(78 days)
The IMPLANET Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The IMPLANET Spine System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), turnor, pseudarthrosis, or revision of a failed fusion attempt.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the IMPLANET Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
The IMPLANET Spine System is a posterior instrumentation system. The polyaxial screws are made of Ti6Al4V titanium alloy compliant with ISO 5832-3. The polyaxial screw is comprised of three sections: the pedicle screw; the head; and a ring that connects the screw to the head. The polyaxial screws are offered in diameters ranging from 5.0 to 7.5 mm and in lengths ranging from 35 to 60 mm.
The monoaxial pedicle screws are made of Ti6A!4V titanium alloy compliant with ISO 583203 and are available in 5.0 mm, 6.0 mm, 7.0 mm, and 8.0 mm diameters. The screws range in length from 35 to 60 mm.
The system includes both straight and pre-bent rods made of Ti6Al4V titanium alloy.
The transverse connectors are composed of Ti6Al4V titanium alloy compliant with ISO 5832-3. These connectors are used to build a transverse connection between two union rods.
The hooks are made of Ti6AI4V titanium alloy and are provided in multiple configurations.
The IMPLANET Spine System has principles of operation substantially similar to other pedicle screw-based systems for the indications listed.
The IMPLANET Spine System components may be used for posterior pedicle screw fixation in pediatric cases: polyaxial and monoaxial screws, rods, transverse connectors and rods.
The purpose of the subject 510(k) notice is the addition of new indications and the inclusion of intermediate screw sizes to the company's cleared system.
The provided text describes the Implanet S.A.'s IMPLANET Spine System, a spinal fixation device. However, it does not contain information related to software or AI/ML performance, acceptance criteria, or studies for such, as the device is a physical implant. The document is a 510(k) summary for a medical device (spinal fixation system) that primarily focuses on mechanical and material properties, biocompatibility, and manufacturing standards.
Therefore, I cannot provide the requested information about acceptance criteria and studies for a digital health device. The document describes:
- Device Description: Physical components like polyaxial and monoaxial pedicle screws, rods, transverse connectors, and hooks, all made of Ti6Al4V titanium alloy.
- Technological Characteristics: Substantially similar to other pedicle screw-based systems.
- Performance Data: This section only refers to bench testing (static and dynamic mechanical tests), sterility validation, shelf life validation, and biocompatibility testing. It specifically states that these tests were performed on the previous version of the device (Calypso System) and some biocompatibility tests were reperformed on the current version.
- Substantial Equivalence: Concludes that the device is substantially equivalent to predicate devices based on intended use, indications, technological characteristics, and principles of operation, with minor differences raising no new safety or effectiveness issues.
Conclusion:
The provided text does not contain any information about:
- Acceptance criteria related to AI/ML performance.
- Reported device performance for AI/ML.
- Sample size for test sets in an AI/ML context.
- Data provenance for AI/ML training/testing.
- Number or qualifications of experts for AI/ML ground truth.
- Adjudication methods for AI/ML ground truth.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness studies with AI assistance.
- Standalone algorithm-only performance studies.
- Type of ground truth used (expert consensus, pathology, outcomes data) in an AI/ML context.
- Sample size or establishment of ground truth for a training set in an AI/ML context.
{0}------------------------------------------------
510(k) SUMMARY
Implanet S.A.'s IMPLANET Spine System
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Implanet S.A. Technopole Bordeaux Montesquieue Allée François Magendie Phone +33 557 995 555 Facsimile: +33 557 995 700
Contact Person: Franck Rigal, Director of Quality and Regulatory Affairs
Date Prepared: September 15, 2013
Name of Device
IMPLANET Spine System
Common or Usual Name
Spinal fixation device
Classification Name
888.3070 - Spinal Pedicle Fixation Orthosis - NKB, OSH, MNI, MNH
888.3050 - Spinal Interlaminal Fixation Orthosis - KWP
Predicate Devices
Implanet S.A.'s Calypso System (K120564)
Medtronic Sofamor Danek's CD HORIZON® Spinal System (K091445)
Intended Use / Indications for Use
The IMPLANET Spine System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The IMPLANET Spine System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), turnor, pseudarthrosis, or revision of a failed fusion attempt.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the IMPLANET Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
OCT 10 2013
{1}------------------------------------------------
Device Description
The IMPLANET Spine System is a posterior instrumentation system. The polyaxial screws are made of Ti6Al4V titanium alloy compliant with ISO 5832-3. The polyaxial screw is comprised of three sections: the pedicle screw; the head; and a ring that connects the screw to the head. The polyaxial screws are offered in diameters ranging from 5.0 to 7.5 mm and in lengths ranging from 35 to 60 mm.
The monoaxial pedicle screws are made of Ti6A!4V titanium alloy compliant with ISO 583203 and are available in 5.0 mm, 6.0 mm, 7.0 mm, and 8.0 mm diameters. The screws range in length from 35 to 60 mm.
The system includes both straight and pre-bent rods made of Ti6Al4V titanium alloy.
The transverse connectors are composed of Ti6Al4V titanium alloy compliant with ISO 5832-3. These connectors are used to build a transverse connection between two union rods.
The hooks are made of Ti6AI4V titanium alloy and are provided in multiple configurations.
The IMPLANET Spine System has principles of operation substantially similar to other pedicle screw-based systems for the indications listed.
The IMPLANET Spine System components may be used for posterior pedicle screw fixation in pediatric cases: polyaxial and monoaxial screws, rods, transverse connectors and rods.
The purpose of the subject 510(k) notice is the addition of new indications and the inclusion of intermediate screw sizes to the company's cleared system.
Technological Characteristics
The IMPLANET Spine System consists of: monoaxial pedicle screws, union rods, transverse connectors, and hooks.
Performance Data
This 510(k) premarket notification incorporates by reference the bench testing performed in support of Implanet S.A.'s Calypso System (K120564), which is listed below. All bench testing confirmed that the product met the necessary specifications. Sterilization and shelf life validation testing conducted for the Calypso System in accordance with recognized industry standards are also incorporated by reference. In addition, the biocompatibility of the device was confirmed in accordance with ISO 10993. A list of the tests performed on the previous version of the device to support substantial equivalence is provided below:
- Static axial gripping capacity, static flexion/extension bending, static axial torque . gripping capacity - ASTM F1798;
- Static compression bending ASTM F1717; .
- Static torsion ASTM F1717; .
- Dynamic compression bending ASTM F1717; .
- Shelf life ASTM F 1980; .
- Implant sterilization validation ISO 11137; t
- Instrument cleaning and sterilization validation ISO 17665; .
- Cytotoxicity ISO 10993; .
- . Acute systemic toxicity - ISO 10993.
{2}------------------------------------------------
In addition, a list of the tests performed again on the current version of the device to support substantial equivalence is provided below:
- Cytotoxicity ISO 10993; ●
- Chemical characterization ISO 10993. .
Substantial Equivalence
When used for pediatric use, the IMPLANET Spine System is substantially similar to Implanet S.A.'s Calypso System and Medtronic Sofamor Danek's CD HORIZON® Spinal System. The IMPLANET Spine System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the IMPLANET Spine System and its predicate devices raise no new issues of safety or effectiveness. Thus, the IMPLANET Spine System is substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized graphic of three curved lines, resembling a person embracing another person. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the graphic. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 10, 2013
Implanet S.A. % Ms. Janice M. Hogan Hogan Lovells US LLP 1835 Market Street, 2016 Floor Philadelphia, Pennsylvania 19103
Re: K132303
Trade/Device Name: IMPLANET Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, KWP. MNI. MNH Dated: July 24, 2013 Received: July 24, 2013
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA `s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 ) CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Janice M. Hogan
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Erin I. Keith
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): _ K132303
Device Name: IMPLANET Spine System
Indications for Use:
The IMPLANET Spine System is intended to provide immobilization of spinal segments in skeletally mature palients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The IMPLANET Spine System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudarthrosis, or revision of a failed fusion attempt.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the IMPLANET Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scollosis. These devices are to be used with autografi and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Amy S. Graf -S
(for RPJ)
(Division Sign-off) Division of Orthopedic Devices 510(k) Number: K132303
Page 1 of 1
N/A