K Number
K110555

Validate with FDA (Live)

Date Cleared
2011-03-24

(24 days)

Product Code
Regulation Number
888.3358
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
  1. Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis.
  2. Rheumatoid arthritis.
  3. Correction of functional deformity.
  4. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
  5. Revision of previously failed total hip arthroplasty.
  6. Dislocation risks.
    The ArComXL™ Active Articulation Head is a single-use implant, intended for uncemented applications.
Device Description

The ArComXL™ Active Articulation belongs to the family of dual mobility acetabular implants: the presence of two articulating surfaces in the same joint device. The ArComXL™ Active Articulation Head fits over a femoral modular head, which articulates within the ArComXL™ Head. The resultant assembly then articulates within the acetabular metal shell. The ArComXL™ Head is designed to be used with several styles of acetabular shells that have been cleared in previous submissions: M²a Magnum™ (K042037), Magnum™ Tri-Spike (K062995), and M²a 38™ Flared Cups and Non-Flared Cups (K011110).
The ArComXL™ Active Articulation Heads are available in sizes 44-66mm (Note: Size 44-66mm references O.D. of mating shell; the actual head sizes are 38-60mm.) and are manufactured from highly cross-linked polyethylene, conforming to ASTM F648. ArComXL™ is not a new material; the material and manufacturing process were cleared in K042051, ArComXL™ Polyethylene Liners, as well as subsequent submissions. The ArComXL™ Active Articulation is designed for both primary and total revision surgeries, where all device components associated with the wear couple are removed and replaced. The system is intended for uncemented applications.

AI/ML Overview

This document describes the Biomet ArComXL™ Active Articulation Head, a hip replacement component. This submission is for a material change from the predicate device.

Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance CriteriaReported Device Performance
Push-In and Lever-OutEquivalence to the Bipolar/Tri-Polar predicate (K991990). Meets or exceeds parameters.Six samples demonstrated equivalence to the Bipolar/Tri-Polar predicate (K991990). Meets parameters.
Wear TestingWear rates less than that of the ArCom™ 36mm liners (K032396). Meets or exceeds parameters.Tested for 5 million cycles, demonstrated equivalence to K032396. Meets parameters.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Push-In and Lever-Out: Six samples.
    • Wear Testing: Not explicitly stated, but typically multiple samples are used in ISO 14242 testing. The document states "the ArComXL™ Active Articulation heads were tested".
  • Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted by Biomet Manufacturing Corp. (the submitter). These are non-clinical tests.

3. Number of Experts and Qualifications for Ground Truth of Test Set

Not applicable. The reported studies are non-clinical mechanical tests (Push-In and Lever-Out, Wear Testing), not clinical studies requiring expert ground truth for interpretation.

4. Adjudication Method for Test Set

Not applicable. This section relates to clinical studies and expert review, which were not performed for the determination of substantial equivalence for this device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. No clinical studies were conducted, and therefore, no MRMC studies comparing human readers with and without AI assistance were performed. The device is a medical implant, not an AI-assisted diagnostic tool.

6. Standalone (Algorithm Only) Performance Study

Not applicable. This device is a physical implant, not an algorithm or software. The performance studies are non-clinical hardware tests.

7. Type of Ground Truth Used

  • Push-In and Lever-Out: Comparison to the performance of a legally marketed predicate device (K991990), establishing functional equivalence.
  • Wear Testing: Comparison to the wear rates of a legally marketed predicate device, ArCom™ 36mm liners (K032396), as per ISO 14242 standards.

8. Sample Size for Training Set

Not applicable. This device is a physical implant; there is no "training set" in the context of an algorithm or AI model development.

9. How Ground Truth for Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

i10555

MAR 2 4 2011

Image /page/0/Figure/2 description: The image shows the logo for Biomet Manufacturing Corp. The word "BIOMET" is displayed in a stylized, outlined font above the words "MANUFACTURING CORP." The logo is simple and professional, likely used for branding purposes.

510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92

Submitter Information
NameBiomet Manufacturing Corp.
Address56 East Bell DriveWarsaw, IN 46581-0857
Phone number(574) 267-6639
Fax number(574) 372-1683
EstablishmentRegistration Number1825034
Name of contactpersonBecky Earl
Date prepared02/23/2011
Name of device
Trade or proprietarynameArComXL™ Active Articulation
Common or usualnameArtificial Hip Replacement Component--Acetabular
Classification nameHip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis (21CFR §888.3358)
Classification panelOrthopedic
Regulation21CFR §888.3358
Product Code(s)LPH (888.3358), LZO (888.3353), KWY (888.3390)
Legally marketeddevice(s) to whichequivalence is claimedE1™ Avantage™ Head (E1™ Active Articulation), K101336
Reason for 510(k)submissionThe ArComXL™ Active Articulation is only a material changefrom the predicate, offering more options to patient,hospital, and surgeon.
Device descriptionThe ArComXL™ Active Articulation belongs to the family ofdual mobility acetabular implants: the presence of twoarticulating surfaces in the same joint device. TheArComXL™ Active Articulation Head fits over a femoralmodular head, which articulates within the ArComXL™Head. The resultant assembly then articulates within theacetabular metal shell. The ArComXL™ Head is designed tobe used with several styles of acetabular shells that have
been cleared in previous submissions: M²a Magnum™ (K042037), Magnum™ Tri-Spike (K062995), and M²a 38™Flared Cups and Non-Flared Cups (K011110).
The ArComXL™ Active Articulation Heads are available insizes 44-66mm (Note: Size 44-66mm references O.D. ofmating shell; the actual head sizes are 38-60mm.) and aremanufactured from highly cross-linked polyethylene,conforming to ASTM F648. ArComXL™ is not a newmaterial; the material and manufacturing process werecleared in K042051, ArComXL™ Polyethylene Liners, as wellas subsequent submissions. The ArComXL™ ActiveArticulation is designed for both primary and total revisionsurgeries, where all device components associated with thewear couple are removed and replaced. The system isintended for uncemented applications.
Intended use of thedeviceThe system is intended for uncemented applications.
Indications for use1. Noninflammatory degenerative joint disease, includingosteoarthritis and avascular necrosis.2. Rheumatoid arthritis.3. Correction of functional deformity.4. Treatment of non-union, femoral neck fracture andtrochanteric fractures of the proximal femur with headinvolvement, unmanageable using othertechniques.5. Revision of previously failed total hip arthroplasty.6. Dislocation risks.The ArComXL™ Active Articulation Head is a single-useimplant, intended for uncemented applications.
Summary of the technological characteristics of the device compared to thepredicate
CharacteristicNew DevicePredicate
DesignThe ArComXL ActiveArticulation™ Head fits overa femoral modular head,which articulates within theK101336
ArComXL™ head. Theresultant assembly thenarticulates within theacetabular metal shell.K042051
MaterialArComXL™, UHMWPE
Size Range44mm to 66mmK101336
PERFORMANCE DATA
SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OFSUBSTANTIAL EQUIVALENCE
Performance Test Summary-New Device
CharacteristicStandard/Test/FDAGuidanceResults Summary
Push-In and Lever-OutNASix samples, identical to finalproduction product andtested in the same manneras the predicatedemonstrated equivalence tothe Bipolar/Tri-Polarpredicate. (K991990)
Wear TestingISO 14242As in the E1™ ActiveArticulation predicate(K101336), the ArComXL™Active Articulation headswere tested for 5 millioncycles and compared to theArCom™ 36mm liners(K032396), the largest-sized,cleared ArCom™ liners.Acceptance criteria called forwear rates less than that ofthe ArCom liners; the testingdemonstrated equivalence toK032396.
Comparative Performance Information Summary
CharacteristicRequirementNew DevicePredicate Device
Push-In and LeverOutMeets or exceedsparametersMeets parametersK101336
Wear TestingMeets or exceedsparametersMeets parametersK101336
Push-In and LeverOutMeets or exceeds parametersMeets parametersK991990
Wear TestingMeets or exceeds parametersMeets parametersK032396
SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OFSUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical Performance Data/Information: Not applicable

P O. Box 567
Wastaw. III 46581-0581
Tot Froe: 201348.9500 ிங்கு 574 267 6639 Ыза Fac 574 767 8137 anter hightel com

p. 1 of 4

{1}------------------------------------------------

K110555

ਹੈ। ਉ ਤੀ The Trans MANUFACTURING CORP.

ડ્રો

PO Box 087
Walker (103) Roma (103) Rosenti (108) Rosental (108) Rosental (100 Friday (100 Friday (100 Friday) Norther (1972) North (1972) National (1988) Nationa

2 of ﮯ

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a sequence of handwritten characters. The sequence appears to be "K110555". The characters are written in a simple, slightly irregular style, typical of handwriting. The image has a clear, white background, which makes the characters stand out.

Image /page/2/Picture/1 description: The image shows the word "BIOMET" in a stylized, outlined font. The letters are blocky and have a three-dimensional appearance due to the outlining. There is a small trademark symbol in the upper right corner of the image.

MANUFACTURING CORP.

P.O. Box 537
Wall Friday (2010-05-2010
Tuall Friday (2010-05-2010)
Tuall Friday (2012-02-2010)
Luxuri (2012-02-2010)
Kustif (2010) (2010) 10:137
World Donel Donn


of U പ്ര 11

{3}------------------------------------------------

t110555

Image /page/3/Picture/1 description: The image shows the word "BIOMET" in a stylized, outlined font. The letters are blocky and have a three-dimensional appearance due to the outlining. The word is presented in all capital letters and appears to be a logo or brand name.

MANUFACTURING CORP.

CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA

No clinical testing was necessary for a determination of substantial equivalence.

The results of mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices.

P O : Box 587
Warsow 1H 46Sa1-0507
Trail Fine: 8JX0 348.8500 011-01-27 17:57 1533 Main Fax: 574 267 8131

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized bird in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The bird in the center is abstract, with flowing lines suggesting wings and a body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

BioMet Manufacturing Corp. % Ms. Becky Earl Regulatory Specialist P.O. Box 587 Warsaw, Indiana 46581-0587

Received: February 28, 2011

MAR 2 4 25.1

Re: K110555

Trade/Device Name: ArComXL" Active Articulation Head Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH, LZO, KWY Dated: February 25, 2011

Dear Ms. Earl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{5}------------------------------------------------

Page 2 - Ms. Becky Earl

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100, 1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, Jean note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

$\lg$ B. $R$ h

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): k1400555

Device Name: ArComXL™ Active Articulation Head

Indications For Use:

    1. Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis.
    1. Rheumatoid arthritis.
    1. Correction of functional deformity.
    1. Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.
    1. Revision of previously failed total hip arthroplasty.
    1. Dislocation risks.

The ArComXL™ Active Articulation Head is a single-use implant, intended for uncemented applications.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for M.Mellecam

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K110555

Page 1 of 1

N/A