K Number
K130603

Validate with FDA (Live)

Date Cleared
2014-03-27

(385 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PATHFAST® cTnl Calibrators are for calibration of the PATHFAST® system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.

Device Description

The PATHFAST cTnl Calibrators are for calibration of the PATHFAST cTnl-II test performed on the PATHFAST instrument. PATHFAST cTnI-II test is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction.

Currently, PATHAST cTnl Calibrators 1 and 2 are provided as lyophilized products of two vials each. Calibrator 1 consists of buffer only and Calibrator 2 consists of cTnI complex in buffer. Four vials of calibrator diluent also are provided for reconstitution of the calibrators. The diluent consists of an aqueous solution with 0.05% sodium azide. This Special 510(k) is being submitted for a change to a liquid Calibrator 1 formulation and a new dropper bottle container. There are no changes to PATHFAST cTnl Calibrator 2. As a result of elimination of the reconstitution step for PATHFAST cTnI Calibrator 1, the number of bottles of Calibrator Diluent is being reduced from four bottles to two bottles.

AI/ML Overview

The provided document describes the PATHFAST® cTnI Calibrator 1, which underwent a Special 510(k) submission for a change in formulation from lyophilized to liquid and a new container type. The purpose of the submission is to demonstrate substantial equivalence to the predicate device (PATHFAST® cTnI Calibrator 1, lyophilized format, K100130).

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly detailed in a quantitative table. Instead, the document states that the modified device "met all specifications for the studies identified in the company's design control system Risk Analysis." The studies performed were:

Study AreaPerformance MetricReported Device Performance
FormulationAssay sensitivity, accuracy, precision, method comparison, sample type and matrix comparisonMet all specifications
FormatReal-time stabilityMet all specifications
ContainerElution and evaporationMet all specifications

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for the test sets used in the validation/verification studies. It only mentions the types of studies performed (assay sensitivity, accuracy, precision, method comparison, sample type and matrix comparison, real-time stability, elution, and evaporation).

The data provenance (country of origin, retrospective/prospective) is not mentioned.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. This device is a calibrator, and its performance is evaluated against established analytical specifications and comparison to a predicate device, rather than through expert-established ground truth on clinical data.

4. Adjudication Method for the Test Set

Not applicable.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This is a medical device calibrator, not an AI-assisted diagnostic tool for human readers.

6. Standalone Performance

Yes, the studies conducted (formulation, format, container evaluations) represent standalone performance of the modified calibrator against its own specifications and in comparison to the predicate device's performance characteristics. The studies directly assess the analytical performance of the calibrator itself.

7. Type of Ground Truth Used

The "ground truth" for the calibrator's performance is established by:

  • Analytical specifications: The device is expected to meet predefined performance targets for sensitivity, accuracy, precision, etc.
  • Comparison to predicate device: The modified calibrator's performance is compared to that of the legally marketed predicate device (lyophilized format) to demonstrate substantial equivalence.
  • Reference materials/methods: Implicitly, the accuracy and precision studies would rely on reference materials or methods to determine true values.

8. Sample Size for the Training Set

Not applicable. This device is a calibrator and does not involve AI/machine learning models that require training sets in the conventional sense. Its purpose is to calibrate an existing diagnostic system, not to perform a diagnostic function itself based on learned patterns.

9. How the Ground Truth for the Training Set Was Established

Not applicable.

{0}------------------------------------------------

Special 510 (k) – PATHFAST® cTnI Calibrator 1

510(k) SUMMARY

February 28, 2014

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K130603

CONTACT:Judi SmithPrecision for Medicine2 Bethesda Metro Center, Suite 850Bethesda, MD 20814on behalf of:Mitsubishi Chemical Medience Corporation2-8, Shibaura 4-chome, Minato-kuTokyo, 108-8559 Japan
NAME OF DEVICE:Trade Name:PATHFAST cTnI Calibrator 1
Common/Classification Names:Calibrator
Regulation Number:862.1250
Product Code:JIT
Class:II
PREDICATE DEVICE:PATHFAST cTnI Calibrator 1 which issupplied with the PATHFAST cTnI-II Test[K100130, 5/16/2011]

DEVICE DESCRIPTION:

INTENDED USE:

The PATHFAST® cTnI Calibrators are for calibration of the PATHFAST® system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.

PRODUCT DESCRIPTION:

The PATHFAST cTnl Calibrators are for calibration of the PATHFAST cTnl-II test performed on the PATHFAST instrument. PATHFAST cTnI-II test is an in vitro diagnostic test for the quantitative measurement of cardiac Troponin I (cTnl) in heparinized or EDTA whole blood and plasma. Measurements of

{1}------------------------------------------------

Special 510 (k) - PATHFAST® cTnI Calibrator 1 510(k) SUMMARY (cont.)

cardiac Troponin I are used as an aid in the diagnosis of acute myocardial infarction.

Currently, PATHAST cTnl Calibrators 1 and 2 are provided as lyophilized products of two vials each. Calibrator 1 consists of buffer only and Calibrator 2 consists of cTnI complex in buffer. Four vials of calibrator diluent also are provided for reconstitution of the calibrators. The diluent consists of an aqueous solution with 0.05% sodium azide. This Special 510(k) is being submitted for a change to a liquid Calibrator 1 formulation and a new dropper bottle container. There are no changes to PATHFAST cTnl Calibrator 2. As a result of elimination of the reconstitution step for PATHFAST cTnI Calibrator 1, the number of bottles of Calibrator Diluent is being reduced from four bottles to two bottles.

SUBSTANTIAL EQUIVALENCE:

The PATHFAST cTnl Calibrator 1 (liquid format) is substantially equivalent to the PATHFAST cTnl Calibrator 1 (lyophilized format) (K100130) in intended use and design. The similarities and differences are listed below.

SIMILARITIES
ItemPATHFAST cTnI Calibrator 1Lyophilized(predicate device)PATHFAST cTnI Calibrator 1Liquid(Modified)
FDA SubmissionNumberK100130N/A
Intended UseThe PATHFAST® cTnICalibrators are for calibration ofthe PATHFAST® system whenused for the quantitativedetermination of cardiac TroponinI in human heparinized or EDTAwhole blood and plasma.No change
FundamentalScientificTechnologycTnI concentration = 0 ng/mL.Provides zero analyte level foruser calibration curve.No change
ValueAssignmentPrimary calibrator, mastercalibrator, stock solution,working calibratorNo change
InstructionsDispense approximately 100 µLof CAL-1 and CAL-2 in samplewells to load on PATHFAST.No change

Substantial Equivalence Comparison Table

{2}------------------------------------------------

Special 510 (k) - PATHFAST® cTnI Calibrator 1 510(k) SUMMARY (cont.)

DIFFERENCES
ItemPATHFAST cTnI Calibrator 1Lyophilized(predicate device)PATHFAST cTnI Calibrator 1Liquid(Modified)
FormulationLyophilized MES pH 6.0,lactose, and enzyme free humanserum, DTTLiquid saline solution with 0.05%sodium azide as preservative
FormatLyophilizedLiquid
Calibrator 1ContainerSilicon-coated glass bottle withsilicone rubber and plastic screwcapPolypropylene droplet bottle withhigh density polyethylene nozzleand polypropylene screw cap
CalibratorQuantityCAL-1 (1mL lyophilized×2vials)CAL-2 (1mL lyophilized×2vials)Calibrator Diluent (1mL×4bottles)CAL-1 (2mL liquid×1 bottle)CAL-2 (1mL lyophilized ×2vials)(no change)Calibrator Diluent (1mL×2bottles)
InstructionsReconstitute each vial of CAL-1and CAL-2 with one bottle(1mL) of Calibrator Diluent.Reconstitute CAL-2 with onebottle (1mL) of CalibratorDiluent.

As part of the company's Design Control system, the following validation/verifications studies were performed.

  • Formulation: assay sensitivity, accuracy, and precision, method comparison, . sample type and matrix comparison
  • Format: real time stability .
  • Container: elution and evaporation .

The results of the validation and verification testing demonstrate that the modified PATHFAST cTnl Calibrator 1 met all specifications for the studies identified in the company's design control system Risk Analysis. Therefore, the modified PATHFAST cTnl Calibrator 1 is substantially equivalent to the current 510(k)-cleared PATHFAST cTnI Calibrator 1.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Aveme Document Control Center - WO66-G60 Silver Spring, MD 20993-0002

March 27, 2014

MITSUBISHI CHEMICAL MEDIENCE CORPORATION C/O JUDI SMITH 2 BETHESDA METRO CENTER SUITE 850 BETHESDA MD 20814

Rc: K130603

Trade/Device Name: PATHFAST cTnl Calibrators Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT Dated: March 18. 2014 Received: March 19. 2014

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassilied in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however. that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Ms. Smith

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Courtney H. Lias -S

Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K130603

Device Name

PATHFAST® cTnI Calibrators

Indications for Use (Describe)

The PATHFAST® cTnl Calibrators are for calibration of the PATHFAST® system when used for the quantitative determination of cardiac Troponin I in human heparinized or EDTA whole blood and plasma.

Type of Use (Select one or both, as applicable)

12 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

A. Chesler -S

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.