K Number
K130330

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-05-02

(80 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RANGE /DENALI/MESA, SMALL STATURE and ARI are cleared for the following indications:

Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skelcially mature patients reveiving fusion by autogenous bone graft having implants attached to the lumbar and sucral spine ( L3 10 sacrum) with removal of the implants after the attainment of a solid fusion.

Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion, in skeletally mature patients, for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.c. fracture or dislocation); spinal stenosis; curvatures ( i.c. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical lixation in pediatric patients. The Runge Spinal System for pediatric use is intended to be used with autografi. Pediatric pediale screw fixation is limited to a posterior approach.

Device Description

The Range Spinal System is a top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation system which consists of pedicle screws, rods, hooks and rod connectors. The purpose of this submission is to add modified rods and hooks and to incorporate previously cleared implants that were not included in the pediatric submission.

Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (What they tested for)Reported Device Performance (How the device performed)
Static/dynamic compression bending and static torsion (worst-case constructs)Constructs representing the worst case components were previously tested in static/dynamic compression bending and static torsion in accordance with ASTM F1717 and were determined to be substantially equivalent to predicate devices. The modified implants were determined to not represent a new worst case.
Substantial Equivalence to Predicate DevicesSubstantially equivalent in design, function, material, and intended use to predicate devices and other devices currently marketed.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document refers to "Constructs representing the worst case components." It does not specify a numerical sample size but implies a selection of representative configurations rather than a large statistical sample of individual implants.
  • Data Provenance: The study was a comparison to previously tested predicate devices. The document does not explicitly state the country of origin or whether the data for the Range Spinal System was retrospective or prospective, but it's implied to be data generated to support the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • This document describes a mechanical performance study, not a clinical study involving ground truth established by medical experts for a test set. Therefore, this section is not applicable to this submission. The "ground truth" here is the performance according to ASTM standards and comparison to predicate devices.

4. Adjudication Method for the Test Set

  • This document describes a mechanical performance study, not a clinical study requiring adjudication of expert interpretations. Therefore, this section is not applicable to this submission. The results were likely validated by engineers and testing personnel.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This is a submission for a spinal fixation system, and the studies described are mechanical performance tests comparing the device to predicates, not clinical studies evaluating human reader performance with or without AI assistance.

6. Standalone Performance Study (Algorithm only without Human-in-the-loop performance)

  • This question is not applicable. The device is a physical spinal implant system, not an algorithm or AI software. The performance studies conducted were mechanical strength and durability.

7. Type of Ground Truth Used

  • The "ground truth" for this submission is based on engineering standards (ASTM F1717) and direct comparative mechanical testing against predicate devices. The performance of the Range Spinal System was measured against established benchmarks for spinal fixation devices.

8. Sample Size for the Training Set

  • This question is not applicable. This is not an AI/machine learning device that requires a training set. The "training" for such a device is the engineering design and manufacturing process to meet specified standards.

9. How the Ground Truth for the Training Set Was Established

  • This question is not applicable. As stated above, this is not an AI/machine learning device.

{0}------------------------------------------------

MAY 0 2 2013

K130330 Page 1 of 2

510(k) Summary for the Range Spinal System: Additional Components

This 510(k) summary for the Range Spinal System is provided us required per Scction 513(i)(3) of the Food, Drug and Cosmetic Act.

  1. Submitter : K2M, Inc. 751 Miller Drive SE, Leesburg, VA 20175

Date Prepared: 04/04/13

  1. Tradename:

Common Name: Classification Name:

Device Product Code: Regulatory Class:

Range Spinal System Spinal Fixation System Pedicle Screw Spinal System (21CFR 888.3070) Spinal Interlaminal Fixation Orthosis (21CFR 888.3050) OSH, MNI, KWP, MNH Class II

Contact Person :

Telephone: 703-777-3155

Nancy Giczen K2M, Inc.

  1. Predicate or legally marketed devices which are substantially equivalent : K2M Range Spinal System (K070229, K112920, K120099, K121016, K121630) .

4. Description of the device:

The Range Spinal System is a top-loading, multiple component, posterior (thoracic-lumbar) spinal fixation system which consists of pedicle screws, rods, hooks and rod connectors. The purpose of this submission is to add modified rods and hooks and to incorporate previously cleared implants that were not included in the pediatric submission.

Materials: The devices are manufactured from Titanium Alloy and Cobalt Chrome per ASTM and ISO standards.

Function: The system functions as an adjunct to fusion to provide immobilization and stabilization of the posterior thoracic and lumbar spine.

5. Intended Use:

RANGE /DENALI/MESA, SMALL STATURE and ARI are cleared for the following indications:

Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the 1.5-SI vertebra in skelcially mature patients reveiving fusion by autogenous bone graft having implants attached to the lumbar and sucral spine ( 1.3 10 sacrum) with removal of the implants after the attainment of a solid fusion.

Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion, in skeletally mature patients, for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.c. fracture or dislocation); spinal stenosis; curvatures ( i.c. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

{1}------------------------------------------------

Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical lixation in pediatric patients. The Runge Spinal System for pediatric use is intended to be used with autografi. Pediatric pediale screw fixation is limited to a posterior approach.

  1. Comparison of the technological characteristics of the device to predicate and legally marketed devices : The design features and sizing of the components were compared to predicate devices and the Range Spinal System was found to be substantially the same as these systems.

7. Comparison of the performance data of the device to predicate and legally marketed devices : Constructs representing the worst case components were previously tested in static/dynamic compression bending and static torsion in accordance with ASTM F1717 and were determined to be substantially equivalent to predicate devices. The modified implants were determined to not represent a new worst case.

8. Conclusion:

There are no significant differences between the proposed Range Spinal System implants and other devices currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The text is in all caps and appears to be in a sans-serif font.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 27, 2013

K2M, Incorporated % Ms. Nancy Giezen 751 Miller Drive SE Leesburg, Virginia 20175

Re: K130330

Trade/Device Name: Range Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: OSH, MNI, MNH, KWP Dated: April 4, 2013 Received: April 5, 2013

Dear Ms. Giezen:

This letter corrects our substantially equivalent letter of May 2, 2013.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Ms. Nancy Giezen

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/8 description: The image shows the name "Erin Toweith" in a stylized font. The first name, "Erin," is written in a simple, bold font. The last name, "Toweith," is written in a more complex, decorative font that makes it harder to read. The overall effect is a visually interesting nameplate.

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K130330

Device Name: Range Spinal System

Indications for Use:

RANGE /DENALI, MESA and SMALL STATURE and ARI are cleared for the following indications:

Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

Non-cervical. non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion, in skeletally mature patients, for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation); spinal stenosis; curvatures ( i.e. scoliosis; and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion.

Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatic patients. The Range Spinal System for pediatric use is intended to be used with autograft. Pediatric pediatric pedicle screw fixation is limited to a posterior approach.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-the-counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS-LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ronald P. Jean -S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130330

N/A