Search Results
Found 2 results
510(k) Data Aggregation
(71 days)
The Teleport microcatheters are indicated for:
supporting and facilitating the placement of guidewires in the coronary and peripheral vasculature.
exchanging guidewires in the coronary and peripheral vasculature.
the delivery of contrast media into the coronary, peripheral, and abdominal vasculature.
The Teleport family of microcatheters are single lumen catheters, offered in two shaft sizes (2.0F and 2.1F) with working lengths of 135cm or 150cm, designed for use in the coronary, peripheral, and abdominal vasculature. The shaft profiles gradually decrease from 2.6F (0.034") to 2.0F (0.0265") for the 2.0F configuration (Teleport), and 2.7F (0.036") to 2.1F (0.0275") for the 2.1F configuration (Teleport Control) respectively. The catheter consists of four sections: body shaft, proximal tip shaft, distal tip shaft, and a radiopaque tip. The distal most 60cm of the outer surface is coated with a hydrophilic polymer to increase lubricity and the lumen of the catheter is lined with a fluoropolymer to facilitate movement of the guidewire. The catheter is compatible with a standard 0.014 inch (0.36mm) guidewire.
The provided text describes a 510(k) summary for the Teleport Microcatheter, focusing on its substantial equivalence to a predicate device. However, it does not contain a detailed study proving the device meets specific acceptance criteria in the format requested. Instead, it lists the types of performance tests conducted and a general statement that "The Teleport microcatheter test results met all acceptance criteria and were similar to the predicate and reference devices."
Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor can I provide information on sample sizes for test sets, data provenance, number or qualifications of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth used, or training set details as these are not present in the provided document.
The document confirms that standalone performance testing was done for the device, as indicated by the list of "Performance Testing" conducted, which did not involve human readers (e.g., flow rate, burst pressure, torque strength).
Ground truth for these engineering and material performance tests would be established through predefined engineering specifications and standards (e.g., a specific flow rate must be achieved, a certain pressure must be withstood before burst).
Ask a specific question about this device
(8 days)
The PTA Balloon Catheter family is intended for balloon dilatation of lesions in the peripheral vasculature, including the iliac, femoral, popliteal, infra-popliteal, and renal arteries. NOT for use in the coronary or cerebral vasculature.
All of the devices in the PTA Balloon Catheter family, including the GliderXtreme PTA Balloon Catheter and the GliderfleX PTA Balloon Catheter consist of a standard balloon dilatation catheter on a braided shaft with or without a hypotube and an atraumatic, tapered and beveled tip. The devices are compatible with commonly used accessories, including standard 0.014" and 0.018" guidewires and 5F or 6F introducer sheath (or 6F or 7F guide catheter). Overall catheter lengths are approximately 100 cm to 150cm.
The distal end has a semi-compliant balloon that expands to known diameters (refer to compliance chart) at specific pressures. The balloon contains radiopaque markers to assist with positioning. The shaft is braid reinforced and has a lubricious hydrophilic coating.
The proximal end of the device is a common PTA catheter connected to a plastic hub and strain relief. The hub is used to inflate the balloon; the luer connector is compatible with standard inflation devices.
All PTA Balloon Catheters in the family are supplied sterile and intended for single use only.
The provided document focuses on the mechanical and physical performance of a medical device (PTA Balloon Catheter family) rather than an AI/ML device. Therefore, the questions related to AI/ML specific criteria (such as 'number of experts', 'adjudication method', 'MRMC study', 'standalone performance', 'training set sample size', and 'ground truth for training set') are not applicable to this document.
Here's the information that can be extracted relevant to the performance of the PTA Balloon Catheter:
1. Table of Acceptance Criteria and Reported Device Performance
The document states that "Bench testing was performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." However, specific quantitative acceptance criteria and the exact reported performance values are not detailed in this summary. Instead, it lists the types of tests performed.
Acceptance Criterion (Type of Test) | Reported Device Performance |
---|---|
Balloon Rated Burst Pressure | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Balloon Inflation and Deflation | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Balloon Fatigue | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Catheter Body Strength (Bond Strength) | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Torsional Strength | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Catheter Diameter, Balloon Profile and Tip Configuration | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Balloon Compliance | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Trackability, Pushability | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Kink Resistance | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
Device Interface Compatibility | "conformance to the requirements for its intended use" and "reliably achieved the desired effect and is safe for its intended use." Specific numeric values not provided. |
The document concludes that "Performance testing demonstrated that the devices reliably achieved the desired effect and is safe for its intended use. No new questions of safety or effectiveness were identified during device testing."
2. Sample size used for the test set and the data provenance
The document mentions "Bench testing was performed" but does not specify the sample sizes used for each test. The tests are in vitro (laboratory bench tests) to assess physical device properties, not clinical data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not an AI/ML device requiring human expert ground truth for interpretation of clinical data. The tests are physical performance tests.
4. Adjudication method for the test set
Not applicable. This is not an AI/ML device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used
The "ground truth" for this device would be established by engineering specifications, standards, and established physical measurement techniques. The tests performed (e.g., burst pressure, fatigue, strength) are quantitative physical performance evaluations against predefined engineering targets, not against clinical expert consensus or pathology.
8. The sample size for the training set
Not applicable. This is not an AI/ML device. There is no "training set."
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device.
Ask a specific question about this device
Page 1 of 1