K Number
K111263

Validate with FDA (Live)

Device Name
CXI TRIFORCE
Manufacturer
Date Cleared
2011-07-25

(82 days)

Product Code
Regulation Number
870.1210
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CXIM TriForce Peripheral Crossing Set is intended to be percutaneously introduced into blood vessels and support a wire guide while performing percutaneous peripheral intervention. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.

Device Description

The CXI™ TriForce is designed to facilitate wire guide exchange, infusion, and wire guide support. Each set contains the following sterile components:

  • Outer 5.0 French Flexor® introducer .
  • Inner 4.0 French CXI™ Support Catheter .
  • Peel-Away® sheath .
    The outer Flexor® introducer is manufactured with a stainless steel coil reinforced nylon construction which provides both flexibility and support to the sheath during use. The outer Flexor introducer is used to support the inner CXI™ Support Catheter during a procedure. The outer Flexor introducers are available in length measurements of 55 or 90 centimeters and can be manufactured with an angled or a straight tip. The inner CXI™ Support Catheter is manufactured with 4.0 French braided flexible kink resistant shaft materials with hydrophilic coating. The flexible shaft material is a laminated material consisting of a nylon layer, stainless steel braiding and an inner polyimide, polyurethane layer, and an innermost polytetrafluoroethylene layer. The inner CXI™ Support Catheters allow acceptance of a 0.035 inch diameter (0.89 millimeters) wire guide. The inner CXI™ Support Catheter will support a wire guide while performing percutaneous peripheral intervention. The CXI™ Support Catheters are available in length measurements of 65 or 100 centimeters and can be manufactured with an angled or a straight tip.
    The tetrafluoroethylene Peel-Away® sheath is included in the set for assistance when inserting the inner CXI™ Support Catheter into the outer Flexor® introducer.
AI/ML Overview

This document describes the acceptance criteria and study proving the device meets those criteria for the CXI™ TriForce Peripheral Crossing Set.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Tensile Strength TestingDevice should not fracture or rupture introducer or catheter during proper clinical use. Meets applicable sections of BS EN ISO 11070 and ISO 10555-1.Predetermined acceptance criteria were met.
Liquid Leakage TestingNo liquid leakage from introducer, introducer valve, or catheter during proper clinical use. Meets applicable sections of ISO 11070 and ISO 10555-1.Predetermined acceptance criteria were met.
Static Burst TestingPressures reached during proper clinical use (maximum pressure at maximum flow rate) are less than the static burst pressure of the catheter, and should not fracture or rupture the catheter.Predetermined acceptance criteria were met.
Dynamic Burst TestingPressures reached during proper clinical use (maximum pressure at maximum flow rate) are less than the dynamic burst pressure of the catheter, and should not fracture or rupture the catheter.Predetermined acceptance criteria were met.
Flow Rate TestingPressure exerted at maximum flow rate during proper clinical use should not fracture or rupture the catheter.Predetermined acceptance criteria were met.
Biocompatibility TestingDevice is biocompatible. Meets applicable sections of ISO 10993-1.Predetermined acceptance criteria were met.

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify the exact sample sizes for each test conducted on the CXI™ TriForce Peripheral Crossing Set. The studies are described as "applicable testing to assure reliable design and performance" and "appropriate engineering tests." The data provenance is not specified, but the tests were performed by Cook Incorporated, a US-based company. These are likely prospective engineering and laboratory tests, not clinical studies on human subjects.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable. The tests performed are engineering and laboratory tests (e.g., tensile strength, burst pressure) on the device itself, not clinical studies requiring expert interpretation of patient data or establishment of ground truth.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods are relevant for studies involving human interpretation or clinical outcomes, not for the direct physical and chemical property testing described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was mentioned or performed. The studies conducted are focused on the device's physical and biological performance against established standards, not on human reader performance, with or without AI assistance.

6. Standalone Performance Study

Yes, a standalone performance study was done in the sense that the device itself was subjected to various engineering and biocompatibility tests to ensure it met predetermined acceptance criteria. These tests were "algorithm only" in the context of device performance, as they measured the intrinsic characteristics of the product.

7. Type of Ground Truth Used

The "ground truth" for these tests consisted of predetermined physical and chemical specifications and standards as outlined in international and national standards (BS EN ISO 11070, ISO 10555-1, ISO 10993-1). The device's performance was measured against these established parameters. For example, the ground truth for tensile strength testing would be a specified maximum force the device must withstand without failure.

8. Sample Size for the Training Set

This information is not applicable. The listed studies are not AI/machine learning studies and therefore do not involve "training sets" in that context. The device performance was evaluated through direct testing against predetermined criteria.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there was no "training set." The performance criteria were established based on recognized engineering standards and regulatory requirements.

{0}------------------------------------------------

K111263 page 1 of 3

1-2

Cook Incorporated CXI™ TriForce Peripheral Crossing Set Traditional 510(k) 13 July 2011

5. 510(k) Summary

JUL 2 5 2011

CXI™ TriForce Peripheral Crossing Set 510(k) Summary 21 CFR 807.92

1. Submitter Information:

Applicant:Cook Incorporated
Address:750 Daniels Way
Bloomington, IN 47404
Phone Number:(800) 468-1379
Fax Number:(812) 332-0281
Contact:Molly Busenbark, MA, RAC
Contact Address:Cook Incorporated
750 Daniels Way
Bloomington, IN 47404
Contact Phone Number:812-339-2235 Ext. 2162
Contact Fax Number:812-332-0281

2. Device Information:

Trade name:CXITM TriForce Peripheral Crossing Set
Common name:Continuous flush catheter
Classification:Class II
Regulation:21 CFR 870.1210
Product Code:KRA

3. Predicate Device:

The CXI™ TriForce Peripheral Crossing Set (hereafter referred to as the CXI™ TriForce) is substantially equivalent to the CXI™ Support Catheter, manufactured by Cook Incorporated, which is cleared under 510(k) number K072724. The CXI™ TriForce is also substantially equivalent to the FineCross™ MG Coronary Micro-Guide Catheter and the Glidesheath™ manufactured by Terumo Medical Corporation, which are cleared under 510(k) numbers K082519 and K102008, respectively.

4. Comparison to Predicate Device:

It has been demonstrated that the CXI™ TriForce is comparable to the predicate devices in terms of design, intended use, materials, fundamental technology, and principal of operation.

5. Device Description:

The CXI™ TriForce is designed to facilitate wire guide exchange, infusion, and wire guide support. Each set contains the following sterile components:

{1}------------------------------------------------

K111263 page 2 of 3

1-3

Cook Incorporated CXI™ TriForce Peripheral Crossing Set Traditional 510(k) 13 July 2011

  • Outer 5.0 French Flexor® introducer .
  • Inner 4.0 French CXI™ Support Catheter .
  • Peel-Away® sheath .

The outer Flexor® introducer is manufactured with a stainless steel coil reinforced nylon construction which provides both flexibility and support to the sheath during use. The outer Flexor introducer is used to support the inner CXI™ Support Catheter during a procedure. The outer Flexor introducers are available in length measurements of 55 or 90 centimeters and can be manufactured with an angled or a straight tip. The inner CXI™ Support Catheter is manufactured with 4.0 French braided flexible kink resistant shaft materials with hydrophilic coating. The flexible shaft material is a laminated material consisting of a nylon layer, stainless steel braiding and an inner polyimide, polyurethane layer, and an innermost polytetrafluoroethylene layer. The inner CXI™ Support Catheters allow acceptance of a 0.035 inch diameter (0.89 millimeters) wire guide. The inner CXI™ Support Catheter will support a wire guide while performing percutaneous peripheral intervention. The CXI™ Support Catheters are available in length measurements of 65 or 100 centimeters and can be manufactured with an angled or a straight tip.

The tetrafluoroethylene Peel-Away® sheath is included in the set for assistance when inserting the inner CXI™ Support Catheter into the outer Flexor® introducer.

6. Intended Use:

The CXI™ TriForce is intended to be percutaneously introduced into blood vessels and support a wire guide while performing percutaneous peripheral intervention. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.

7. Technological Characteristics:

The proposed CXI™ TriForce consists of a Flexor® introducer, a CXI™ Support Catheter, and a Peel-Away® sheath. The CXI™ TriForce is substantially equivalent to the CXI™ Support Catheter (K072724), the FineCross™ MG Coronary Micro-Guide Catheter (K082519), and the Glidesheath™ (K102008). No new technological aspects are being introduced with the proposed device.

The proposed CXI™ TriForce was subjected to applicable testing to assure reliable design and performance under the testing parameters.

8. Test Data:

The following tests were performed to demonstrate that the CXI™ TriForce meets applicable design and performance requirements and supports a determination of substantial equivalence. Additionally, appropriate engineering tests were also performed on aged product to ensure that the CXI™ TriForce meets the performance requirements throughout the duration of shelf life.

  • Tensile Strength Testing shows the tensile strength during proper clinical use . should not fracture or rupture the introducer or catheter. In conformance with the

{2}------------------------------------------------

K111263 page 3 of 3

applicable sections of BS EN ISO 11070 and ISO 10555-1, the predetermined acceptance criteria were met.

  • Liquid Leakage Testing shows there would be no liquid leakage from the . introducer, introducer valve, or catheter during proper clinical use. In conformance with the applicable sections of ISO 11070 and ISO 10555-1, the predetermined acceptance criteria were met.
  • . Static Burst - Testing shows the pressures reached during proper clinical use (maximum pressure at maximum flow rate) are less than the static burst pressure of the catheter, and should not fracture or rupture the catheter. The predetermined acceptance criteria were met.
  • . Dynamic Burst - Testing shows the pressures reached during proper clinical use (maximum pressure at maximum flow rate) are less than the dynamic burst pressure of the catheter, and should not fracture or rupture the catheter. The predetermined acceptance criteria were met.
  • Flow Rate Testing shows the pressure exerted at the maximum flow rate during . proper clinical use should not fracture or rupture the catheter. The predetermined acceptance criteria were met.
  • Biocompatibility Testing shows the device is biocompatible. In conformance with . the applicable sections of ISO 10993-1, the predetermined acceptance criteria were met.

In conclusion, the results of these tests provide reasonable assurance that the device is as safe and as effective as the predicate devices and supports a determination of substantial equivalence.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circle's perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Cook Incorporated c/o Ms. Molly Busenbark 750 Daniels Way Bloomington, IN 47404

JUL 25 2011

Re: K111263

Trade/Device Name: CXI TriForce Peripheral Crossing Set Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II (two) Product Code: KRA Dated: July 13, 2011 Received: July 14, 2011

Dear Ms. Busenbark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Br
Di

Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K111263 page 161

4. Indications for Use Statement

510(k) Number (if known): K | | 26 3

CXI™ TriForce Peripheral Crossing Set Device Name:

Indications for Use:

The CXIM TriForce Peripheral Crossing Set is intended to be percutaneously introduced into blood vessels and support a wire guide while performing percutaneous peripheral intervention. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) diovascular Devices

510(k) Number K111263

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).