K Number
K130083

Validate with FDA (Live)

Manufacturer
Date Cleared
2013-07-17

(184 days)

Product Code
Regulation Number
878.4493
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Demetech Absorbable Poliglecaprone 25 Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

Device Description

DemeCAPRONE (Poliglecaprone 25) is a synthetic monofilament absorbable surgical suture composed of Poly (glycolic-co-caprolactone) copolymer (PGCL) and is supplied un-dyed and dyed with D&C Violet #2 below 0.1wt%. DemeCAPRONE (Poliglecaprone 25) synthetic absorbable suture is available in sizes 6-0 through 1 (metric sizes 0.7 - 4). DemeCAPRONE (Poliglecaprone 25) Surgical Suture meets the requirements established by the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures except for diameter.

AI/ML Overview

The provided document describes the 510(k) summary for DemeCAPRONE (Poliglecaprone 25) Synthetic Monofilament Absorbable Suture. This document focuses on demonstrating substantial equivalence to predicate devices for a surgical suture, not an AI/ML powered medical device. Therefore, much of the requested information regarding AI/ML device studies (e.g., sample sizes for training/test sets, expert adjudication, MRMC studies, standalone performance) is not applicable to this submission.

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text, while acknowledging the inherent differences for a sutures submission:

1. A table of acceptance criteria and the reported device performance:

The acceptance criteria for DemeCAPRONE (Poliglecaprone 25) suture are based on demonstrating equivalence to legally marketed predicate devices and conformance to established standards, primarily the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures, and ISO 10993 for biocompatibility.

Acceptance Criteria (Set by USP & ISO 10993)Reported Device Performance (DemeCAPRONE)
Material Composition: Composed of Poly (glycolic-co-caprolactone) copolymer (PGCL)Same as predicate devices. Composed of Poly (glycolic-co-caprolactone) copolymer.
Physical Form: Sterile, flexible, monofilament threadSame as predicate devices. Sterile, flexible, monofilament thread.
Tensile Strength (< 881>): Meets or exceeds USP performance requirements.Meets or exceeds the performance requirements defined in the United States Pharmacopeia for "Tensile Strength" < 881 >.
Needle Attachment (< 871>): Meets or exceeds USP performance requirements.Meets or exceeds the performance requirements defined in the United States Pharmacopeia and the current edition USP for "Needle Attachment" < 871 >.
Suture Length Requirement: Meets 95% of stated label length (per USP).Meets or exceeds the performance requirements defined in the United States Pharmacopeia for "Suture Length Requirement" (95% of stated label length).
Sterility: Meets the performance requirements defined in the United States Pharmacopeia (current edition U.S.P.).Meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. for sterility.
Packaging & Labeling: Packed in a sterile, single/double manner with labeling conforming to 21 CFR and current USP.Is packaged in a same or equivalent manner with sterile single or double packaging having labeling conforming to 21 CFR and Current edition of USP.
Biocompatibility (ISO 10993): Biologically compatible.Is biologically compatible when tested as per ISO-10993. (Specific tests performed: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity, Genotoxicity - Bacterial Reverse Mutation and Chromosomal Aberration, Bone Marrow Micronucleus, Subchronic Toxicity (4-week, following subcutaneous implantation), Muscle Implantation (12-week)).
Toxicity (ISO 10993): Non-toxic.Is tested and proved to be non-toxic when tested as per ISO-10993 for toxicity.
Absorption Profile (In-vivo and In-vitro): Demonstrates substantial equivalence to predicate devices.In-vivo and in-vitro resorption testing was conducted to further demonstrate substantial equivalence to the predicate devices. (Specific data not provided in the summary, but implied to be acceptable for equivalence).
Extractable Color: Assured conformance to USP methods.Functionality testing assured that the device conformed with extractable color to methods outlined in USP 35. (Specific data not provided in the summary, but implied to be acceptable).
Suture Diameter: Assured conformance to USP methods (except where noted).Functionality testing assured that the device conformed with suture diameter to methods outlined in USP 35, except for diameter. (This is a noted exception, but still considered acceptable due to overall equivalence).

2. Sample sizes used for the test set and the data provenance:

  • Sample Size for Test Set: The document does not specify exact sample sizes for each non-clinical test (e.g., number of sutures tested for tensile strength, length, etc.). However, it states that "Non-clinical testing was conducted on the device per FDA's Special Control Guidance Document: Surgical Sutures, to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and invivo and in-vitro resorption." This implies that the sample sizes used were in accordance with the specified standards and guidance documents.
  • Data Provenance: The studies were non-clinical (laboratory and animal studies for biocompatibility, in-vivo/in-vitro resorption). The location of the testing facilities is not explicitly stated, but the applicant is Demetech Corporation, Miami Lakes, FL, USA.
  • Retrospective or Prospective: The testing would have been prospective, as new tests were performed on the DemeCAPRONE suture to demonstrate its properties.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This question is not applicable for this type of device submission. Ground truth, in the context of AI/ML, refers to human expert annotations. For a surgical suture, the "ground truth" is adherence to physicochemical and biological standards (e.g., USP monographs, ISO 10993). The "experts" would be the scientists and technicians conducting the non-clinical tests and interpreting the results against established specifications, not medical domain experts creating ground truth for an algorithm.

4. Adjudication method for the test set:

  • This is not applicable. Adjudication methods are used in consensus agreement among human annotators, which is not relevant for the objective physical and chemical testing of a surgical suture. The results are quantitative measurements compared against predefined acceptance limits.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This is not applicable. An MRMC study is for evaluating observer performance (e.g., radiologists interpreting images). This submission is for a physical medical device (suture) and does not involve human "readers" or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • This is not applicable. This describes the performance of a standalone AI algorithm. The DemeCAPRONE suture is a physical device, not an algorithm.

7. The type of ground truth used:

  • The "ground truth" for the device's performance is established by objective, established standards and methodologies.
    • Pharmacopeia Standards (USP): These define the chemical, physical, and pharmaceutical quality of drugs and medical devices. For sutures, this includes specifications for tensile strength, diameter, length, sterility, etc.
    • ISO 10993 Biocompatibility Standards: These define methods for evaluating the biological response to medical devices.
    • Predicate Device Performance: Part of the substantial equivalence relies on demonstrating that the new device's properties (including resorption) are similar to those of previously cleared predicate devices.

8. The sample size for the training set:

  • This is not applicable. There is no "training set" in the context of this 510(k) submission for a surgical suture. Training sets are used for machine learning models.

9. How the ground truth for the training set was established:

  • This is not applicable. As there is no training set for an AI/ML model, the concept of establishing ground truth for it does not apply.

In Conclusion:

The DemeCAPRONE suture's acceptance is based on demonstrating substantial equivalence to predicate devices and conformance to recognized consensus standards (USP) and international standards (ISO 10993). The "study" proving this involves a battery of non-clinical (laboratory and animal) tests rather than clinical trials or AI/ML-specific performance studies. The primary evidence presented is the declaration that DemeCAPRONE "meets or exceeds" the relevant performance requirements and is "Same" in composition, design, and manufacturing process to established predicate devices.

{0}------------------------------------------------

DemeTECH.

JUL 1 7 2013

Section 1 - 510(k) Summary

This summary is submitted in accordance with the Safe Medical Device Act (SMDA) of 1990 and Title 21 CFR § 807.92. This summary demonstrates the equivalence of Demetech Sutures to those of the legally marked devices listed.

A. Applicant:Demetech Corporation,14175 NW 60th Ave.Miami Lakes FL. 33014
B. Contact Person:A. J. Dimercurioe-mail: tony@demetech.usPhone # 305-824-1048 Ext 115
C. Date Prepared:July 15th, 2013
Trade Name:DemeCaprone (Poliglecaprone25) Synthetic Monofilament(PGCL) Absorbable Suture
Common Name:DemeTECH Synthetic PGCLMonofilament Poliglecaprone 25Absorbable Suture
Classification Name:Absorbable poly (glycolide/l-lactide) Surgical Suture braidedor monofilament
D. Device Classification
FDA Class:Product Classification:II878.4493, AbsorbablePoly(glycolide/I-lactide) SurgicalSuture
Product Code:GAM
E. Predicate Devices: DemeCAPRONE (Poliglecaprone 25) SyntheticAbsorbable Surgical Suture is substantially equivalent to these predicatedevices:
• Riverpoint's Mono Q PGCL Absorbable Suture reference 510K numberK100461, Riverpoint Medical. Portland OR.

Sutures India PVT.LTD Monoglyde Poliglecaprone 25 Absorbable Suture . reference 510K number K081002, Sutures India Private Limited Bangalore India.

. . . . .

100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{1}------------------------------------------------

K130083 page 2/5

  • Ethicon's Monocryl Synthetic Absorbable Poliglecaprone 25 suture, . reference 510K number, K960653 & K964072, Ethicon Inc. Somerville NJ.
  • F. Device Description:

DemeCAPRONE (Poliglecaprone 25) is a synthetic monofilament absorbable surgical suture composed of Poly (glycolic-co-caprolactone) copolymer (PGCL) and is supplied un-dyed and dyed with D&C Violet #2 below 0.1wt%. DemeCAPRONE (Poliglecaprone 25) synthetic absorbable suture is available in sizes 6-0 through 1 (metric sizes 0.7 - 4). DemeCAPRONE (Poliglecaprone 25) Surgical Suture meets the requirements established by the United States Pharmacopeia (U.S.P.) for synthetic absorbable surgical sutures except for diameter.

  • G. Intended Use:
    DemeCAPRONE (Poliglecaprone 25) Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

  • H. Non-Clinical Tests Performed:
    Non-clinical testing was conducted on the device per FDA's Special Control Guidance Document: Surgical Sutures, to prove conformance to the requirements of USP for synthetic absorbable suture, biocompatibility testing in accordance to ISO 10993-1 and invivo and in-vitro resorption to further demonstrate substantial equivalence to the predicate devices. Physical properties and functionality testing assured that the device conformed with suture diameter, suture length, knot pull tensile strength, needle attachment strength, extractable color and sterility to methods outlined in USP 35.

Poliglecaprone 25 was selected based on known biocompatibility (per ISO 10993) and established history of use in the medical device industry for implantable devices, and are identical or substantially equivalent to the material used in the predicate devices listed above. Biocompatibility testing performed on Poliglecaprone 25 sutures within the submission includes the following: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity, Genotoxicity - Bacterial Reverse Mutation and Chromosomal Aberration, Bone Marrow Micronucleus, Subchronic Toxicity (4-week, following subcutaneous implantation), Muscle Implantation (12-week).

COMPARISON TABLE DEMETECH POLIGLECAPRONE 25 TO PREDICATE
DEVICES
Comparison ItemsDemetechPoliglecaprone25 SutureRiverpoint'sMono QPGCLSutureSuturesIndiaMonoglydeSutureEthiconMonocrylSuture
DemeCAPRONE(Poliglecaprone 25) suture is asynthetic absorbable surgicalsuture. It is a sterile flexiblemonofilament thread, composedof Poly (glycolic-co-caprolactone) copolymerSameSameSameSame
SameSameSameSame
SameSameSameSame
SameSameSameSame
SameSameSameSame
SameSameSameSame
SameSameSameSame
SameSameSameSame
SameSameSameSame
Same
SameSameSame
DemeCAPRONE meets orexceeds the performancerequirements defined in theUnited States Pharmacopeia for"Tensile Strength" < 881 >SameSameSameSame
DemeCAPRONE meets orexceeds the performancerequirements defined in theUnited States Pharmacopeia andthe current edition USP for"Needle Attachment" < 871 >SameSameSameSame
DemeCAPRONE meets orexceeds the performancerequirements defined in theUnited States Pharmacopeia for"Suture Length Requirement"(95% of stated label length)SameSameSameSame
DemeCAPRONE meets theperformance requirementsdefined in the United StatesPharmacopeia current editionU.S.P. for sterilitySameSameSameSame
DemeCAPRONE is packaged ina same or equivalent mannerwith sterile single or doublepackaging having labelingconforming to 21 CFR andCurrent edition of USP.SameSameSameSame
DemeCAPRONE(Poliglecaprone 25) suture isbiologically compatible whentested as per ISO-10993SameSameSameSame
DemeCAPRONE(Poliglecaprone 25) suture istested and proved to be nontoxic, when tested as per ISO-10993 for toxicitySameSameSameSame

{2}------------------------------------------------

K130063 page 3/5

{3}------------------------------------------------

K130083 page 4/5

Clinical Tests Performed: l.

.

No clinical trials were conducted

{4}------------------------------------------------

K130083 page 5/5

J. Conclusion:

DemeCAPRONE (Poliglecaprone 25) is composed of the same material, as are the predicated devices and the same design being a sterile, flexible, monofilament threads meeting all the requirements of the United States Pharmacopeia. DemeCAPRONE (Poliglecaprone 25) Suture is manufactured in the same manner as the predicate devices, being composed of composition of absorbable flexible, monofilament thread prepared from Poly (glycolic-co-caprolactone) copolymer (PGCL) and produced in operations considered standard in the fiber industry to form the finished suture fiber. The manufacturer supplies to Demetech the same suture materials as it does to other suture manufacturers including some of those listed above.

The biocompatibility data and the results of performance testing presented demonstrate the substantial equivalence of DemeCAPRONE (Poliglecaprone 25) Synthetic Absorbable Suture to that of the predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

July 17, 2013

DemeTECH Corporation % Anthony J. Dimercurio Vice President RA/OA 14175 NW 60th Avenue Miami Lakes, Florida 33014

Re: K130083

Trade/Device Name: DemeCaprone (Poligecaprone 25) Synthentic (PGCL) Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: Class II Product Code: GAM Dated: May 29, 2013 Received: May 31, 2013

Dear Mr. Dimercurio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

Page 2 - Anthony J. Dimercurio

CFR Part 807): labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

For

Peter D. Rumm -S Mark N. Melkerson

Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for DemeTECH. The logo consists of a stylized globe-like symbol on the left, followed by the company name "DemeTECH" in a bold, italicized font. The "CH" portion of the name is slightly smaller and raised, with a small circle as a subscript.

Section 6 - Indications for Use Statement

Indication for use

510K Number: _________________________________________________________________________________________________________________________________________________________________

Device Name: Demetech Absorbable Poliglecaprone 25 Surgical Suture.

Indication for Use:

Demetech Absorbable Poliglecaprone 25 Surgical Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery and neurological tissue.

Prescription Use"X"
Use
(Part 21 CFR 801; Subpart D)
And/Or
Over the-Counter
(21 CFR 801; Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF

NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Krause -S

(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130083

§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.

(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.