(143 days)
The Implanet Calypso System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The Implanet Calypso System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis, or revision of a failed fusion attempt.
The Calypso System is a posterior instrumentation system consisting of monoaxial and polyaxial pedicle screws, union rods, transverse connectors, and hooks. The implants in the system are composed of Ti6Al4V titanium alloy described by ISO 5832-3.
The Implanet S.A.'s Calypso System is a spinal fixation device. The provided text details the performance data for the device, focusing on its mechanical strength and biological compatibility to demonstrate substantial equivalence to predicate devices. It does not describe an AI medical device. Therefore, several of the requested categories (e.g., number of experts for ground truth, adjudication method, MRMC study, sample size for training set, how ground truth for training set was established) are not applicable as they pertain to AI/ML device studies.
Here's a summary of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Type | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Mechanical Testing | ||
| Static Axial Gripping Capacity | Met necessary specifications (ASTM F1798) | Confirmed that the product met the necessary specifications. |
| Static Flexion/Extension Bending | Met necessary specifications (ASTM F1798) | Confirmed that the product met the necessary specifications. |
| Static Axial Torque Gripping Capacity | Met necessary specifications (ASTM F1798) | Confirmed that the product met the necessary specifications. |
| Static Compression Bending | Met necessary specifications (ASTM F1717) | Confirmed that the product met the necessary specifications. |
| Static Torsion | Met necessary specifications (ASTM F1717) | Confirmed that the product met the necessary specifications. |
| Dynamic Compression Bending | Met necessary specifications (ASTM F1717) | Confirmed that the product met the necessary specifications. |
| Biocompatibility | ||
| Cytotoxicity | In accordance with ISO 10993 | Biocompatibility "confirmed" in accordance with ISO-10993. |
| Acute Systemic Toxicity | In accordance with ISO 10993 | Biocompatibility "confirmed" in accordance with ISO-10993. |
| Sterilization & Shelf Life Validation | ||
| Shelf Life | In accordance with ASTM 1980 | Conducted in accordance with recognized industry standards. |
| Implant Sterilization Validation | In accordance with ISO 11137 | Conducted in accordance with recognized industry standards. |
| Instrument Cleaning & Sterilization Validation | In accordance with ISO 17665 | Conducted in accordance with recognized industry standards. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the "sample size" in terms of number of devices tested for each mechanical test. It mentions "bench testing," implying a set of physical samples were tested.
- Data Provenance: The tests are bench tests performed by Implanet S.A. No country of origin for data or retrospective/prospective status is relevant for these types of mechanical and biological tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. Ground truth for mechanical and biological testing is established by adherence to recognized international standards (e.g., ASTM, ISO) and laboratory measurements, not by expert consensus interpreting complex data like medical images.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation (e.g., imaging studies) to resolve disagreements. Here, objective measurements against pre-defined standards are used.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. This is a spinal fixation device, not an AI medical device, so an MRMC study is not relevant to its regulatory submission.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
No. This is a physical medical device, not an algorithm.
7. Type of Ground Truth Used
The ground truth used for this device's evaluation is defined by established industry standards and specifications (e.g., ASTM F1798, ASTM F1717, ISO 10993, ISO 11137, ISO 17665). The measurements taken during bench testing are compared directly against these quantitative and qualitative criteria.
8. Sample Size for the Training Set
Not applicable. This device does not involve machine learning, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no training set for an AI/ML model, there is no ground truth to establish for such a set.
{0}------------------------------------------------
510(k) SUMMARY
lmplanet S.A.'s Calypso System
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Implanet S.A. Technopole Bordeaux Montesquieu Allée Francois Magendie 33650 Martillac France Phone: +33 557 995 555 Facsimile: +33 557 995 700
Contact Person: Franck Rigal, Director of Quality and Regulatory Affairs
Date Prepared: July 12, 2012
Name of Device
Calypso System
Common or Usual Name
Spinal fixation device
Classification Name
888.3070 - Pedicle Screw Spinal System
888.3050 - Spinal interlaminal fixation orthosis
Predicate Devices
Aesculap Implant Systems, Inc.'s S4 Spinal System (K100623)
Medtronic Sofamor Danek USA, Inc.'s CD HORIZON Spinal System (K113174)
Intended Use / Indications for Use
The Implanet Calypso System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The Implanet Calypso System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis, or revision of a failed fusion attempt.
JUL 16 2012
{1}------------------------------------------------
Technological Characteristics
The Calypso System is a posterior instrumentation system consisting of monoaxial and polyaxial pedicle screws, union rods, transverse connectors, and hooks. The implants in the system are composed of Ti6Al4V titanium alloy described by ISO 5832-3.
Performance Data
In support of this 510(k) Premarket Notification, Implanet S.A. has conducted bench testing to demonstrate that the Calypso System provides adequate mechanical strength for its intended use. All bench testing confirmed that the product met the necessary specifications. In addition, the biocompatibility of the device has been confirmed in accordance with ISO-10993, and the company has conducted sterilization and shelf life validation in accordance with recognized industry standards. A list of the tests performed to support substantial equivalence is provided below:
- Static axial gripping capacity, static flexion/extension bending, static axial torque gripping . capacity- ASTM F1798
- Static compression bending ASTM F1717 .
- Static torsion ASTM F1717 .
- Dynamic compression bending ASTM F1717 ●
- Cytotoxicity ISO 10993 ●
- Acute systemic toxicity ISO 10993 ●
- . Shelf life - ASTM 1980
- Implant sterilization validation ISO 11137 ●
- Instrument cleaning and sterilization validation ISO 17665
Substantial Equivalence
The Calypso System is very similar to the Aesculap Implant Systems, Inc.'s S4 Spinal System and Medtronic Sofamor Danek USA, Inc.'s CD HORIZON Spinal System. The Calypso System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the Calypso System and its predicate devices raise no new issues of safety or effectiveness. Performance data, including mechanical testing in static compression bending, static torsion, and dynamic compression bending, as well as biocompatibility and sterility testing, demonstrate that the minor differences between the Calvoso System and the predicates do not adversely impact its performance. Thus, the Calypso System is substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The background is white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Hogan Lovells US, LLP % Ms. Janice Hogan 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K120564
Trade/Device Name: Calypso System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, MNI, MNH, KWP Dated: June 21, 2012 Received: June 21, 2012
JUL 16 2012
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Ms. Janice Hogan
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement
K120564 510(k) Number (if known):_
Device Name: Calypso System
Indications for Use:
The Implanet Calypso System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion of the thoracic, lumbar and/or sacral spine. The Implanet Calypso System is intended for posterior, non-cervical pedicle and non-pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal deformities (i.e., scoliosis, kyphosis and/or lordosis), tumor, pseudoarthrosis, or revision of a failed fusion attempt.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Oil) Division of Surgical, Orthopedic, and Restorative Devices
Page I of I
KI20564 10(k) Number_
18
N/A