT and the EMPI 300V. As detailed in the tables above, the differences between the subject and predicate devices do not adversely impact the FlexMT+ safety and effectiveness for it's substantial equivalence to the predicates.
N/A