Search Filters

Search Results

Found 130 results

510(k) Data Aggregation

    K Number
    K233265
    Manufacturer
    Date Cleared
    2024-10-11

    (378 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPTIME® R sutures are intended for use in general soft tissue approximation including ophthalmic procedures, but not in cardiovascular or neurological procedures.

    Device Description

    OPTIME®R sutures are intended to be used in suturing in open or laparoscopic surgery. These sutures are available with a disposable needle attachment. OPTIME®R is a sterile, multiple-stranded (multifilament), synthetic bioabsorbable thread made from a polymerization of polyglycolic acid (PGA) used in suturing when short-term support and rapid strength loss are needed. To perform suturing on soft tissue wounds involves a series of precise steps. The surgeon grasps the needle, pushes it through tissue, pulls it through and repeats for continuous suturing. Further, knots are tied, excess thread is cut, and used needles are single use and safely disposed in accordance with hospital procedure.

    AI/ML Overview

    The provided text is an FDA 510(k) Pre-Market Notification for a medical device called OPTIME® R, which is an absorbable surgical suture. The document primarily focuses on establishing substantial equivalence to a predicate device (Safil Quick Aesculap, K031286) based on technological characteristics and performance testing.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The document states that the OPTIME® R suture was subjected to the requirements of the United States Pharmacopeia (U.S.P) monograph for Synthetic Absorbable Sutures. The following tests and their corresponding USP standards are mentioned:

    Acceptance Criteria / TestReported Device Performance
    Diameter (following USP 43 )Meets USP performance requirements for absorbable sutures
    Tensile strength (following USP 43 )Meets USP performance requirements for absorbable sutures
    Needle attachment (following USP 43 )Meets USP performance requirements for absorbable sutures
    Length (following USP 43 - Absorbable surgical suture)Meets USP performance requirements for absorbable sutures
    Resorption propertiesEssentially complete at 42 days (matches predicate's absorption time)
    Biocompatibility (following ISO 10993)Biocompatibility tests performed and results indicate compliance
    Sterilization methodsGamma sterilization (matches predicate)
    Shelf-life5 years
    Ideal storageStored in original packaging,
    Ask a Question

    Ask a specific question about this device

    K Number
    K232372
    Manufacturer
    Date Cleared
    2024-07-12

    (339 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ADVANTIME® synthetic absorbable surgical sutures are intended for use in general soft tissue approximation and/or ligation, but not for use cardiovascular surgery, microsurgery, ophthalmic surgery, and neurological tissue.

    Device Description

    ADVANTIME® sutures are intended to be used in suturing and/or tissue ligation in open or laparoscopic surgery. These sutures are available with a disposable needle attachment.

    ADVANTIME® is a single use, sterile, single-stranded (monofilament), synthetic bioabsorbable thread made of a poliglecaprone (a copolymer of glycolide and E- caprolactone of low tissue reactivity). Sutures are commonly used in digestive, urological, aesthetic, gynecological and intradermal surgery. To perform suturing on soft tissue wounds involves a series of precise steps. The surgeon grasps the needle, pushes it through tissue, pulls it through and repeats for continuous suturing. The same applies to double-armed needles. Further, knots are tied, excess thread is cut, and used needles are single use and safely disposed in accordance with hospital procedure.

    AI/ML Overview

    The provided document is a 510(k) summary for a surgical suture device (ADVANTIME Absorbable Suture) and does not describe an AI/ML powered device. Therefore, the questions regarding acceptance criteria, study details, and performance measures (such as sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone performance of an algorithm) related to AI/ML devices are not applicable.

    The document discusses the substantial equivalence of the ADVANTIME Absorbable Suture to a predicate device based on physical and biological performance testing for a medical device that is not an AI/ML device.

    Here's the information that is relevant from the provided text regarding this medical device:

    1. Acceptance Criteria and Reported Device Performance (Non-AI/ML Device)

    The device was subjected to the requirements of the United States Pharmacopeia (U.S.P.) monograph for Synthetic Absorbable Sutures.

    Acceptance Criteria (USP 43)Reported Device Performance
    Diameter ()Meets USP requirements with minor variations (within FDA guidance allowance)
    Tensile strength ()Meets USP requirements
    Needle attachment ()Meets USP requirements
    Length (Absorbable surgical suture)Meets USP requirements
    Extractable color (Absorbable surgical suture)Meets USP requirements
    Resorption profilesEvaluated for each material; data provided in-vitro
    StabilityTesting completed
    Biocompatibility (ISO 10993-1)All tested aspects met (Cytotoxicity, Sensitization, Irritation, Material mediated pyrogenicity, Acute Systemic Toxicity, Sub-acute/sub-chronic/chronic toxicity, Genotoxicity, Carcinogenicity, Chemical characterization and toxicological risk assessment, Implantation, Degradation, Sterilization residue testing)

    2. Sample size used for the test set and the data provenance: Not applicable as this is not an AI/ML device study. Performance was evaluated based on physical and biological tests according to USP standards and ISO 10993-1.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is not an AI/ML device. Ground truth for physical properties is defined by USP standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable as this is not an AI/ML device. Performance is objectively measured against USP standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the physical characteristics, the ground truth is defined by the United States Pharmacopeia (USP) standards and test methodologies. For biological evaluation, the ground truth is established by standardized testing protocols detailed in ISO 10993-1.

    8. The sample size for the training set: Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221527
    Device Name
    StitchKit
    Date Cleared
    2023-02-14

    (264 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StitchKit® Suture Delivery Canister facilitates minimally invasive robotic surgery by transporting suture to the operative field and removing used needles after suturing. The suture contained within the device is intended for soft tissue approximation where use of the specific absorbable or non-absorbable sutures contained within it is appropriate.

    Device Description

    StitchKit® is a sterile, single-use plastic canister that is provided pre-loaded with suturing materials with attached needles. The device facilitates endoscopic robotic surgery by introducing multiple strands of suture to the surgical site at one time and allowing for the safe retrieval of the needles. The canister is sized to be passed through a ≥12 mm trocar or passed through an incision sized for an 8 mm trocar under direct endoscopic visualization. As suturing is completed with each strand, the used needle is placed into a compartment within the canister for safekeeping until the entire canister is removed either through the ≥ 12 mm trocar or through the 8mm trocar incision in tandem with the 8mm trocar, also under direct endoscopic visualization using the attached retrieval string. It is supplied sterile.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called StitchKit®. This document primarily focuses on demonstrating substantial equivalence to a predicate device, specifically for a minor modification related to the device's insertion and removal method (allowing use through an 8mm trocar incision in addition to a 12mm trocar).

    Crucially, this document does NOT describe the acceptance criteria or a study proving the device meets performance criteria for a complex AI/ML medical device. The device in question is a suture delivery canister, a physical surgical tool, not an AI or imaging device. Therefore, many of the requested points related to AI/ML device validation (like expert adjudication, MRMC studies, ground truth for training/test sets, effect size of human reader improvement with AI assistance, and standalone AI performance) are not applicable to this submission.

    The "clinical data" mentioned (retrospective case review n=422) is cited in the context of supporting the modification to the instructions for use, not for establishing the core performance of an AI algorithm. The performance data section refers to "functional performance test consisting of simulated surgical testing in an animal model" to demonstrate compatibility with the new insertion/removal technique, which is a physical device test, not an AI/ML study.

    Therefore, I cannot fulfill most of your request for an AI/ML device, as the provided document pertains to a physical surgical device and its minor instruction for use modification.

    However, I can extract the safety and efficacy information relevant to this device (the StitchKit® suture delivery canister) as presented in the document, acknowledging that it doesn't align with the detailed AI/ML validation questions.


    Based on the provided document for the StitchKit® Suture Delivery Canister (a non-AI device):

    1. A table of acceptance criteria and the reported device performance

    The document does not specify formal, quantitative acceptance criteria for device performance in the typical sense of accuracy, sensitivity, or specificity, as it's a physical tool. Instead, the "performance" demonstrated relates to its functional compatibility with a new surgical technique.

    Acceptance Criteria (Implied)Reported Device Performance
    Compatibility with 8 mm trocar site incision insertion/removalDemonstrated through "functional performance test consisting of simulated surgical testing in an animal model." No specific numerical performance metric provided beyond "compatible."
    Maintenance of Substantial Equivalence to PredicateConcluded based on: unchanged Indication for Use and Technological Characteristics, functional test data, clinical data (retrospective case review), and comparison to predicate device including risk analysis.
    Functionality in transporting suture and removing used needles(Presumed to be maintained from the predicate device and inherent to the device's design, as no change to core functionality is described.)

    2. Sample size used for the test set and the data provenance

    • Test Set (for the modification validity):
      • Sample Size: n=422 (Cited as a "retrospective case review clinical study").
      • Data Provenance: Conducted by "an academic medical center." The country of origin is not specified but is implicitly in the US due to FDA submission. The study was retrospective.
    • Functional Performance Test: Conducted in an "animal model." Sample size for this specific test is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This is a physical device, not an AI/ML diagnostic or image analysis device where "ground truth" would be established by experts interpreting medical data. The "ground truth" for the retrospective case review would likely be the surgical outcomes or successful use reported in the patient records.

    4. Adjudication method for the test set

    • Not Applicable. As above, no expert adjudication process is described or relevant for this type of device and study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. No MRMC study was done, as this is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. No standalone algorithm performance was evaluated, as this is not an algorithm-based device.

    7. The type of ground truth used

    • For the retrospective case review (n=422), the "ground truth" would be derived from clinical records and surgical outcomes related to the use of the device in actual patient cases, rather than expert interpretation of images or pathology. The document doesn't explicitly state how "success" or "performance" was quantified from these cases, but it supports the safety/effectiveness of the device, particularly with the proposed instruction modification.
    • For the animal model functional test, the "ground truth" was direct observation of the device's physical handling and compatibility with the 8mm trocar site.

    8. The sample size for the training set

    • Not Applicable. There is no "training set" as this is not an machine learning model.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221744
    Manufacturer
    Date Cleared
    2022-11-15

    (152 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate.

    Device Description

    The STRATAFIX™ Spiral MONOCRYL™ Plus Unidirectional Knotless Tissue Control Device is an antibacterial monofilament, synthetic absorbable single-use surgical suture device composed of a copolymer of glycolide and (epsilon) s-caprolactone. The device contains IRGACARE®: MP (triclosan), a broad-spectrum antibacterial agent, at no more than 2360 ug/m. The colorant employed is D&C Violet No.2, in

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device, specifically a surgical suture. It does not contain information about an AI/ML medical device, which would typically include acceptance criteria, study details (sample size, data provenance, ground truth establishment, expert qualifications, etc.), or details of human reader studies (MRMC).

    Therefore, I cannot provide the requested information regarding acceptance criteria and the study that proves the device meets them, as the provided text pertains to a traditional medical device (surgical suture) and not an AI/ML device. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing rather than the detailed clinical validation studies common for AI/ML products.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212810
    Date Cleared
    2022-08-05

    (336 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MONOFIX PGCL, knotless wound closure device is indicated for use in soft tissue approximation where the use of absorbable suture is appropriate. MONOFIX PGCL, knotless wound closure device should not be used to approximate tissue under tension and the use of barbed suture has not been shown to be safe in the closure of hernias, laparotomy incisions, bowel anastomosis or suture lines, vascular suture lines, and vaginal cuff closure in controlled clinical or preclinical studies for minimally invasive or open surgery.

    Device Description

    MONOFIX PGCL, knotless wound closure device, comprised of Poly(glycolide-co-carprolactone)(PGA-PCL) is a synthetic absorbable monofilament undyed-suture. It is available sterile, developed by Samyang Holdings Corp., Ltd. The MONOFIX PGCL, knotless wound closure device consists of a Barb-type suture and a surgical needle, with a suture at one end and a stopper shaped end for tissue anchoring at the stopper is attached to the end of the suture and has a triangular shape with one side length of 1.8 to 3.3 mm and a thickness of 0.7 to 1.0 mm. The device is designed with small uni-directional barbs in size 0.18-0.55 mm along the length and stopper that eliminate the need to tie knots during approximation. The device is available in lengths of 15 to 60cm and diameter sizes 1 through 3-0 with needles in 1/2 circle, 5/8 circle, 1/2 curved and straight shapes attached to each end.

    AI/ML Overview

    This appears to be an FDA 510(k) clearance letter and a summary of the predicate device comparison for a surgical suture, not a diagnostic AI/ML device. Therefore, the requested information regarding acceptance criteria, study details (sample size, data provenance, expert consensus, MRMC studies, standalone performance), and ground truth establishment, which are typical for AI/ML device submissions, are not applicable in this context.

    The document describes the non-clinical tests performed to demonstrate the device's compliance with established standards for surgical sutures. These include:

    • Bench tests: To verify design specifications, focusing on:
      • USP Suture Diameter
      • USP Suture-Needle attachment
      • USP Tensile Strength Surgical Suture
      • USP Monographs: Absorbable Surgical Suture
      • EP Monographs: Sutures, Sterile Synthetic Absorbable Monofilament
    • Sterilization validation: ISO 11135:2014, ISO 10993-7
    • Sterile barrier system testing: ASTM F 88
    • Accelerated aging stability testing: ASTM F 1980
    • Biocompatibility testing: A comprehensive suite of ISO 10993 standards (e.g., ISO 10993-5, -10, -11, -4, -12, -3, -6) and ASTM F756-17.
    • In vivo tensile strength: Performed to evaluate tensile strength retention rate.

    The "Acceptance Criteria" in this document would refer to the specific thresholds and performance requirements defined by these standards (e.g., minimum tensile strength, maximum/minimum diameter, acceptable levels of residuals, successful sterilization). The "Reported Device Performance" would be the results from the tests confirming the device met these criteria.

    Therefore, it is impossible to provide the table and study details as requested because this document does not pertain to an AI/ML device. The study outlined is a non-clinical, benchtop and biocompatibility assessment, not a clinical performance study with human readers or AI algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211792
    Device Name
    StitchKit
    Date Cleared
    2021-07-16

    (36 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StitchKit® Suture Delivery Canister facilitates minimally invasive robotic surgery by transporting suture to the operative field and removing used needles after suturing. The suture contained within the device is intended for soft tissue approximation where use of the specific absorbable or non-absorbable sutures contained within it is appropriate.

    Device Description

    StitchKit® is a suture delivery canister which facilitates endoscopic robotic surgery by introducing multiple strands of different suturing materials to the surgical site at one time and allowing for the safe retrieval of the needles. It is sized to be passed through a ≥12 mm trocar. As suturing is completed with each strand, the used needle is placed into a disposal compartment within the canister for safekeeping until the entire canister is removed through the trocar using the attached retrieval string. It is supplied sterile. Each StitchKit® contains strands of existing legally-marketed Suturing Materials, cleared for use within the device.

    AI/ML Overview

    The provided document is a 510(k) summary for the StitchKit® Suture Delivery Canister, which is a Class II medical device. This document details the device's indications for use, technological characteristics, and substantiates its substantial equivalence to predicate devices, rather than presenting a study to prove acceptance criteria in the typical sense of a human-in-the-loop AI effectiveness study.

    The device itself is a delivery system for existing sutures and not an AI-powered diagnostic or assistive device. Therefore, many of the requested criteria related to AI performance, AI effect size on human readers, expert ground truth, and training data are not applicable to this submission.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the StitchKit® are based on demonstrating substantial equivalence to legally marketed predicate devices. The performance data provided focuses on the physical and chemical properties of the device and its included sutures.

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance / Compliance
    Technological Characteristics- Canister is legally marketed and unchanged.
    • Sutures are legally marketed and have not been modified, only trimmed to length and packaged within the StitchKit® canister. | - Canister is confirmed to be legally marketed and unchanged.
    • Sutures are confirmed to be legally marketed, not modified other than trimming and packaging for the StitchKit®. |
      | Performance Testing | - Comparative Physico-Chemical analysis
    • LAL Pyrogen testing
    • Material Mediated Pyrogen testing
    • Stability Evaluations (including USP suture tensile strength, diameter, and needle attachment testing)
    • Comparative in-vitro simulated biodegradation testing for absorbable sutures | - Testing performed to verify substantial equivalence to predicate devices.
    • Specific results or thresholds are not provided in this summary, but the conclusion states that the device "has been shown to be substantially equivalent to its predicate devices" based on this data. |
      | Indications for Use | - Facilitates minimally invasive robotic surgery by transporting suture and removing used needles.
    • Suture intended for soft tissue approximation where use of specific absorbable/non-absorbable sutures is appropriate. | - The device's design directly supports these indications. No performance metrics are given here for success rate of needle removal or ease of transport, rather it's implicit in its design that it fulfills these functions. |

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of an AI study. Instead, it refers to various forms of performance testing on the device and its components. The document doesn't provide specific sample sizes for these tests (e.g., number of sutures tested for tensile strength or number of canisters tested for pyrogenicity). The provenance of the data (e.g., country of origin, retrospective/prospective) is also not specified, as this type of information is typically not included in a 510(k) summary for a physical device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. As this is not an AI or diagnostic device, there is no "ground truth" established by experts in the context of image interpretation or clinical diagnosis. The performance criteria are based on objective physical/chemical tests.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    Not applicable. As there are no expert interpretations or subjective assessments requiring adjudication, no such method was used.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. The StitchKit® is a physical surgical delivery device and does not involve AI or human readers for diagnostic assistance. Therefore, an MRMC study and AI effect size are irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device does not involve an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance validation is based on established engineering and material science standards (e.g., USP standards for suture properties, LAL pyrogen testing standards, physicochemical analysis). It's not clinical "ground truth" derived from patient outcomes or expert consensus on a diagnosis.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI component requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no AI component requiring a training set and associated ground truth.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202950
    Device Name
    StitchKit COMBO
    Manufacturer
    Date Cleared
    2021-02-23

    (146 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    StitchKit® Suture Delivery Canister facilitates minimally invasive robotic surgery by transporting suture to the operative field and removing used needles after suturing. The suture contained within the StitchKit® device is intended for soft tissue approximation where use of the specific absorbable or non-absorbable sutures contained within it is appropriate.

    Device Description

    StitchKit® COMBO is a suture delivery canister which facilitates endoscopic robotic surgery by introducing multiple strands of different suturing materials to the surgical site at one time and allowing for the safe retrieval of the needles. It is sized to be passed through a ≥12 mm trocar. As suturing is completed with each strand, the used needle is placed into a disposal compartment within the canister for safekeeping until the entire canister is removed through the trocar using the attached retrieval string. It is supplied sterile in a foil pouch. Each StitchKit® version contains strands of existing legallymarketed Suturing Materials, cleared for use within the StitchKit®.

    AI/ML Overview

    This looks like a 510(k) summary for a medical device called "StitchKit® COMBO Suture Delivery Canister". Based on the provided text, the device is explicitly a suture delivery system and not an AI or imaging device, so the requested information about AI performance, ground truth, expert adjudication, etc., is not applicable.

    Here's a breakdown of why the requested information cannot be provided given this document:

    • No AI component: The document describes a "suture delivery canister which facilitates minimally invasive robotic surgery by transporting suture to the operative field and removing used needles after suturing." There is no mention of artificial intelligence, machine learning, image analysis, or any diagnostic capabilities.
    • Physical device, not a diagnostic algorithm: The device handles physical sutures and needles. The performance data listed (knot-pull tensile testing, diameter, needle attachment, biodegradation testing, pyrogen testing, physico-chemical analysis, stability evaluations) are all related to the physical properties and safety of the surgical sutures and the delivery system itself, not to the performance of a software algorithm.

    Therefore, since the device is not an AI/ML medical device, the following points from your request are not applicable and cannot be found in the provided text:

    1. A table of acceptance criteria and the reported device performance (The "performance data" mentioned refers to physical product testing, not AI performance metrics)
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    8. The sample size for the training set
    9. How the ground truth for the training set was established

    In summary, this document describes a traditional medical device (suture delivery system) and does not provide any information relevant to the performance or validation of an AI/ML component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201996
    Manufacturer
    Date Cleared
    2020-12-21

    (154 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MONOCRYL™ Plus Antibacterial Sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.

    Device Description

    MONOCRYL™ Plus Antibacterial Suture is a sterile, monofilament, synthetic, absorbable surgical suture composed of a copolymer of glycolide and (epsilon) s-caprolactone. The empirical formula of the polymer is (C2H2O2)m (C6H102)n.

    MONOCRYL™ Plus Antibacterial Suture is available undyed and dyed with D&C Violet No. 2

    MONOCRYL™ Plus Antibacterial Suture contains Irgacare®† MP (triclosan), a broad spectrum antibacterial agent, at no more than 2360 µg/m.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called MONOCRYL™ Plus Antibacterial Suture. It seeks to demonstrate substantial equivalence to a previously cleared device. Therefore, the information provided focuses on the equivalence of the modified device to its predicate, rather than detailing a de novo study to establish acceptance criteria and prove performance against them as would be typical for a novel device or AI/software.

    Based on the provided text, here's an analysis of the "acceptance criteria" and "study" as they relate to this specific 510(k) submission:

    Understanding the Context:

    This submission (K201996) is for a modified device (MONOCRYL™ Plus Antibacterial Suture) that is largely identical to a predicate device (MONOCRYL™ Plus Antibacterial Poliglecaprone – 25 with 510(k) number K050845). The only reported changes are to the labeling, specifically updating in vitro effectiveness against an additional microorganism and including new descriptors/icons.

    For a 510(k) involving minor changes, the "acceptance criteria" revolve around demonstrating that the modified device remains as safe and effective as the predicate, or that the changes do not raise new questions of safety or effectiveness. The "study" in this context refers to the data presented to support this claim, rather than a clinical trial for a new product.

    1. A table of acceptance criteria and the reported device performance

    Since this is a submission for a modified device essentially claiming no change in technological characteristics, the "acceptance criteria" are implicitly that the modified device retains the performance attributes of the predicate. The "reported device performance" is essentially a reaffirmation of the predicate's performance, along with specific in vitro data for the antibacterial claims.

    Acceptance Criteria (Implicit for Substantial Equivalence)Reported Device Performance (Summary from Document)
    For Suture Material:
    - Same technological characteristics as predicate (composition, structure, absorbability)"The modified MONOCRYL™ Plus Suture has the same technological characteristics as the predicate device... it is a sterile, monofilament synthetic absorbable suture composed of a copolymer of glycolide and (epsilon) s-caprolactone."
    - Compliance with USP standards for absorbable surgical sutures"complies with the requirements of the United States Pharmacopoeia (USP) for absorbable surgical sutures except for a slight oversize in diameter." (This exception is noted as existing for the predicate as well, implying continued equivalence.)
    For Antibacterial Efficacy:
    - Contains Irgacare®† MP (triclosan) at specified concentration"MONOCRYL™ Plus Antibacterial Suture contains Irgacare®† MP (triclosan), a broad spectrum antibacterial agent, at no more than 2360 µg/m."
    - Inhibit colonization of specified bacteria"In vitro efficacy studies were conducted to demonstrate that bacteria will not colonize MONOCRYL™ Plus Antibacterial suture. The in vitro data indicated that MONOCRYL™ Plus suture provides an antibacterial effect sufficient to inhibit colonization of the suture by Staphylococcus aureus, Staphylococcus epidermidis, Methicillin-Resistant S. aureus, Methicillin-Resistant S. epidermidis, Klebsiella pneumoniae, Escherichia coli and Enterobacter cloacae."
    For Intended Use/Indications:
    - Same intended use and indications for use as predicate"The modified device has the same intended use and indications for use as the predicate device." (General soft tissue approximation/ligation, not for cardiovascular, neurological, microsurgery, or ophthalmic surgery).

    2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "in vitro efficacy studies" for the antibacterial claims. Details on the specific sample sizes (e.g., number of suture samples, number of bacterial cultures) are not provided in this summary. The data provenance is described as "in vitro," implying laboratory studies, rather than patient data. No information on country of origin or retrospective/prospective nature is given, as it's not a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This type of information is generally relevant for AI/radiology devices where expert consensus is used to label images. For a physical medical device like a suture, "ground truth" for material properties and antibacterial efficacy is established through standardized laboratory testing and validated analytical methods (e.g., chemical analysis, microbiology assays). The document does not specify the number or qualifications of experts involved in running these in vitro tests, as this is typically handled by qualified laboratory personnel following established protocols.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies, particularly for imaging interpretation, to resolve discrepancies between readers. This document describes laboratory in vitro studies for which such an adjudication method is not relevant.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done. If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    Not applicable. This device is a physical surgical suture, not an AI or imaging diagnostic tool. Therefore, MRMC studies and human reader improvement with AI assistance are not relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This document is about a physical medical device, not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the antibacterial claims, the "ground truth" would be established by the results of the microbiological in vitro assays which directly measure the inhibition of bacterial colonization on the suture material. For the material properties, the "ground truth" is established by chemical and physical testing against established standards (e.g., USP monographs). This is empirical data from laboratory studies, not consensus of human experts on subjective interpretations.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that requires training data.

    9. How the ground truth for the training set was established

    Not applicable, as explained in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200392
    Device Name
    REXSIN
    Manufacturer
    Date Cleared
    2020-07-07

    (140 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.

    Device Description

    Synthetic Absorbable PGA Suture With or Without Needle (REXSIN) are produced and provided by SM ENG Co., Ltd. REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue. Suture of REXSIN is Trisorb manufactured by SAMY ANG Biopharmaceuticals Corporation. SM Eng is receiving bulk Trisorb from Samyang, and then go through cutting, adhesion of need and thread, winding and sterilization process according to SM Eng procedure, which becomes REXSIN suture. REXSIN is a sterilized medical device composed with absorbable, braided, coated suture with needle, stainless steel SUS 304. REXSIN Suture is a synthetic absorbable, braided, coated suture composed of a Polyglycolic acid (PGA). The suture is available undyed (natural) or dyed (D&C Violet No.2). Approximately over 65% of tensile strength remain after 2week. Complete absorption in tissues takes around 90 days. REXSIN Sutures meets requirements established by the United States Pharmacopoeia (USP) for Absorbable Surgical Sutures. REXSIN Sutures is available in 8-0, 7-0, 6-0, 5-0, 2-0, 0, 1 and 2, which are the sizes identified in the currently recognized United States.

    AI/ML Overview

    The provided text is related to a 510(k) Pre-market Notification for a medical device (surgical suture). It describes the device, its intended use, and non-clinical tests conducted to demonstrate substantial equivalence to a predicate device. However, it does not contain information about acceptance criteria for an AI/ML powered device, a study proving device performance against such criteria, or details regarding ground truth, expert consensus, or specific effect sizes for human readers with AI assistance.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance.
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method.
    • MRMC comparative effectiveness study results or effect sizes.
    • Standalone algorithm performance.
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    This document describes a traditional medical device (suture) and its non-clinical testing for substantial equivalence, not an AI/ML-powered diagnostic or decision-support system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200140
    Date Cleared
    2020-03-27

    (66 days)

    Product Code
    Regulation Number
    878.4493
    Reference & Predicate Devices
    Why did this record match?
    Product Code :

    GAM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Silhouette Instalift device is indicated for use in mid-face suspension surgery to temporarily fixate the cheek sub dermis in an elevated position.

    Device Description

    Silhouette Instalift Sutures are absorbable, sterile sutures. It is manufactured from a USP size 3-0 Poly/Glycolide/I-lactide suture material and an implantable grade of bioabsorbable PLGA resin (100% Poly (L-lactideco-glycolide) (co-monomer ratio: L-Lactide =82mol%: glycolide=18mol%). The 30 centimeter suture (+- 10 %) (SMS 28-PLG-3.0.1- NA), 27.5 centimeter suture (+- 10 %) (SMS 29-PLG-3.0.1-NA) or 26.8 centimeter suture (+- 10 %) (SMS 30-PLG-3.0.1-NA) are attached to two 12 centimeter straight stainless steel needles. All products are supplied sterile (EO) for single use only. Silhouette Instaliff Sutures elicit a minimal acute inflammatory reaction in tissue that is followed by gradual encapsulation.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called Silhouette Instalift. It primarily discusses a change to the device's labeling and asserts substantial equivalence to a previously cleared device. It does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria.

    The document states:

    • "No changes have been made to the intended use or fundamental scientific technology and therefore the supporting information on sterilization, biocompatibility, and clinical data remain unchanged from the original application."
    • "The only modification made to the predicate device is the modification to the labelling of the device."

    Therefore, I cannot provide the requested information such as acceptance criteria, reported device performance, sample sizes for test/training sets, data provenance, number or qualifications of experts, adjudication methods, MRMC study results, standalone performance, or ground truth details, because this information is explicitly stated as unchanged from the original application and is not provided in this specific 510(k) summary document.

    To obtain this information, one would need to refer to the original 510(k) applications (K142061 and K163676) mentioned in this document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 13