Intended Use

The TIGER® Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic. lumbar. and sacral spine (T)-S1/Ileum): degenerative disc discase (defined as discogenic back pain with degeneration of disc confirmed by history of and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment. fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

Device Description

The TIGER® Spine System is a multiple component, posterior spinal fixation system which consists of pedicle screws, cannulated pedicle screws, posted pedicle screws, rods, crossconnectors, locking cap screws, hooks, and lateral offset connectors. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are made from titanium alloy described by such standards as ASTM F136.

AI/ML Overview

This document describes a 510(k) premarket notification for the TIGER® Spine System, a medical device for spinal fixation. Let's break down the acceptance criteria and the supporting study, focusing on the requested information.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes physical performance testing rather than algorithm-based acceptance criteria for a medical imaging device. However, we can infer the "acceptance criteria" as meeting the specified ASTM standards and the "reported device performance" as having been tested to those standards.

Acceptance Criteria (Inferred from Study Type)Reported Device Performance
Compliance with ASTM F1717-11aTIGER® Posted Spine System constructs were tested in static axial compression, static torsion, and dynamic axial compression per ASTM F1717-11a.
Compliance with ASTM F1798-97TIGER® Spine System Hook constructs were tested in static axial and torsion grip per ASTM F1798-97.
Compliance with ASTM F1717-11aTIGER® Spine System Cannulated Screw constructs were tested in static axial compression, static torsion, and dynamic axial compression per ASTM F1717-11a.
Substantial equivalence to predicate devices"The overall technology characteristics and mechanical performance data lead to the conclusion that TIGER® Spine System is substantially equivalent to the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

This information is not available in the provided text. The document refers to mechanical testing of physical device components, not a study involving human subjects or image data. Therefore, concepts like a "test set" and "data provenance (e.g. country of origin of the data, retrospective or prospective)" are inapplicable in this context.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable/available. The study described is mechanical testing of medical device components, not a clinical study involving experts establishing ground truth for diagnostic or prognostic purposes.

4. Adjudication Method for the Test Set

This information is not applicable/available as the study did not involve a test set requiring adjudication in the context of expert review.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document describes mechanical performance testing of spinal fixation components, not a study evaluating human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

No, a standalone study in the context of an algorithm's performance was not done. This device is a physical spinal fixation system, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the mechanical performance testing was the engineering specifications and established test methods defined by ASTM standards (ASTM F1717-11a and ASTM F1798-97). The device components were tested to see if they met the performance requirements outlined in these standards for static axial compression, static torsion, dynamic axial compression, and torsion grip.

8. The Sample Size for the Training Set

This information is not applicable/available. The device is a physical product, not an AI or algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/available as there is no training set for a physical medical device like the TIGER® Spine System.

{0}------------------------------------------------

510(K) SUMMARY

Submitter's Name:CoreLink, LLC
Submitter's Address:7606 Forsyth BlvdClayton, MO 63105
Submitter's Telephone:314-863-5401
Submitter's Fax:636-680-5715
Authorized Contact Name:Meredith May, Authorized Contact Person
Contact's Telephone:719-337-7579
Contact's Email:MMay@EmpiricalTesting.com
Date Summary wasPrepared:12-Apr-13
Trade or Proprietary Name:TIGER® Spine System
Common or Usual Name:Orthosis, Spinal Pedicle FixationOrthosis, Spondylolisthesis Spinal FixationOrthosis , Spinal Interlaminal Fixation
Classification:Class III per 21 CFR §888.3070 and §888.3050
Product Codes:MNI, MNH, NKB, KWP
Classification Panel:Orthopedic and Rehabilitation Devices Panel
Predicate Devices:TIGER® Spine System (K110321, K120696. K113058,and K121728)Depuy Expedium Offset (K101070)Depuy Monarch (K024348 and K010576)EBI Omega 21 (K014137, K001357, K992333, K991721,K990303, and K973683)Medtronic Sofamor Danek TSRH-3D (K021170)

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The TIGER® Spine System is a multiple component, posterior spinal fixation system which consists of pedicle screws, cannulated pedicle screws, posted pedicle screws, rods, crossconnectors, locking cap screws, hooks, and lateral offset connectors. All of the components are available in a variety of sizes to match more closely to the patient's anatomy. All components are made from titanium alloy described by such standards as ASTM F136.

CHANGE FROM PREDICATE:

The purpose of this submission is to make modifications/additions to the components of the TIGER® Spine System cleared in K110321, K120696, K113058, and K121728. The standard

Tiger® Spine System Line Extension

{1}------------------------------------------------

construct is modified by the addition of hooks, cross-connectors, cannulated screws, and posted screws.

TECHNOLOGICAL CHARACTERISTICS:

The intended use and technological features of the modifications/additions to the components of the TIGER® Spine System do not substantially differ from the legally marketed predicate devices, which are the TIGER® Spine System (K110321. K120696, K113058, and K121728), Depuy Expedium Offset (K101070). Depuy Monarch (K024348 and K010576), EBI Omega 21 (K014137, K001357, K992333, K991721, K990303, and K973683), and Medtronic Sofamor Danek TSRH-3D (K021170). The predicate devices and the subject additions to the Tiger system are designed for posterior stabilization to provide immobilization and stabilization of spinal segments as an adjunct to fusion.

INDICATIONS FOR USE

The TIGER® Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic. lumbar. and sacral spine (T)-S1/Ileum): degenerative disc discase (defined as discogenic back pain with degeneration of disc confirmed by history of and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment. fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

PERFORMANCE DATA

The TIGER® Posted Spine System constructs were tested in static axial compression, static torsion, and dynamic axial compression per ASTM F1717-11a. The TIGER® Spine System Hook constructs were tested in static axial and torsion grip per ASTM F1798-97. The TIGER® Spine System Cannulated Screw constructs were tested in static axial compression, static torsion, and dynamic axial compression per ASTM F1717-11a.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that TIGER® Spine System is substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that converge at the bottom.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

CoreLink, LLC % Ms. Meredith May Empirical Testing Corporation 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K131250

Trade/Device Name: TIGER® Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH. KWP Dated: July 18, 2013 Received: July 31, 2013

Dear Ms. May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please notc: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

September 10, 2013

{3}------------------------------------------------

Page 2 - Ms. Meredith May

forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours,

Mark N. Melkerson -S

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{4}------------------------------------------------

Indications for Use Statement

Device Name: TIGER® Spine System

510(k) Number: K131250

The TIGER® Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine (T)-S VIlleum): degenerative disc disease (defined as discogenic back pain with degeneration of disc confirmed by history of and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scollosis, spinal tumor, and failed previous fusion (pscudoarthrosis).

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Ronald P. Jean -S

(Division Sign-Off) Division of Orthopedic Devices 510 (k) Number: Kl31250