(90 days)
The indications for the use of the Short and Long Soft Tlssue Attachment Sleeves include:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis
- Rheumatoid arthritis
- Revision where other devices or treatments have failed
- Correction of functional deformity
- Treatment of acute fracture of the humeral head unmanageable using other treatment methods
The sleeves can be used for oncology applications.
The sleeves are designed for use with the Biomet Mosaic™ Humeral System (K020045), to provide the option for soft tissue stabilization and attachment.
The Biomet Mosaic™ Humeral System (K020045) is for use with bone cernent.
This particular design utilizes a taper junction between the sleeve and the segment component of the Biomet Proximal Humeral Replacement System (K020045). For secure placement, the sleeve has a larger proximal taper junction. Because the soft tissue will naturally be pulling proximally, if the taper junction does move, the soft tissue will only tighten the taper.
There are two sizes that vary only in length and suture hole availability. The short sleeve has 4 suture holes available and the holes are arranged perpendicular to the axis of the segment. The long sleeve has 5 suture holes that are arranged perpendicular to the axis of the segment and 4 suture holes placed parallel to the axis of the segment. Both sleeves have the same outside and inside diameters and inside mating taper.
The sleeves are designed for use in conjunction with the Biomet Modular Proximal Humeral Replacement System (K020045), now called the Mosaic™ Humeral System to provide the option for soft tissue stabilization and attachment to the prosthetic device
Here's an analysis of the provided information regarding the acceptance criteria and study for the Short and Long Soft Tissue Attachment Sleeve:
Acceptance Criteria and Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Substantial Equivalence | Device must be substantially equivalent to legally marketed predicate devices for its intended use and technology. | The Short and Long Soft Tissue Attachment Sleeve components (materials, design, sizes, and indications) are similar or identical to the predicate devices (LIMA RPS Shoulder (K913282) and Biomet Modular Proximal Humeral Replacement System (K020045), now known as the Mosaic™ Humeral System). The device also uses a taper junction design which, if it moves, is designed to tighten the soft tissue, ensuring secure placement. |
| Safety (Non-Clinical) | Device must not present new risks compared to legally marketed predicate devices. | Mechanical testing with Engineering Justifications determined that the Short and Long Soft Tissue Attachment Sleeve components presented no new risks and were, therefore, substantially equivalent to the predicate device. |
| Effectiveness (Non-Clinical) | Device must be effective for its intended use and comparable to predicate devices. | The mechanical testing and engineering justifications implicitly support effectiveness by demonstrating the device's ability to maintain secure placement and function as intended for soft tissue attachment, similarly to the predicate devices. |
| Indications for Use | Device's indications for use must align with, or be substantially equivalent to, predicate devices. | The stated indications for use are: Non-inflammatory degenerative joint disease (including osteoarthritis and avascular necrosis), rheumatoid arthritis, revision where other devices or treatments have failed, correction of functional deformity, treatment of acute fracture of the humeral head unmanageable using other treatment methods, and oncology applications. The device is specifically designed for use with the Biomet Mosaic™ Humeral System (K020045) for soft tissue stabilization and attachment. These are consistent with the function of similar humeral replacement and soft tissue attachment systems. |
| Biocompatibility/Material | Materials used should be safe and compatible with the human body. (Implicit, though not explicitly detailed in the provided text for acceptance criteria) | The mention of "materials" being similar or identical to predicate devices implies that the materials are considered biocompatible and safe based on prior approvals of the predicate devices. Specific biocompatibility testing results are not detailed in the provided summary. |
Study Information
Based on the provided document, the device approval was based on a demonstration of substantial equivalence to predicate devices, rather than a traditional clinical study with defined acceptance criteria for outcomes like accuracy, sensitivity, or specificity.
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable. No clinical test set (patient population) was used for evaluation.
- Data Provenance: Not applicable. The evaluation was based on non-clinical (mechanical) testing and comparison to existing device specifications.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable. Ground truth from experts in a clinical context was not required as no clinical testing was performed. The "ground truth" here is the established safety and effectiveness of the predicate devices and the engineering principles applied during mechanical testing.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. No test set requiring adjudication was used.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a medical device (prosthesis component), not an AI-assisted diagnostic or clinical decision support tool. Therefore, an MRMC study is irrelevant.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the substantial equivalence determination was based on:
- Predicate Device Performance: The established safety and effectiveness records of the legally marketed predicate devices (LIMA RPS Shoulder (K913282) and Biomet Modular Proximal Humeral Replacement System (K020045)).
- Engineering Justifications: Data derived from mechanical testing, interpreted by engineers, to demonstrate that the device's design, materials, and functional characteristics were comparable and did not introduce new risks.
- The "ground truth" for the substantial equivalence determination was based on:
-
The sample size for the training set:
- Not applicable. There was no "training set" in the context of an algorithm or diagnostic model. The "training" for the device design would implicitly come from the design and manufacturing knowledge for the predicate devices.
-
How the ground truth for the training set was established:
- Not applicable, as there was no training set in the typical sense for algorithms. The device's design and manufacturing process would have been informed by established engineering principles and the performance history of similar devices.
Summary of the Study:
The "study" conducted for the Short and Long Soft Tissue Attachment Sleeve was non-clinical mechanical testing combined with engineering justifications. This study aimed to demonstrate substantial equivalence to existing legally marketed predicate devices. The core findings were:
- The device's materials, design, sizes, and indications are "similar or identical" to the predicate devices.
- Mechanical testing and engineering justifications concluded that the device presented "no new risks" compared to the predicate device.
- No clinical testing was performed or provided as a basis for substantial equivalence.
Therefore, the device met the acceptance criteria by successfully demonstrating that it was as safe and effective as the predicate devices through non-clinical means, rather than through a traditional clinical trial establishing efficacy against specific measurable outcomes in patients.
{0}------------------------------------------------
022079
SEP 2 4 2002
Summary of Safety and Effectiveness
| Applicant/Sponsor: | Biomet Orthopedics, Inc.P.O. Box 587Warsaw, Indiana 46581-0587 |
|---|---|
| Contact Person: | Kacy Arnold, RN, MBATelephone: (574) 372-1644Fax: (574) 372-1683 |
| Proprietary Name: | Short and Long Soft Tissue Attachment Sleeve |
| Common Name: | Short and Long Soft Tissue Attachment Sleeve |
| Classification Name: | Prosthesis, shoulder non-constrained, metal / polymercemented (21 CFR 888.3650) |
Legally Marketed Devices to Which Substantial Equivalence is Claimed:
-
- LIMA RPS Shoulder (K913282)
-
- Biomet Modular Proximal Humeral Replacement System (K020045), now known as the Mosaic™ Humeral System
Device Description: This particular design utilizes a taper junction between the sleeve and the segment component of the Biomet Proximal Humeral Replacement System (K020045). For secure placement, the sleeve has a larger proximal taper junction. Because the soft tissue will naturally be pulling proximally, if the taper junction does move, the soft tissue will only tighten the taper.
There are two sizes that vary only in length and suture hole availability. The short sleeve has 4 suture holes available and the holes are arranged perpendicular to the axis of the segment. The long sleeve has 5 suture holes that are arranged perpendicular to the axis of the segment and 4 suture holes placed parallel to the axis of the segment. Both sleeves have the same outside and inside diameters and inside mating taper.
The sleeves are designed for use in conjunction with the Biomet Modular Proximal Humeral Replacement System (K020045), now called the Mosaic™ Humeral System to provide the option for soft tissue stabilization and attachment to the prosthetic device
{1}------------------------------------------------
022079
0.2/2
Indications:
-
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
-
- Rheumatoid arthritis
-
- Revision where other devices or treatments have failed
-
- Correction of functional deformity
-
- Treatment of acute fracture of the humeral head unmanageable using other treatment methods.
The sleeves can be used for oncology applications.
The sleeves are designed for use with the Biomet Modular Proximal Humeral Replacement System (K020045), now called the Mosaic™ Humeral System to provide the option for soft tissue stabilization and attachment.
The Biomet Proximal Humeral Replacement System (K020045) is for use with bone cement.
| Summary of Technologies: | The Short and Long Soft Tissue AttachmentSleeve components (materials, design, sizesand indications) are similar or identical to thepredicate devices. |
|---|---|
| Non-Clinical Testing: | Mechanical testing with EngineeringJustifications determined that the Short andLong Soft Tissue Attachment Sleevecomponents presented no new risks and weretherefore, substantially equivalent to thepredicate device. |
| Clinical Testing: | No clinical testing was provided as a basis forsubstantial equivalence. |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FP 2 4 2002
Ms. Kacy Arnold, RN, MBA Regulatory Affairs Specialist Biomet, Inc. P.O. Box 587 Warsaw. Indiana 46581-0578
Re: K022079
Trade/Device Name: Short and Long Soft Tissue Attachment Sleeves Regulation Number: 21 CFR 888.3650 Regulation Name: Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWT Dated: June 14, 2002 Received: June 26, 2002
Dear Ms. Arnold:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Kacy Arnold, RN, MBA
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark M. Milliman
Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510 (k) Number (if known):
K 022079
Device Name: Short and Long Soft Tissue Attachment Sleeves
Indications for Use:
The indications for the use of the Short and Long Soft Tlssue Attachment Sleeves include:
- include: necrosis
-
- Rheumatoid arthritis
-
- Revision where other devices or treatments have failed
-
- Correction of functional deformity
-
- Treatment of acute fracture of the humeral head unmanageable using other treatment methods
The sleeves can be used for oncology applications.
The sleeves are designed for use with the Biomet Mosaic™ Humeral System (K020045), to provide the option for soft tissue stabilization and attachment.
The Biomet Mosaic™ Humeral System (K020045) is for use with bone cernent.
Mark N. Millerson
i Pestorative ( %: and Neurological Boulees
510(k) Number K022074
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter-Use_ (Optional Format 1-2-96)
N/A