(84 days)
The Nanovis Spinal System is intended for posterior, non-cervical (i.e., T1-S1) pedicle and non-pedicle fixation to provide immobilization and stabilization in skeletally mature patients as an adjunct to fusion for the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, dislocation, spinal stenosis, scoliosis, kyphosis, lordosis, spinal tumor, pseudarthrosis and failed previous fusion.
The Nanovis Spinal System consists of rods, hooks, monoaxial and polyaxial screws with set screws, and crosslinks with fastening set screws. Rods are available either straight or pre-contoured in a variety of lengths. Hooks are offered in a variety shapes and sizes. Solid and cannulated monoaxial and polyaxial screws are available in standard and reduction versions in a variety of diameter-length combinations to accommodate differing patient anatomy.
The Nanovis Spinal System is a medical device and its approval is based on a demonstration of substantial equivalence to predicate devices rather than the typical performance study structure found in AI/ML device submissions. Therefore, many of the requested categories related to human-AI performance, sample sizes for training/test sets, and ground truth establishment are not applicable.
Here's the information extracted from the provided text regarding the Nanovis Spinal System:
Acceptance Criteria and Device Performance
The acceptance criteria for the Nanovis Spinal System are based on demonstrating mechanical performance that is "as well as or better than the predicate device" through ASTM F1717 testing.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical performance "as well as or better than" the predicate device for: | "The mechanical test results demonstrated that the Nanovis Spinal System performs as well as or better than the predicate device." |
| - Static compression bending | (Met) |
| - Static torsion | (Met) |
| - Dynamic compression bending of the worst-case construct | (Met) |
Study Proving Device Meets Acceptance Criteria
The study conducted was a mechanical testing study performed according to ASTM F1717.
-
Sample size used for the test set and data provenance:
- Sample Size: Not explicitly stated as a number of samples in the classic sense (e.g., cases, images). The testing refers to "the worst case Nanovis construct." This implies that specific configurations of the device were chosen for testing, likely representative samples of the manufactured components.
- Data Provenance: Not applicable in the context of clinical data. This refers to mechanical testing performed on physical prototypes/products.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Mechanical testing uses engineering standards (ASTM F1717) and physical measurements as the 'ground truth,' not expert clinical opinion.
-
Adjudication method for the test set:
- Not applicable. Adjudication is relevant for subjective assessments, typically in clinical studies or image review. Mechanical tests yield objective, quantifiable results.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is a mechanical device, not an AI/ML diagnostic or assistive tool.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This refers to a physical medical device (spinal implant system).
-
The type of ground truth used:
- The "ground truth" for this device's performance evaluation was established by the specified mechanical testing protocols and acceptance criteria of ASTM F1717. This standard defines the methods for assessing the static and fatigue properties of metallic spinal implant constructs.
-
The sample size for the training set:
- Not applicable. This device does not involve a "training set" in the context of AI/ML or statistical modeling.
-
How the ground truth for the training set was established:
- Not applicable. As above, there is no "training set" in this context.
{0}------------------------------------------------
JAN 1 9 2012
Image /page/0/Picture/2 description: The image shows the word "NANOVIS" in a stylized font. The letters are black and the font is sans-serif. The "N" is stylized with a curved line on the left side. The word is centered in the image and the background is white.
| 510(k) Summary | |
|---|---|
| Date: | 21 October 2011 |
| Sponsor: | Nanovis, LLC5865 East State Rd. 14Columbia City, Indiana 46725 USA(877) 907-6266(260) 625-3834 |
| Contact Person: | Matthew Hedrick, CEO & Chief Operating Officer |
| Proposed Trade Name: | Nanovis Spinal System |
| Device Classification | Class III, Class II |
| Classification Name: | Pedicle screw and hook spinal system |
| Regulation: | 888.3070, 888.3050 |
| Device Product Code: | NKB, MNI, MNH, KWP |
| Device Description: | The Nanovis Spinal System consists of rods, hooks, monoaxial andpolyaxial screws with set screws, and crosslinks with fastening setscrews. Rods are available either straight or pre-contoured in a varietyof lengths. Hooks are offered in a variety shapes and sizes. Solid andcannulated monoaxial and polyaxial screws are available in standardand reduction versions in a variety of diameter-length combinations toaccommodate differing patient anatomy. |
| Intended Use: | The Nanovis Spinal System is intended for posterior, non-cervical (i.e.,T1-S1) pedicle and non-pedicle fixation to provide immobilization andstabilization in skeletally mature patients as an adjunct to fusion for thetreatment of the following acute and chronic instabilities or deformities:degenerative disc disease (DDD, defined as back pain of discogenicorigin with degeneration of the disc confirmed by history andradiographic studies), spondylolisthesis, fracture, dislocation, spinalstenosis, scoliosis, kyphosis, lordosis, spinal tumor, pseudarthrosisand failed previous fusion. |
| Materials: | The Nanovis Spinal System components are manufactured fromtitanium alloy (Ti-6Al-4V ELI) as described by ASTM F136. |
| Predicate Devices: | CD HORIZON® (K031655/K041460),Expedium™ 5.5 Ti Spine System (K041119/K051024)Moss® Miami (K992168/K022623) and theSynergy™ VLS (K950099/K974749). |
Pag 1 of 2
Page 5
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a sequence of handwritten characters, which appear to be a combination of letters and numbers. The sequence reads as 'KII3173'. The characters are written in a simple, clear style, making them easily readable. The image seems to be a close-up of the written sequence.
Technological Characteristics: The Nanovis Spinal System possesses the same technological characteristics as the predicates. These include
- basic design (rod-based fixation system having monoaxial and . polyaxial screws),
- material (titanium alloy), .
- sizes (rod and screw sizes are encompassed by those offered by . the predicate systems) and
- intended use (as described above). .
The fundamental scientific technology of the Nanovis Spinal System is the same as the previously cleared device.
Static compression bending and torsion, and dynamic compression Performance Data: bending of the worst case Nanovis construct was performed according to ASTM F1717. The mechanical test results demonstrated that the Nanovis Spinal System performs as well as or better than the predicate device.
In comparison to the predicate devices, the Nanovis Spinal System Conclusion: has
- . the same intended use (as described above),
- the same technological characteristics (as described above) .
- and so does not raise new questions of safety and effectiveness.
Therefore the Nanovis Spinal System can be found substantially equivalent to the predicate devices.
Pg 1 of 2 Page 6
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight. The symbol is composed of three curved lines that converge at the bottom.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 1 9 2012
Nanovis. LLC % Backroads Consulting, Inc. Karen E. Warden, Ph.D. 8202 Sherman Road Chesterland, Ohio 44026-2141
Re: K113173
Trade/Device Name: Nanovis Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI, K WP Dated: October 21, 2011 Received: October 27, 2011
Dear Dr. Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Karen E. Warden, Ph.D.
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Radiological Health
Mark N. Melkerson
Director
Der LIN
Division of Surgical, Orthopedic
and Restorative Devices
Office of Device Evaluation
Center for Devices and
Enclosure
{4}------------------------------------------------
Indications for Use Statement
510(k) Number:_K113133
Device Name: Nanovis Spinal System
Indications for Use:
The Nanovis Spinal System is intended for posterior, non-cervical (i.e., T1-S1) pedicle and non-pedicle fixation to provide immobilization and stabilization in skeletally mature patients as non adjunct to fusion for the treatment of the following acute and chronic instabilities or deformities: degenerative disc disease (DDD, defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, dislocation, spinal stenosis, scoliosis, kyphosis, lordosis, spinal tumor, fracture. pseudarthrosis and failed previous fusion.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113173
N/A