    Intended Use

    The Nano FortiFix® System is intended to provide immobilization of spinal segments in sketally mature patients as an adjunct to fusion in the non-cervical spine for the following indications: degenerative disc disease, spinal stenosis, spondylolisthesis, spinal deformities (i.e., scoliosis, and/ or lordosis), trauma (i.e., fracture or dislocation), tumor, pseudarthrosis and/or failed previous fusion. The Nano FortiFix® System can be used in an open approach or a percutaneous approach with MIS instrumentation.

    Device Description

    Nano FortiFix® is a posterior pedicle screw system consisting of rods, polyaxial and monoaxial pedicle screws, connectors and fasteners in a variety of sizes to accommodate differing anatomic requirements. The Nano FortiFix® pedicle screw shaft is available having a nanosurface. The Nano FortiFix® nano pedicle screw has a micro- and nano-roughened surface that demonstrates the requirements for nanotechnology. The surface of the nano screw threads has been deliberately manipulated to produce nanoscale dimensions which exhibit specific properties. These threads are electrochemically treated to possess a controlled nanotopography composed of nanotube arrays having a pore size diameter between 30-90 nanometers. Calcium and phosphate are incorporated into the nanotube surface.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Nano FortiFix® System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What was measured?)Reported Device Performance (Was it met?)
    Mechanical Performance
    * Static Compression Strength (ASTM F1717)Demonstrated performance substantially equivalent to predicate devices.
    * Dynamic Compression Strength (ASTM F1717)Demonstrated performance substantially equivalent to predicate devices.
    * Static Torsion Strength (ASTM F1717)Demonstrated performance substantially equivalent to predicate devices.
    Nanotechnology Adherence/Performance
    * Nanosurface characteristics (pore size diameter 30-90 nanometers)The device's nano pedicle screw has a micro- and nano-roughened surface with electrochemically treated nanotopography composed of nanotube arrays having a pore size diameter between 30-90 nanometers, and calcium and phosphate incorporated into the nanotube surface. This confirms the stated nanotechnology characteristics.
    * Mineralization of Extracellular Matrix (Osteoblasts & Mesenchymal Stem Cells)In vitro study results demonstrated that the Nano FortiFix nanosurface develops statistically significantly greater mineralization in both osteoblast and mesenchymal stem cell cultures compared to other surfaces. (This indicates it met the performance criteria for functional benefit related to the nanosurface.)
    Biocompatibility/Safety
    * Bacterial Endotoxin Testing (AAMI ST72:2011)Bacterial endotoxin testing was conducted in accordance with AAMI ST72:2011 and met the specified testing limit.
    Substantial EquivalenceThe Nano FortiFix® System possesses the same intended use and technological characteristics (performance, basic design, implant grade materials, and sizes) as the predicate devices. Therefore, Nano FortiFix® is substantially equivalent for its intended use.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • For mechanical testing (ASTM F1717), the text mentions "worst case construct" but does not specify the numerical sample size (e.g., number of constructs tested for each type of loading or the number of repetitions for each test).
      • For in vitro evaluations (mineralization), the text does not specify the numerical sample size (e.g., number of cell cultures, experimental replicates).
      • For bacterial endotoxin testing, the text does not specify the numerical sample size (e.g., number of samples tested).
    • Data Provenance: The studies were conducted by the manufacturer as part of their 510(k) submission. The exact country of origin for the studies is not explicitly stated, but the sponsor's location (Nanovis Spine, LLC, Columbia City, Indiana, USA) suggests a U.S. origin. All studies appear to be prospective in nature, designed specifically to evaluate the Nano FortiFix® System.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable to the type of studies described. The studies are primarily in vitro (cell culture and mechanical testing) and material/biological safety analyses (bacterial endotoxin). There is no "ground truth" derived from human experts in the context of diagnostic performance or clinical outcomes. The "ground truth" for these tests comes from adherence to established standards (ASTM F1717, AAMI ST72:2011) and scientific principles for cell culture and material properties.

    4. Adjudication method for the test set

    • This information is not applicable as the studies are not based on interpreting patient data where adjudication would be necessary. The results are quantitative measurements from laboratory tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a thoracolumbosacral pedicle screw system, an implantable medical device, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • No, a standalone algorithm performance study was not done. This device is a physical implant, not an algorithm.

    7. The type of ground truth used

    • For mechanical testing: The ground truth is established by the specified performance requirements of the ASTM F1717 standard and the comparative performance of legally marketed predicate devices. The "truth" is whether the device meets or exceeds these established benchmarks.
    • For nanotechnology/mineralization studies: The ground truth is the quantitative measurement of mineralization in cell cultures and the successful demonstration of the specified nanotopography. The comparison to "other surfaces" provides the baseline.
    • For bacterial endotoxin testing: The ground truth is defined by the "specified testing limit" within the AAMI ST72:2011 standard.

    8. The sample size for the training set

    • Not applicable. This device is a physical implant, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for a machine learning model, there is no ground truth established for a training set.