K Number
K081002

Validate with FDA (Live)

Date Cleared
2008-06-09

(62 days)

Product Code
Regulation Number
878.4493
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MONOGLYDE IS A SYNTHETIC ABSORBABLE (POLIGLECAPRONE 25) SURGICAL SUTURE; STERILE FLEXIBLE STRAND, PREPARED AND SYNTHESIZED FROM THE MONOMERS, POLYGLYCOLIC ACID 75% AND CAPROLACTONE 25% AND IS INDICATED FOR USE IN GENERAL SOFT TISSUE APPROXIMATION, BUT NOT FOR USE IN OPHTHALMIC SURGERY, CARDIO-VASCULAR AND NEUROLOGICAL TISSUES

Device Description

MONOGLYDE is synthetic absorbable surgical suture, (Monofilament Poliglecaprone 25). Monoglyde is a sterile flexible monofilament thread, composed of Caprolactone. The sutures are inert, non collageneous and non antigenic. MONOGLYDE is dyed with D&C Violet #2 and it is available undyed also.

AI/ML Overview

The provided text describes Sutures India Pvt. Ltd.'s premarket notification (510k) for ABSORBABLE (POLIGLECAPRONE 25) SUTURE, named MONOGLYDE. The submission aims to demonstrate substantial equivalence to predicate devices, namely Monocryl (K930772) and Unicaprone (K072646).

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

The acceptance criteria for the MONOGLYDE suture are based on meeting the performance requirements defined in the United States Pharmacopeia (USP) current edition (USP 29) for various parameters, and showing equivalence to predicate devices. The reported device performance is that MONOGLYDE meets or exceeds these specifications.

Comparison Parameters (Acceptance Criteria - based on USP 29 & Predicate Equivalence)MONOGLYDE Reported Device Performance
Absorbable (Poliglecaprone 25) suture is a synthetic Absorbable surgical suture. It is a sterile flexible monofilament thread, composed of Poliglecaprone.Same (as predicate)
The sutures are inert, noncollageneous, and nonantigenic.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is dyed with D&C Violet #2 and being monofilament it is coated.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is supplied for single use only.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is sterilized by E.O. gas method.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is packaged in the same or equivalent manner, and has the same or equivalent labeling claims as that of the predicate devices.Same (as predicate)
Absorbable (Poliglecaprone 25) suture meets the official monograph of the United States Pharmacopeia current edition U.S.P. 29 for extractable color.Meets / Exceeds In-house specifications and is equivalent (to predicate)
Absorbable (Poliglecaprone 25) suture meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. 29 for Diameter<861>.Meets / Exceeds In-house specifications and is equivalent (to predicate)
Absorbable (Poliglecaprone 25) suture meets the performance requirements defined in the current edition of U.S.P. 29 for Tensile strength<881>.Meets / Exceeds In-house specifications and is equivalent (to predicate)
Absorbable (Poliglecaprone 25) suture meets the performance requirements defined in the current edition of U.S.P. 29 for Needle attachment<871>.Meets / Exceeds In-house specifications and is equivalent (to predicate)
Absorbable (Poliglecaprone 25) suture meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. 29 for finished suture Length Requirement (95% of stated label length).Meets / Exceeds In-house specifications and is equivalent (to predicate)
Absorbable (Poliglecaprone 25) suture meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. 29 for sterility.Meets / Exceeds In-house specifications and is equivalent (to predicate)
Absorbable (Poliglecaprone 25) suture is packed in a same or equivalent manner with sterile single or double packing having labeling conforming to 21CFR & U.S.P. 29.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is biologically compatible when tested as per ISO-10993.Same (as predicate)
Absorbable (Poliglecaprone 25) suture is tested and proved to be non toxic, when tested as per ISO-10993 for toxicity.Same (as predicate)

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the sample sizes for the test sets used for each specific test (e.g., diameter, tensile strength, etc.). It only mentions that "Testing of suture diameter, suture length, knot pull tensile strength, needle attachment strength, extractable color and sterility to methods outlined in USP 29 demonstrates Sutures India's MONOGLYDE, Absorbable (Poliglecaprone 25) suture meets or exceeds In-house specifications".

The data provenance is from India (Sutures India Pvt. Ltd. is located in Bangalore, India). It is unclear from the provided text whether the data is retrospective or prospective. Given the context of a 510(k) submission for a medical device seeking market clearance, these tests would typically be performed prospectively as part of the device development and validation process.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. The acceptance criteria are based on established standards (USP 29, ISO-10993) and equivalence to predicate devices, rather than expert interpretation of a unique test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not applicable and not provided. The study involves objective measurements against established standards, not subjective interpretations requiring an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging or AI-assisted interpretation, where human readers evaluate cases. This submission is for a surgical suture, which is a physical device, not an interpretative AI tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

A standalone algorithm performance study was not done. As explained above, this submission is for a surgical suture, not a software algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" or reference standard for this device is based on established industry standards and regulatory monographs:

  • United States Pharmacopeia (USP) current edition (USP 29): Used for physical and chemical properties like diameter, tensile strength, length, needle attachment, extractable color, and sterility. These are objective, quantitative measurements defined by the pharmacopeia.
  • ISO-10993: Used for biological compatibility and toxicity testing. These are also standardized test methods with defined pass/fail criteria.
  • Comparative data against predicate devices: The device demonstrates equivalence in material, design, manufacturing, colorant, packaging, and labeling claims to legally marketed predicate devices.

8. The sample size for the training set:

There is no training set in the context of this 510(k) submission. This is not an AI/ML device that requires training data. The "training" for a physical device would refer to the manufacturing processes and quality control to ensure consistency, but specific data sets for "training" an algorithm are not applicable.

9. How the ground truth for the training set was established:

As there is no training set mentioned or relevant to this type of device submission, this question is not applicable.

{0}------------------------------------------------

SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE (POLIGLECAPRONE 25) SUTURE

JUN - 9 2008

PAGE No: 17.0-1 SECTION No: 17.0

510k SUMMARY as required by: 21CFR 807.92 A. APPLICANT INFORMATION

Name: SUTURES INDIA PVT. LTD
Address: 472 D 13 th Cross, 4 th Phase,Peenya Industrial Area,Bangalore-560058. India
PHONE NO.: 91-80-41868000 (30 lines)
FAX NO.: 91-80-41171056
E mail: sales@suturesin.com
Web Address: www.suturesin.com
B. Contact Person: L.G.ChandrasekharMANAGING DIRECTOR
C. Date Prepared: 1.4.2008
D. DEVICE NAME:
  • : MONOGLYDE Trade Name .
  • : Absorbable Surgical Suture, Synthetic (Poliglecaprone 25) Common name
  • : Absorbable (Poliglecaprone 25) suture Classification Name .

F. PREDICATE DEVICES

  • A. Monocryl, Absorbable (Poliglecaprone 25) sutures, 510(k) Number K930772, Ethicon Incorporation, Somerville, NJ 088760151
    B. Unicaprone Absorbable (Poliglecaprone 25) Suture, 510(k) Number K072646, United Medical Industries Co.Ltd (UNIMED), Riyadh, SA 11553

{1}------------------------------------------------

SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE (POLIGLECAPRONE 25) SUTURE

KO81002

5

PAGE No: 17.0-2 SECTION No: 17.0

F. DESCRIPTION OF THE DEVICE

MONOGLYDE is synthetic absorbable surgical suture, (Monofilament Poliglecaprone 25). Monoglyde is a sterile flexible monofilament thread, composed of Caprolactone. The sutures are inert, non collageneous and non antigenic.

MONOGLYDE is dyed with D&C Violet #2 and it is available undyed also.

G. INTENDED USE OF THE DEVICE

Sutures India's MONOGLYDE Absorbable Poliglecaprone suture, is indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures.

{2}------------------------------------------------

K0 81002
page 3 of 5

SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE (POLIGLECAPRONE 25) SUTURE SECTION No: 17.0

COMPARISON TABLE SUTURES INDIA'S "MONOGLYDE" ABSORBABLE POLIGLECAPRONE SUTURE TO PREDICATE DEVICES

.

Comparison ParametersMonoglydeMonocrylUnicaprone
Absorbable (Poliglecaprone 25) suture is a synthetic Absorbable surgical suture. It is a sterile flexible monofilament thread, composed of Poliglecaprone.SameSameSame
The sutures are inert, noncollageneous and nonantigenic.SameSameSame
Absorbable (Poliglycaprone 25) suture is dyed with D&C Violet #2 and being monofilament it is coatedSameSameSame
Absorbable (Poliglecaprone 25) suture is indicated for use in soft tissue approximation, including use in ophthalmic procedures, but not for use in cardio vascular and neurological procedures.SameSameSame
Absorbable (Poliglecaprone 25) suture is supplied for single use only.SameSameSame
Absorbable (Poliglecaprone 25) suture is sterilized by E.O. gas methodSameSameSame
Absorbable (Poliglecaprone 25) suture is packaged in the same or equivalent manner, and has the same or equivalent labeling claims as that of the predicate devices including indications, warnings, cautions and precautionsSameSameSame
Absorbable (Poliglecaprone 25) suture meets the official monograph of the United States Pharmacopeia current edition U.S.P. 29 for extractable color.SameSameSame
Absorbable (Poliglecaprone 25) suture meets the performance requirements defined in the United States Pharmacopeia current edition U.S.P. 29 for Diameter<861>SameSameSame
Absorbable (Poliglecaprone 25)suture meets the performance requirements defined in the current edition of U.S.P. 29 for Tensile strength<881>SameSameSame
Comparison ParametersMonoglydeMonocrylUnicaprone
Absorbable (Poliglecaprone 25) suturemeets the performance requirements definedin the ia current edition of U.S.P. 29 forNeedle attachment<871>SameSameSame
Absorbable (Poliglecaprone 25) suturemeets the performance requirements definedin the United States Pharmacopeia currentedition U.S.P. 29 for finished suture LengthRequirement (95% of stated label length)SameSameSame
Absorbable (Poliglecaprone 25) suturemeets the performance requirements definedin the United States Pharmacopeia currentedition U.S.P. 29 for sterilitySameSameSame
Absorbable (Poliglecaprone 25) suture ispacked in a same or equivalent manner withsterile single or double packing havinglabeling conforming to 21CFR & U.S.P. 29SameSameSame
Absorbable (Poliglecaprone 25) suture isbiologically compatible when tested as perISO-10993SameSameSame
Absorbable (Poliglecaprone 25) suture istested and proved to be non toxic, whentested as per ISO-10993 for toxicitySameSameSame

{3}------------------------------------------------

K081002
page 4 of 5

:

:

{4}------------------------------------------------

K081002
Page 5 of 5

SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510k) FOR ABSORBABLE (POLIGLECAPRONE 25) SUTURE

PAGE No: 17.0-4 SECTION No: 17.0

CONCLUSION

Sutures India's MONOGLYDE Absorbable (Poliglecaprone 25) suture is composed of the same material, as that of the predicate devices and has the same design, as do the predicate devices. The suture is manufactured in a manner typical of the industry and equivalent to that used to produce predicate devices. Further the subject device is offered with the same colorant D&C Violet No.2 at a concentration that conforms to the requirements of Title 21 CFR § 74.3602, as are of the predicate devices.

Testing of suture diameter, suture length, knot pull tensile strength, needle attachment strength, extractable color and sterility to methods outlined in USP 29 demonstrates Sutures India's MONOGLYDE, Absorbable (Poliglecaprone 25) suture meets or exceeds In-house specifications and are equivalent in terms of the above mentioned predicate devices.

For SUTURES INDIA PVT. LTD,

  1. G. CHANDRASEKHAR

L.G. ChandraselthaECTOR Managing Director

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, facing to the left. The eagle is stylized with three lines forming its body and wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 9 2008

Sutures India Private Limited % L.G. Chandrasekhar 472 D 13th Cross, 4th Phase Peenya Industrial Area Bangalore-560058 India

Re: K081002

Trade/Device Name: MONOGLYDE Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Regulatory Class: II Product Code: GAM Dated: May 22, 2008 Received: May 27, 2008

Dear L.G. Chandrasekhar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{6}------------------------------------------------

Page 2 - L.G. Chandrasekhar

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Mullan

Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number K081002

Device Name: MONOGLYDE

ABSORBABLE SURGICAL SUTURE (SYNTHETIC) (MONOFILAMENT POLIGLECAPRONE 25)

Indications For Use:

MONOGLYDE IS A SYNTHETIC ABSORBABLE (POLIGLECAPRONE 25) SURGICAL SUTURE; STERILE FLEXIBLE STRAND, PREPARED AND SYNTHESIZED FROM THE MONOMERS, POLYGLYCOLIC ACID 75% AND CAPROLACTONE 25% AND IS INDICATED FOR USE IN GENERAL SOFT TISSUE APPROXIMATION, BUT NOT FOR USE IN OPHTHALMIC SURGERY, CARDIO-VASCULAR AND NEUROLOGICAL TISSUES

Neil R.P. O'Flynn for man

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

1081002 510(k) Number__

Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1

§ 878.4493 Absorbable poly(glycolide/l-lactide) surgical suture.

(a)
Identification. An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.