(177 days)
The Comprehensive® Segmental Revision System is intended for use in cases of:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- Rheumatoid arthritis.
- Revision where other devices or treatments have failed.
- Correction of functional deformity.
- Oncology applications including bone loss due to tumor resection.
When used in a proximal or total humeral replacement, the Comprehensive® Segmental Revision System is also intended for:
Treatment of acute or chronic fractures with humeral head (shoulder) involvement, which are unmanageable using other treatment methods.
When used as a distal or total humeral replacement, the Comprehensive® Segmental Revision System is also intended for:
Treatment of acute or chronic fractures with humeral epicondyle (elbow) involvement, which are unmanageable using other treatment methods.
The Comprehensive® Segmental Revision System is intended for use with or without bone cement in the proximal shoulder.
The Comprehensive® Segmental Revision System is intended for use with bone cement in distal humeral and total humeral applications.
Tissue Attachment Augments provide the option for tissue stabilization and attachment.
The Comprehensive® SRS address the needs of the upper extremity, especially in cases where there is marked bone loss. Applications of the system include proximal humeral (shoulder) replacements distal humeral (elbow) replacements and total humeral replacements. Components of the system include humeral heads, proximal humeral bodies, intercalary segments, humeral stems, total humeral couplers, distal bodies with a modular flange, and modular tissue attachment augments.
The provided text describes a 510(k) submission for the "Comprehensive® Segmental Revision System (SRS)," a shoulder, elbow, and total humeral replacement prosthesis. The submission details the device's characteristics and its substantial equivalence to predicate devices based on non-clinical testing. Crucially, the document explicitly states, "No clinical data submitted" and "No clinical data was necessary for a determination of substantial equivalence."
Therefore, many of the requested elements regarding acceptance criteria and studies (especially clinical ones) cannot be extracted from this document, as they were not conducted or submitted for this device.
Here's a breakdown of the information that can be provided based on the input:
1. A table of acceptance criteria and the reported device performance
The document lists "Performance Data" which are summaries of non-clinical tests. It does not explicitly state numerical acceptance criteria for these tests, but implies that the device "performed within the intended use" and was comparable to the predicate devices.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Structural integrity at weakest point of construct | Engineering analysis to determine weakest point of the construct (presumably met acceptable design specifications) |
| Range of motion | Engineering analysis to determine range of motion (presumably within physiological limits for the intended use) |
| Strength of smaller diameter long stems | Engineering analysis to justify smaller diameter long stems (presumably demonstrating adequate strength) |
| Stem strength compared to predicate | Cantilever fatigue testing to compare stem strength to predicate (results indicated comparable strength, leading to substantial equivalence) |
| Augment stability post-cyclic loading | Cyclic loading followed by screw torque out to confirm augment stability (results indicated stability) |
| Flange loading determination | Engineering analysis for flange loading determination (presumably met design specifications) |
| Humeral flange fatigue | Cyclic fatigue testing of humeral flange (results indicated acceptable fatigue life) |
| Static axial separation of SRS taper junction | Static Axial Separation of SRS taper junction (results indicated acceptable separation force) |
| Static axial separation of Comprehensive® taper junction | Static Axial Separation of Comprehensive® taper junction (results indicated acceptable separation force) |
| Elbow condyle strength | Shear testing to determine elbow condyle strength (results indicated acceptable strength) |
| MR compatibility | MR Compatibility to ASTM F2182-09 (results indicated compliance with the standard) |
| Overall safety and effectiveness (compared to predicate devices) | The results of analysis and mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices. (This is the overarching conclusion, implying all acceptance criteria were met by showing equivalence to legally marketed devices.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample size: Not specified for any of the non-clinical tests. Non-clinical mechanical testing typically involves a small number of samples depending on the test type (e.g., typically N=3 to N=6 or more for fatigue testing per ISO standards).
- Data provenance: The tests are non-clinical, so country of origin of patient data is not applicable. The tests were performed as part of a 510(k) submission in the US. The design of these tests suggests they are prospective in nature within a laboratory setting.
- Retrospective or prospective: The non-clinical tests are, by nature, prospective in that they were designed and executed to evaluate specific performance aspects of the newly designed device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. As "No clinical data submitted" there was no ground truth established by experts for a test set in the clinical sense. The "ground truth" for non-clinical tests would be the established engineering and materials science principles and standards against which the device performance was measured. These would typically be evaluated by engineers and materials scientists.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable, as no clinical test set with expert adjudication was involved.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. No MRMC study was conducted. This device is a physical implant, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a physical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the non-clinical tests, the "ground truth" was based on established engineering principles, material properties standards (e.g., relevant ASTM standards like F2182-09 for MR Compatibility), and performance characteristics of predicate devices. The goal was to demonstrate that the new device performed equivalently or acceptably according to these engineering benchmarks, not against clinical outcomes or expert diagnoses.
8. The sample size for the training set
- Not applicable. As no clinical data was submitted and the device is a physical implant, there was no "training set" in the context of machine learning or AI models. The design and testing of the device would have involved an iterative process where engineering data might inform design changes, but this is not analogous to a machine learning training set.
9. How the ground truth for the training set was established
- Not applicable, as there was no training set in the context of machine learning or AI models.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a blurry, black and white image of what appears to be Arabic text. The text is written in a calligraphic style, with the letters connected to each other. Due to the blurriness of the image, it is difficult to read the text. The background of the image is white.
DEC 1 5 2011
・・
re
Image /page/0/Picture/3 description: The image shows the logo for BIOMET MANUFACTURING CORP. The word "BIOMET" is in a stylized font with the letters connected. Below the logo, the words "MANUFACTURING CORP." are printed in a simple, sans-serif font. The text is black on a white background.
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| SUBMITTER INFORMATION | ||
|---|---|---|
| Name | Biomet Manufacturing Corp. | |
| Address | 56 East Bell DriveWarsaw, IN 46582 | |
| Phone number | (574) 267-6639 | |
| Fax number | (574) 371-1027 | |
| Establishment Registration Number | 1825034 | |
| Name of contact person | Patricia Sandborn BeresSenior Regulatory Specialist | |
| Date prepared | December 5, 2011 | |
| DEVICE INFORMATION | ||
| Name of device | ||
| Trade or proprietary name | Comprehensive® Segmental Revision System (SRS) | |
| Common or usual name | Shoulder, elbow and total humeral replacement prosthesis | |
| Classification name | Shoulder joint metal/polymer non-constrained cemented prosthesis Shoulder joint metal/polymer semi-constrained cemented prosthesis Shoulder joint metal/polymer/metal non-constrained or semi-constrained porous coated uncemented prosthesis Elbow joint metal/polymer constrained cemented prosthesis | |
| Classification panel | Orthopedics | |
| Regulation | 21 CFR§888.3650/3660/3670/3150 | |
| Product Code(s) | KWT, KWS, MBF, JDC | |
| Legally marketed device(s) towhich equivalence is claimed | K043505- Discovery® - Mosaic® Total Humeral System K042321- Mosaic® Non-Modular Proximal Body and EAS Offset Modular Humeral Head K033280 -- Discovery® Elbow - Mosaic® Proximal Humeral Replacement System K020045 - 3-Piece Proximal Humeral Replacement System K022079 - Short and Long Tissue Attachment Sleeves K925613 - Proximal Humeral Replacement System K030710 - Bio-Modular® Shoulder System | |
| Reason for 510(k) submission | Update of existing product line | |
| Device description | The Comprehensive® SRS address the needs of the upper extremity, especially in cases where there is marked bone loss. Applications of the system include proximal humeral (shoulder) replacements distal humeral (elbow) replacements and total humeral replacements. Components of the system include humeral heads, proximal humeral bodies, intercalary segments, humeral stems, total humeral couplers, distal bodies with a modular flange, and modular tissue attachment augments. | |
| Indications for use | The Comprehensive® Segmental Revision System is intended for usein cases of:1. Non-inflammatory degenerative joint disease includingosteoarthritis and avascular necrosis.2. Rheumatoid arthritis.3. Revision where other devices or treatments have failed.4. Correction of functional deformity.5. Oncology applications including bone loss due to tumor resection.When used in a proximal or total humeral replacement, theComprehensive® Segmental Revision System is also intended for:Treatment of acute or chronic fractures with humeral head(shoulder) involvement, which are unmanageable using othertreatment methods.When used as a distal or total humeral replacement, theComprehensive® Segmental Revision System is also intended for:Treatment of acute or chronic fractures with humeral epicondyle(elbow) involvement, which are unmanageable using othertreatment methods. | |
| Intended use of the device | The Comprehensive® Segmental Revision System is intended for usewith or without bone cement in the proximal shoulder.The Comprehensive® Segmental Revision System is intended for usewith bone cement in distal humeral and total humeral applications.Tissue Attachment Augments provide the option for tissuestabilization and attachment. | |
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICECOMPARED TO THE PREDICATE | ||
| Characteristic | Comprehensive® SRS | Predicate |
| Materials | Ti-6Al-4VCo-Cr-Mo | K020045K020045 |
| Principal of Operation | Proximal humeral, distal humeralor total humeral replacement | K925613, K020045, K033280,K043505 |
| Proximal Bodies | Resection Levels - 42-71mm | K020045 |
| Humeral Stems | Taper Head Attachment | K030710 |
| Diameters - 4-20mm | K020045, K033280, K925613 | |
| Lengths - 75-200mm | K020045, K033280, K925613 | |
| Porous Coating Proximally | K020045, K033280, K030710 | |
| Intercalary Segments and Couplers | Diameter - 19mm | K020045, K033280, K043505 |
| Lengths - 30-120mm | K020045, K033280, K043505 | |
| Tissue Attachment | Augments | K022079 |
| Humeral Heads | Diameters - 40-54mm | K030710 |
| Extended Articular Surface | K042321, K030710 | |
| Distal Humeral Bodies | Resection Height - 50, 60, 70mm | K033280, K043505 |
{1}------------------------------------------------
111746
p.2/3
510(k) Summary
Comprehensive® Segmental Revision System (SRS)
Page 2 of 3
{2}------------------------------------------------
$\rho.^{3}/_{3}$
510(k) Summary
Comprehensive® Segmental Revision System (SRS)
Page 3 of 3
| PERFORMANCE DATA |
|---|
| Summary Of Non-Clinical Tests Conducted For Determination Of Substantial Equivalence |
| Engineering analysis to determine weakest point of the construct |
| Engineering analysis to determine range of motion |
| Engineering analysis to justify smaller diameter long stems |
| Cantilever fatigue testing to compare stem strength to predicate |
| Cyclic loading followed by screw torque out to confirm augment stability |
| Engineering analysis for flange loading determination |
| Cyclic fatigue testing of humeral flange |
| Static Axial Separation of SRS taper junction |
| Static Axial Separation of Comprehensive® taper junction |
| Shear testing to determine elbow condyle strength |
| MR Compatibility to ASTM F2182-09 |
| Summary Of Clinical Tests Conducted For Determination Of Substantial Equivalence And/Or Clinical Information |
| No clinical data submitted |
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA |
| No clinical data was necessary for a determination of substantial equivalence. |
| The results of analysis and mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices. |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corporation % Ms. Patricia Sandborn Beres Senior Regulatory Specialist 56 East Bell Drive Warsaw, Indiana 46582
DEC 1 5 2011
Re: K111746
Trade/Device Name: Comprehensive® Segmental Revision System (SRS) Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT, KWS, MBF, JDC Dated: December 12, 2011 Received: December 13, 2011
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Patricia Sandborn Beres
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincercly yours,
Eunel Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: _Comprehensive® Segmental Revision System (SRS)
İndications For Use:
The Comprehensive® Segmental Revision System is intended for use in cases of:
-
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
-
- Rheumatoid arthritis.
. ")
-
- Revision where other devices or treatments have failed.
-
- Correction of functional deformity.
-
- Oncology applications including bone loss due to tumor resection.
When used in a proximal or total humeral replacement, the Comprehensive® Segmental Revision System is also intended for:
Treatment of acute or chronic fractures with humeral head (shoulder) involvement, which are unmanageable using other treatment methods.
When used as a distal or total humeral replacement, the Comprehensive® Segmental Revision System is also intended for:
Treatment of acute or chronic fractures with humeral epicondyle (elbow) involvement, which are unmanageable using other treatment methods.
The Comprehensive® Segmental Revision System is intended for use with or without bone cement in the proximal shoulder.
The Comprehensive® Segmental Revision System is intended for use with bone cement in distal humeral and total humeral applications.
Tissue Attachment Augments provide the option for tissue stabilization and attachment.
AND/OR
Prescription Use __ X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use __ NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
510(k) Number K111746
N/A