Intended Use

The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENALI®) Spinal Systems, all of which are cleared for the following indications;

Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: Trauma ( i.e. fracture or dislocation ): spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Device Description

The Everest Spinal System is a top-loading, multiple component, posterior (thoraciclumbar) spinal fixation system which consists of pedicle screws, rods, locking set screws, and hooks.

Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine.

The purpose of this 510(k) submission is to add minimally invasive implants.

AI/ML Overview

Here's an analysis of the provided text regarding the Everest Spinal System, focusing on acceptance criteria and the study proving device performance:

Summary of Acceptance Criteria and Device Performance for the Everest Spinal System

It's important to note that the provided document is a 510(k) summary for a spinal fixation system, and not a diagnostic AI device. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are geared towards mechanical and design equivalence to predicate devices, rather than diagnostic accuracy metrics like sensitivity, specificity, or AUC which are typical for AI-powered diagnostic tools.

1. Table of Acceptance Criteria and the Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Mechanical PerformanceStatic CompressionPerformed "equally to or better than" predicate devices in accordance with ASTM F1717
Static TorsionPerformed "equally to or better than" predicate devices in accordance with ASTM F1717
Dynamic CompressionPerformed "equally to or better than" predicate devices in accordance with ASTM F1717
Design FeaturesDesign of components"substantially the same as"predicate systems
Sizing of ComponentsSizing of components"substantially the same as"predicate systems
Material CompositionMaterials usedManufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards (Implicitly accepted as equivalent to predicate devices through regulatory compliance)
Intended UseIndications for UseThe system's indications for use are identical to those of the predicate K2M Range Spinal Systems (MESA and DENALI), demonstrating equivalence.

2. Sample Size Used for the Test Set and the Data Provenance

Given this is a mechanical testing and design comparison, the concept of a "test set" and "data provenance" (country of origin, retrospective/prospective) as understood for AI/diagnostic studies is not directly applicable.

  • Test Set Sample Size: The document does not specify a numerical sample size for the mechanical tests. It would typically involve a certain number of devices or components tested according to ASTM F1717.
  • Data Provenance: Not applicable in the context of mechanical testing. The testing would be conducted in a laboratory environment following standardized procedures.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This category is not applicable for the Everest Spinal System 510(k) summary. "Ground truth" in this context refers to an objective standard, which for mechanical performance is established by the ASTM F1717 standard itself and the measured physical properties of the tested devices, not expert human assessment.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human experts in diagnostic studies. Here, the "truth" is determined by direct mechanical measurements against a standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a surgical implant, not an AI-assisted diagnostic tool. Therefore, MRMC studies for human reader performance with or without AI assistance are irrelevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. The Everest Spinal System is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance is established by:

  • Mechanical Measurement Results: Directly measured values from static compression, static torsion, and dynamic compression tests performed according to ASTM F1717.
  • Predicate Device Performance: The accepted performance characteristics of the legally marketed predicate devices (K2M Everest Spinal System, K2M Range Spinal System, NuVasive SpherX, Globus Resolve). The new device's performance is compared against these established benchmarks.
  • Design and Sizing Equivalence: Direct comparison of the physical design and dimensions of the new components to those of the predicate devices.
  • Material Standards: Compliance with ASTM and ISO standards for Titanium Alloy and Cobalt Chrome.

8. The Sample Size for the Training Set

Not applicable. As this is a physical medical device undergoing mechanical testing and design comparison, there is no "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set was Established

Not applicable. There is no training set for this type of device submission.

{0}------------------------------------------------

510(k) Summary for the Everest Spinal System

This 510(k) summary for the Everest Spinal System is provided as required per Section 513(i)(3) of the Food, Drug and Cosmetic Act.

1. Submitter :

K2M, Inc. 751 Miller Drive SE. Suite FI Leesburg, VA 20175

Contact Person :

Nancy Giezen K2M, Inc. Telephone: 703-777-3155

Date Prepared: 09/26/13

OCT 0 3 2013

2. Tradename:Everest Spinal System
Common Name:Spinal Fixation System
Classification Name:Spinal Interlaminal Fixation Orthosis (21CFR 888.3050)Pedicle Screw Spinal System (888.3070)
Device Product Code:KWP, MNI
Regulatory Class:Class II
  1. Predicate or legally marketed devices which are substantially equivalent :

4. Description of the device:

The Everest Spinal System is a top-loading, multiple component, posterior (thoraciclumbar) spinal fixation system which consists of pedicle screws, rods, locking set screws, and hooks.

Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine.

The purpose of this 510(k) submission is to add minimally invasive implants.

{1}------------------------------------------------

5. Intended Use:

The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENALI®) Spinal Systems, all of which are cleared for the following indications;

Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: Trauma ( i.e. fracture or dislocation ): spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

6. Comparison of the technological characteristics of the device to predicate and legally marketed devices :

The minimally invasive components of the Everest Spinal System were mechanically tested and compared to predicate devices. The subject device performed equally to or better than these systems in static compression, static torsion and dynamic compression in accordance with ASTM F1717. The design features and sizing of the components were also compared and found to be substantially the same as these systems.

There are no significant differences between the Everest Spinal System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all capital letters and is evenly spaced around the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2013

K2M, Incorporated Ms. Nancy Giezen 751 Miller Drive Southeast Leesburg, Virginia 20175

Re: K132361

Trade/Device Name: Everest Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, KWP Dated: August 6, 2013 Received: August 7, 2013

Dear Ms. Giezen:

.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Nancy Giezen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Erin Keith

for

Mark N. Melkerson Director Division of Orthonedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K132361

Device Name: Everest Spinal System

Indications for Use:

The EVEREST Spinal System may be used in conjunction with the RANGE® (MESA® and DENAL 90 Spinal Systems, all of which are cleared for the following indications:

Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: Trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scollosis. kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-the-counter Use (21 CFR 801 Subpart C)