K Number
K120776

Validate with FDA (Live)

Manufacturer
Date Cleared
2012-09-11

(181 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw retained or cement retained to the abutment.

Device Description

BIOMET 3i Encode Patient Specific Abutment for Nobel Active Connection can be compared to current Encode Patient Specific Abutment for Straumann Bone level Connection. The change to this device is only to the connection that is intended to interface which is identical to the interface found on the Nobel Active Internal Connection Implant. This feature does not modify the intended functionality of the implant/abutment system. This device is also intended for dental laboratories to provide limited design input to match the anatomical requirements of the patient.

AI/ML Overview

The device is the Encode Patient Specific Abutment for Nobel Active Connection. The study conducted was Design Verification Testing according to ISO 14801:2007 "Dentistry -- Dynamic Fatigue Test for Endosseous Dental Implants".

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Meets mechanical properties recommendations by FDA as outlined in "Guidance for Industry and FDA Staff: - Special Controls Class II -Standards: Root Form Endosseous Dental Abutment/ Implant"Bench Testing conducted in accordance with the FDA Guidance document demonstrates that the proposed device meets the mechanical properties recommendations by FDA.
Meets the requirements of ISO 14801:2007 "Dentistry -- Dynamic Fatigue Test for Endosseous Dental Implants"All testing conducted met the acceptance criteria and evaluated the worst-case scenario, including 30° pre-angled abutments as compared to predicate BIOMET 3i designs commercially in the marketplace.
Substantially equivalent to predicate devices (K071370 and K112730) in terms of intended use, indications for use, technological characteristics, and principles of operation.The device has the same intended use, indications for use, and similar technological characteristics and principles of operation as the predicate devices. The major technological difference (connection interface) was also evaluated and found to be safe and effective.

2. Sample Size and Data Provenance for Test Set:

The document does not specify the exact sample size used for the Design Verification Testing (test set). It mentions "all testing conducted" and "worst-case scenario including 30° pre-angled abutments". The data provenance is internal testing conducted by BIOMET 3i, presumably in the USA. The data is prospective in the sense that the testing was performed specifically for this submission to demonstrate compliance.

3. Number of Experts and Qualifications for Ground Truth:

This section is not applicable. The study involved bench testing with physical devices and standards, not clinical data requiring expert human assessment for ground truth.

4. Adjudication Method:

This section is not applicable, as the study involves bench testing of physical devices against predefined standards, not human assessment of data.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This section is not applicable. This is a 510(k) submission for a medical device (dental implant abutment) and involves bench testing to demonstrate mechanical performance and substantial equivalence, not a clinical study involving human readers or AI assistance.

6. Standalone (Algorithm Only) Performance Study:

This section is not applicable. This is a physical medical device, not a software algorithm.

7. Type of Ground Truth Used for Test Set:

The ground truth for the test set was defined by regulatory standards and guidelines, specifically:

  • ISO 14801:2007 "Dentistry -- Dynamic Fatigue Test for Endosseous Dental Implants"
  • FDA Guidance Document: "Guidance for Industry and FDA Staff: - Special Controls Class II -Standards: Root Form Endosseous Dental Abutment/ Implant"

8. Sample Size for Training Set:

This section is not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The study references predicate devices and prior BIOMET 3i designs, which broadly serve as a basis for comparison, but not a formally defined training set in the context of an algorithm.

9. How Ground Truth for Training Set Was Established:

This section is not applicable for the reasons stated above.

{0}------------------------------------------------

K120776

SEP 11 2012

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR & 807.93

Submitter: BIOMET 3i 4555 Riverside Drive Palm Beach Gardens, FL 33410

Establishment Reg. 1038806 Number:

Prepared By: Mayank Choudhary Regulatory Affairs Specialist BIOMET 3i 4555 Riverside Drive Palm Beach Gardens, FL 33410 Tel. 561-776-6831 Fax. 561-514 6316 Email mayank.choudhary@BIOMET.com

Jacquelyn Hughes Contact for Questions/Review: Director Clinical and Regulatory Affairs BIOMET 3! 4555 Riverside Drive Palm Beach Gardens, FL 33410 Tel. 561-776-6700 Fax. 561-514 6316 Email jacquelyn.hughes@BIOMET.com

Encode Patient Specific Abutment for Nobel Active Connection Trade/Proprietary Name:

Common/Usual Dental Implant Abutment Name:

Classification Abutment, implant, dental, endosseous/ Dental Panel Name/ FDA Reviewing Branch:

Device Class II - 21 CFR $872.3630 / NHA Classification/Code:

K112730 - Encode Patient Specific Abutment for Straumann Bone Level Connection Predicate Device K071370 - Nobel Active Internal Connection Implant Manufacturer:

{1}------------------------------------------------

Purpose of the SPECIAL 510(k) notice:

The reason for this Special 510k submission is to request clearance for a modification to a device that has been cleared under the 510(k) process referred to herein as Encode Patient Specific Abutment for Nobel Active Connection. Dental Implant Abutment are referenced under 21 CFR §872.3630 and are considered Class II devices.

Device Description: BIOMET 3i Encode Patient Specific Abutment for Nobel Active Connection can be compared to current Encode Patient Specific Abutment for Straumann Bone level Connection. The change to this device is only to the connection that is intended to interface which is identical to the interface found on the Nobel Active Internal Connection Implant. This feature does not modify the intended functionality of the implant/abutment system. This device is also intended for dental laboratories to provide limited design input to match the anatomical requirements of the patient.

Indications for Use:

BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentulous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw retained or cement retained to the abutment.

Bellatek™ Patient Specific Abutment (EDA and EDAX)
ManufacturerImplantPlatform (mm)
Biomet 3iBiomet3i Certain Osseotite, Nanotite3.4
4.1
5
6
Biomet 3iBiomet3i Ex-Hex Osseotite, Nanotite3.4
4.1
5
6
Nobel BiocareInternal Connection Nobel Replace Implant3.5
4.3
5
6
Internal Connection Nobel Active Implant3.5

{2}------------------------------------------------

4.3
5.0
3.3
StraumannNobel Active Implant4.1
4.8

Technological Characteristics: The predicates and Encode Patient Specific Abutment for Nobel Active Connection have a number of very similar and equivalent design/ technological characteristics, as follows:

Element ofComparisonProposed BIOMET 3i PatientSpecific Abutment for NobelActive ConnectionPredicate OEM Nobel Active InternalConnection ImplantsPredicate Encode PatientSpecific Abutments forStraumann Bone Level
510(k) NumberTBDK071370K112730
IndicationsThe 3i Patient-Specific DentalAbutment is intended for use asan accessory to an endosseousdental implant to support aprosthetic device in a partiallyor edentulous patient. It isintended for use to supportsingle and multiple toothprostheses, in the mandible ormaxilla. The prostheses can bescrew or cement retained tothe abutment.Nobel Biocare's NobelActive implantsare endosseous implant intended tobe surgically placed in the bone of theupper or lower jaw arches to providesupport for prosthetic devices, suchas an artificial tooth, in order torestore patient esthetics and chewingfunction. Nobel Biocare's NobelActiveimplants are indicated for single ormultiple unit restorations in splintedor non-splinted applications. NobelBiocare's NobelActive implants maybe placed immediately and put intoimmediate function provided thatinitial stability requirements detailedin the manual are satisfied.The 3i Patient-SpecificDental Abutment isintended for use as anaccessory to anendosseous dentalimplant to support aprosthetic device in apartially or edentulouspatient. It is intended foruse to support single andmultiple toothprostheses, in themandible or maxilla. Theprostheses can be screwor cement retained to theabutment.
MatingConnectionAnti-rotational, internalengagement featurecompatible with Nobel ActiveImplantsAnti-rotational, internal engagementfeature for Nobel Active InternalConnection ImplantsAnti-rotational, internalengagement featurecompatible withStraumann Bone Level
StatisticaltoleranceanalysisYesOEM confidential dataYes

{3}------------------------------------------------

CompatiblePlatforms3.5mm, 4.3mm, 5.0mm3.5mm, 4.3mm, 5.0mm3.3m, 4.1mm, 4.8mm
MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI
BiocompatibleYesYesYes
Screw MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI

Performance Data: BIOMET 31 has conducted Design Verification Testing according to ISO 14801:2007 "Dentistry -- Dynamic Fatigue Test for Endosseous Dental Implants" on the Encode Patient Specific Abutment for Nobel Active Connection under this submission. All testing conducted met the acceptance criteria and evaluated the worst case scenario including 30° pre-angled abutments as compared to predicate BIOMET 31 designs commercially in the marketplace. Performance testing data indicates that changes to the predicate device are safe and effective for its intended use and it demonstrate to be substantially equivalent. Bench Testing conducted outlined in the FDA Guidance referenced above demonstrates that the proposed device meets the mechanical properties recommendations by FDA.

N/A Clinical Data:

The following FDA Guidance Document for this type of product was utilized in this Performance submission: "Guidance for Industry and FDA Staff: - Special Controls Class II -Standards: Root Form Endosseous Dental Abutment/ Implant". Also, Testing was conducted following ISO standard 14801:2007 Dentistry -- Implants - "Dynamic fatigue test for endosseous dental implants". The test articles met all predetermined acceptance criteria.

Substantial The Encode Patient Specific Abutment for Nobel Replace Connection included in this submission have the same intended use, indications for use, technological Equivalence: characteristics, and principles of operation as previously cleared Encode Patient Specific abutment for Straumann Bone Level connection per the 510(k) number referenced in the Predicate Devices section above.

· Refer to the following substantial equivalence data table:

Element ofComparisonProposed BIOMET 3i PatientSpecific Abutment for NobelActive ConnectionPredicate OEM Nobel ActiveImplantsPredicate Encode PatientSpecific Abutments forStraumann Bone Level
510(k) NumberTBDK071370K112730
IndicationsThe 3i Patient-Specific DentalAbutment is intended for use asNobel Biocare's NobelActive implantsare endosseous implant intended toThe 3i Patient-SpecificDental Abutment is

3/13/2012

{4}------------------------------------------------

an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prostheses can be screw or cement retained to the abutment.be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. Nobel Biocare's NobelActive implants are indicated for single or multiple unit restorations in splinted or non-splinted applications. Nobel Biocare's NobelActive implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.intended for use as an accessory to an endosseous dental implant to support a prosthetic device in a partially or edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prostheses can be screw or cement retained to the abutment.
MatingConnectionAnti-rotational, internal engagement feature compatible with Nobel Active ImplantsAnti-rotational, internal engagement feature for Nobel Active Internal Connection ImplantsAnti-rotational, internal engagement feature compatible with Straumann Bone Level
StatisticaltoleranceanalysisYesOEM confidential dataYes
CompatiblePlatforms3.5mm, 4.3mm, 5.0mm3.5mm, 4.3mm, 5.0mm3.3mm, 4.1mm, 4.8mm
MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI
BiocompatibleYesYesYes
Screw MaterialTi-6Al-4V ELITi-6Al-4V ELITi-6Al-4V ELI

Conclusion: Encode Patient Specific Abutment for Nobel Active Connection and predicate designs have the same intended use, indications for use, similar technological characteristics, and principles of operation. The major technological difference between the Encode Patient Specific Abutment for Nobel Active Connection and its predicates is:

The change is only to the connection that is intended to interface which is identical to the interface found on the Nobel Active Internal Connection Implant.

The information demonstrates that the proposed device is substantially equivalent to the predicate device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains an image of an eagle in the center. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are written around the edge of the circle. The eagle is facing to the right and has its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 11 2012

Biomet 3i Ms. Tamara West Senior Regulatory Affairs Specialist 4555 Riverside Drive Palm Beach Gardens, Florida 33410

Re: K120776

Trade/Device Name: Encode Patient Specific Abutment for Nobel Active Connection Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: August 13, 2012 Received: August 15, 2012

Dear Ms. West:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Ms. West

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use .

.

510(k) Number (if known): K120776

Device Name: Encode Patient Specific Abutment for Nobel Active Connection

Indications for Use: BIOMET 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support a prosthetic device in a partially or completely edentylous patient. A dental abutment is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be screw retained or cement retained to the abutment.

Bellatek™ Patient Specific Abutment (EDA and EDAX)
ManufacturerImplantPlatform (mm)
Biomet 3iBiomet3i Certain Osseotite, Nanotite3.4
4.1
5
6
Biomet 3iBiomet3i Ex-Hex Osseotite, Nanotite3.4
4.1
5
6
Nobel BiocareInternal Connection Nobel Replace Implant3.5
4.3
5
6
Nobel BiocareInternal Connection Nobel Active Implant3.5
4.3
5.0

{8}------------------------------------------------

StraumannNobel Active Implant3.3
4.1
4.8

Prescription Use _____________________________________________________________________________________________________________________________________________________________ Ov er-The-Counter Use _

(Part 21 CFR 801 Subpart D)

(Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runne

(Division Sign-Off) (Division Sign-Off)
Division of Anestheslology, General Hospital
Sheether Control Dental Devices Division of Anesthesions .
Infection Control, Dental Devices

510(k) Number:

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)