(93 days)
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft and/or allograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, Scheuermann's disease, and/or lordosis,), tumor, stenosis, pseudoarthrosis, or failed previous fusion.
The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the above indications.
For pediatric patients, the Polaris System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to fusion to treat adolescent idiopathic scoliosis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by tumor and/or trauma. Pedicle screw fixation is limited to a posterior approach.
The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, Nextgen Altius OCT Spine System. The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
The Polaris Spinal System is a non-cervical spinal fixation device made from titanium alloy (Ti-6Al-4V) per ASTM F 136, unalloyed titanium per ASTM F 67, stainless steel per ASTM F 138 or ASTM F 1314 and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM 1537. The system includes screws, various types and sizes of rods, locking nuts, hooks. lateral connectors, plugs. fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This line extension is to expand the indications for use for Polaris Spinal System using previous cleared system components.
This document describes a 510(k) submission for the Polaris Spinal System, a medical device. The information provided heavily focuses on the comparison to predicate devices to establish substantial equivalence rather than detailing specific device performance or clinical studies with acceptance criteria for a novel AI or diagnostic device.
Therefore, many of the requested categories for a study proving device meets acceptance criteria are not directly applicable or are not explicitly stated in the provided text. The document is primarily a regulatory submission demonstrating equivalence, not a detailed performance study report.
Here's an analysis based on the provided text, highlighting what is available and what is not:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric / Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical Strength (Static Compression Bending, Static Torsion, Dynamic Compression Bending per ASTM F1717) | Implicit historical data: "The implants included in the submission were previously tested for static compression bending, static torsion and dynamic compression bending per ASTM F1717." |
| Safety and Effectiveness for Pediatric Pedicle Screw Fixation (especially Adolescent Idiopathic Scoliosis) | "The assessment concluded that pedicle screw fixation is safe and effective for use in pediatric patients for the treatment of adolescent idiopathic scoliosis." (Based on clinical literature assessment) |
| Substantial Equivalence (Intended Use, Indications, Design, Materials, Manufacturing Methods, Sterility, Fundamental Technology, Operational Principles) | The Polaris Spinal System is deemed "substantially equivalent" to multiple predicate devices (Revere" 4.5 Stabilization System, CD Horizon" Spinal System, TSRH* Spinal System, and previous Polaris Spinal System versions) based on these characteristics. |
Missing Information/Not Applicable: The document does not specify quantitative acceptance criteria (e.g., "must achieve X N-m of torque" or "failure rate must be below Y%"). The "reported device performance" for mechanical strength is a statement that it was "previously tested" and "already demonstrated," indicating compliance with prior standards or predicate device performance, but no specific values or criteria are provided in this summary.
2. Sample Size Used for the Test Set and Data Provenance
The document does not describe a "test set" in the context of a clinical study with a specific sample size.
- Mechanical Testing: While mechanical tests per ASTM F1717 were performed, the sample size for these tests is not mentioned.
- Clinical Literature Assessment: This involved reviewing "published clinical data." The number of patients or studies included in this assessment is not specified, nor is the country of origin, or whether it was retrospective or prospective. It's a review of existing data, not a new data collection.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not applicable. The document does not describe a study involving expert-established ground truth for a test set. The clinical literature assessment likely involved review by the regulatory affairs team, but not in the context of establishing "ground truth" for a device performance study.
4. Adjudication Method for the Test Set
This information is not applicable as no "test set" requiring adjudication is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was done or mentioned. This device is a surgical implant, not a diagnostic imaging AI, so such a study would not typically apply in this direct comparison context.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)
This information is not applicable. The Polaris Spinal System is a physical implant, not an algorithm. Standalone performance for mechanical devices relates to their physical integrity and function under simulated use, which is addressed by the ASTM F1717 testing.
7. Type of Ground Truth Used
- Mechanical Testing: The "ground truth" for mechanical testing would be the physical properties and performance metrics defined by the ASTM F1717 standard.
- Clinical Efficacy (for pediatric pedicle screw fixation): The "ground truth" was derived from a "clinical literature assessment," meaning conclusions drawn from existing published clinical data and expert consensus in the medical community about the safety and effectiveness of pedicle screw fixation for the stated conditions.
8. Sample Size for the Training Set
This information is not applicable, as there is no "training set" in the context of a medical device like this. Training sets are relevant for machine learning algorithms.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable, as there is no "training set" for this device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Biomet Spine. The logo is in black and white, with the word "BIOMET" in large, bold letters. Below the word "BIOMET" is the word "SPINE" in smaller letters. The logo is simple and clean, and it is likely used to represent the company's brand.
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
| Preparation Date: | May 31, 2013 |
|---|---|
| Applicant/Sponsor: | Biomet Spine399 Jefferson RoadParsippany, NJ 07054 |
| Contact Person: | Vivian KellyRegulatory Affairs Project ManagerPhone: 973-299-9300 x2214Fax: 973-887-1347 |
| Trade name: | Polaris Spinal System |
| Common Name: | Non-cervical spinal fixation system |
| Classification Name(Product Code): | Posterior, noncervical, nonpedicle use (KWP)Anterior/anterolateral noncervical use (KWQ)Noncervical pedicle applications (MNI, MNH and NKB)Pedicle screw spinal system, adolescent idiopathic scoliosis (OSH) |
Device Panel - Regulation No.: Orthopedic - 21 CFR 888.3050, 888.3060 and 888.3070
Device Description:
The Polaris Spinal System is a non-cervical spinal fixation device made from titanium alloy (Ti-6Al-4V) per ASTM F 136, unalloyed titanium per ASTM F 67, stainless steel per ASTM F 138 or ASTM F 1314 and Cobalt Chrome Alloy (Co-28Cr-6Mo) per ASTM 1537. The system includes screws, various types and sizes of rods, locking nuts, hooks. lateral connectors, plugs. fixation washers, rod connectors/dominos and various cross connectors. Various instruments are also available for use by the surgeon to facilitate implantation of the device. This line extension is to expand the indications for use for Polaris Spinal System using previous cleared system components.
Indications for Use:
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft and/or allograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, Scheuermann's disease, and/or lordosis,), tumor, stenosis, pseudoarthrosis, or failed previous fusion.
The Ballista/Ballista II instruments are intended to be used with Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista II instruments to provide the surgeon with a percutaneous approach for posterior spinal surgery for the above indications.
{1}------------------------------------------------
For pediatric patients, the Polaris System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to fusion to treat adolescent idiopathic scoliosis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by tumor and/or trauma. Pedicle screw fixation is limited to a posterior approach.
The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, Nextgen Altius OCT Spine System. The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
Summary of Technologies:
The technological characteristics of the Polaris Spinal System remain the same as, or similar to. the predicate devices in regards to intended use, indications, design, materials, manufacturing methods, sterility, fundamental technology, and operational principles.
Performance Data:
The implants included in the submission were previously tested for static compression bending, static torsion and dynamic compression bending per ASTM F1717. Because the system's mechanical strength has already been demonstrated, additional testing is not required to demonstrate substantial equivalence for expanded indications and an engineering rationale was provided.
To support substantial equivalence, a clinical literature assessment was conducted using published clinical data for pedicle screw fixation for adolescent idiopathic scoliosis. The assessment concluded that pedicle screw fixation is safe and effective for use in pediatric patients for the treatment of adolescent idiopathic scoliosis.
Substantial Equivalence:
The Polaris Spinal System is substantially equivalent to the Revere" 4.5 Stabilization System, Globus Medical Inc. (K113395), the CD Horizon" Spinal System. Medtronic (K091445. K111457 & K 113395) and the TSRH* Spinal System, Medtronic (K 110070) for pediatric pedicle screw use and the Polaris Spinal System (K123549 & K090203) in regards to material, design and operational principles.
Conclusion:
The Polaris Spinal System is substantially equivalent to the predicate systems as spinal fixation devices in regards intended use, indications, design, materials, manufacturing methods, sterility, fundamental technology, and operational principles. Furthermore, preclinical data, the clinical literature assessment review and other supporting information sufficiently demonstrate the substantial equivalence of the subject device to the Polaris Spinal System and the other named predicate systems, which have been cleared for pedicle screw fixation for adolescent idiopathic scoliosis. Scheuermann's Disease, and spondylolysis.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 4, 2013
Biomet Spine Ms. Vivian Kelly Regulatory Affairs Project Manager 399 Jefferson Road Parsippany, New Jersey 07054
Re: K131615
Trade/Device Name: Polaris Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: August 12, 2013 Received: August 15, 2013
Dear Ms. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Vivian Kelly
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K131615
Device Name: Polaris Spinal System
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system for use with autograft and/or allograft. The Polaris Spinal System is indicated for the following conditions: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiouraphic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, Schevermann's disease, and/or lordosis.), tumor, stenosis, pseudoarthrosis, or failed previous fusion.
The Ballista/Ballista II instruments are intended to be used with Ballista/Ballista II/Polaris 5.5mm implants. Cannulated screws and percutaneous rods may be used with the Ballista/Ballista II instruments to provide the surgeon with a perculaneous approach for posterior spinal surgery for the above indications.
For pediatric patients, the Polaris System may be used for posterior, non-cervical pedicle screw fixation as an adjunct to freat adolescent idiopathic scolinsis and is also indicated for treatment of the following conditions: spondylolisthesis/spondylolysis and fractures caused by tumor and/or trauma. Pedicle screw fixation is limited to a posterior approach.
The Polaris Spinal System may be used with the instruments in the AccuVision Minimally Invasive Spinal Exposure System to provide the surgeon with a minimally invasive approach for posterior spinal surgery.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, Nextgen Altius OCT Spine System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
Prescription Use N (Part 21 CFR 801 Subpart D)
ANDAOR
()ver-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Ronald P. Jean -S
(Division Sign-Off) Division of Orthonedic Devices 510(k) Number: K131615
N/A