K Number
K133369

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-01-14

(74 days)

Product Code
Regulation Number
888.3070
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TIGER® Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (TT Sifleum): degenerative discogenic back pain with degeneration of disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective widence of neurological impairment, frecture, dislocation, scoliosis, kyphosis, spinal turnor, and failed previous fusion (pseudarthrosis).

Device Description

The TIGER® Spine System is composed of rods, connectors, and pedicle screws. It can be used for single or multiple level fixations.

AI/ML Overview

The provided text describes a Special 510(k) Summary for the TIGER® Spine System. It focuses on the substantial equivalence of the device to existing predicate devices based on non-clinical testing. Crucially, the document explicitly states that no clinical studies were performed. Therefore, this device submission does not contain information regarding acceptance criteria for a device's performance based on clinical data, nor does it describe a study that proves the device meets such criteria through clinical outcomes.

Instead, the document details non-clinical testing performed to demonstrate mechanical safety and performance in comparison to predicate devices.

Here's a breakdown of the requested information based only on the provided text, acknowledging the absence of clinical performance data:


1. A table of acceptance criteria and the reported device performance

Since this is a non-clinical submission based on substantial equivalence to predicate devices, the "acceptance criteria" are implicitly meeting or exceeding the performance of the predicate device(s) in specified mechanical tests. The "reported device performance" is the conclusion that the TIGER® Spine System is equivalent.

Acceptance Criteria (Implicit for Substantial Equivalence via Non-Clinical Testing)Reported Device Performance (Conclusion)
Meet or exceed performance of predicate devices in:The TIGER® Spine System is equivalent to predicate device(s) in terms of mechanical safety and performances.
- Static and dynamic compression per ASTM F1717
- Static torsion per ASTM F1717
- Static axial and torsional grip per ASTM F1798
- Dimensional comparison of components

2. Sample size used for the test set and the data provenance

The document does not specify the sample size (number of devices or test repetitions) used for the non-clinical mechanical tests. It also does not discuss data provenance as it pertains to patient data, as no clinical studies were performed. The tests are laboratory-based mechanical tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable. No ground truth based on expert clinical assessment (e.g., radiologists) was established because no clinical studies were performed. The "ground truth" for the non-clinical tests is based on established ASTM standards and comparative analysis to predicate devices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This question is not applicable. No clinical test set requiring adjudication of expert opinions was created.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. No MRMC study was performed, as no clinical studies involving human readers or AI assistance were conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. No algorithm or standalone software performance was evaluated, as this submission is for a physical spinal implant system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests, the "ground truth" is based on established engineering standards (ASTM F1717, ASTM F1798) and the performance characteristics of legally marketed predicate devices. This is a technical, rather than clinical, ground truth.

8. The sample size for the training set

This question is not applicable. There was no training set, as no algorithms requiring training were involved in this submission.

9. How the ground truth for the training set was established

This question is not applicable. There was no training set.


Summary of Study (Based on Provided Document):

The submission describes a non-clinical study involving mechanical and dimensional testing of the TIGER® Spine System. The study's purpose was to demonstrate substantial equivalence to existing predicate devices (e.g., Expedium™/Viper™ Spine System, Scient'x Polyaxial LP) for the addition of new components. The study involved:

  • Tests performed: Static and dynamic compression (ASTM F1717), static torsion (ASTM F1717), static axial and torsional grip (ASTM F1798), and dimensional comparison.
  • Conclusion: The results indicated that the TIGER® Spine System is equivalent to the predicate device(s) in terms of principles of operation, technology, materials, and indications of use.
  • Sample Size and Ground Truth: Not explicitly detailed for the specific tests, but generally refers to standard testing protocols. The "ground truth" for this equivalence claim is the established performance and safety profile of the predicate devices.
  • Clinical Studies: Explicitly stated as not performed. Therefore, no clinical acceptance criteria or performance studies were conducted or reported in this submission.

{0}------------------------------------------------

Special 510(k) Summary for the Tiger ® Spine System

JAN 1 4 2014

In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following Special 510(k) summary is submitted for the Tiger ® Spine System

1. GENERAL INFORMATION

Date Prepared: +December 17, 2013
Trade Name: TIGER® Spine System
Common Name: pedicle screw system
Classification Name: orthosis, spinal pedicle fixation
orthosis, spondylolisthesis spinal fixation
Class: III
Product Code: MNI
MNH
KWP
NKB
CFR section: 21 CFR section 888.3070
21 CFR section 888.3050
Device panel: Orthopedic
Legally Marketed TIGER® Spine System (K110321, K113058, K120696, K121728, K131250)
Predicate Device: Expedium™/Viper™ Spine System (K033901, K051024, K062174, K063156,
K063741, K080313, K081898, K090648, K101993)
Scient'x Polyaxial LP (K062912)
Submitter: Corelink, LLC
7606 Forsyth Blvd
Clayton, MO 63105
Contact: J.D. Webb
1001 Oakwood Blvd
Round Rock, TX 78681
512-388-0199
e-mail: jdwebb@orthomedix net

2. DEVICE DESCRIPTION

The TIGER® Spine System is composed of rods, connectors, and pedicle screws. It can be used for single or multiple level fixations.

Change from Predicate:

The purpose of this premarket notification is the addition of new components to the TIGER® Spine System.

Materials:

Ti-6Al-4V alloy per ASTM F136

3. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES

The TIGER® Spine System is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances.

{1}------------------------------------------------

4. INTENDED USE

The TIGER® Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacraliliac spine (TI Si/Ileum): degenerative disc disease (defined as discogenic back pain with deqeneration of disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

5. NON-CLINICAL TEST SUMMARY

The following tests were conducted:

  • . Static and dynamic compression per ASTM F1717
  • Static torsion per ASTM F1717 .
  • . Static axial and torsional grip per ASTM F1798
  • Dimensional comparison of components .

The results of this testing indicate that the TIGER® Spine System is equivalent to predicate device(s).

6. CLINICAL TEST SUMMARY

No clinical studies were performed

7. CONCLUSIONS NONCLINICAL AND CLINICAL

CoreLink, LLC considers the TIGER® Spine System to be equivalent to the predicate device(s) listed above. This conclusion is based upon the devices' similarities in principles of operation, technology, materials and indications of use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure embracing a sphere.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 14, 2014

Corelink, LLC % Mr. J.D. Webb The OrthoMedix Group, Incorporated 1001 Oakwood Boulevard Round Rock, Texas 78681

Re: K133369

Trade/Device Name: TIGER® Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH, KWP Dated: December 17, 2013 Received: December 19, 2013

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. J.D. Webb

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Lori A. Wiggins

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

. DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

K133369

Device Name

TIGER® Spine System

Indications for Use (Describe)

The TIGER® Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral/iliac spine (TT Sifleum): degenerative discogenic back pain with degeneration of disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective widence of neurological impairment, frecture, dislocation, scoliosis, kyphosis, spinal turnor, and failed previous fusion (pseudarthrosis).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Zane W詞Watt -S

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

N/A