The Everest Spinal System if used in conjunction with the Range (Mesa and Denali) Spinal Systems are:
Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: Trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disc disease (DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
The Everest Spinal System is a top-loading, multiple component, posterior (thoracic-lumba) spinal fixation system which consists of pedicle screws, rods, locking set screws, and hooks. The purpose of this submission is to increase the range of screw sizes and to add connectors to enable use with Range Spinal System conponents.
Materials: The devices are manufactured from Titanium Alloy and Chrome per ASTM and ISO standards.
Function: The system functions as an adjunct to fusion to provide immobilization of the posterior thoracic and lumbar spine.
The provided 510(k) summary for the "Everest Spinal System" is for a physical medical device (spinal fixation system) and not for an AI/ML-driven device or software. Therefore, the questions regarding acceptance criteria, study design, ground truth, and reader studies that are typically applicable to AI/ML device performance evaluation do not directly apply in this context.
This summary focuses on demonstrating substantial equivalence to existing predicate devices based on mechanical performance and material compatibility.
Here's an analysis of the provided information, framed as closely as possible to your request, but highlighting the differences for a physical device:
Acceptance Criteria and Device Performance (for a physical medical device)
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Mechanical Performance | - Static Compression Strength (referencing ASTM F1717) | "The modified Everest Spinal System constructs performed equally to or better than the predicates in static compression..." |
| Mechanical Performance | - Static Torsion Strength (referencing ASTM F1717) | "The modified Everest Spinal System constructs performed equally to or better than the predicates in... static torsion..." |
| Mechanical Performance | - Dynamic Compression Strength (referencing ASTM F1717) | "The modified Everest Spinal System constructs performed equally to or better than the predicates in... dynamic compression in accordance with ASTM F1717." |
| Design & Materials | - Design features and sizing comparison to predicates | "The design features and sizing of the components were also compared and the Everest Spinal System was found to be substantially the same as these systems." |
| Design & Materials | - Materials: Titanium Alloy and Chrome per ASTM and ISO standards | Confirmed to be manufactured from Titanium Alloy and Chrome per ASTM and ISO standards. |
| Intended Use | - Adjunct to fusion, posterior stabilization in thoracic-lumbar spine for specific indications (trauma, stenosis, curvatures, tumor, pseudoarthrosis, failed fusion, severe spondylolisthesis L5-S1) | The device is intended for these uses, consistent with predicates. |
Summary of Device Evaluation and Study for Substantial Equivalence:
The study presented is not an AI/ML algorithm validation study, but rather a comparative engineering and design study to demonstrate substantial equivalence of a physical medical device.
-
Sample size used for the test set and the data provenance:
- This is not applicable in the context of an AI/ML study. For a physical device, "test set" refers to physical prototypes or components of the Everest Spinal System. The document doesn't specify the exact number of components tested but implicitly suggests sufficient samples were used to conduct the required ASTM F1717 mechanical tests.
- Data Provenance: The tests were conducted according to ASTM and ISO standards, implying a controlled laboratory environment. The "country of origin of the data" would be where these tests were performed, but this detail is not provided. It is a prospective generation of test data from manufactured components.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- This is not applicable. The "ground truth" for a mechanical device is established by objective, verifiable physical measurements and engineering standards (ASTM F1717, ISO standards). There are no human "experts" establishing a subjective ground truth in this type of study.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. Mechanical tests are typically evaluated against predefined quantitative limits set by standards like ASTM F1717, not through human adjudication.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a physical device, and no human "readers" or AI assistance are involved in its performance evaluation.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For this physical device, the "ground truth" is defined by established engineering performance standards (ASTM F1717) and material specifications (ASTM and ISO standards for Titanium Alloy and Chrome). The performance of the device is directly compared to these objective standards and to the performance of existing legally marketed predicate devices under the same test conditions.
-
The sample size for the training set:
- Not applicable. There is no AI/ML model being trained.
-
How the ground truth for the training set was established:
- Not applicable. No AI/ML model training occurred.
Conclusion for this Device:
The K2M Everest Spinal System demonstrated substantial equivalence by:
- Showing that its mechanical performance (static compression, static torsion, dynamic compression) was "equally to or better than" the predicate devices when tested according to ASTM F1717.
- Confirming its design features and sizing were "substantially the same" as predicate systems.
- Verifying its materials (Titanium Alloy and Chrome) met the relevant ASTM and ISO standards.
- Confirming its intended use aligns with the predicate devices.
This allows the FDA to conclude there are no significant differences that would adversely affect the product's use, thus granting 510(k) clearance based on substantial equivalence.
{0}------------------------------------------------
APR - 4 2012
510(k) Summary for the Everest Spinal System
This 510(k) summary for the Everest Spinal System is provided as reguired per Section 513(i)(3) of the Food. Drug and Cosmetic Act.
- Submitter :
K2M, Inc. 751 Miller Drive SE. Suite F1 Leesburg, VA 20175 Contact Person : Nancy Giezen K2M, Inc. Telephone: 703-777-3155
Date Prepared: 03/02/12
- Tradename: Common Name: Classification Name:
Device Product Code: Regulatory Class:
Everest Spinal System Spinal Fixation System Spinal Interlaminal Fixation Orthosis (21CFR 888.3050), Pedicle screw spinal system (21CFR 888.3070). KWP, MNI, MNH Class II
- Predicate or legally marketed devices which are substantially equivalent :
- K2M Everest Spinal System (K081107) .
- K2M Range Spinal System (K070229, K072914, K080792, K081381, K100851, K112037, K112920) .
- . K2M Caspian Spinal System (K081107, K092540, K093534, K103232, K113654)
- DePuy Mountaineer (K110353 .
4. Description of the device:
The Everest Spinal System is a top-loading, multiple component, posterior (thoracic-lumba) spinal fixation system which consists of pedicle screws, rods, locking set screws, and hooks. The purpose of this submission is to increase the range of screw sizes and to add connectors to enable use with Range Spinal System conponents.
Materials: The devices are manufactured from Titanium Alloy and Chrome per ASTM and ISO standards.
Function: The system functions as an adjunct to fusion to provide immobilization of the posterior thoracic and lumbar spine.
5. Intended Use:
The Everest Spinal System if used in conjunction with the Range (Mesa and Denali) Spinal Systems are:
Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: Trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumbar screw stabilization as an adjunct to fusion for the following indications: degenerative disc disease (DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and
Ps 1 of 2
{1}------------------------------------------------
radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
-
- Comparison of the technological characteristics of the device to predicate and legally marketed devices : Proposed components of the Everest Spinal System were tested and compared to predicate devices. The modified Everest Spinal System constructs performed equally to or better than the predicates in static compression, static torsion and dynamic compression in accordance with ASTM F1717. The design features and sizing of the components were also compared and the Everest Spinal System was found to be substantially the same as these systems.
7. Conclusion:
There are no significant differences between the Everest Spinal System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized, curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
K2M. Incorporated % Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive Southeast, Suite F1 Leesburg, Virginia 20175
APR - 4 2012
Re: K120656
Trade/Device Name: Everest Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNI, MNH, KWP Dated: March 02, 2012 Received: March 05, 2012
Dear Ms. Giezen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Drederal statutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de rece related daverse overals) (1 CFR Part 820); and if applicable, the electronic form in the quant) 2) router (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Nancy Giezen
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please i go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin L. Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
K120656 510(k) Number (if known):
Device Name: Everest Spinal System
Indications for Use:
The Everest Spinal System if used in conjunction with the Range (Mesa and Denali) Spinal Systems are:
Non-cervical, pedicle screw fixation devices for posterior stabilization as an adjunct to fusion for the following indications: Trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the kyphono, and of lordone); cannot, por assates 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous boose graft having implants attached to the lumbar and sacral spine ( L3 to sacrum) with removal of the implants after the attainment of a solid fusion.
Non-cervical, non-pedicle spinal fixation devices intended for posterior or anterolateral thoracolumber screw stabilization as an adjunct to fosion for the following indications: degenerative disc disease (DDD ) (defined stabilization as in adjunct to rasson for degeneration of the disc confirmed by history and radiographic studies as back pan of thisolgonio origin was acgr dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, <yphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS-LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
K120656 510(k) Number.